Despite multiple published reviews, the optimum method of correction and stabilisation of Blount's disease remains controversial. The purpose of this study is to evaluate the clinical and radiological outcomes of acute correction of late-onset tibial vara by percutaneous proximal tibial osteotomy with
Introduction. Temporary spanning fixation aims to provide bony stability whilst allowing access and resuscitation of traumatised soft-tissues. Conventional monolateral fixators are prone to half-pin morbidity in feet, variation in construct stability and limited weight-bearing potential. This study compares traditional delta-frame fixators to a circular trauma frame; a virtual tibial ring block spanned onto a fine-wire foot ring fixation. Materials and Methods. The two cohorts were compared for demographics and fracture patterns. The quality of initial reduction and the maintenance of reduction until definitive surgery was assessed by two authors and categorised into four domains. Secondary measures included fixator costs, time to definitive surgery and complications. Results. Fifty-six delta-frames and 48 circular fixators were statistically matched for demographics and fracture pattern. Good or excellent initial reduction was achieved in 51 (91%) delta-frames and 48 (100%) circular fixators (p=0.022). Loss of reduction was observed in 15 (27%) delta-frames and 3 (6%) circular fixators (p<0.001). Post-fixator dislocation occurred in five (9%) delta-frames and one (2%) circular fixator (p=0.147). Duration in spanned fixation was equivalent (11.5 and 11.6 days respectively, p=0.211). Three (5%) delta-frames and 12 (25%) circular fixators were used as definitive fixation. The mean hardware cost was £3,116 for delta-frames and £2,712 for circular fixators. Conclusions. Temporary
Biomechanical studies involving all-wire and hybrid types of circular frame have shown that oblique tibial fractures remain unstable when they are loaded. We have assessed a range of techniques for enhancing the fixation of these fractures. Eight models were constructed using Sawbones tibiae and standard Sheffield ring fixators, to which six additional fixation techniques were applied sequentially. The major component of displacement was shear along the obliquity of the fracture. This was the most sensitive to any change in the method of fixation. All additional fixation systems were found to reduce shear movement significantly, the most effective being push-pull wires and arched wires with a three-hole bend. Less effective systems included an additional half pin and arched wires with a shallower arc. Angled pins were more effective at reducing shear than transverse pins. The choice of additional fixation should be made after consideration of both the amount of stability required and the practicalities of applying the method to a particular fracture.
Introduction. The vast majority of orthopaedic surgeons use C-arm fluoroscopy in the operating theatre when building a circular external fixator. In the absence of previous research in this area, we hypothesised that the surgeon who builds a circular external fixator is exposed to a greater amount of radiation purely as a result of the presence of the metallic fixator in the x-ray beam. The aim of our study therefore was to investigate how the presence of a circular external fixator affects the radiation dose to the surgeon and the surgical assistant. Materials & Methods. A simulated environment was created using a radiolucent operating table, an acrylic lower limb phantom (below knee segment), various configurations of metalic
Introduction. Circular external fixators are fundamental to lower limb reconstruction, primarily in situations with a high risk of infection such as open fractures. During the Covid-19 pandemic, use of circular frames in our unit decreased, following departmental approval, due to resource management and in keeping with BOA guidelines as we opted to “consider alternative techniques for patients who require soft tissue reconstruction to avoid multiple operations”. These alternatives included the use of internal fixation (plate osteosynthesis and intramedullary nailing) as a measure to reduce the number of hospital attendances for patients and to conserve resources. This change in practice has continued in part following the pandemic with the increased use of internal fixation in cases previously deemed unsuitable for such techniques. We present our experience of this treatment strategy in the management of complex lower limb injuries, focusing on outcomes and consider the lessons learnt. Materials & Methods. Data of patients with complex lower limb injuries treated before, during and after the pandemic were collected from our in-house trauma database, theatre records and follow up clinics. The rationale for choosing other techniques over a circular frame, the type of alternative technique used, the cost of such alternatives, the need for soft tissue reconstruction, time to recovery, complications and amputation rates were compared among groups. Results. These data suggest comparable outcomes between circular frames and alternative techniques can be achieved. A notable reduction in the number of circular frames applied during the review period was observed. Furthermore, frame fixation was associated with more frequent outpatient review and the associated implications for resource management. Conclusions. Conclusion: The Covid-19 pandemic has posed great challenges to the Trauma and Orthopaedic community, forcing us to be flexible by adopting alternative treatment methods to traditional
We have carried out a retrospective review of 20 patients with segmental fractures of the tibia who had been treated by
Hig energy fractures of the lower limb are often associated with tibial or femoral bone loss, skin exposition with vascular and nervous injuries (Gustilo et al.). The surgical procedure is a real challenge, consisting in a temporary stabilization of the fracture associated with a plastic and/or vascular reconstruction. Once the skin and vascular injuries are recovered, the orthopaedic surgeon can remove the temporary stabilization performing a
Purpose of Study. Bicondylar tibial plateau fractures are serious injuries to a major weight bearing joint. These injuries are often associated with severe soft tissue injuries that complicate the surgical management. This retrospective study evaluates the management of these high-energy injuries with the use of limited open reduction and fine wire
Abstract. OBJECTIVES. To determine if force measured using a strain gauge in
Purpose of the study:. Is
Introduction. Distal tibial fractures are notoriously difficult to treat and a lack of consensus remains on the best approach. This study examined clinical and functional outcomes in such patients treated definitively by
Ilizarov fixators are reliant on tensioned fine wires for stability. The tension in the wires is generated using specific tensioning devices. Loss of wire tension over time may lead of loss a stability and complications. A series of in vitro experiments were undertaken to explore wire tensioner accuracy, the impact of fixation bolt torque and initial tension on loss of tension in ilizarov constructs under static and dynamic loads. Medical grade materials were applied to a synthetic bone analogue using surgical instruments in all experiments. Bolt torque was fixed at 6, 10 or 14 Nm using a torque limiting wrench. Wire tension was assessed using a strain measurement bridge. Wires were tensioned to 90, 110 and 130kg as measured by a commercial dynamometric tensioner. Static and dynamic testing was undertaken using an instron testing machine. Cyclical loads from 50–750N were applied for 5000 cycles.Introduction
Materials & Methods
Purpose:. This study attempts to establish whether biomechanical manipulation through distraction can result in fracture union. Method:. A retrospective clinical audit of 15 patients with delayed or hypertrophic non-unions treated successfully with closed distraction in
Statement of Purpose. To compare the functional outcome of Distal Tibial Metaphyseal fracture treated with Circular frame compared vs. Locking Plate. Methods and Results. Distal Tibial Metaphyseal fractures were retrospectively identified over an 18 month period. Each fracture was assessed individually using radiographs. All paediatric, compound, tibial plateau and intra-articular fractures were excluded from the study. Other methods of fixation including intramedullary nailing were also excluded. The remaining fractures were assigned to either the
Introduction and Aims: We report a series of 41 corrections in 36 adult patients performed for complex deformities of the foot and ankle using
Background:. Various surgical flaps have been described for the reconstruction of post traumatic soft tissue defects of the heel. These techniques are not all familiar to orthopaedic surgeons. The presented technique is based on the use of a reverse flow island sural flap combined with a circular external fixator. Material:. This presentation is a review of 8 cases performed between 2003 and 2012. Results:. All cases were males with an average age of 39 years (16–56) with a follow up period between 2 and 10 years. The soft tissue defects were located around the heel and Achilles tendon as a result of road traffic accidents involving motor cycles. Six patients had associated bony injuries of the foot and ankle related to the soft tissue defect. All flaps remained viable, and cover was easily achieved with direct closure. Three cases required debridement of subsequent superficial necrosis and additional split skin grafting. Although the flaps are insensate per design, there were no associated problems of late onset pressure ulcers. The
Schatzker V & VI tibial plateau fractures are serious life-changing injuries often resulting in significant complications including post-traumatic arthritis. Reported incidence of secondary TKA following ORIF of all tibial plateau fractures is 7.3% and 13% for Schatzker V & VI tibial. This study reports a 15-year single centre experience of CEF of Schatzker V & VI fractures including PROMs and incidence of secondary TKA. This study was approved by the local Institutional board. All patients from 2007 – 2022 with Schatzker V or VI fractures treated with CEF were identified from a departmental limb reconstruction registry and included in this retrospective study. Patients’ demographics were collected from electronic institutional patient system. Further data was collected for secondary intervention, adverse events, and alignment at discharge. All deceased patients at the time of the study were excluded. Each participant completed a questionnaire about secondary intervention, EQ-5D-3L and Oxford Knee Score (OKS).Introduction
Materials & Methods
Treatment of tibial plateau fractures Schatzker type V and VI or with soft tissues injuries is still remains under discussion. The purpose of this study is to evaluate the results of treatment with circular frame and closed reduction in 25 patients (15 males and 10 females) with tibial plateau fractures, with a mean age of 42 years old (20 – 76 years). Five fractures were classified as Schatzker type II and III and 20 as type V and VI. Reduction was obtained in 22 cases under foot traction and in 3 arthroscopically. Bone grafts inserted through a hole (• 1 cm) in the inner cortex of the tibia metaphysis under fluoroscopy. Eight unstable knees needed bridging the joint for 4 weeks. In 2 cases a cannulated interfragmentary screw was used. Full weight bearing was allowed 3 months after injury when the device was removed. Follow up ranged from 1 to 10 years (mean 5 years). All fractures were united and there was no infection. Full range of the knee motion was achieved in 23 patients while 2 needed an open arthrolysis. There were 2 malunions which were treated with one valgus osteotomy and one TKR. Asymptomatic arthritis appeared in 6 patients. According to Knee Society Score (KSS) the results were classified as excellent in 12, good in 8, fair in 3 and poor in 2 patients. Circular frames are a satisfactory alternative method for the treatment of tibial plateau fractures either in severe soft tissues injuries or in very complex cases
31 cases of high-energy proximal tibial fractures were retrospectively analysed. The series included 22 cases of Schatzker VI and 9 cases of proximal tibial extraarticular fractures. There were 7 females and 24 males, with average age being 45years(26–94). There were 12 open fractures ( 1 Gustilo grade1, 10 grade 3b and one 3c); while 19 cases had Tcherne’s grade 2–3 injury. 4 patients developed compartment syndromes requiring fasciotomy. All fractures were treated with preliminary ligamentotaxis using a unilateral external fixator. In addition, Open fractures underwent radical debridement with the one case of 3c requiring vascular reconstruction. CT scan was then done to assess the joint incongruity ,anatomy of the fracture, and to aid in decision making. The fractures were then fixed using percutaneous techniques and a circular external fixator. Minimal open reduction was resorted to in cases with significant joint depression. In all, 26 cases were managed using percutaneous techniques alone while 5 required minimal open reduction and screw fixation. Bone grafting was done in 6 cases and 11 required a plastics procedure for soft tissue reconstruction. The results were assessed using the radiological Rasmussen’s criteria and the clinically using he IKSS knee score. At a mean follow-up of 31mths, the mean time to metaphyseal union was 18weeks (6–25weeks);. 28 patients had good to excellent clinical scores, while 3 had a fair result. The radiologic assessment graded 12 cases as excellent and 19 as good. Complications included 2 cases with flap edge necrosis, 2 with severe pin tract sepsis, 1 with proximal DVT and one case with septic arthritis. We conclude that the above treatment protocol yields promising results, preserving good knee function without prejudicing future need for arthroplasty.