Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_5 | Pages 27 - 27
23 Apr 2024
Howard A Harwood P Benton A Merrel C Culmer P Bolton W Stewart T
Full Access

Introduction

Ilizarov fixators are reliant on tensioned fine wires for stability. The tension in the wires is generated using specific tensioning devices. Loss of wire tension over time may lead of loss a stability and complications. A series of in vitro experiments were undertaken to explore wire tensioner accuracy, the impact of fixation bolt torque and initial tension on loss of tension in ilizarov constructs under static and dynamic loads.

Materials & Methods

Medical grade materials were applied to a synthetic bone analogue using surgical instruments in all experiments. Bolt torque was fixed at 6, 10 or 14 Nm using a torque limiting wrench. Wire tension was assessed using a strain measurement bridge. Wires were tensioned to 90, 110 and 130kg as measured by a commercial dynamometric tensioner. Static and dynamic testing was undertaken using an instron testing machine. Cyclical loads from 50–750N were applied for 5000 cycles.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_16 | Pages 17 - 17
1 Dec 2021
Shuttleworth M Vickers O Isaac G Culmer P Williams S Kay R
Full Access

Abstract

Objectives

Dual mobility (DM) hip implants whereby the polyethylene liner is “free-floating” are being used increasingly clinically. The motion of the liner is not well understood and this may provide insight into failure mechanisms; however, there are no published methods on tracking liner motion while testing under clinically relevant conditions. The aim was to develop and evaluate a bespoke inertial tracking system for DM implants that could operate submerged in lubricant without line-of-sight and provide 3D orientation information.

Methods

Trackers (n=5) adhered to DM liners were evaluated using a robotic arm and a six-degree of freedom anatomical hip simulator. Before each set of testing the onboard sensor suites were calibrated to account for steady-state and non-linearity errors. The trackers were subjected to ranges of motion from ±5° to ±25° and cycle frequencies from 0.35Hz to 1.25Hz and the outputs used to find the absolute error at the peak angle for each principle axis. In total each tracker was evaluated for ten unique motion profiles with each sequence lasting 60 cycles.