Hip

RE-REVISION RISK OF MODULAR AND MONOBLOC REVISION STEMS AFTER REVISION TOTAL HIP ARTHROPLASTY

The International Hip Society (IHS) 2024 Closed Meeting, Athens, Greece, 15–18 May 2024.



Abstract

Implant fracture of modular revision stems is a major complication after total hip arthroplasty revision (rTHA). Studies looking at specific modular designs report fracture rates of 0.3% to 0.66% whereas fractures of monobloc designs are only reported anecdotally. It is unclear whether the overall re-revision rate of modular designs is higher and if, whether stem fractures or other revision reasons are responsible for this elevation.

All revisions within 5 years after implantation of a revision stems (n0=13,900; n5=2506) were analysed using Cox regression with design (modular: n=17, monobloc: n=27), BMI, Sex and Elixhauser Score as independent variables. One stage and two stage revisions were analysed separately (1-stage: modular n= 7,102; monobloc n= 4,542; 2-stage: 1,551 / 704). The revision volume of the hospitals was also considered (low: <20 revisions, medium: 21–50 revisions, high: >50 revisions).

For the 1-stage revisions, the re-revision risk after 4 years was 14,3% [13.2%, 15.5%] for monobloc and 17.4% [16.40%, 18.40%] for modular stems (p< 0.001). Stem fracture was the reason for re-revision in 2.4% of the modular (fracture rate 0.42%) and 0.6% of the monobloc revisions. The difference in re-revision rates between the designs was mainly due to differences in dislocation and stem loosening. For the 2-stage revisions, the revision risks for either design were similar (21.7% [18,5%, 25.4%] vs. 23.0% [20.8%, 25.4%]; p=0.05). Patient characteristics influenced the comparison between the two designs in the 1-stage group but very little in the 2-stage group.

Modular revision stem fractures only contribute very minor to re-revision risk. In 2-stage revisions, no difference in overall re-revision rates between designs was observed. This might indicate that the differences observed for 1-stage procedures are due to differences between the patient cohorts, not reflected by the parameters available or surgeon choice.


Email: