Abstract. OBJECTIVES. Staphylococcus aureus is one of the most common pathogens in orthopaedic biomaterial-associated infections. The transition of planktonic S. aureus to its biofilm phenotype is critical in the pathogenesis of biomaterial-associated infections and the development of antimicrobial
Introduction:. Angular mismatch of the modular junction between the head and the trunion has been recognized as a contributing factor to fretting and corrosion of hip prostheses. Excessive angular-mismatch can lead to relative motion at the taper interface, and tribo-corrosion of the head-neck junction secondary to disruption of the passive oxide layer. Although manufacturing standards have been adopted to define acceptable
Background. After surgical correction of thoracic scoliosis, an improvement in the cardio-respiratory adaptation to exercise would be expected because of the correction of the rib cage associated with the spinal deformity. This work intended to evaluate the physiologic responses to incremental exercise in patients undergoing surgical correction of adolescent idiopathic scoliosis (AIS). The hypothesis of this study was that the exercise limitations described in patients with AIS could be related with the physical deconditioning instead of being linked to the severity of the vertebral deformity. Methods. Cross-sectional study of the exercise
Introduction: Existing fatigue studies of ACL fixation have two disadvantages. There is no agreed standard protocol, making comparison of various studies difficult and average results are presented, disregarding data spread. This may be over-optimistic, because approximately half the fixations will not achieve the average level. The effect of data spread can be summarised using the one-sided 80/80 lower
Aim. The use of piperacillin/tazobactam with vancomycin as empirical antimicrobial therapy (EAT) for prosthetic joint infection (PJI) has been associated with an increased risk of acute kidney injury (AKI), leading to propose cefepim as an alternative since 2017 in our reference center. The present study compared microbiological efficacy and
Purpose of the study: Because of its specific properties, linezolid (LZD) is an alternative to glycopeptides for the treatment of BJI causes by resistant Gram-positive cocci. The efficacy and
Axial loading of the foot/ankle complex is an important injury mechanism in vehicular trauma, responsible for severe injuries such as calcaneus, talus and tibial pilon fractures. Axial loading may be applied to the leg externally, by the toepan and/or pedals, as well as internally by active muscle tension applied through the Achilles tendon during pre-impact bracing. In order to evaluate the effect of active muscle tension on the injury-tolerance of the foot/ankle complex, axial impact tests were performed on isolated lower legs, with and without experimentally stimulated muscle tension applied through the Achilles’ tendon. Acoustic emission was used to determine the exact time of fracture during the tests. The primary fracture mode was calcaneal fracture in both groups, but tibial pilon fractures occurred more frequently with the addition of Achilles tension. A linear regression model was developed that describes the expected axial loading injury
The aim of this study was to compare the pain caused by the application
of a tourniquet after exsanguination of the upper limb with that
occurring after simple elevation. We used 26 healthy volunteers (52 arms), each of whom acted as
their own matched control. The primary outcome measure was the total pain experienced by
each volunteer while the tourniquet was inflated for 20 minutes.
This was calculated as the area under the pain curve for each individual
subject. Secondary outcomes were pain at each time point; the total
pain experienced during the recovery phase; the ability to tolerate the
tourniquet and the time for full recovery after deflation of the
tourniquet. Aims
Patients and Methods
Aims. The optimum clearance between the bearing surfaces of hip arthroplasties is unknown. Theoretically, to minimize wear, it is understood that clearances must be low enough to maintain optimal contact pressure and fluid film lubrication, while being large enough to allow lubricant recovery and reduce contact patch size. This study aimed to identify the relationship between diametrical clearance and volumetric wear, through the analysis of retrieved components. Methods. A total of 81 metal-on-metal Pinnacle hips paired with 12/14 stems were included in this study. Geometrical analysis was performed on each component, using coordinate and roundness measuring machines. The relationship between their as-manufactured diametrical clearance and volumetric wear was investigated. The Mann-Whitney U test and unpaired t-test were used, in addition to calculating the non-parametric Spearman's correlation coefficient, to statistically evaluate the acquired data. Results. The hips in this study were found to have had a median unworn diametrical clearance of 90.31 μm (interquartile range (IQR) 77.59 to 97.40); 32% (n = 26) were found to have been below the manufacturing
Aims. To achieve expert clinical consensus in the delivery of hydrodilatation for the treatment of primary frozen shoulder to inform clinical practice and the design of an intervention for evaluation. Methods. We conducted a two-stage, electronic questionnaire-based, modified Delphi survey of shoulder experts in the UK NHS. Round one required positive, negative, or neutral ratings about hydrodilatation. In round two, each participant was reminded of their round one responses and the modal (or ‘group’) response from all participants. This allowed participants to modify their responses in round two. We proposed respectively mandating or encouraging elements of hydrodilatation with 100% and 90% positive consensus, and respectively disallowing or discouraging with 90% and 80% negative consensus. Other elements would be optional. Results. Between 4 August 2020 and 4 August 2021, shoulder experts from 47 hospitals in the UK completed the study. There were 106 participants (consultant upper limb orthopaedic surgeons, n = 50; consultant radiologists, n = 52; consultant physiotherapist, n = 1; extended scope physiotherapists, n = 3) who completed round one, of whom 97 (92%) completed round two. No elements of hydrodilatation were “mandated” (100% positive rating). Elements that were “encouraged” (≥ 80% positive rating) were the use of image guidance, local anaesthetic, normal saline, and steroids to deliver the injection. Injecting according to patient
Local antimicrobial therapy is an integral aspect of treating orthopaedic device related infection (ODRI), which is conventionally administered via polymethylmethacrylate (PMMA) bone cement. PMMA, however, is limited by a suboptimal antibiotic release profile and a lack of biodegradability. In this study, we compare the efficacy of PMMA versus an antibioticloaded hydrogel in a single- stage revision for chronic methicillin-resistant Staphylococcus aureus (MRSA) ODRI in. sheep. Antibiofilm activity of the antibiotic combination (gentamicin and vancomycin) was determined in vitro. Swiss alpine sheep underwent a single-stage revision of a tibial intramedullary nail with MRSA infection. Local gentamicin and vancomycin therapy was delivered via hydrogel or PMMA (n = 5 per group), in conjunction with systemic antibiotic therapy. In vivo observations included: local antibiotic tissue concentration, renal and liver function tests, and quantitative microbiology on tissues and hardware post-mortem. There was a nonsignificant reduction in biofilm with an increasing antibiotic concentration in vitro (p = 0.12), confirming the antibiotic
Fracture related infection remains a major challenge in musculoskeletal trauma surgery. Despite best practice, treatment strategies suffer from high failure rates due to antibiotic resistance and
Introduction. Acquiring adaptive soft-tissue balance is one of the most important factors in total knee arthroplasty (TKA). However, there have been few reports regarding to alteration of tolerability of varus/valgus stress between before and after TKA. In particular, there is no enough data about mid-flexion stability. Based on these backgrounds, it is hypothesized that alteration of varus/valgus
Introduction. Fretting corrosion of the modular taper junction in total hip arthroplasty has been studied in several finite element (FE) studies. Manufacturing
Staphylococcus aureus is the main cause of osteomyelitis and forms biofilm and staphylococcal abscess communities (SACs) in humans. While S. aureus has several toxins with specificity for human targets and working with human host cells would be preferred, for SACs no in vitro models, two-dimensional (2D) or three-dimensional (3D), have been described in literature to date. Advanced 3D in vitro cell culture models enable the incorporation of human cells and resemble in vivo tissue more closely than conventional 2D cell culture. Therefore, the aim of this study was to develop an in vitro model of SACs by using a 3D system. The model should allow for studies into antibiotic
Aims. Due to their radiolucency and favourable mechanical properties, carbon fibre nails may be a preferable alternative to titanium nails for oncology patients. We aim to compare the surgical characteristics and short-term results of patients who underwent intramedullary fixation with either a titanium or carbon fibre nail for pathological long-bone fracture. Methods. This single tertiary-institutional, retrospectively matched case-control study included 72 patients who underwent prophylactic or therapeutic fixation for pathological fracture of the humerus, femur, or tibia with either a titanium (control group, n = 36) or carbon fibre (case group, n = 36) intramedullary nail between 2016 to 2020. Patients were excluded if intramedullary fixation was combined with any other surgical procedure/fixation method. Outcomes included operating time, blood loss, fluoroscopic time, and complications. Fisher’s exact test and Mann-Whitney U test were used for categorical and continuous outcomes, respectively. Results. Patients receiving carbon nails as compared to those receiving titanium nails had higher blood loss (median 150 ml (interquartile range (IQR) 100 to 250) vs 100 ml (IQR 50 to 150); p = 0.042) and longer fluoroscopic time (median 150 seconds (IQR 114 to 182) vs 94 seconds (IQR 58 to 124); p = 0.001). Implant complications occurred in seven patients (19%) in the titanium group versus one patient (3%) in the carbon fibre group (p = 0.055). There were no notable differences between groups with regard to operating time, surgical wound infection, or survival. Conclusion. This pilot study demonstrates a non-inferior surgical and short-term clinical profile supporting further consideration of carbon fibre nails for pathological fracture fixation in orthopaedic oncology patients. Given enhanced accommodation of imaging methods important for oncological surveillance and radiation therapy planning, as well as high
Staphylococcus aureus is responsible for 60–70% infections of surgical implants and prostheses in Orthopaedic surgery, with cumulative treatment costs for all prosthetic joint infections estimated to be ∼ $1 billion per annum (UK and North America). Its ability to develop resistance or
Aims. The purpose of this investigation was to determine the relationship between height, weight, and sex with implant size in total knee arthroplasty (TKA) using a multivariate linear regression model and a Bayesian model. Methods. A retrospective review of an institutional registry was performed of primary TKAs performed between January 2005 and December 2016. Patient demographics including patient age, sex, height, weight, and body mass index (BMI) were obtained from registry and medical record review. In total, 8,100 primary TKAs were included. The mean age was 67.3 years (SD 9.5) with a mean BMI of 30.4 kg/m. 2. (SD 6.3). The TKAs were randomly split into a training cohort (n = 4,022) and a testing cohort (n = 4,078). A multivariate linear regression model was created on the training cohort and then applied to the testing cohort . A Bayesian model was created based on the frequencies of implant sizes in the training cohort. The model was then applied to the testing cohort to determine the accuracy of the model at 1%, 5%, and 10%
INTRODUCTION. Precise determination of material loss is essential for failure analysis of retrieved hip cups. To determine wear, the measured geometry of the retrieval hast to be compared to its pristine geometry, which usually is not available. There are different approaches to generate reference geometries to approximate the pristine geometry that is commonly assumed as sphere. However, the geometry of press fit cup retrievals might not be spherical due to deformation caused by excessive press-fitting. The effect of three different reference geometries on the determined wear patterns and material loss of pristine and worn uncemented metal-on-metal hip cups was determined. METHODS. The surfaces of two cups (ASR, DePuy, Leeds; one pristine, one a worn retrieval) were digitized using a coordinate measurement machine (CRYSTA-Apex S574, Mitutoyo; 3 µm accuracy). Both cups were measured undeformed and while being deformed between a clamp. Three different methods for generating reference geometries were investigated (PolyWorks|Inspector 2018, InnovMetric). Method 1: A sphere with the nominal internal cup dimensions was generated. Method 2: A sphere was fitted to the measured data points after removing those from worn areas (deviation > 3 µm is defined as wear) to eliminate the influence of manufacturing
Aim. Bone regeneration following the treatment of Staphylococcal bone infection or osteomyelitis is challenging due to the ability of Staphylococcus aureus to invade and persist within bone cells, which could possibly lead to antimicrobial