Introduction. The advantages of metal on metal (MOM) hip replacement are decreased wear rate, preservation of bone stock, anatomical restoration and enhanced stability. Large amounts of metal wear particles and metal ions are released which may induce adverse reactions including local soft tissue toxicity, hypersensitivity reactions, bone loss and risk of carcinogenesis. Aseptic loosening can be the result of a peri prosthetic osteolysis generated as a result of a biological response to particulate wear debris. Our aim in this study was to determine whether a steeply inclined acetabular component would give rise to a higher concentration of metal ions. Patients and methods. Between April 2003 and June 2006, 22 patients had MOM hip replacement for osteoarthritis by a single Surgeon. There were 12 male and 10 female patients. The average age at the time of surgery was 56 years (Range: 44–69 years). We divided the 22 patients into 2 groups, one group (A) of 11 patients with the acetabular inclination angle more than 50 degrees and the other group (B) of 11 patients with the angle less than 50 degrees. The inclination of the acetabular cup was measured using a standard AP radiograph of the pelvis. The patients had metal ion levels (blood chromium and serum cobalt) measured at an average follow up of 3.2 years (Range 2.4 to 5 years). Results. Mean blood chromium level in the group A (146 nM/L) was significantly higher (p=0.005) than in Group B (92 nM/L). Mean