Advertisement for orthosearch.org.uk
Results 1 - 9 of 9
Results per page:
Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_5 | Pages 31 - 31
23 Apr 2024
Bandopadhyay G Lo S Yonjan I Rose A Roditi G Drury C Maclean A
Full Access

Introduction

The presence of pluripotent mesenchymal cells in the periosteum along with the growth factors produced or released following injury provides this tissue with an important role in bone healing. Utilising this property, vascularised periosteal flaps may increase the union rates in recalcitrant atrophic long bone non-union. The novel chimeric fibula-periosteal flap utilises the periosteum raised on an independent periosteal vessel, thus allowing the periosteum to be inset freely around the osteotomy site, improving bone biology.

Materials & Methods

Ten patients, with established non-union, underwent fibula-periosteal chimeric flaps (2016–2022) at the Canniesburn Plastic Surgery Unit, UK. Preoperative CT angiography was performed to identify the periosteal branches. A case-control approach was used. Patients acted as their own controls, which obviated patient specific risks for non-union. One osteotomy site was covered by the chimeric periosteal flap and one without. In two patients both the osteotomies were covered using a long periosteal flap.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_18 | Pages 6 - 6
1 Nov 2017
Halai M Sharp E Drury C MacLean A
Full Access

Limb length deficiency, secondary to trauma or infection, is a common reason for referral to our tertiary service. After experiencing troubles with the Intramedullary Skeletal Kinetic Distractor (ISKD), we changed implant to the magnet operated Precice nail. We evaluated the safety and reliability of this novel device and compared it to our early ISKD results.

To minimise variables, we selected femurs only. In total, we reviewed medical and radiographic records of 20 cases (8 ISKD, 12 Precise) from 2010–2015. At each postoperative visit, the accuracy and precision of distraction and complications were recorded. Accuracy reflected how close the measured lengthening was to the prescribed distraction at each postoperative visit. Precision reflected how close the repeated measurements were to each other over the course of total lengthening period. No patients were lost to follow-up (1–3.5 years).

With the Precice nail (2012–2015), 11/12 were male and 10/12 were caused by trauma. The mean age was 34. Mean total lengthening was 38mm (range, 29–53mm), with an accuracy of 98percnt; and precision of 92percnt;. All patients achieved target lengthening at a rate of 1mm lengthening per day. In one case, the Precice nail fractured and this was revised successfully. With the ISKD group (2010–2012), 8/8 had complications (magnet jamming, nail breakage, equinus contractures and claw toes), with 25percnt; achieving accurate lengthening and precision of 38percnt;.

Our results so far have justified our change in implant choice and, in our experience, support the Precice nail as being safe and precise.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_31 | Pages 14 - 14
1 Aug 2013
Drury C Elias-Jones C Tait G
Full Access

Arthritis of the glenohumeral joint accompanied by an irreparable tear of the rotator cuff can cause severe pain, disability and loss of function, particularly in the elderly population. Anatomical shoulder arthroplasty requires a functioning rotator cuff, however, reverse shoulder arthroplasty is capable of addressing both rotator cuff disorders and glenohumeral deficiencies. The Aequalis Reversed Shoulder Prosthesis design is based on two bio-mechanical principles by Grammont; a medialized center of rotation located inside the glenoid bone surface and second, a 155 degree angle of inclination. Combined, they increase the deltoid lever arm by distalizing the humerus and make the prosthesis inherently stable.

24 consecutive primary reverse total shoulder arthroplasties were performed by a single surgeon for arthritis with rotator cuff compromise and 1 as a revision for a failed primary total shoulder replacement between December 2009 and October 2012. Patients were assessed postoperatively with the use of the DASH score, Oxford shoulder score, range of shoulder motion and plain radiography with Sirveaux score for scapular notching.

Mean age at the time of surgery was 72.5 years (range 59 to 86). Average follow up time was 19.4 months (range 4 to 38). Functional outcome scores from our series were comparable with patients from other follow up studies of similar prosthesis design. All patients showed improvement in range of shoulder movement postoperatively. Complications included one dislocation, one acromion fracture and one humeral shaft fracture. No cases of deep infection were recorded. Overall, the short-term clinical results were promising for this series of patients and indicate reverse shoulder arthroplasty as an appropriate treatment for this group of patients.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_30 | Pages 58 - 58
1 Aug 2013
Drury C Bailey O
Full Access

The Exeter Trauma Stem (ETS) is a monoblock unipolar prosthesis currently in use throughout various orthopaedic departments. It can be a useful procedure for specialty trainees in developing modern cementation techniques in hip arthroplasty. We propose that in order for this procedure to be a valid training tool that, as well as having a standardized surgical approach and operative technique, outcomes should be easily assessed and should be similar if performed by either a trainee or consultant.

All ETS procedures carried out at our institution from January 2009 until September 2011 were reviewed retrospectively. Patient demographics and operative details were recorded from patient notes. Radiographic evaluation involved the Barrack cementation grading system, Dorr's criteria, stem alignment and leg length measurement.

There was no significant difference in operative time between ETS performed by consultant or by specialty trainee. On postoperative x-ray, cement mantles were Barrack grade A or B in 55.6% (trainees) versus 61.9% (consultants). Stem alignment was neutral in 50% and varus in 50% of cases for trainees, versus 28.6% and 71.4% of cases for consultants. In total, 69.2 % of patients had lengthening of the operated limb with a mean increase of 10.4mm (2–25) for trainees and 9.3mm (2–18) for consultants.

Both trainees and consultants can attain a good cement mantle. However, from our results stem alignment is less accurate by trainees with half being placed in varus. Our results highlight the difficulty of obtaining correct leg length positioning with the ETS with trainees and consultants having similar discrepancies.

The ETS is a useful procedure for orthopaedic trainees to attain adequate skills in modern cementation techniques with similar post-operative radiographic outcomes to consultants. The tendency for trainees to be less accurate with stem positioning could be improved with supervision or careful pre-operative templating.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_30 | Pages 1 - 1
1 Aug 2013
Halai M Jayaram P Drury C Gregori A Murray D Oroko P Periasamy K
Full Access

Aluminia ceramic on ceramic (COC) bearing surfaces have been used for 35 years in total hip arthroplasty (THA). Studies report 85% survival at a minimum follow-up of 18.5 years. Nonetheless, an audible noise is a finding associated with COC bearings with incidence rates of 2–10%. This study aims to determine the prevalence of noise and evaluate its effect on patients.

All patients who had a COC THA from August 2003 to December 2010 were contacted and asked to complete a standardised questionnaire. This asked about the presence and characteristics of a noise and if associated with activities, pain and whether this phenomenon should be mentioned preoperatively.

Four consultant surgeons performed 282 consecutive primary COC THAs in 258 patients. (Male=122, Female=136 mean age 68.5; age range 28–88). In all cases, the same brand of ceramic acetabular component and stems were implanted. 11.0% had a noise, of which 5.5% had a squeak. Pain was experienced in 38.7% of patients in hips that made a noise. There was no trauma and one dislocation in this group. In this study, 85% of noises occur during weight-bearing although no patients have reduced daily activities as a result of the noises. Of all the patients, 55.0% stated they would have preferred to have known about a noisy hip possibility before consenting but none would have refused consent.

Squeaking has not been a problem here despite the prevalence being higher than most in the literature. The authors recommend that squeaking should be discussed preoperatively. A checklist for Orthopaedic Trainees is being drafted to enable trainees to counsel patients appropriately, allowing patients a better opportunity to give informed consent.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 499 - 499
1 Sep 2012
Bell S Young P Drury C Blyth M Jones B Mclean A
Full Access

Introduction

Total knee arthroplasty is an established and successful operation. In up to 10% of patients who undergo total knee arthroplasty continue to complain of pain [1]. Recently computerised tomography (CT) has been used to assess the rotational profile of both the tibial and femoral components in painful total knee arthroplasty.

Methods

We reviewed 56 painful total knee replacements and compared these to 59 pain free total knee replacements. Datum gathered from case notes and radiographs using a prospective orthopaedic database to identify patients. The age, sex, preoperative Oxford score and BMI, postoperative Oxford score and treatments recorded. The CT information recorded was limb alignment, tibial component rotation, femoral component rotation and combined rotation.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXIII | Pages 14 - 14
1 Jul 2012
Bell S Young P Drury C Jones CB Blyth M MacLean A
Full Access

Total knee arthroplasty (TKA) is an established and successful operation. However patient satisfaction rates vary from 81 to 89% 1,2,3. Pain following TKA is a significant factor in patient dissatisfaction 1. Many causes for pain following total knee arthroplasty have been identified 4 but rates of unexplained pain vary from 4 to 13.1% 5,6. Recently computerised tomography (CT) has been used to assess the rotational profile of both the tibial and femoral components in painful TKA

We reviewed 57 patients with an unexplained painful following TKA and compared these to a matched control group of 60 patients with TKA. Datum gathered from case notes and radiographs using a prospective database to identify patients. The CT information recorded was limb alignment, tibial component rotation, and femoral component rotation and combined rotation.

The two matched cohorts of patients had similar demographics. A significant difference in tibial, femoral and combined component rotation was identified between the groups. The following mean rotations were identified for the painful and control groups respectively. Tibial rotation was 3.46 degrees internal rotation (IR) compared to 2.50 degrees external rotation (ER)(p=0.001). Femoral rotation was 2.30 IR compared to 0.36 ER(p=0.02). Combined rotation was 7.08 IR compared to 2.85 ER(p=0.001).

This is the largest study presently in the literature. We have identified significant internal rotation in a patient cohort with unexplained painful TKA when compared to a matched control group. Internal rotation of the tibial component, femoral component and combined rotation was identified as a factor in unexplained pain following TKA.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXVIII | Pages 1 - 1
1 Jun 2012
Bell S Young P Drury C Jones B Blyth M MacLean A
Full Access

Total knee arthroplasty is an established and successful operation. In up to 13% of patients who undergo total knee arthroplasty continue to complain of pain. Recently computerised tomography (CT) has been used to assess the rotational profile of both the tibial and femoral components in painful total knee arthroplasty.

We reviewed 56 painful total knee replacements and compared these to 56 matched patients with pain free total knee replacements. Patients with infection, aseptic loosening, revision arthroplasties and gross coronal malalignment were excluded. Datum gathered from case notes and radiographs using a prospective orthopaedic database to identify patients. The age, sex, preoperative and postoperative Oxford scores, visual analogue scores and treatments recorded. The CT information recorded was limb alignment, tibial component rotation, femoral component rotation and combined rotation.

The two cohorts of patients had similar demographics. The mean limb alignments were 1.7 degrees varus and 0.01 degrees valgus in the painful and control groups respectively. A significant difference in tibial component rotation was identified between the groups with 3.2 degrees of internal rotation in the painful group compared to 0.5 degrees of external rotation in the control group (p=0.001). A significant difference in femoral component rotation was identified between the groups with 3.8 degrees of internal rotation in the painful group compared to 1.1 degrees of external rotation in the control group (p=0.001). A significant difference in the combined component rotation was identified between the groups with 6.8 degrees of internal rotation in the painful group compared to 1.7 degrees of external rotation in the control group (p=0.001).

We have identified significant internal rotation in a patient cohort with painful total knee arthroplasty when compared to a control group with internal rotation of the tibial component, femoral component and combined rotation. This is the largest comparison series currently in the literature.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXVIII | Pages 20 - 20
1 Jun 2012
Holloway N Drury C Ritchie I
Full Access

Metal-on-metal (MOM) hip arthroplasty, including resurfacing, has become the subject of recent research and debate. There is the perceived benefit of improved wear rates of bearing surfaces leading to superior durability and performance of these types of implant. An associated feature of MOM bearing surfaces is the generation of metal ions. These can have local and systemic cytotoxic effects. An immunoloigical response has been suggested, however, metal wear debris may cause direct damage to cellular DNA. Studies have shown that release of these ions is related to bearing diameter and component alignment. However, little is known about the relationship between metal ion levels and implant survivorship. The MHRA has published guidelines on the follow-up of patients with MOM implants including measurement of serum ion levels and cross sectional imaging.

Between February 2001 and November 2009, 135 patients (164 hips) had MOM resurfacing arthroplasty at our institution. We report a retrospective analysis of the data generated by review of these patients. Of the 135 patients, 91 were identified for clinical review. Each patient had serum metal ion levels measured, plain AP radiographs of the pelvis examined and, in the presence of raised metal ions, a Metal Artefact Reduction Sequence (MARS) MRI performed. 27 patients (35 hips) had raised metal ion levels (Cobalt and Chromium). Patients with raised metal ion levels had a mean acetabular cup inclination of 52.7 degrees compared with a mean inclination of 48.6 degrees in patients with normal ion levels (p<0.05). MARS MRI in the raised ion group revealed 9 patients with appearances suggestive of ALVAL. A number of these patients had hip revision surgery with the remainder awaiting potential revision.

These findings reflect current evidence suggesting a relationship between sub-optimal component position and raised metal ion levels and an increased rate of ALVAL.