Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_14 | Pages 4 - 4
10 Oct 2023
Russell H Tinning C Raza A Duff S Preiss RA
Full Access

The Thompson hemiarthroplasty is a common treatment option for acute neck of femur fractures in the elderly population. Our department noted a significant number of patients returning with thigh pain, radiographic loosening and femoral osteolysis following cemented implantation of the titanium alloy version of the Thompson hemiarthroplasty. We are not aware of any previous reports documenting complications specific to the titanium Thompson implant and a retrospective cohort study was therefore initiated following clinical governance approval. 366 titanium alloy Thompson prostheses were implanted for hip fracture treatment between 2017 and 2020. As of February 2023, 6 of these have been revised at our hospital. 5 were revised for symptomatic femoral osteolysis and 1 presented with an acute periprosthetic fracture. All revised cases were determined to be aseptic. 32 living patients were excluded from recall on compassionate grounds due to permanent nursing home residence. 47 living patients were identified of which 33 attended for xray. 28 deceased and/or nursing home resident patients who had pelvis x-rays in the previous 12 months were also included in the analysis. Including the 6 index hips already revised, a total of 61 hip xrays were analysed, of which 19 hips (31.1%) showed radiographic evidence of femoral osteolysis or loosening. We conclude that there is a concerning incidence of femoral osteolysis and implant loosening associated with the titanium Thompson implant. We have discontinued use of the implant and reported our experience to the MHRA. We encourage other Scottish Health-Boards who use this implant to consider enhanced follow-up


Aims. The Intraosseous Transcutaneous Amputation Prosthesis (ITAP) may improve quality of life for amputees by avoiding soft-tissue complications associated with socket prostheses and by improving sensory feedback and function. It relies on the formation of a seal between the soft tissues and the implant and currently has a flange with drilled holes to promote dermal attachment. Despite this, infection remains a significant risk. This study explored alternative strategies to enhance soft-tissue integration. Materials and Methods. The effect of ITAP pins with a fully porous titanium alloy flange with interconnected pores on soft-tissue integration was investigated. The flanges were coated with fibronectin-functionalised hydroxyapatite and silver coatings, which have been shown to have an antibacterial effect, while also promoting viable fibroblast growth in vitro. The ITAP pins were implanted along the length of ovine tibias, and histological assessment was undertaken four weeks post-operatively. Results. The porous titanium alloy flange reduced epithelial downgrowth and increased soft-tissue integration compared with the current drilled flange. The addition of coatings did not enhance these effects. Conclusion. These results indicate that a fully porous titanium alloy flange has the potential to increase the soft-tissue seal around ITAP and reduce susceptibility to infection compared with the current design. Cite this article: Bone Joint J 2017;99-B:393–400


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 136 - 136
1 Sep 2012
El-Husseiny M Pendegrass C Elnikety S Haddad F Blunn G
Full Access

Introduction. Following amputation, residual stumps used to attach the external prostheses can be associated with sores, infection and skin necrosis. These problems could be overcome by off loading the soft tissues. Intraosseous transcutaneous amputation prostheses (ITAP) attach external implants directly to residual bone reducing these complications. However, a tight seal at the skin implant interface is crucial in preventing epithelial down-growth and infection. Fibronectin (Fn) and laminin 332 (Ln), enhance early cell growth and adhesion of keratinocytes. Silanization to titanium alloy (Ti) allows these proteins to bond to the metal directly. We hypothesize that silanized dual coatings of fibronectin and laminin (SiFnLn) will be more durable than absorbed proteins and that keratinocyte adhesion will be increased compared with Ti controls and single silanized proteins. Methods. 10 mm diameter Ti alloy discs were polished, sterilized and silanized. The kinetics of silanized single and dual protein coating attachment onto titanium alloy was quantified using radio-labelled Fn(125I-Fn) and Ln(125I-Ln). Coating durability was assessed when soaked in fetal calf serum (FCS) for 0, 1, 24, 48, 72hrs. Data was compared to un-silanized Ti discs with the same amount of adsorbed proteins. In order to study cell attachment 20 × 103 keratinocytes were seeded on the discs (n = 6): silanized (Si), silanized fibronectin (SiFn), silanized laminin (SiLn), silanized dual coating (SiFnLn) for 1, 4 and 24hrs. Adhesion of cells was assessed using mouse vinculin antibody for 2hrs and alexafluor for 1hr which stains focal adhesions responsible for attaching cells to surfaces. Axiovision Image Analysis software was used to measure cell area, vinculin markers per cell unit and per unit cell area on 15 cells per disc. Data was analysed in SPSS and significance was assumed at the 0.05 level. Results. Silanized dual coatings bonded to Ti alloy in significantly larger quantities compared with adsorbed coatings (all p values < 0.05). When proteins were combined on silanized discs the same amount of each protein was attached as when used as a single coating (i.e. non competitive binding). Keratinocytes cultured on silanized dual coatings were significantly larger, produced more vinculin markers per unit cell and per cell area compared with single coatings at all time points. Conclusion. This study has demonstrated that silanized using dual proteins on Ti alloy enhances early keratinocyte growth and attachment in vitro. It also shows that there is non-competitive binding of laminin to Ti alloys in presence of fibronectin. This may lead to improved epidermal attachment to ITAP creating a tight seal at the implant interface, which will prevent migration of the epithelium and subsequent infection in vivo


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 135 - 135
1 Sep 2012
El-Husseiny M Pendegrass C Haddad F Blunn G
Full Access

Introduction. Intraosseous transcutaneous amputation prostheses (ITAP) provide an alternative means of attaching artificial limbs for amputees. Conventional stump-socket devices are associated with soft tissue complications including; pressure sores and tissue necrosis. ITAP resolves these problems by attaching the exo-prosthesis transcutaneously to the skeleton. The aim of this study is to increase the attachment of dermal fibroblasts to titanium alloy in vitro. Fibronectin (Fn) and laminin 332 (Ln) enhance early cell growth and adhesion. We hypothesize that silanized dual coatings of fibronectin and laminin (SiFnLn) will be more durable when compared with adsorbed dual coating (AdFnLn), and will enhance early fibroblast growth and adhesion compared to single coatings. Methods. The kinetics of dual single and dual protein coating attachment onto titanium alloy was quantified on silanized 10mm diameter discs using radiolabelled Fn (125I-Fn) and Ln (125I-Ln). Sixty discs were polished, sterilized and silanized. Coating durability was assessed when soaked in fetal calf serum (FCS) for 0, 1, 24, 48 and 72hrs. Data was compared to un-silanized Ti discs with the same coatings. Five thousand human dermal fibroblasts were seeded on discs (n = 6) of Ti polished alone (Pol), Ti with adsorbed fibronectin (AdFn), Ti with adsorbed laminin (AdLn), Ti adsorbed dual coating (AdFnLn), Ti silanized (Si), Ti silanized with fibronectin (SiFn), Ti silanized with laminin (SiLn), Ti silanized with a dual coating (SiFnLn) for 24hrs. In order to measure cell adhesion fibroblasts were fixed, vinculin stained using mouse vinculin antibody and alexa fluor. Axiovision Image Analysis software was used to measure cell area, vinculin focal adhesion markers per cell and per unit cell area. Data was analysed in SPSS and significance was assumed at the 0.05 level. Results. Silanized dual coatings bonded to Ti alloy in significantly larger quantities compared with adsorbed coatings at all time points (all p values < 0.05). Fibroblasts cultured on dual coatings were significantly larger, produced more vinculin markers per cell, and per unit cell area compared with single coatings. Cells on SiFnLn were larger with more numerous vinculin markers per cell, and per unit cell area compared with AdFnLn (p<0.05). Conclusion. This study has demonstrated that covalently bonding both fibronectin and laminin to Ti alloy provides a durable, dual coating that enhances early fibroblast growth and attachment compared with either protein coating alone in vitro. Our study showed that there is non-competitive binding of laminin on Ti surfaces in the presence of fibronectin. Dual coatings may be applied to the skin-penetrating region of transcutaneous devices to improve the skin seal and this may have positive implications for the development of ITAP


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 355 - 355
1 Sep 2012
Philippot R Camilleri JP Boyer B Farizon F
Full Access

The concept of stainless steel dual mobility cups in total hip arthroplasty has demonstrated very low long-term instability rates and a 98% survival rate after 12 years. We systematically implanted titanium alloy acetabular cups during a one year period. The purpose of our retrospective study was to report the 18-year clinical outcome data in a homogeneous and continuous series of 103 primary total hip replacements after implantation of a cementless titanium cup. All patients were implanted with NOVAE Ti (SERF) cups made of titanium alloy combined with a retentive polyethylene liner and a 22.2 mm cobalt chrome prosthetic head. Mean patient age at the time of surgery was 53 years. All patients were clinically and radiographically evaluated. The overall 18-year actuarial cup survival rate with a 95% confidence interval was 87.4%. At last follow-up, there was no evidence of implant instability whereas acetabular aseptic loosening was reported in one case and high wear of the retentive liner in 9. The results of this investigation confirmed the long-term stability of dual-mobility implants. The main limitation of this system was early wear of the polyethylene liner in contact with the titanium metal back and reaction with third body along with loss of liner retentivity. In our study, titanium demonstrated favourable osteointegration properties but poor tribologic characteristics, therefore suggesting its interest at the bone-cup interface only


Bone & Joint Open
Vol. 5, Issue 6 | Pages 457 - 463
2 Jun 2024
Coviello M Abate A Maccagnano G Ippolito F Nappi V Abbaticchio AM Caiaffa E Caiaffa V

Aims

Proximal femur fractures treatment can involve anterograde nailing with a single or double cephalic screw. An undesirable failure for this fixation is screw cut-out. In a single-screw nail, a tip-apex distance (TAD) greater than 25 mm has been associated with an increased risk of cut-out. The aim of the study was to examine the role of TAD as a risk factor in a cephalic double-screw nail.

Methods

A retrospective study was conducted on 112 patients treated for intertrochanteric femur fracture with a double proximal screw nail (Endovis BA2; EBA2) from January to September 2021. The analyzed variables were age, sex, BMI, comorbidities, fracture type, side, time of surgery, quality of reduction, pre-existing therapy with bisphosphonate for osteoporosis, screw placement in two different views, and TAD. The last follow-up was at 12 months. Logistic regression was used to study the potential factors of screw cut-out, and receiver operating characteristic curve to identify the threshold value.


Bone & Joint Open
Vol. 4, Issue 8 | Pages 584 - 593
15 Aug 2023
Sainio H Rämö L Reito A Silvasti-Lundell M Lindahl J

Aims

Several previously identified patient-, injury-, and treatment-related factors are associated with the development of nonunion in distal femur fractures. However, the predictive value of these factors is not well defined. We aimed to assess the predictive ability of previously identified risk factors in the development of nonunion leading to secondary surgery in distal femur fractures.

Methods

We conducted a retrospective cohort study of adult patients with traumatic distal femur fracture treated with lateral locking plate between 2009 and 2018. The patients who underwent secondary surgery due to fracture healing problem or plate failure were considered having nonunion. Background knowledge of risk factors of distal femur fracture nonunion based on previous literature was used to form an initial set of variables. A logistic regression model was used with previously identified patient- and injury-related variables (age, sex, BMI, diabetes, smoking, periprosthetic fracture, open fracture, trauma energy, fracture zone length, fracture comminution, medial side comminution) in the first analysis and with treatment-related variables (different surgeon-controlled factors, e.g. plate length, screw placement, and proximal fixation) in the second analysis to predict the nonunion leading to secondary surgery in distal femur fractures.


Bone & Joint Open
Vol. 2, Issue 10 | Pages 825 - 833
8 Oct 2021
Dailey HL Schwarzenberg P Webb, III EB Boran SAM Guerin S Harty JA

Aims

The study objective was to prospectively assess clinical outcomes for a pilot cohort of tibial shaft fractures treated with a new tibial nailing system that produces controlled axial interfragmentary micromotion. The hypothesis was that axial micromotion enhances fracture healing compared to static interlocking.

Methods

Patients were treated in a single level I trauma centre over a 2.5-year period. Group allocation was not randomized; both the micromotion nail and standard-of-care static locking nails (control group) were commercially available and selected at the discretion of the treating surgeons. Injury risk levels were quantified using the Nonunion Risk Determination (NURD) score. Radiological healing was assessed until 24 weeks or clinical union. Low-dose CT scans were acquired at 12 weeks and virtual mechanical testing was performed to objectively assess structural bone healing.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_16 | Pages 14 - 14
1 Apr 2013
Penzkofer U Schmidt R Bachmaier S Augat P
Full Access

Background. The inherently high stiffness of locked plate constructs is increasingly recognized as a potential cause of deficient healing observed in patients with periarticular locked plating systems. The objective of this study is to perform a biomechanical comparison of distal femur locked plating systems. Methods. Biomechanical study using bone substitutes in a distal femur fracture model (OTA/AO 33-A3). Four different locked plate fixation systems were compared (AxSOS, LCP, PERI-LOC, POLYAX). Eight bone implant constructs of each plating system were evaluated in a multiple testing model to examine static failure, stiffness under static and cyclic loading and cyclic fatigue. Results. The implant systems made of titanium alloy (POLYAX & LCP) supported smaller torsional stiffness values under static loads and smaller axial stiffness values under cyclic loads compared with the implant systems made of stainless steel (PERI-LOC & AxSOS). All bone implant constructs reached the cyclic failure criterion of 10 degrees displacement at the fracture area within the third load level (>50000 cycles) except for the LCP which failed earlier. Conclusion. The tested four different locked plating systems differ significantly in terms of stiffness and load to failure. Two of the clinical available systems differed almost 100% in stiffness values. And one system differed almost 100% in fatigue strength. For clinical use, this knowledge is essential for the practicing orthopaedic surgeon


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 523 - 523
1 Sep 2012
Fontaine C Wavreille G Leroy M Dos Remedios C Chantelot C
Full Access

In rheumatoid arthritis (RA), non constrained or semi-constrained prostheses can be used. The authors used the Kudo III, IV or V or iBP prostheses 54 times from 1994 to 2003. After initial satisfactory results, they had to change one or both implants for several reasons: humeral stem fracture (5 cases), unipolar humeral loosening (1 case), ulnar loosening without laxity (8 cases), polyethylene wear (11 cases), due to progressive ulnar collateral ligament lengthening and progressive valgus deformity, without or with metallosis, due to contact between Cr-Co humeral component and titanium alloy ulnar component, chronic infection (1 case). When the local conditions were satisfactory (bone stock, ligament balance), the fractured or loosened component was changed. When the conditions were bad (poor bone stock, ligament misbalance, metallosis), both implants were removed; posterior humeral and/or medial or lateral ulnar window were used to removed the uncemented stems still osteointegrated. All the bipolar operations used the Coonrad-Morrey prosthesis, but the last case a Discovery prosthesis. The operative tricks are described, the management of the extensor apparatus is discussed, the clinical outcomes (especially the extensor apparatus function, most often weak) and the radiographic outcomes are presented


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 236 - 236
1 Sep 2012
Queiroz M Barros F Daniachi D Polesello G Guimarães R Ricioli W Ono N Honda E
Full Access

Introduction. One of the most common complications of ceramic on ceramic hip replacement is squeaking. The association of Accolade stem and Trident acetabular system has been reported to have squeaking incidence of up to 35,6%. There is doubt if this phenomenon occurs due to: the stem titanium alloy, the V40 femoral neck, the recessed liner of the trident cup or even the mal-seating of the trident insert on the cup. Objectives. Primary: The purpose of the present study was to determine the incidence of squeaking in association with the use of Exeter stem and Trident ceramic acetabular system. Secondary: Analysis of the correlation of the cup abduction angle and squeaking. Methods. During the period from March 2004 to December 2008, two surgeons performed 87 total hip arthroplasties in 77 patients with use of a ceramic-on-ceramic bearing (Exeter stem, alumina head, Trident ceramic acetabular system). Seventy six patients (86 THA) were available for review after at least 18 months follow-up. The incidence of squeaking and other noises was analyzed. Cup abduction angle was measured. The Pearson correlation coefficient was used to determine if a correlation existed between the cup abduction angle and squeaking. Results. The incidence of squeaking was 2,63% (2 patients). Both patients reported a “click” noise in hyperextension of the hip. The mean abduction angle was 44 degress (35–60), and 48 degrees (46 and 50) on the squeaking group. There was no statistically significant difference in the in the mean cup inclination between squeaky and quiet hips. Conclusion. The incidence of squeaking in association with the use of Exeter stem and Trident ceramic acetabular system was 2,63%. There was no correlation of the cup abduction angle and squeaking


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 496 - 496
1 Sep 2012
Huber M Zweymueller K Lintner F
Full Access

Background. Continual implant stability is an important factor for the long-term success of cementless hip replacements. The increasing lifespan of patients causes a higher frequency of osteoporosis which may result in implant loosening due to bone loss. This study aimed to evaluate stability of long living implants in patients with advanced age. Patients and methods. Nine cementless stems made of Titanium-alloy including adjacent bone tissue obtained post mortem were evaluated by radiologic-microradigraphical, histological and morphometrical analysis. The percentage of the surface area covered by bone (BICI=bone implant contact index) was determined. The age of seven women and two men ranged between 81 and 92 years. The time in situ ranged between 10 and 20 years. From the entire length of the femora bearing implants 5 transverse segments were excised, dehydrated, embedded in methylmethacrylate. After the grinding procedure, the sections were evaluated by light microscopy and morphometrical analysis. The autopsy findings were recorded. Atherosclerosis and their related diseases were evident in all cases. Results. The femora of all female patients revealed features of high bony atrophy with concomitant transformation of the corticalis into spongy bone, whereas in male patients minor to moderate atrophic bone changes in the proximal femoral area without implication of the corticalis could be observed. All of the cementless stems made of Titanium-alloy showed osteointegration. The stabilization of the implant resulted in the forceps-like encasement of the edges of the implant within the cortical anchoring and by the development of compensatory bony hypertrophy. The BICI ranged between 35 und 63 percent. Conclusion. Elderly patients provided with cementless hip replacments revealed stable implants in spite of marked bone atrophy and an implantation period up to 20 years. Simultaneously, severe atherosclerosis and their related diseases, which may contribute to bone loss, were evident. The present findings may result from the favoring properties of cementless endoprostheses made of titanium alloy, cortical prosthesis anchoring, and self regulating bone processes. Pharmacologic and therapeutic consequences together with geriatric assessment should be required to preserve functionality and mobility


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 413 - 413
1 Sep 2012
Henriksen S Ding M Overgaard S
Full Access

Introduction. The combined incubation of a composite scaffold with bone marrow stromal cells in a perfusion bioreactor could make up a novel hybrid graft material with optimal properties for early fixation of implant to bone. The aim of this study was to create a bioreactor activated graft (BAG) material, which could induce early implant fixation similar to that of allograft. Two porous scaffold materials incubated with cells in a perfusion bioreactor were tested in this study. Methods and Materials. Two groups of 8 skeletally mature female sheep were anaesthetized before aspiration of bone marrow from the iliac crest. For both groups, mononuclear cells were isolated, and injected into a perfusion bioreactor (Millenium Biologix AG, Switzerland). Scaffold granules Ø∼900–1500 μm, ∼88% porosity) in group 1, consisted of hydroxyapatite (HA, 70%) with -tricalcium-phosphate (−TCP, 30%) (Danish Technological Institute, Denmark). The granules were coated with poly-lactic acid (PLA) 12%, in order to increase the mechanical strength of the material (Phusis, France). Scaffold granules Ø∼900–1400 μm, 80% porosity) in group 2 consisted of pure HA/-TCP (Fin Ceramica, Italy). For both groups, cells were incubated in the bioreactor for 2 weeks. Fresh culture medium supplemented with dexamethasone and ascorbic-acid was added every third or fourth day. Porous titanium alloy implants with diameter=length=10mm (Biomet, USA) were inserted bilaterally in each of the distal femurs of the sheep; thus 4 implants in each sheep. The concentric gap (2 mm) surrounding the implant was filled with 1) BAG (autogenous), 2) granules, 3) granules+bone marrow aspirate (BMA, autologous) or 4) allograft. The sheep were euthanized after 6 weeks. Distal femurs were removed and implant-bone samples were divided in two parts. The superficial part was used for mechanical testing and micro-CT scanning, and the profound part for histomorphometry. Push-out tests were performed on an 858 Bionix MTS hydraulic materials testing machine. Shear mechanical properties between implant and newly generated bone were calculated to assess implant fixation. Results were assessed by One-way ANOVA. P-values less than 0.05 were considered significant. Results. One sheep in group 1 had to be euthanized after 4 weeks (excluded). One implant in each group was loosened and could not undergo push-out test (excluded). Group 1: No significant differences regarding failure energy (kJ/m2, p=0.44) or ultimate shear strength (MPa, p=0.17) could be seen. Shear stiffness (MPa) was significantly higher for the allograft group (p=0.04). Group 2: No significant differences regarding failure energy (p=0.11) or shear stiffness (p=0.52) could be seen. Ultimate shear strength was significantly higher for allograft (p=0.04). Results from μ-CT scanning and histomorphometry are pending. Discussion and Conclusion. The present study shows a possible effect of bioreactor activated bone substitute on early implant fixation. We are currently working on bone microarchitecture surrounding implant and histomorphometry. These results will aid in determining if BAG could make up a promising alternative for allograft as bone graft material


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 134 - 134
1 Sep 2012
Beaulé P White C Carsen S Rasuli K Doucette S
Full Access

Introduction. Modern cementless press fit stems rely on early fixation and stability for osteointegration and longterm success with early migration increasing the risk of failure. The Ein Bild Roentegen Analyse Femoral Component Analysis (EBRA-FCA) methods allow accurate measurement of femoral stem subsidence without the need for Tantalum markers. The degree of subsidence of femoral stems in the first two years has been shown to be highly predictive of failure when using the cut off value of 1.5 mm. We aimed to measure the early migration pattern of a titanium alloy, tapered, plasma and hydroxyapatite coated femoral stem and any factors associated with subsidence. Methods. Between January 2005-June 2007, 387 Accolade cementless femoral stems (Stryker, Allendale NJ) were implanted at our institution. Seventy-seven had a minimum of two years post operative follow up and a complete set of pre and postoperative radiographs for analysis. Our group inlcuded 45 females with a mean age of 71.4 years, and 32 males with a mean age of 68.5 years. The primary diagnosis was degenerative osteoarthritis in 71 patients, avascular necrosis in two, and post fracture in four patients. The average BMI was 27.1. We measured the canal index to assess bone quality and the canal calcar index to assess the proximal femoral morphology. Immediate postoperative radiographs were assessed for canal fill of the prosthesis and implantation varus/valgus angles. The EBRA-FCA software was used to obtain migration curves for each stem. Results. The mean follow up was 29.3months (24–48). The mean canal index was 0.55 (0.36–0.68) with a mean canal calcar index of 0.54 (0.39–0.79). The average canal fill index at the midpoint of the stem was <0.8 in 37 stems and >80 in 40 stems. The average subsidence at 24 months was 2 mm and this had risen to 2.4 mm by 36 months postoperatively. When analysed using a Kaplan Meier curve using 1.5 mm as an end point we found a survivorship of 63.4% (52.3–74.5) at 24 months and this had worsened to 41.6% (26.6–56.5) by 36 months. Multivariate and univariate regression analysis of measured variables did not reveal any significant hazard for any factor other than the larger stem sizes doing worse. Discussion. Although several cementless tapered stem designs have had an excellent track record, our migration analysis of the Accolade stem is somewhat concerning. Thirty three percent of stems had reached the 1.5 mm subsidence point by two years. This is of concern as work has previously shown this to predict failure of stems with aseptic loosening at ten years with an accuracy of 79%. If these stems go on to fail at the predicted rate this would represent an unacceptably high level of failure. Our data raises serious concerns about the overall clinical performance of this stem design due to poor initial stability and integration


The Bone & Joint Journal
Vol. 99-B, Issue 1_Supple_A | Pages 50 - 59
1 Jan 2017
Carli AV Negus JJ Haddad FS

Aims

Periprosthetic femoral fractures (PFF) following total hip arthroplasty (THA) are devastating complications that are associated with functional limitations and increased overall mortality. Although cementless implants have been associated with an increased risk of PFF, the precise contribution of implant geometry and design on the risk of both intra-operative and post-operative PFF remains poorly investigated. A systematic review was performed to aggregate all of the PFF literature with specific attention to the femoral implant used.

Patients and Methods

A systematic search strategy of several journal databases and recent proceedings from the American Academy of Orthopaedic Surgeons was performed. Clinical articles were included for analysis if sufficient implant description was provided. All articles were reviewed by two reviewers. A review of fundamental investigations of implant load-to-failure was performed, with the intent of identifying similar conclusions from the clinical and fundamental literature.


Bone & Joint Research
Vol. 2, Issue 5 | Pages 79 - 83
1 May 2013
Goffin JM Pankaj P Simpson AHRW Seil R Gerich TG

Objectives

Because of the contradictory body of evidence related to the potential benefits of helical blades in trochanteric fracture fixation, we studied the effect of bone compaction resulting from the insertion of a proximal femoral nail anti-rotation (PFNA).

Methods

We developed a subject-specific computational model of a trochanteric fracture (31-A2 in the AO classification) with lack of medial support and varied the bone density to account for variability in bone properties among hip fracture patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 665 - 677
1 May 2011
Sköldenberg OG Salemyr MO Bodén HS Lundberg A Ahl TE Adolphson PY

Our aim in this pilot study was to evaluate the fixation of, the bone remodelling around, and the clinical outcome after surgery of a new, uncemented, fully hydroxyapatite-coated, collared and tapered femoral component, designed specifically for elderly patients with a fracture of the femoral neck.

We enrolled 50 patients, of at least 70 years of age, with an acute displaced fracture of the femoral neck in this prospective single-series study. They received a total hip replacement using the new component and were followed up regularly for two years.

Fixation was evaluated by radiostereometric analysis and bone remodelling by dual-energy x-ray absorptiometry. Hip function and the health-related quality of life were assessed using the Harris hip score and the EuroQol-5D.

Up to six weeks post-operatively there was a mean subsidence of 0.2 mm (−2.1 to +0.5) and a retroversion of a mean of 1.2° (−8.2° to +1.5°). No component migrated after three months. The patients had a continuous loss of peri-prosthetic bone which amounted to a mean of 16% (−49% to +10%) at two years. The mean Harris hip score was 82 (51 to 100) after two years.

The two-year results from this pilot study indicate that this new, uncemented femoral component can be used for elderly patients with osteoporotic fractures of the femoral neck.