Aim of this study was to identify reoperation rates in patients with short oblique and transverse fractures around a well fixed cemented polished taper slip stem and to determine any associations with treatment failure.

Retrospective cohort study of 31 patients with AO transverse or short oblique Vancouver B1 PFFs around THA (total hip arthroplasty) cemented taper slip stems: 12 male (39%); mean age 74±11.9 (range 44–91); mean BMI 28.5±1.4 (range 16–48); and median ASA 3. Patient journeys were assessed, re-interventions reviewed. The primary outcome measure was reoperation.

Time from primary THA to fracture was 11.3±7.8yrs (0.5–26yrs). Primary surgical management was fixation in 27/31 and rTHA (revision total hip arthroplasty) in 4/31. 10 of 31 (32%) patients required reoperation, 9 within 2 years of fracture: 1 following rTHA and 8 following ORIF. The commonest mode of failure was non-union (n=6). No significant associations with reoperation requirement were identified. Kaplan-Meier free from reoperation was 67.4% (49.8–85.0 95% CI) at 2 years and this was unaffected by initial management with ORIF or rTHA (Log rank 0.898). Of those reoperated, 6/10 required multiple reoperations to obtain either bony union or a stable revision construct and 13% ultimately required proximal femoral endoprostheses. The relative risk of 1 year mortality was 1.6 (0.25 to 10.1 95%CI) among patients who required reoperation compared to those who did not.

These are difficult fractures to manage, should not be underestimated and patients should be counselled that there is a 30% risk of reoperation and 20% of requiring multiple reoperations.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_5 | Pages 8 - 8
13 Mar 2023
Powell-Bowns M Oag E Martin D Moran M Scott C
Full Access

The aim of the study was to report the survival of open reduction and internal fixation (ORIF) of Vancouver B fractures associated with the Exeter Stem (ES) at a minimum of 5 years.

This retrospective cohort study assessed 129 consecutive patients with Vancouver B type fractures treated with ORIF from 2008-2016 at a minimum of 5 years. Patient records were examined, and the following recorded: details of primary prosthesis, details of injury, Vancouver classification, details of operative management, complications, and requirement for reoperation. Data was analysed using SPSS. Survival analysis was undertaken using the endpoint ‘reoperation for any reason’.

Mean age at fracture was 78.2 (SD10.6, 46-96) and 54 (43%) were female. Vancouver subclassifications were: 24% B1, 70.5% B2 and 5.5% B3. For all Vancouver B fractures, Kaplan Meier analysis demonstrated a 5 year survival free from reoperation of 88.8% (82.0-94.7 95%CI). Fourteen patients required reoperation, most commonly within the first year for non-union and plate fracture (5.4%). Five-year survival for any reoperation differed significantly according to fracture type (p=0.016) and was worst in B1s: B1 76.6% (61.3-91.9); B2 92.6% 986.9-98.3); and 100% of B3. Univariate analysis identified B1 type (p=0.008) and a transverse fracture pattern (p=0.003) to be significantly associated with the need for reoperation.

Adopting a strategy of fixation of all Vancouver B fractures involving the ES where the fracture was anatomically reducible and the bone cement interface was well-fixed was associated with a 5 year survival, free from reoperation of 88.8%.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 2 - 2
1 Sep 2012
Wuestemann T Bastian A Parvizi J Nessler J Kolisek F Nevelos J
Full Access

Introduction

The origins of the uncemented tapered wedge hip stem design currently offered by several orthopaedic device companies can be linked back to the cemented Straight Mueller type stem design first used in 1977. The design, a wedge shape with a taper angle of 6 degrees, maintains a single medial curvature for all sizes and increases laterally in the width to accommodate different size femurs. Although evolutionary improvements have been made over the years the basic body geometry of the stem has stayed mainly unchanged with excellent clinical survivorship. Over the past decade, the demographics of hip replacement have changed, with a large increase in younger male patients in the age range of 40 to 60 years. In this study the femoral fit of a novel tapered stem, designed to fit a wide array of patient types, is compared to a standard predicate tapered stem design.

Methods

A bone morphology study was performed on a patient population of 556 patients using three dimensional digital data from CT-scans. To characterize the fit of the stem designs we analyzed the ratio of a distal (60mm below lesser trochanter) and a proximal (10mm above lesser trochanter) cross section. The same measurements were taken with the standard tapered stem design and the novel tapered stem design, with a given constant implantation height of 20mm above the lesser trochanter. The fit of the stems was classified as Type 1, where there was both proximal and distal engagement, Type 2, proximal engagement only, Type 3, distal engagement only. The distal and proximal engagement, Type 1, was specified with a maximum engagement difference of 2mm proximal to distal.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_25 | Pages 7 - 7
1 May 2013
Patil S Goudie S Keating JF Patton S
Full Access

Vancouver B fractures around a cemented polished tapered stem (CTPS) are often treated with revision arthroplasty. Results of osteosynthesis in these fractures are poor as per current literature. However, the available literature does not distinguish between fractures around CTPS from those around other stems.

The aim of our study was to assess the clinical and radiological outcome of open reduction and internal fixation in Vancouver B fractures around CTPS using a broad non-locking plate.

Patients treated with osteosynthesis between January 1997 and July 2011 were retrospectively reviewed. All underwent direct reduction and stabilisation using cerclage wires before definitive fixation with a broad DCP. Bicortical screw fixation was obtained in the proximal and distal fragments. We defined failure of treatment as revision for any cause.

101 patients (42 men and 59 women, mean age 79) were included. 70 had minimum follow-up of 6 months. 63 of these went on to clinical and radiological union. Three developed infected non-union. 7 had failure of fixation. Lack of anatomical reduction was the commonest predictor of failure followed by inadequate proximal fragment fixation and infection. 14 patients dropped at least 1 mobility grade from their preoperative status.

This is the largest series of a very specific group of periprosthetic fractures treated with osteosynthesis. Patients who develop these fractures are often frail and “high risk” for major revision surgery. We recommend osteosynthesis for patients with Vancouver B periprosthetic fractures around CTPS provided these fractures can be anatomically reduced and adequately fixed.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_7 | Pages 8 - 8
1 May 2019
Moriarty P Wong L Kearney D Harty J
Full Access

Revision hip arthroplasty presents a challenge in the setting of femoral bone loss. Tapered stems are susceptible to subsidence which leads to leg length inequality, hip instability and may necessitate repeat femoral revision surgery. The purpose of this study was to compare radiographic outcomes in two modular tapered revision systems with different distal stem geometries. We sought to establish the minimum postoperative stem bi-cortical contact length that predicts subsidence for tapered stems. This study examined revision total hip arthroplasties between 2009 and 2016 in a European university affiliated major trauma center. Modular stem A has a taper of 3 degrees whereas modular stem B has a taper of 2 degrees. Radiologic assessment compared x-rays at two time points: immediately post-surgery and most recent x-ray available at a minimum follow up of two years. Leg length discrepancy, subsidence and postoperative bi cortical contact was assessed. Descriptive summary statistics calculated clinical factors (i.e. age, gender, Paprosky classification). 122 arthroplasties were completed. Complete data was available for 112. Revisions were carried out for Paprovski grade 3a/ 3b femoral deficits. Post-operative bi-cortical contact of the proximal stem < 20mm was associated with higher subsidence rates (P = 0.047). Subsidence rates for implant A and B system were 4.27mm (0.12–25.62mm) and 3.43 mm (0.3–11.1 mm) respectively. Significant subsidence was noted in 9.8% (n=8) in implant A and 5.2%(n=2) in implant B. We conclude that immediate postoperative bi-cortical stem contact of <20mm was associated with significantly higher subsidence rates in this study