Several short- and mid-term studies have shown minimal liner wear of highly cross-linked polyethylene (HXLPE) in total hip arthroplasty (THA), but the safety of using thinner HXLPE liners to maximize femoral head size remains uncertain. The objective of this study was to analyze clinical survival and radiological wear rates of patients with HXLPE liners, a 36 mm femoral head, and a small acetabular component with a minimum of ten years’ follow-up. We retrospectively identified 55 patients who underwent primary THA performed at a single centre, using HXLPE liners with 36 mm cobalt-chrome heads in acetabular components with an outer diameter of 52 mm or smaller. Patient demographic details, implant details, death, and all-cause revisions were recorded. Cox regression and Kaplan-Meier survival was used to determine all-cause and liner-specific revision. Of these 55 patients, 22 had a minimum radiological follow-up of seven years and were assessed radiologically for linear and volumetric wear.Aims
Methods
The increased The proximal penetration of the femoral head five years post-operatively
was measured for 15 patients using radiostereometric analysis (RSA).Aims
Patients and Methods
Dislocation remains among the most common complications
of, and reasons for, revision of both primary and revision total
hip replacements (THR). Hence, there is great interest in maximising
stability to prevent this complication. Head size has been recognised
to have a strong influence on the risk of dislocation post-operatively.
As femoral head size increases, stability is augmented, secondary
to an increase in impingement-free range of movement. Larger head
sizes also greatly increase the ‘jump distance’ required for the
head to dislocate in an appropriately positioned cup. Level-one
studies support the use of larger diameter heads as they decrease
the risk of dislocation following primary and revision THR. Highly cross-linked
polyethylene has allowed us to increase femoral head size, without
a marked increase in wear. However, the thin polyethylene liners
necessary to accommodate larger heads may increase the risk of liner
fracture and larger heads have also been implicated in causing soft-tissue
impingement resulting in groin pain. Larger diameter heads also
impart larger forces on the femoral trunnion, which may contribute
to corrosion, metal release, and adverse local tissue reactions.
Alternative large bearings including large ceramic heads and dual
mobility bearings may mitigate some of these risks, and several
of these devices have been used with clinical success. Cite this article:
Introduction. Recent studies on
Previous reports on the outcomes of isolated head and liner exchange in revision total hip arthroplasty have found high rates of instability following these surgeries. Most reports have studied constructs using ≤28mm femoral heads. The purpose of this study was to determine if modern techniques with the use of larger head sizes can improve the rate of instability after head and liner exchange. We identified 138 hips in 132 patients who underwent isolated head and liner exchange for polyethylene wear/osteolysis (57%), acute infection (27%), metallosis (13%), or other (2%). All patients underwent revision with either 32mm (23%), 36mm (62%), or 40mm (15%) diameter heads. Crosslinked polyethylene was used in all revisions. Lipped and/or offset liners were used in 104 (75%) hips. Average follow up was 3.5 (1.0–9.1) years. Statistical analyses were performed with significance set at p<0.05. Revision-free survivorship for any cause was 94.6% and for aseptic causes was 98.2% at 5 years. 11 (8%) hips experienced a complication with 7 (5%) hips requiring additional revision surgery. Following revision, 4 (3%) hips experienced dislocation, 5 (4%) hips experienced infection, and 1 (1%) hip was revised for trunnionosis. No demographic or surgical factors significantly affected outcomes. Our study shows that isolated head and liner exchange using
The benefit of dual mobility cup (DMC) for primary total hip arthroplasties (THA) is still controversial. This study aimed to compare 1) the complications rate, 2) the revisions rate, 3) the survival rate after monobloc DMC compared to
Aims. The purpose of this study was to compare two different types
of metal-on-metal (MoM) bearing for total hip arthroplasty (THA):
one with a
Aims. The aim of this study was to evaluate blood metal ion levels, leucocyte profiles, and serum cytokines in patients with a total hip arthroplasty (THA) involving modular dual-mobility components. Patients and Methods. A total of 39 patients were recruited, with clinical follow-up of up to two years. Outcome was assessed using the Harris Hip Score (HHS, the 12-Item Short-Form Health Survey (SF-12), the Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), and a visual analogue scale (VAS) for pain. Blood concentrations of cobalt (Co), chromium (Cr), and serum cytokines were measured. Subpopulations of leucocytes were analyzed by flow cytometry. Results. The clinical performance was good. Blood Co levels (ref 1.0 µg/l) were mildly elevated in seven patients at three months, and two patients at two years’ follow-up. The preoperative Cr levels were normal except for one patient with a detectable Cr (1.2 µg/l). Cr levels were detectable in three patients at three months, two patients at one year, and three patients at two years’ follow-up. No patients had symptoms suggestive of failure. Although flow cytometry showed constant circulating leucocyte profiles, there was a significant reduction of serum interleukin (IL)-4, IL-5, and interferon gamma (IFNγ) postoperatively compared with the preoperative levels (p < 0.05). Conclusion. These results suggest that THA using modular dual-mobility components is safe. This allows an opportunity to use a
Introduction. Porous coated cups have a low modulus of elasticity that enhances press-fit and a surface that promotes osseointegration as well as permitting modularity. Monoblock ceramic acetabular cups represent a subtype of uncemented cup permitting the use of
Aims. We have previously demonstrated raised cobalt and chromium levels in patients with
Introduction. Prior to the introduction of alternative bearing surfaces, patients were typically counseled to expect that their total hip arthroplasty (THA) using conventional polyethylene would last for 10 years. With the introduction of crosslinked polyethylene and hard-on-hard bearing surfaces, revisions related to bearing surface wear were expected to decrease. We examined six different bearing surfaces used at our institution over three decades to evaluate how the overall survivorship, reasons for revision and Harris Hip Scores have changed with time. Methods. We identified six cohorts of patients with 754 primary hips done between 1983 and 2007. With the exception of 81 Birmingham hip resurfacings (BHR), all femoral components were straight, extensively porous-coated cylindrical (EPC) stems (AML and Prodigy). All cups were porous coated. In addition to the BHRs, the bearing surfaces included 223 conventional polyethylene (CPE) in a non-modular shell, 114 CPE in a modular shell, 116 crosslinked polyethylene (XLPE), 130 metal-on-metal (MOM), and 90 ceramic-on-ceramic (COC). The mean follow-up for all hip replacements is 13.0±6.0 years. Kaplan-Meier survivorship using revision for any reason as an endpoint with log rank testing was used to evaluate differences among groups. Results. Although there were no differences in survivorship at 10-year follow-up among the groups (p=0.53), the XLPE liners demonstrated improved survivorship at 15-years compared to both CPE groups owing to the absence of wear-related revisions (97% versus 83% for non-modular and 85% for modular cups respectively, p=0.001 and p=0.008). Revisions for femoral loosening have only occurred among 0.6% (4/673) of EPC stems. Revisions for cup loosening have occurred among 4% (10/223) of the non-modular cups but there are none among the other groups. The incidence of dislocation was reduced with the MOM, BHR and COC bearings that used 36-mm or
This study aims to assess the relationship between history of pseudotumour formation secondary to metal-on-metal (MoM) implants and periprosthetic joint infection (PJI) rate, as well as establish ESR and CRP thresholds that are suggestive of infection in these patients. We hypothesized that patients with a pseudotumour were at increased risk of infection. A total of 1,171 total hip arthroplasty (THA) patients with MoM articulations from August 2000 to March 2014 were retrospectively identified. Of those, 328 patients underwent metal artefact reduction sequence MRI and had minimum two years’ clinical follow-up, and met our inclusion criteria. Data collected included demographic details, surgical indication, laterality, implants used, history of pseudotumour, and their corresponding preoperative ESR (mm/hr) and CRP (mg/dl) levels. Multivariate logistic regression modelling was used to evaluate PJI and history of pseudotumour, and receiver operating characteristic curves were created to assess the diagnostic capabilities of ESR and CRP to determine the presence of infection in patients undergoing revision surgery.Aims
Methods
Introduction. The use of
Iliopsoas impingement occurs in 4% to 30% of patients after undergoing total hip arthroplasty (THA). Despite a relatively high incidence, there are few attempts at modelling impingement between the iliopsoas and acetabular component, and no attempts at modelling this in a representative cohort of subjects. The purpose of this study was to develop a novel computational model for quantifying the impingement between the iliopsoas and acetabular component and validate its utility in a case-controlled investigation. This was a retrospective cohort study of patients who underwent THA surgery that included 23 symptomatic patients diagnosed with iliopsoas tendonitis, and 23 patients not diagnosed with iliopsoas tendonitis. All patients received postoperative CT imaging, postoperative standing radiography, and had minimum six months’ follow-up. 3D models of each patient’s prosthetic and bony anatomy were generated, landmarked, and simulated in a novel iliopsoas impingement detection model in supine and standing pelvic positions. Logistic regression models were implemented to determine if the probability of pain could be significantly predicted. Receiver operating characteristic curves were generated to determine the model’s sensitivity, specificity, and area under the curve (AUC).Aims
Methods
This study aimed to examine the changing trends in the causes for early revision of total hip replacements (THR) in one country over fifteen years to assess whether changes in arthroplasty practices have changed revision patterns. The cause for revision within one year was extracted from the New Zealand Joint Registry for the period of January 1999 to December 2013. The results were grouped into three periods of 5 years (Period 1=1999–2003, Period 2=2004–2008 and Period 3 = 2009–2013). The causes were compared across the three periods and were adjusted for age, gender and ASA grade. The rate of early revisions (within one year) rose from 0.86% to 1.30% per year. There was significant rise in revision for deep infection (14% increased to 21% of all causes, p<0.001) and femoral fracture (5% increased to20%, p<0.001), whereas revision for dislocation decreased (55% decreased to 35%, p<0.001). Adjusting for age and gender femoral fracture and deep infection rates remained significant for both (p<0.05). Adjusting for age, gender as well as ASA only remained significant for infection (P<0.05). The reduction in revision for dislocation followed the registry trend for the use of
Radiostereometric analysis (RSA) is the most accurate radiological method to measure in vivo wear of highly cross-linked polyethylene (XLPE) acetabular components. We have previously reported very low wear rates for a sequentially irradiated and annealed X3 XLPE liner (Stryker Orthopaedics, USA) when used in conjunction with a 32 mm femoral heads at ten-year follow-up. Only two studies have reported the long-term wear rate of X3 liners used in conjunction with larger heads using plain radiographs which have poor sensitivity. The aim of this study was to measure the ten-year wear of thin X3 XLPE liners against larger 36 or 40 mm articulations with RSA. We prospectively reviewed 19 patients who underwent primary cementless THA with the XLPE acetabular liner (X3) and a 36 or 40 mm femoral head with a resultant liner thickness of at least 5.8 mm. RSA radiographs at one week, six months, and one, two, five, and ten years postoperatively and femoral head penetration within the acetabular component were measured with UmRSA software. Of the initial 19 patients, 12 were available at the ten-year time point.Aims
Methods
Dislocation remains a leading cause of failure following revision total hip arthroplasty (THA). While dual-mobility (DM) bearings have been shown to mitigate this risk, options are limited when retaining or implanting an uncemented shell without modular DM options. In these circumstances, a monoblock DM cup, designed for cementing, can be cemented into an uncemented acetabular shell. The goal of this study was to describe the implant survival, complications, and radiological outcomes of this construct. We identified 64 patients (65 hips) who had a single-design cemented DM cup cemented into an uncemented acetabular shell during revision THA between 2018 and 2020 at our institution. Cups were cemented into either uncemented cups designed for liner cementing (n = 48; 74%) or retained (n = 17; 26%) acetabular components. Median outer head diameter was 42 mm. Mean age was 69 years (SD 11), mean BMI was 32 kg/m2 (SD 8), and 52% (n = 34) were female. Survival was assessed using Kaplan-Meier methods. Mean follow-up was two years (SD 0.97).Aims
Methods
The purpose of this study was to compare the
amount of acetabular bone removed during hip resurfacing (HR) and cementless
total hip replacement (THR), after controlling for the diameter
of the patient’s native femoral head. Based on a power analysis,
64 consecutive patients (68 hips) undergoing HR or THR were prospectively
enrolled in the study. The following data were recorded intra-operatively:
the diameter of the native femoral head, the largest reamer used,
the final size of the acetabular component, the size of the prosthetic
femoral head and whether a decision was made to increase the size
of the acetabular component in order to accommodate a larger prosthetic femoral
head. Results were compared using two-sided, independent samples
Student’s t-tests. A statistically significant
difference was seen in the mean ratio of the size of the acetabular
component to the diameter of the native femoral head (HR: 1.05 (. sd. 0.04) versus THR:
1.09 (. sd. 0.05); p <
0.001) and largest acetabular reamer
used to the diameter of the native femoral head (HR: 1.03 (. sd. 0.04) versus THR:
1.09 (. sd. 0.05); p <
0.001). The ratios varied minimally
when the groups were subdivided by gender, age and obesity. The
decision to increase the size of the acetabular component to accommodate
a