Introduction and Objective. Curative resection of proximal humerus tumours is now possible in this era of limb salvage with endoprosthetic replacement considered as the preferred reconstructive option. However, it has also been linked with mechanical and non-mechanical failures such as
Introduction and Objective. Some periprosthetic femoral fractures (PFFs) present history and radiographic aspect consistent with an atypical femoral fracture (AFF), fulfilling the criteria for AFF except that PFFs by themselves are excluded from the diagnosis of AFFs. The aim of this study was to evaluate in a single Institution series of PFFs if any of them could be considered a periprosthetic atypical femoral fracture (PAFF), and their prevalence. Materials and Methods. Surgical records were searched for PFFs around a primary hip stem from January 2013 to December 2019. Cases were classified according to Vancouver classification. Demographic and medical history were extracted. Fisher's exact test was used for statistical analysis. Results. One-hundred-fifteen PFFs were identified, 59 of them were type B1 and 16 were type C. Radiographs and medical records were available for all patients. Twenty-four patients (32%) have been treated with bisphosphonates (BPs) for longer than 4 years. Four patients presented a fracture with characteristics of PAFF. When enlarged to all PFFs of the series, no other PAFF was found: prevalence of PAFFs was 5.3% for type B1 and C cases and 3.5% for all surgically treated PFFs. Statistical significative difference between PAFFs and PFFs was found for prolonged BPs assumption and for the level of fracture clear of the
The Exeter Stem (Howmedica, UK) has been in use for over 35 years. Over the years it has undergone several modifications with the most recent being a highly polished, tapered stem in 1986. The manufacturers quote a rate of 0.0006%. In the current literature there are 16 (or less) instances of fractures of the Orthinox stem. We present a case of fracture of an Orthinox Exeter Stem 9 years after insertion. Our patient, BB, presented, aged 62, with symptoms & signs consistent with OA right hip. THR was performed through a lateral approach utilising a trochanteric osteotomy. A size 0 37.5 stem was inserted. Radiographs were very satisfactory. She suffered a post operative DVT/PTE from which she recovered uneventfully. She was independently mobile at 6 month review and was discharged at the 2 year stage pain free. Aged 71, BB presented to outpatient clinic with a several month history of generalised groin pain. She had a Trendelenberg gait. Considerable pain was experienced on axial compression of the limb. Radiographs revealed a midstem fracture with cement loosening proximally. No trauma was reported. She underwent revision surgery through a posterior approach. Acetabular component was rigidly fixed. This was revised to a pressfit Trident (Zimmer, UK) cup with screws & polyethylene liner. An extended trochanteric osteotomy was used to remove the broken stem. An uncemented Restoration (Stryker, UK) stem was inserted with a 28mm head. Post-operative recovery was unremarkable and at 6 months osteotomy has healed. The stem was sent to Stryker UK Laboratories for analysis. They reported the stem broke in fatigue with the origin on the antero-lateral surface. No material or manufacturing defects seen. Dimensionally correct. Fracture may be due to abnormal bending stresses secondary to proximal loosening and firm distal fixation. Our case demonstrates a set of circumstances that led to inevitable fatigue and