Abstract
Summary Statement
Novel radiopaque UHMWPE sublaminar cables may be a promising alternative to gliding pedicle screws or titanium sublaminar cables within a growth-guidance system for the surgical treatment of early onset scoliosis.
Introduction
Growth-guidance or self-lengthening rod systems are an alternative to subcutaneous growing rods and the vertical expandable prosthetic titanium rib for the treatment of early onset scoliosis. Their main perceived advantage over growing rods is the marked decrease in subsequent operative procedures. The Shilla growth-guidance system and a modern Luque trolley are examples of such systems; both depend on gliding pedicle screws and/or sliding titanium sublaminar wires. However, the unknown consequences of metal-on-metal wear debris are reason for concern especially in young patients. In this study, instrumentation stability, residual growth in the operated segment after surgery and biocompatibility of the novel radiopaque UHMWPE cables as an alternative to gliding pedicles screws or titanium sublaminar wires were assessed in an immature sheep model.
Materials and methods
Twelve immature sheep were treated with segmental sublaminar spinal instrumentation: dual CoCr rods were held in place by pedicle screws at the most caudal instrumented level (L5) and novel radiopaque UHMWPE (Bi2O3 additive) woven cables were placed at 5 thoracolumbar levels. Lateral radiographs were taken at 4-week intervals to evaluate growth of the instrumented segment. Four age-matched, unoperated animals served as radiographic control. After 24 weeks follow-up, the animals were sacrificed and the spines were harvested for histological evaluation and CT analysis.
Results
No neurological deficits and no complications occurred during the initial postoperative period. One animal died during follow-up due to unknown cause. At sacrifice, none of the cables had loosened and the instrumentation remained stable. Substantial growth occurred in the instrumented segment (L5-T13) in the intervention group. No significant difference in growth of the operated segment was found between the intervention and control groups. Histological analysis showed fibrous encapsulation of the novel radiopaque UHMWPE sublaminar cable in the epidural space, with no evidence of chronic inflammation.
Discussion
Novel radiopaque UHMWPE cables may be a promising alternative to gliding pedicle screws or titanium sublaminar cables within a growth-guidance system. UHMWPE cables may improve growth results due to the smooth surface properties of the UHMWPE cable and address concerns regarding the consequences of metal-on-metal wear debris.