Tapered fluted titanium stems are increasingly
used for femoral revision arthroplasty. They are available in modular and
non-modular forms. Modularity has advantages when the bone loss
is severe, the proximal femur is mis shapen or the surgeon is unfamiliar
with the implant, but it introduces the risk of fracture of the
stem at the junction between it and the proximal body segment. For
that reason, and while awaiting intermediate-term results of more recently
introduced designs of this junction, non-modularity has attracted
attention, at least for straightforward revision cases. We review the risks and causes of fracture of tapered titanium
modular revision stems and present an argument in favour of the
more selective use of modular designs. Cite this article:
Implant fracture of modular revision stems is a major complication after total hip arthroplasty revision (rTHA). Studies looking at specific modular designs report fracture rates of 0.3% to 0.66% whereas fractures of monobloc designs are only reported anecdotally. It is unclear whether the overall re-revision rate of modular designs is higher and if, whether stem fractures or other revision reasons are responsible for this elevation. All revisions within 5 years after implantation of a revision stems (n. 0. =13,900; n. 5. =2506) were analysed using Cox regression with design (modular: n=17, monobloc: n=27), BMI, Sex and Elixhauser Score as independent variables. One stage and two stage revisions were analysed separately (1-stage: modular n= 7,102; monobloc n= 4,542; 2-stage: 1,551 / 704). The revision volume of the hospitals was also considered (low: <20 revisions, medium: 21–50 revisions, high: >50 revisions). For the 1-stage revisions, the re-revision risk after 4 years was 14,3% [13.2%, 15.5%] for monobloc and 17.4% [16.40%, 18.40%] for modular stems (p< 0.001). Stem fracture was the reason for re-revision in 2.4% of the modular (fracture rate 0.42%) and 0.6% of the monobloc revisions. The difference in re-revision rates between the designs was mainly due to differences in dislocation and stem loosening. For the 2-stage revisions, the revision risks for either design were similar (21.7% [18,5%, 25.4%] vs. 23.0% [20.8%, 25.4%]; p=0.05). Patient characteristics influenced the comparison between the two designs in the 1-stage group but very little in the 2-stage group. Modular
The Exeter V40 cemented polished tapered stem system has demonstrated excellent long-term outcomes. This paper presents a systematic review of the existing literature and reports on a large case series comparing implant fractures between the Exeter V40 series; 125 mm and conventional length stem systems. A systematic literature search was performed adhering to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) criteria. In parallel, we performed a retrospective single centre study of Exeter V40 femoral stem prosthetic fractures between April 2003 and June 2020.Aims
Methods
We implanted 51 Metal-Cancellous Cementless Lübeck (MCCL) prostheses into 45 patients with dysplastic hips and followed 49 hips (96.1%) for five to nine years. One had needed