Radiostereometric Analysis (RSA) is a well developed imaging technique used to estimate implant fixation of orthopaedic implants in randomized clinical trials. The precision of RSA depends on a number of factors including image quality related to the individual modality properties. This study assesses the precision of RSA with a novel Digital Radiography (DR) system compared to a CR imaging system using different imaging techniques. Additionally, the study assesses the precision of locating beads embedded in a modified spine pedicle screw. A modified titanium spinal pedicle screw 4.5 mm diameter, 35 mm length, marked with two 1.0 mm tantalum beads, one inside the head and one near the screw tip was inserted into a bovine tibia segment. Six additional 1.0 mm tantalum beads were inserted into the bone segment – superiorly, distally and adjacent to the pedicle screw. The phantom was placed on a standard clinical diagnostic imaging bed above a custom RSA carbon fiber calibration cage (Halifax Biomedical Inc.). A pair of DR or CR imaging plates were placed below the calibration cage and irradiated 15 times at 100, 125 kV at 2.5 mAs. To determine precision, the standard deviation of 3D vector distances between beads was determined using RSA for each of the different imaging parameters.Purpose
Method
Radiostereometric Analysis (RSA) is an imaging method that is increasingly being utilized for monitoring fixation of orthopaedic implants in randomized clinical trials. Extensive RSA research has been conducted over the last 35+ years using standard clinical x-ray acquisition modalities that irradiate screen/film media or Computed Radiography (CR) plates. The precision of RSA can depend on a number of factors including modality image quality. This study assesses the precision of RSA with a novel Digital Radiography (DR) system compared to a CR imaging system using different imaging techniques. Additionally, the study assesses the precision of locating beads embedded in a modified spine pedicle screw.Introduction
Objective