Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
Children's Orthopaedics

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_3 | Pages 9 - 9
1 Jan 2013
Abram S Stebbins J Theologis T Wainwright A
Full Access

Purpose

The purpose of this study was to assess the accuracy of three-dimensional camera technology when monitoring deformity correction by an Ilizarov frame and to compare it to manual measurements.

Methods and Results

A model consisting of an Ilizarov frame built around an artificial tibia and fibula was used with retro-reflective markers placed on the frame and bones to allow for the positions of each to be detected by the camera system. Measurements made by the camera system were compared to measurements taken manually. In the assessment of frame lengthening, the camera system average error was 2% (SD 2%) compared to 7% (SD 6%) for manual measurement. In the assessment of bone lengthening, the camera system average error was 4% (SD 4%) compared to 34% (SD 8%) for manual measurement. The technology also demonstrated good accuracy in the measurement of angular deformity changes.