header advert
Results 1 - 2 of 2
Results per page:
Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 305 - 305
1 Jul 2014
Roth A Bogie R Willems P Welting T Arts C van Rhijn L
Full Access

Summary Statement

Novel radiopaque UHMWPE sublaminar cables may be a promising alternative to gliding pedicle screws or titanium sublaminar cables within a growth-guidance system for the surgical treatment of early onset scoliosis.

Introduction

Growth-guidance or self-lengthening rod systems are an alternative to subcutaneous growing rods and the vertical expandable prosthetic titanium rib for the treatment of early onset scoliosis. Their main perceived advantage over growing rods is the marked decrease in subsequent operative procedures. The Shilla growth-guidance system and a modern Luque trolley are examples of such systems; both depend on gliding pedicle screws and/or sliding titanium sublaminar wires. However, the unknown consequences of metal-on-metal wear debris are reason for concern especially in young patients. In this study, instrumentation stability, residual growth in the operated segment after surgery and biocompatibility of the novel radiopaque UHMWPE cables as an alternative to gliding pedicles screws or titanium sublaminar wires were assessed in an immature sheep model.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 287 - 287
1 Sep 2012
Bogie R Voss L Welting T Willems P Arts J Van Rhijn L
Full Access

INTRODUCTION

Surgical correction of spinal deformities in the growing child can be applied with or without fusion. Sublaminar wiring, first described by Luque, allows continuation of growth of the non-fused spine after correction of the deformity. Neurological complications and wire breakage are the main clinical problems during the introduction and removal of currently used sublaminar wires. In this pilot study a posterior hybrid construction with the use of a medical-grade UHMWPE (Dyneema Purity®) sublaminar wire was assessed in an ovine model. We hypothesized that such a hybrid construction can safely replace current titanium laminar wires, while providing sufficient stability of the non-fused spinal column with preservation of growth.

MATERIALS AND METHODS

This study included 6 Tesselaar sheep, age 7±2months. Two pedicle screws (Legacy system, Medtronic) were placed at lumbar level. Four consecutive laminae were attached to two titanium bars (4.5 mm) using 3 mm diameter UHMWPE (Dyneema Purity®) on the left side and 5 mm diameter on the right side. The sublaminar wires were fixed with a double loop sliding knot and tightened with a tensioning device. As a control, in one animal titanium sublaminar wires (Atlas cable, Medtronic) were applied. After sacrifice the spine of the animals was harvested. Radiographs were taken and CT scans were performed. The vertebrae were dissected and placed in formaldehyde for macroscopic and histological evaluation.