Introduction. Our unit has extensive experience with the use of Ilizarov circular frames for acute fracture and nonunion surgery. We have observed and analysed fracture healing patterns which question the role of relative stability in fracture healing and we offer limb
Despite multiple published reviews, the optimum method of correction and stabilisation of Blount's disease remains controversial. The purpose of this study is to evaluate the clinical and radiological outcomes of acute correction of late-onset tibial vara by percutaneous proximal tibial osteotomy with circular external fixation using two simple rings. This technique was developed to minimise cost in a context of limited resources. This study was conducted between 2016 and 2020. We retrospectively reviewed the clinical notes and radiographs of 30 patients (32tibiae) who had correction of late-onset tibia by proximal tibial osteotomy and Ilizarov external fixator. All cases were followed up to 2 years. The mean proximal tibial angle was 65.7° (±7.8) preoperatively and 89.8° (±1.7) postoperatively. The mean
Background. Ilizarov hip reconstruction is a salvage option traditionally done with fixators, which often face limited compliance. We aimed at reviewing our early results of a novel technique with internal devices. Material and Methods. Retrospective evaluation of 7 patients was done that underwent Milch Batchelor procedure with dual plates followed by a second stage femoral lengthening by retrograde magnetic nail. The mean patient age was 13.8 years. A mean valgus of 410 (range, 300 to 550) and extension of 220 (range, 100 to 300) were created proximally. Results. At their mean recent follow-up of 19 months, all extremities were aligned with a mean Pelvic
Background. Computer-assisted navigation systems for total knee arthroplasty (TKA) were introduced to improve implantation accuracy and to optimize ligament balancing. Several comparative studies in the literature confirmed an effect on the component position and other studies could not confirm these results. For ligament balancing most studies found no significant influence on the clinical outcome using a navigation system for TKA. In the literature there were no reports of mid-term results after navigated TKA. With our study we wanted to show if the use of a navigation system for TKA will have an influence on the component's position and on the clinical results at 5-year follow up. Methods. We enrolled 200 patients in a prospective randomized study with a minimum follow up of 5 years. 100 TKA were operated on without using a navigation system (Group A) and 100 surgeries (Group B) were done with computer assistance. Radiological investigation by standard radiographs including a long-leg X-ray was performed with a follow up rate in Group A of 86.2% versus 80.2% in Group B. We measured the
Kinematic alignment has increased in popularity over the last few years in an attempt to improve clinical outcomes following total knee arthroplasty (TKA). In our unit kinematic alignment has been used with patient-specific cutting guides as part of on-going clinical trials. We performed a retrospective analysis on all the TKA which had been planned to be implanted outside of the
The objective of this study was to compare differences in alignment following robotic arm-assisted bi-unicompartmental knee arthroplasty (Bi-UKA) and conventional total knee arthroplasty (TKA). This was a prospective, randomised controlled trial of 70 patients. 39 TKAs were implanted manually, as per standard protocol at our institution, and 31 Bi-UKA patients simultaneously received fixed-bearing medial and lateral UKAs, implanted using robotic arm-assistance. Preoperative and 3-month postoperative CT scans were analysed to determine hip knee ankle angle (HKAA), medial distal femoral angle (MDFA), and medial proximal tibial angle (MPTA). Analysis was repeated for 10 patients by a second rater to validate measurement reliability by calculating the intra-class correlation coefficient (ICC). Mean change in HKAA towards neutral was 2.7° in TKA patients and 2.3° in Bi-UKA patients (P=0.6). Mean change in MDFA was 2.5° for TKA and 1.0° for Bi-UKA (P<0.01). Mean change in MPTA was 3.7° for TKA and 0.8° for Bi-UKA (P<0.01). Mean postoperative MDFA and MPTA for TKAs were 89.8° and 89.6° respectively, indicating orientation of femoral and tibial components perpendicular to the
Aim. The purpose of this study was to analyze the patellofemoral kinematics in neglected adult developmental dysplasia of the hip patients with patellofemoral symptoms and tried to clarify the affect of the severity of dislocation, the amount of limb length discrepancy, the deviation of
Electromagnetic navigation versus conventional Total Knee Arthroplasty: Clinical improvements Optical and electromagnetic (EM) tracking systems are widely used commercially. However in orthopaedic applications optical systems dominate the market. Optical systems suffer from deficiencies due to line of sight. EM trackers are smaller but are affected by metal. The accuracy of the two tracker systems has been seen to be comparable1. Recent advancements in optical navigated TKA have shown improved overall limb alignment, implant placement and reduce outliers when compared to conventional TKA2-4. This study is the first RCT to compare EM and conventional TKA. Two groups of 100 patients underwent TKA using either the EM navigation system or the conventional method. Frontal, sagittal and rotational alignment was analysed from a CT scan. Clinical scores including Oxford Knee Score (OKS) and Knee/Function American Knee Society Score (AKSS) were recorded pre-op, and at 3 and 12 months post-op. 3 month data presented includes 180 patients (n = 90). The 12 months data presented includes 140 (n = 70). The two groups had similar mean
Introduction. Femorotibial malalignment exceeding ±3° is a recognised contributor of early mechanical failure after total knee replacement (TKR). The angle between the mechanical and anatomical axes of the femur remains the best guide to restore alignment. We investigated where the femoral head lies relative to the pelvis and how its position varies with respect to recognised demographic and anatomic parameters. We have tested the hypothesis of the senior author that the position of the centre of the femoral head varies very little, and if its location can be identified, it could serve to outline the
Background. Humerus fracture non-union is a challenge for which a wide range of treatments exist. We present our experience of managing these by hybrid Ilizarov frame fixation, without bone graft or debridement of the non-union site. Methods. Case notes review of a consecutive series of 20 patients treated for aseptic humeral non-union between 2004 and 2016. Eighteen patients had previous plate or intramedullary nail fixations, and 2 had no prior surgery. During Ilizarov application, any existing metalwork preventing dynamisation of the fracture site was removed through minimal incisions before compression of the fracture site was then achieved. Only 3 patients had open debridement or osteotomy of the non-union site, otherwise all other patients had no debridement of their non-union. Results. Bony union was achieved in 17 patients (85%), with a further 2 achieving a functional fibrous union. The remaining patient subsequently had successful open surgery. Union rates were 66% (2/3) and 88% (15/17) for the debridement/osteotomy and non-debridement groups respectively. Mean frame duration was 193 days. One patient was treated for pin-site infection. The Chertsey Outcome Score for Trauma was used to assess patient reported outcomes. Conclusions. Simple changes to the physical properties acting upon a humeral non-union, such as adequate control of its strain environment and restoration of the
The 8-plate (Orthofix, SRL, Italy) is a titanium extraperiosteal plate with 2 screws which acts as a hinge at the outer limits of the physis. It has been used for correction of both angular and sagittal deformity around the knee. To our knowledge this is the first study describing the use of 8-plates in leg length discrepancy (LLD) correction. We aimed to evaluate outcomes of temporary 8-plate epiphysiodesis in LLD, and to assess the complications associated with its usage. This retrospective study included 30 patients between 2007 and 2010 whom underwent 8-plate epiphysiodesis to address LLD. Leg length measurements were recorded using erect full leg length scanograms and comparison made between pre-operative, interval and final scanograms. Any deviations of the
Accurate measurement of the alignment of the tibia is important both clinically and in research. The conventional method of measuring the angle of malunion after a fracture of the shaft of the tibia is potentially inaccurate because the
Background. Adequate rotation of femoral component in total knee arthroplasty(TKR) is mandatory for preventing numerous adverse sequelae. The transepicondylar axis has been a well-accepted reference for femoral component rotation in the measured resection technique. In this technique, measured resection is performed referenced off the tibial cut - perpendicular to the tibial
Introduction. Patellar tracking in total knee replacements has been extensively studied, but little is known about patellar tracking in isolated patellofemoral replacements. We compared patellar tracking and the position of the patellar groove in the natural knee, followed by implantation of the femoral component of a PFR (patella unresurfaced) and after implantation of the femoral & patellar component of the PFR. Methods. Computer navigation was used to track the patella in eight whole lower extremities of four cadavers in the natural knee, in the same knee with the femoral component of the PFR (PFR-P) and with the femoral and patellar component of the PFR (PFR+P, patella resurfaced) (Depuy Sigma PFR). The form and position of the trochlea in the natural knee and the patellar groove of the femoral component was also analysed. Values are means+/−SD, two tailed Student's t-test for paired samples. Results. With a PFR-P the patella had a slightly more lateral tilt (0.8+/−0.8° to 2.8+/−2.5° at 40–100° of flexion, p<0.05 vs. Nat), this was more pronounced with the PFR+P (2.0+/−0.7° to 4.9+/−1.8° at 20–90° flexion, p<0.05 vs. Nat., p<0.05 vs. PFR-P at 20–80° flexion). No differences in patella rotation were seen between the three groups. In the PFR-P group the patella tracked a little more medially compared to the natural knee (0.6+/− 0.7mm to 1.3+/−2.6mm, p<0.05 at 20°,80°,90° flexion). The difference was more pronounced after patella resurfacing (PFR+P) (2.1+/−2.0mm to 3.0+/−2.2mm, p<0.05 vs. Nat. at 10°–100°, p<0.05 vs. PFR-P from 10–100°). When analysed relative to the patellar groove of the trochlea/femoral component the patella in the natural knee tracked slightly lateral to the groove (2.0+/−1.7mm to 2.9+/−2.0mm at 50–100° p<0.05), so did the patella of PFR-P (2.0+/−2.3mm to 2.3+/−2.3mm at 60–90° flexion, p<0.05), whilst the PFR+P tracked right on the groove (0.6+/−3.7mm medially to 0.6+/−2.9mm laterally, p<0.05 vs Nat at 10–30° & 70–100°, p<0.05 vs. PFR-P at 10–100°). Distance from the patellatot the epicondylar axis was slightly larger in the PFR-P group (0.6+/− 0.7mm to 1.3+/−1.4mm, p<0.05 vs. Natu at 20,80 & 90°. This was more pronounced with patellar resurfacing (2.1+/−2.0 to 3.0+/−2.2mm, p<0.05 vs. Nat at 10–100°, p<0.05 vs. PFR-P at 20–100°) The patella groove on the natural knee and the implanted femoral component of the implanted PFR had the same radius, inclination relative to the femoral
External fixators are the traditional fixation method of choice for contaminated open fractures. However, patient acceptance is low due to the high profile and therefore physical burden of the constructs. An externalised locking compression plate is a low profile alternative. However, the biomechanical differences have not been assessed. The objective of this study was to evaluate the axial and torsional stiffness of the externalised titanium locking compression plate (ET-LCP), the externalised stainless steel locking compression plate (ESS-LCP) and the unilateral external fixator (UEF). A fracture gap model was created to simulate comminuted mid-shaft tibia fractures using synthetic composite bones. Fifteen constructs were stabilised with ET-LCP, ESS-LCP or UEF (five constructs each). The constructs were loaded under both axial and torsional directions to determine construct stiffness.Objectives
Methods
This study aimed to determine the long-term functional,
clinical and radiological outcomes in patients with Schatzker IV
to VI fractures of the tibial plateau treated with an Ilizarov frame.
Clinical, functional and radiological assessment was carried out
at a minimum of one year post-operatively. A cohort of 105 patients
(62 men, 43 women) with a mean age of 49 years (15 to 87) and a
mean follow-up of 7.8 years (1 to 19) were reviewed. There were
18 type IV, 10 type V and 77 type VI fractures. All fractures united
with a mean time to union of 20.1 weeks (10.6 to 42.3). No patient
developed a deep infection. The median range of movement (ROM) of
the knee was 110o and the median Iowa score was 85. Our study demonstrates good long-term functional outcome with
no deep infection; spanning the knee had no detrimental effect on
the ROM or functional outcome. High-energy fractures of the tibial plateau may be treated effectively
with a fine wire Ilizarov fixator. Cite this article:
Radiological evidence of post-traumatic osteoarthritis
(PTOA) after fracture of the tibial plateau is common but end-stage arthritis
which requires total knee arthroplasty is much rarer. The aim of this study was to examine the indications for, and
outcomes of, total knee arthroplasty after fracture of the tibial
plateau and to compare this with an age and gender-matched cohort
of TKAs carried out for primary osteoarthritis. Between 1997 and 2011, 31 consecutive patients (23 women, eight
men) with a mean age of 65 years (40 to 89) underwent TKA at a mean
of 24 months (2 to 124) after a fracture of the tibial plateau.
Of these, 24 had undergone ORIF and seven had been treated non-operatively.
Patients were assessed pre-operatively and at 6, 12 and >
60 months
using the Short Form-12, Oxford Knee Score and a patient satisfaction
score. Patients with instability or nonunion needed total knee arthroplasty
earlier (14 and 13.3 months post-injury) than those with intra-articular
malunion (50 months, p <
0.001). Primary cruciate-retaining implants
were used in 27 (87%) patients. Complication rates were higher in
the PTOA cohort and included wound complications (13% Total knee arthroplasty undertaken after fracture of the tibial
plateau has a higher rate of complications than that undertaken
for primary osteoarthritis, but patient-reported outcomes and satisfaction
are comparable. Cite this article:
We investigated a new intramedullary locking
nail that allows the distal interlocking screws to be locked to
the nail. We compared fixation using this new implant with fixation
using either a conventional nail or a locking plate in a laboratory
simulation of an osteoporotic fracture of the distal femur. A total
of 15 human cadaver femora were used to simulate an AO 33-A3 fracture
pattern. Paired specimens compared fixation using either a locking
or non-locking retrograde nail, and using either a locking retrograde
nail or a locking plate. The constructs underwent cyclical loading
to simulate single-leg stance up to 125 000 cycles. Axial and torsional
stiffness and displacement, cycles to failure and modes of failure
were recorded for each specimen. When compared with locking plate
constructs, locking nail constructs had significantly longer mean
fatigue life (75 800 cycles ( The new locking retrograde femoral nail showed better stiffness
and fatigue life than locking plates, and superior fatigue life
to non-locking nails, which may be advantageous in elderly patients. Cite this article:
Coronal plane fractures of the posterior femoral
condyle, also known as Hoffa fractures, are rare. Lateral fractures are
three times more common than medial fractures, although the reason
for this is not clear. The exact mechanism of injury is likely to
be a vertical shear force on the posterior femoral condyle with
varying degrees of knee flexion. These fractures are commonly associated
with high-energy trauma and are a diagnostic and surgical challenge. Hoffa
fractures are often associated with inter- or supracondylar distal
femoral fractures and CT scans are useful in delineating the coronal
shear component, which can easily be missed. There are few recommendations
in the literature regarding the surgical approach and methods of
fixation that may be used for this injury. Non-operative treatment
has been associated with poor outcomes. The goals of treatment are
anatomical reduction of the articular surface with rigid, stable
fixation to allow early mobilisation in order to restore function.
A surgical approach that allows access to the posterior aspect of
the femoral condyle is described and the use of postero-anterior
lag screws with or without an additional buttress plate for fixation
of these difficult fractures. Cite this article:
Filling the empty holes in peri-articular locking
plates may improve the fatigue strength of the fixation. The purpose of
this A locking/compression plate was applied to 33 synthetic femurs
and then a 6 cm metaphyseal defect was created (AO Type 33-A3).
The specimens were then divided into three groups: unplugged, plugged
with locking screw only and fully plugged holes. They were then
tested using a stepwise or run-out fatigue protocol, each applying
cyclic physiological multiaxial loads. All specimens in the stepwise group failed at the 770 N load
level. The mean number of cycles to failure for the stepwise specimen
was 25 500 cycles ( In conclusion, filling the empty combination locking/compression
holes in peri-articular distal femur locking plates at the level
of supracondylar comminution does not increase the fatigue life
of the fixation in a comminuted supracondylar femoral fracture model
(AO 33-A3) with a 6 cm gap.