Advertisement for orthosearch.org.uk
Results 1 - 20 of 85
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 738 - 745
1 Jun 2011
Davda K Lali FV Sampson B Skinner JA Hart AJ

We retrospectively analysed concentrations of chromium and cobalt ions in samples of synovial fluid and whole blood taken from a group of 92 patients with failed current-generation metal-on-metal hip replacements. We applied acid oxidative digestion to our trace metal analysis protocol, which found significantly higher levels of metal ion concentrations in blood and synovial fluid than a non-digestive method. Patients were subcategorised by mode of failure as either ‘unexplained pain’ or ‘defined causes’. Using this classification, chromium and cobalt ion levels were present over a wider range in synovial fluid and not as strongly correlated with blood ion levels as previously reported. There was no significant difference between metal ion concentrations and manufacturer of the implant, nor femoral head size below or above 50 mm. There was a moderately positive correlation between metal ion levels and acetabular component inclination angle as measured on three-dimensional CT imaging. Our results suggest that acid digestion of samples of synovial fluid samples is necessary to determine metal ion concentrations accurately so that meaningful comparisons can be made between studies


Bone & Joint Research
Vol. 8, Issue 3 | Pages 146 - 155
1 Mar 2019
Langton DJ Natu S Harrington CF Bowsher JG Nargol AVF

Objectives. We investigated the reliability of the cobalt-chromium (CoCr) synovial joint fluid ratio (JFR) in identifying the presence of a severe aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL) response and/or suboptimal taper performance (SOTP) following metal-on-metal (MoM) hip arthroplasty. We then examined the possibility that the CoCr JFR may influence the serum partitioning of Co and Cr. Methods. For part A, we included all revision surgeries carried out at our unit with the relevant data, including volumetric wear analysis, joint fluid (JF) Co and Cr concentrations, and ALVAL grade (n = 315). Receiver operating characteristic curves were constructed to assess the reliability of the CoCr JFR in identifying severe ALVAL and/or SOTP. For part B, we included only patients with unilateral prostheses who had given matched serum and whole blood samples for Co and Cr analysis (n = 155). Multiple regression was used to examine the influence of JF concentrations on the serum partitioning of Co and Cr in the blood. Results. A CoCr JFR > 1 showed a specificity of 83% (77% to 88%) and sensitivity of 63% (55% to 70%) for the detection of severe ALVAL and/or SOTP. In patients with CoCr JFRs > 1, the median blood Cr to serum Cr ratio was 0.99, compared with 0.71 in patients with CoCr JFRs < 1 (p < 0.001). Regression analysis demonstrated that the blood Cr to serum Cr value was positively associated with the JF Co concentration (p = 0.011) and inversely related to the JF Cr concentration (p < 0.001). Conclusion. Elevations in CoCr JFRs are associated with adverse biological (severe ALVAL) or tribocorrosive processes (SOTP). Comparison of serum Cr with blood Cr concentrations may be a useful additional clinical tool to help to identify these conditions. Cite this article: D. J. Langton, S. Natu, C. F. Harrington, J. G. Bowsher, A. V. F. Nargol. Is the synovial fluid cobalt-to-chromium ratio related to the serum partitioning of metal debris following metal-on-metal hip arthroplasty? Bone Joint Res 2019;8:146–155. DOI: 10.1302/2046-3758.83.BJR-2018-0049.R1


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 20 - 20
23 Jun 2023
Macheras G Papadakis S Argyrou C Kateros K
Full Access

Pre-operative definitive diagnosis of infection in painful total hip arthroplasty (THA) is not always easy to be established, making the intra-operative decision-making process crucial in management of revision hip surgery. Calprotectin is a promising point-of-care novel biomarker that has displayed high accuracy in detecting PJIs. From November 2020 to December 2022, 105 patients with painful primary THA were treated with revision THA in 3 orthopaedic departments. Pre-operatively, 23 were considered infected and treated with two-stage revision THA. The remaining 82 were likely infected according to the 2019 EBJIS criteria. The suspicion of low-grade infection was based on clinical (rest and/or night pain), laboratory (CRP, ESR, WBC – normal or slightly elevated) and radiological evaluation (loosening). Hip aspiration under CT imaging was performed in these cases and 34 of them yielded positive culture and were treated with two-stage revision. Aspiration was ineffective in the remaining 48 cases (33 negative, 15 unsuccessful attempts). Intra-operatively, calprotectin was measured with lateral flow immunoassay test in these patients. Cases with calprotectin levels ≥ 50 mg/L were treated with 2-stage revision THA; otherwise, they were considered not-infected and one-stage revision was performed. Synovial fluid and tissue samples were collected for analysis. Implants were sent for sonication fluid cultures. Calprotectin was positive (≥ 50 mg/L) in 27 cases and negative in 21 cases. There was 1 false negative case with positive tissue cultures. Out of the 27 positive cases, 25 had positive tissue cultures and sonication. However, 2 cases with high calprotectin levels (>200 mg/L) were not infected. The false positive result was attributed to severe metallosis. Calprotectin sensitivity was 96.2%, specificity 90.9%, PPV 92.6%, NPV 95.2%, AUC 0.935. The results of this ongoing study indicate that calprotectin seems to be a valuable tool in facilitating the intra-operative decision-making process in cases that low-grade infection is suspected and diagnosis cannot be established pre-operatively


Bone & Joint Research
Vol. 5, Issue 2 | Pages 52 - 60
1 Feb 2016
Revell PA Matharu GS Mittal S Pynsent PB Buckley CD Revell MP

Objectives. T-cells are considered to play an important role in the inflammatory response causing arthroplasty failure. The study objectives were to investigate the composition and distribution of CD4+ T-cell phenotypes in the peripheral blood (PB) and synovial fluid (SF) of patients undergoing revision surgery for failed metal-on-metal (MoM) and metal-on-polyethylene (MoP) hip arthroplasties, and in patients awaiting total hip arthroplasty. Methods. In this prospective case-control study, PB and SF were obtained from 22 patients (23 hips) undergoing revision of MoM (n = 14) and MoP (n = 9) hip arthroplasties, with eight controls provided from primary hip osteoarthritis cases awaiting arthroplasty. Lymphocyte subtypes in samples were analysed using flow cytometry. Results. The percentages of CD4+ T-cell subtypes in PB were not different between groups. The CD4+ T-cells in the SF of MoM hips showed a completely different distribution of phenotypes compared with that found in the PB in the same patients, including significantly decreased CD4+ T-central memory cells (p < 0.05) and increased T-effector memory cells (p < 0.0001) in the SF. Inducible co-stimulator (ICOS) was the only co-stimulatory molecule with different expression on CD4+ CD28+ cells between groups. In PB, ICOS expression was increased in MoM (p < 0.001) and MoP (p < 0.05) cases compared with the controls. In SF, ICOS expression was increased in MoM hips compared with MoP hips (p < 0.05). Conclusions. Increased expression of ICOS on CD4+ T-cells in PB and SF of patients with failed arthroplasties suggests that these cells are activated and involved in generating immune responses. Variations in ICOS expression between MoM and MoP hips may indicate different modes of arthroplasty failure. Cite this article: Professor P. A. Revell. Increased expression of inducible co-stimulator on CD4+ T-cells in the peripheral blood and synovial fluid of patients with failed hip arthroplasties. Bone Joint Res 2016;5:52–60. doi: 10.1302/2046-3758.52.2000574


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_1 | Pages 39 - 39
1 Jan 2018
Houdek M Wyles C Jannetto P Berry D Lewallen D
Full Access

Serum and blood cobalt (Co) and chromium (Cr) ion levels are used to monitor patients at risk for adverse reaction to metal debris (ARMD) following metal-on-metal (MoM) total hip arthroplasty (THA). However, these levels often do not correlate with the degree of local soft-tissue reaction and damage observed at the time of revision. The purpose of this study was to analyze synovial fluid metal ion concentrations in patients with a failed THA in the setting of an ARMD and determine if these levels can be more predictive of soft-tissue destruction than serum or whole blood levels. Synovial, blood and serum samples were prospectively collected from patients undergoing revision THA with ARMD (n=29) and those undergoing aseptic revision without ARMD (n=29). There was no difference in mean age (P=0.50), BMI (P=0.18), sex distribution (P=0.18), serum creatinine (P=0.74), or time to revision THA (P=0.13) between the cohorts. In the AMRD cohort, the components included MoM THA (n=18), hip resurfacing (n=5), dual-modular taper THA (n=4) and MoM and dual-modular taper THA (n=2). At the time of revision THA, 26 (90%) patients in the metal reaction cohort had gross evidence of metallosis in the soft-tissues, the remaining 3 (10%) had evidence of corrosion of the dual taper neck or MoM bearing. In the non ARMD cohort the bearing surfaces included metal-on-polyethylene (n=19) and ceramic-on-polyethylene (n=10). The indications for revision included isolated acetabular loosening (n=11), isolated femoral component loosening (n=11), polyethylene wear (n=5), recurrent dislocation (n=1) and combined femoral and acetabular component loosening (n=1). None had a clinical diagnosis or gross evidence of taper corrosion. Pre-revision, 21 (72%) patients in the metal reaction group had periarticular fluid collections or a mass on MRI. Mean cyst size was 202.9±71.6 cm3 and masses were grouped into Type I (cyst wall <3 mm, n=10), Type II (cyst wall ≥3 mm, n=8) and Type III (mainly solid, n=3). At the time of revision THA, the mean Co levels were elevated in patients with ARMD compared to those without in synovial fluid (1,833 ppb vs. 12.3 ppb, P=0.008), whole blood (22.6 ppb vs 0.5 ppb, P=0.005)) and serum (19.6 vs. 0.6, P=0.001). Likewise, mean Cr levels were significantly elevated in patients with an ARMD compared to those undergoing revision without in synovial fluid (3,128 ppb vs. 10.3 ppb, P=0.01), whole blood (8.9 ppb vs. 0.5 ppb, P=0.009) and serum (14.1 ppb vs. 0.5 ppb, P=0.005). The synovial fluid Co levels were the most accurate test for detecting pseudotumor (AUC 0.951) and adverse local tissue reaction (AUC 0.826). At a synovial fluid Cr level of 110 ppb, the synovial fluid metal ion analysis was 94% sensitive and 86% specific for pseudotumor formation. In this prospective study, synovial fluid analysis of metal ion levels was more accurate in predicting the presence and extent of pseudo-tumor or ALTR compared to blood or serum analysis. The addition of synovial aspiration with metal ion analysis may provide another helpful data point when risk stratifying these patients for need for revision THA


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_5 | Pages 37 - 37
1 Jul 2020
Poznalska G Auckland C Heddon S Phillips J Wilson M
Full Access

Joint replacement is a life-enhancing, cost-effective surgical intervention widely used to treat disabling joint pain mainly caused by osteoarthritis. Hip and knee joint replacements are common, highly successful operations bringing many patients relief from pain, and improve mobility.

Prosthetic joint infections (PJI) are often difficult to diagnose; diagnosis often relying on a combination of clinical findings, microbiological data and histological evaluation of periprosthetic tissue. The majority of recent studies demonstrate a higher sensitivity for the culture of sonication fluid (62–94%) than periprosthetic tissue (55–88%).

The Royal Devon & Exeter NHS Foundation Trust provided a specialist multidisciplinary team. We evaluated the use of sonication for identifying pathogens during revision hip and knee replacement in our unit with the intention of applying for further grants to study this technology in more detail; it was a diagnostic feasibility study. The target recruitment was 50 patients; 25 consecutive patients undergoing revision hip arthroplasty for any reason, and 25 undergoing revision knee arthroplasty for any reason. The majority of patients were identified in outpatient clinics.

Our results show that tissue culture remains more sensitive than sonication. The sensitivity for direct sonication was 75% compared with 80% for tissue culture; sensitivity of enriched sonication was 80 %, compared to the unit results for tissue culture of 83%. The combined sensitivity of tissue culture was better as a higher number of tissue specimens were obtained.

The different methodology might explain reported differences between this and other studies; however, our study does not support the use of sonication as the only tool in the diagnosis of PJI.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_11 | Pages 41 - 41
7 Jun 2023
Jeyapalan R Baker P Jameson S Rangarajan S Kubelka I
Full Access

Elevated synovial leukocyte count is a minor criterion derived from the musculoskeletal infection society (MSIS) widely used in clinical practice for diagnosis of prosthetic joint infection. There is evidence to suggest analysis within 1 hour, preferentially within 30 minutes, of aspiration reduces the risk of ex vivo cell lysis occurring during prolonged transport. Multiple site working is more common practice and the availability of a lab on site to perform these tests is not always possible. We aimed to assess whether we could safely perform synovial leukocyte counts within our cold site in the diagnosis of prosthetic joint infection. We reviewed all orthopaedic synovial fluid aspirates within the lower limb arthroplasty unit from April 2021 – April 2022 performed at South Tees NHS Foundation Trust. We assessed time from aspirate to the lab using electronic data resources. This information was compared with the labs ability to perform a synovial leukocyte count to determine the impact of delays on testing. 110 patients (34.5% hips and 63.6% knees) were identified between two sites. Time from aspirate to lab ranged from 0 mins to 26 hrs 34 mins. Mean time to processing was 3hrs 10 mins. 50% of all samples had a synovial leukocyte count performed. 67% of patients had a cell differential performed. There was no difference in the ability to perform a synovial leukocyte count between samples process in < 2hours vs > 6 hours. We conclude that it is safe practice to perform joint aspirates for the work up of periprosthetic joint infections in sites where no laboratory is immediately available as the delay to processing synovial fluid does not alter the ability to perform a synovial leukocyte count. This study will provide evidence to enable the work up of periprosthetic joint infections in cold centres and therefore reduce the delay in diagnosis and proceeding management


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 867 - 874
1 Jul 2022
Ji B Li G Zhang X Xu B Wang Y Chen Y Cao L

Aims. Periprosthetic joint infections (PJIs) with prior multiple failed surgery for reinfection represent a huge challenge for surgeons because of poor vascular supply and biofilm formation. This study aims to determine the results of single-stage revision using intra-articular antibiotic infusion in treating this condition. Methods. A retrospective analysis included 78 PJI patients (29 hips; 49 knees) who had undergone multiple prior surgical interventions. Our cohort was treated with single-stage revision using a supplementary intra-articular antibiotic infusion. Of these 78 patients, 59 had undergone more than two prior failed debridement and implant retentions, 12 patients had a failed arthroplasty resection, three hips had previously undergone failed two-stage revision, and four had a failed one-stage revision before their single-stage revision. Previous failure was defined as infection recurrence requiring surgical intervention. Besides intravenous pathogen-sensitive agents, an intra-articular infusion of vancomycin, imipenem, or voriconazole was performed postoperatively. The antibiotic solution was soaked into the joint for 24 hours for a mean of 16 days (12 to 21), then extracted before next injection. Recurrence of infection and clinical outcomes were evaluated. Results. A total of 68 patients (87.1%) were free of infection at a mean follow-up time of 85 months (24 to 133). The seven-year infection-free survival was 87.6% (95% confidence interval (CI) 79.4 to 95.8). No significant difference in infection-free survival was observed between hip and knee PJIs (91.5% (95% CI 79.9 to 100) vs 84.7% (95% CI 73.1 to 96.3); p = 0.648). The mean postoperative Harris Hip Score was 76.1 points (63.2 to 92.4) and Hospital for Special Surgery score was 78. 2 (63.2 to 92.4) at the most recent assessment. Polymicrobial and fungal infections accounted for 14.1% (11/78) and 9.0% (7/78) of all cases, respectively. Conclusion. Single-stage revision with intra-articular antibiotic infusion can provide high antibiotic concentration in synovial fluid, thereby overcoming reduced vascular supply and biofilm formation. This supplementary route of administration may be a viable option in treating PJI after multiple failed prior surgeries for reinfection. Cite this article: Bone Joint J 2022;104-B(7):867–874


Bone & Joint Research
Vol. 7, Issue 6 | Pages 388 - 396
1 Jun 2018
Langton DJ Sidaginamale RP Joyce TJ Bowsher JG Holland JP Deehan D Nargol AVF Natu S

Objectives. We have encountered patients who developed large joint fluid collections with massive elevations in chromium (Cr) and cobalt (Co) concentrations following metal-on-metal (MoM) hip arthroplasties. In some cases, retrieval analysis determined that these ion concentrations could not be explained simply by the wear rates of the components. We hypothesized that these effects may be associated with aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL). Patients and Methods. We examined the influence of the ALVAL grade on synovial fluid Co and Cr concentrations following adjustment for patient and device variables, including volumetric wear rates. Initially restricting the analysis to include only patients with one MoM hip resurfacing device, we performed multiple regression analyses of prospectively collected data. We then repeated the same statistical approach using results from a larger cohort with different MoM designs, including total hip arthroplasties. Results. In the resurfacing cohort (n = 76), the statistical modelling indicated that the presence of severe ALVAL and a large fluid collection were associated with greater joint fluid Co concentrations after adjustment for volumetric wear rates (p = 0.005). These findings were replicated in the mixed implant group (n = 178), where the presence of severe ALVAL and a large fluid collection were significantly associated with greater fluid Co concentrations (p < 0.001). Conclusion. The development of severe ALVAL is associated with elevations in metal ion concentrations far beyond those expected from the volumetric loss from the prosthetic surfaces. This finding may aid the understanding of the sequence of events leading to soft-tissue reactions following MoM hip arthroplasties. Cite this article: D. J. Langton, R. P. Sidaginamale, T. J. Joyce, J. G. Bowsher, J. P. Holland, D. Deehan, A. V. F. Nargol, S. Natu. Aseptic lymphocyte-dominated vasculitis-associated lesions are related to changes in metal ion handling in the joint capsules of metal-on-metal hip arthroplasties. Bone Joint Res 2018;7:388–396. DOI: 10.1302/2046-3758.76.BJR-2018-0037


Bone & Joint Research
Vol. 10, Issue 10 | Pages 693 - 703
1 Oct 2021
Wang X Wang D Xia P Cheng K Wang Q Wang X Lin Q Song J Chen A Li X

Aims. To evaluate the effect of ultrasound-targeted simvastatin-loaded microbubble destruction (UTMDSV) for alleviation of the progression of osteoarthritis (OA) in rabbits through modulation of the peroxisome proliferator-activated receptor (PPARγ). Methods. In vitro, OA chondrocytes were treated with ultrasound (US), US-targeted microbubble destruction (UTMD), simvastatin (SV), and UTMDSV on alternate days for four weeks. Chondrocytes were also treated with PPARγ inhibitor, PPARγ inhibitor+ UTMDSV, and UTMDSV. The cholesterol efflux rate and triglyceride levels were measured using an assay kit and oil red O staining, respectively. In vivo, the OA rabbits were treated with a single intra-articular injection of UTMD, SV, and UTMDSV every seven days for four weeks. Cartilage histopathology was assessed by safranin-O staining and the Mankin score. Total cholesterol (TC) and high-density lipoprotein-cholesterol (HDL-C) in rabbit knee synovial fluid were detected by enzyme-marker assay. Aggrecan, collagen II, and PPARγ expression levels were analyzed by Western blotting (WB). Results. In vitro, UTMDSV significantly increased the cholesterol efflux rate and aggrecan, collagen II, and PPARγ levels in OA chondrocytes; these effects were blocked by the PPARγ inhibitor. In vivo, UTMD. SV. significantly increased aggrecan, collagen II, PPARγ, and HDL-C levels, while TC levels and Mankin scores were decreased compared with the UTMD, SV, OA, and control groups. Conclusion. UTMDSV promotes cartilage extracellular matrix synthesis by modulating the PPARγ-mediated cholesterol efflux pathway in OA rabbits. Cite this article: Bone Joint Res 2021;10(10):693–703


The Bone & Joint Journal
Vol. 98-B, Issue 7 | Pages 925 - 933
1 Jul 2016
Sidaginamale RP Joyce TJ Bowsher JG Lord JK Avery PJ Natu S Nargol AVF Langton DJ

Aims. We wished to investigate the influence of metal debris exposure on the subsequent immune response and resulting soft-tissue injury following metal-on-metal (MoM) hip arthroplasty. Some reports have suggested that debris generated from the head-neck taper junction is more destructive than equivalent doses from metal bearing surfaces. . Patients and Methods. We investigated the influence of the source and volume of metal debris on chromium (Cr) and cobalt (Co) concentrations in corresponding blood and hip synovial fluid samples and the observed agglomerated particle sizes in excised tissues using multiple regression analysis of prospectively collected data. A total of 199 explanted MoM hips (177 patients; 132 hips female) were analysed to determine rates of volumetric wear at the bearing surfaces and taper junctions. . Results. The statistical modelling suggested that a greater source contribution of metal debris from the taper junction was associated with smaller aggregated particle sizes in the local tissues and a relative reduction of Cr ion concentrations in the corresponding synovial fluid and blood samples. Metal debris generated from taper junctions appears to be of a different morphology, composition and therefore, potentially, immunogenicity to that generated from bearing surfaces. Conclusion. The differences in debris arising from the taper and the articulating surfaces may provide some understanding of the increased incidence of soft-tissue reactions reported in patients implanted with MoM total hip arthroplasties compared with patients with hip resurfacings. Cite this article: Bone Joint J 2016;98-B:925–33


Prosthetic joint infections (PJIs) with prior multiple failed surgeries for re-infection (PMFSR) represented a huge challenge for surgeons because of poor vascular supply and biofilm formation. This study aims to determine the results of single-stage revision using intra-articular antibiotic infusion for treatment of PJIs with PMFSR. A retrospective analysis included 78 PJIs (29 hips; 49 knees) with PMFSR, who were treated with single-stage revision using intra-articular antibiotic infusion. Previous failure was defined as infection recurrence requiring surgical intervention. Besides intravenous pathogen-sensitive agents, a intra-articular infusion of vancomycin, imipenem or voriconazole was performed post-operatively. The antibiotic solution was soaked into the joint for 24 hours then extracted before next injection. Recurrence of infection and clinical outcomes were evaluated. Of these 78 patients, 59 had more than twice failed debridement and implant retentions, 12 patients had a failed arthroplasty resection, 3 hips underwent a failed two-stage revision and 4 cases had a failed one-stage revision before their single-stage revision. A total of 68 patients (87.1%) were free of infection at a mean follow-up time of 85 months. No significant difference in infection control rate was observed between hip and knee PJIs (89.7% versus 85.7%). The mean postoperative Harris hip score was 76.1 points and HSS score was 78. 2 at the most recent assessment. Single-stage revision with intra-articular antibiotic infusion can provide high antibiotic concentration in synovial fluid therefore overcome poor vascular supply and biofilm formation, which can be a viable option in treating PJI with PMFSR


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_14 | Pages 37 - 37
1 Nov 2021
Girard J
Full Access

The management of prosthetic joint infection (PJI) has been widely performed for total hip arthroplasties (THA), but none has compared it with hip resurfacing arthroplasty (RSA). We also carried out a retrospective case-control study comparing the surgical treatment of PJI by surgical debridement and implant retention between RSA and THA in order to clarify whether there was a difference in terms of (1) successful healing of PJI (2) functional scores after recovery (3) risk factors for recurrence of PJI. Our hypothesis was that simple debridement with prosthesis retention regardless of the timeframe allowed to obtain a higher success rate for RSA compared to THA. From 2010 to 2018, a single-center case-control study based on 3056 RSA found 13 PJI were age-matched (based on the 139 THA PJI treated) with 15 THA PJI (mean age of 53 years old (47–58) for THA and 59 (45–66) for RSA (p=0.34)). We compared their survival (absence of infectious recurrence) and the means differences between the 2 groups (demographical, clinical and biological data). There was no difference between the 2 groups concerning: age (p=0.3), BMI (p=0.4), initial diagnosis (p=0.4), operating time for primary surgery (p=0.3), the presence of a postoperative hematoma (p=0.4), the type of bacteria (p=0.5), the total duration of antibiotic therapy (p=0.9) and the type of antibiotic therapy (p=0.6). Early postoperative infections (less than 6 weeks) occurred in 7/13 RSA cases (54%) compared to 11/15 THA cases (73%). At the mean follow-up of 5 years (2–7), the success rate without recurrence was significantly higher in the RSA group 100% versus 66.7% (10/15) for the THA group (p=0.044). At the last follow-up, the Oxford Hip Score was higher in the RSA versus THA group's (14 versus 22 p=0.004). Simple surgical debridement an RSA without changing implants after PJI can be done regardless of the time to onset of infection. This is secondary to the absence of metaphyseal bone invasion and the low content of joint fluid


The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 173 - 176
1 Feb 2015
Omar M Ettinger M Reichling M Petri M Guenther D Gehrke T Krettek C Mommsen P

The aim of this study was to assess the role of synovial C-reactive protein (CRP) in the diagnosis of chronic periprosthetic hip infection. We prospectively collected synovial fluid from 89 patients undergoing revision hip arthroplasty and measured synovial CRP, serum CRP, erythrocyte sedimentation rate (ESR), synovial white blood cell (WBC) count and synovial percentages of polymorphonuclear neutrophils (PMN). Patients were classified as septic or aseptic by means of clinical, microbiological, serum and synovial fluid findings. The high viscosity of the synovial fluid precluded the analyses in nine patients permitting the results in 80 patients to be studied. There was a significant difference in synovial CRP levels between the septic (n = 21) and the aseptic (n = 59) cohort. According to the receiver operating characteristic curve, a synovial CRP threshold of 2.5 mg/l had a sensitivity of 95.5% and specificity of 93.3%. The area under the curve was 0.96. Compared with serum CRP and ESR, synovial CRP showed a high diagnostic value. According to these preliminary results, synovial CRP may be a useful parameter in diagnosing chronic periprosthetic hip infection. Cite this article: Bone Joint J 2015; 97-B:173–6


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_11 | Pages 7 - 7
1 Jun 2016
Sidaginamale R O'Hare J Natu S Joyce T Lord J Nargol A Langton D
Full Access

Background. Some reports have suggested that debris generated from the head neck taper junction is more destructive than equivalent doses from metal bearing surfaces. Methods. Part 1. We examined the relationship between the source (taper/bearing) and volume of metal debris on Cr and Co concentrations in corresponding blood and hip synovial fluid samples and the observed agglomerated particle sizes in excised tissues using regression analysis of prospectively collected data at a single revision unit. Part 2. We investigated variables most strongly associated with macroscopic soft tissue injury as documented at revision surgery using ordinal logistic regression. Independent variables included source and volume of CoCr exposure, Cr and Co joint fluid concentrations, joint fluid grade, ALVAL (Aseptic Lymphocytic Vasculitis Associated Lesion) grade, presence of vascular hyalinisation, agglomerated particle size, implant type, patient sex and age. Results. A total of 199 explanted MoM hips were analysed. Multiple regression statistical modelling suggested that a greater source contribution of metal debris from the taper junction was associated with smaller aggregated particle sizes in the local tissues and a relative reduction of Cr ion concentrations in the corresponding synovial fluid and blood samples. There was an association between increasing Co concentrations in the joint fluid and an increasing ALVAL score (p<0.001). In contrast, higher Cr concentrations were inversely related to ALVAL (p<0.001). The ALVAL response was itself strongly associated with larger fluid collections (p<0.001). Vascular hyalinisation and larger fluid collections were significantly associated with macroscopic tissue injury (coefficient 2.22, p<0.001 for fluid grade and 4.35, p<0.001 for hyalinisation). Discussion. Metal debris generated from taper junctions appears to be associated with a different biochemical environment than debris generated from bearing surfaces. We believe that this finding may provide some explanation as to the poor performance of MoM THRs compared to the equivalent resurfacing devices and the confusion surrounding the significance of blood metal testing. Conclusion. Chromium does seem to inhibit the development of ALVAL


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 372 - 377
1 Apr 2019
Zagra L Villa F Cappelletti L Gallazzi E Materazzi G De Vecchi E

Aims. Leucocyte esterase (LE) has been shown to be an accurate marker of prosthetic joint infection (PJI), and has been proposed as an alternative to frozen section (FS) histology for intraoperative diagnosis. In this study, the intraoperative assessment of LE was compared with FS histology for the diagnosis of prosthetic hip infection. Patients and Methods. A total of 119 patients undergoing revision total hip arthroplasty (THA) between June 2015 and December 2017 were included in the study. There were 56 men and 63 women with a mean age of 66.2 years (27 to 88). Synovial fluid was collected before arthrotomy for the assessment of LE using enzymatic colourimetric strips. Between five and six samples were stained with haematoxylin and eosin for FS histology, and considered suggestive of infection when at least five polymorphonuclear leucocytes were found in five high-power fields. Results. The sensitivity and specificity of the LE assay were 100% and 93.8%, respectively; the positive (PPV) and the negative (NPV) predictive values were 79.3% and 100%, respectively. The mean time between the collection of the sample and the result being known was 20.1 minutes (. sd. 4.4). The sensitivity and specificity of FS histology were 78.3% and 96.9%, respectively; the PPV and the NPV were 85.7% and 94.9%, respectively. The mean time between the collection of the sample and the result being known was 27.2 minutes (. sd. 6.9). Conclusion. The sensitivity of LE assay was higher, with similar specificity and diagnostic accuracy, compared with FS histology. The faster turnaround time, its ease of use, and low costs make LE assay a valuable alternative to FS histology. We now use it routinely for the intraoperative diagnosis of PJI. Cite this article: Bone Joint J 2019;101-B:372–377


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_1 | Pages 20 - 20
1 Jan 2018
Renz N Mudrovcic S Trampuz A Perka C
Full Access

Cutibacterium species (formerly Propionibacterium species) are increasingly recognized as causative pathogens of low-grade periprosthetic joint infections (PJI). The clinical manifestation of infections caused by this low virulent microorganism is nonspecific and the interpretation remains challenging. In this prospective cohort study from 01/2012 to 07/2017 we analyzed the clinical and diagnostic characteristics of microbiologically proven hip PJI caused by Cutibacterium species. PJI was defined by growth of Cutibacteria in ≥2 periprosthetic tissue samples or in sonication fluid of the removed implant (>50 CFU/ml) at revision surgery. If Cutibacteria grew only in synovial fluid at least one other positive microbiological specimen or non-microbiological criterion was required. We included 26 patients suffering from Cutibacterium hip PJI, among them 19 were males (73%). The majority of PJI (24, 92%) presented delayed (3–24 months) or late (>24 months) after implantation. Sinus tract was present in 4 patients (16%) and radiological implant loosening in 16 patients (62%). Among non-microbiological diagnostic tests, increased synovial fluid leukocyte count showed the highest sensitivity (82%), followed by tissue histology (71%) and serum C-reactive protein (58%). After 7 days of incubation Cutibacterium grew in synovial fluid, periprosthetic tissue and sonication fluid culture in 20%, 42% and 32%, respectively, and in 43%, 76% and 83%, respectively, after 14 days of incubation. We conclude that Cutibacterium PJI was diagnosed late in the disease course and presented with subtle clinical signs. Prolonged culture incubation and implant sonication improved the poor performance of conventional microbiological tests. Due to lack of reliable diagnostic tests, Cutibacterium remains difficult to detect making the diagnosis challenging


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_9 | Pages 33 - 33
1 May 2018
Raza A Diament M Kulbelka I Baker P Webb J Port A Jameson S
Full Access

Introduction. Periprosthetic joint infection (PJI) can be difficult to diagnose. A variety of techniques have been described. The efficacy of the alpha-defensin laboratory test was examined and compared with other established modalities in the diagnostic workup of ‘real world’ arthroplasty patients. Methods. This was a retrospective review of 210 episodes (86 hips, 124 Knees) in 172 patients at one centre, and included samples from acute admissions, elective aspirations, and planned revisions. MSIS (musculoskeletal infection society) major and minor criteria were used for diagnosing PJI. Each patient was investigated using a standardised protocol with inflammatory markers, synovial fluid analysis for white cell count (SWCC) and polymorphonuclear leukoctyes percentage (PMN %), and synovial fluid/tissue culture. Synovial fluid was also tested for alpha-defensin. Results. Fifty-two (24.8%) episodes were defined as ‘PJI’ using MSIS criteria, and 158 ‘non-PJI’. Alpha-defensin had 71.2% (95% CI 56.9–82.9) sensitivity and 94.3% (89.5–97.4) specificity. The positive predictive value was 80.4% (68.1–88.8) with negative predictive value of 90.9% (6.6–93.9). Thirty-seven (of 52 PJI episodes) were ‘culture positive’ (identical microorganism on ≥2 samples). Eighty (of 158 non-PJI) satisfied none of the major or minor MSIS criteria, and none of these patients subsequently had a PJI. In this sub-set of 117, alpha-defensin had 64.9% (47.5–79.8) sensitivity and 98.7% (92.9–99.9) specificity, whilst the sensitivities of CRP (>20), SWCC (>3000) and PMN (>80%) were 94.6% (81.8–99.3), 86.5% (71.2–95.5) and 83.8% (68.0–93.8) respectively. There were 93 episodes where at least 1 minor criterion was satisfied and/or a sinus was present, but were not ‘culture positive’. In these, alpha-defensin had 85.7% (57.2–98.2) sensitivity and 88.4% (78.4–94.9) specificity. Discussion and conclusion. The alpha-defensin laboratory test has a lower sensitivity than previously reported, limiting its use for diagnosing PJI. SWCC and PMN % have similar sensitivity and are cheaper. We no use the alpha-defensin test


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_12 | Pages 27 - 27
1 Jun 2017
Della Valle C Tetreault M Estrera K Kayupov E Brander C
Full Access

Patients with a painful or failed total joint arthroplasties should be evaluated for periprosthetic joint infection (PJI). The purpose of this study was to determine if patients referred to a tertiary care centre had been evaluated for PJI according to the American Academy of Orthopaedic Surgeons (AAOS) clinical practice guidelines. 113 Patients with painful hip (43) or knee (70) arthroplasties were referred to a single provider by orthopaedic surgeons outside our practice between 2012 and 2014. We retrospectively evaluated the workup by referring physicians, including measurement of serum erythrocyte sedimentation rate (ESR) and C-reactive protein (CRP), performance of a joint aspiration if these values were abnormal, and obtainment of synovial fluid white blood cell (WBC) count, differential, and cultures. Sixty-two of the 113 patients (55%) did not have a workup that followed AAOS guidelines. Serum ESR and CRP were ordered for 64 of the 113 patients (57%). Of 25 patients with elevated inflammatory markers warranting aspiration, 15 (60%) had an aspiration attempted, with synovial fluid WBC, differential, and cultures obtained in 9 of 12 (75%) aspirations that yielded fluid. Of the 62 patients with an incomplete infection workup, 11 (18%) had a bone scan, 6 (10%) a CT scan, and 3 (5%) an MRI. Twelve of the 113 patients (11%) were ultimately diagnosed with PJI, with 5 undiagnosed prior to referral. The AAOS guidelines to evaluate for PJI are frequently not being followed. Improving awareness of these guidelines may avoid unnecessary and costly evaluations and delay in the diagnosis of PJI


Bone & Joint Open
Vol. 4, Issue 8 | Pages 628 - 635
22 Aug 2023
Hedlundh U Karlsson J Sernert N Haag L Movin T Papadogiannakis N Kartus J

Aims

A revision for periprosthetic joint infection (PJI) in total hip arthroplasty (THA) has a major effect on the patient’s quality of life, including walking capacity. The objective of this case control study was to investigate the histological and ultrastructural changes to the gluteus medius tendon (GMED) in patients revised due to a PJI, and to compare it with revision THAs without infection performed using the same lateral approach.

Methods

A group of eight patients revised due to a PJI with a previous lateral approach was compared with a group of 21 revised THAs without infection, performed using the same approach. The primary variables of the study were the fibril diameter, as seen in transmission electron microscopy (TEM), and the total degeneration score (TDS), as seen under the light microscope. An analysis of bacteriology, classification of infection, and antibiotic treatment was also performed.