Filling the empty holes in peri-articular locking
plates may improve the fatigue strength of the fixation. The purpose of
this in vitro study was to investigate the effect
of plugging the unused holes on the fatigue life of peri-articular distal
femoral plates used to fix a comminuted supracondylar fracture model. A locking/compression plate was applied to 33 synthetic femurs
and then a 6 cm metaphyseal defect was created (AO Type 33-A3).
The specimens were then divided into three groups: unplugged, plugged
with locking screw only and fully plugged holes. They were then
tested using a stepwise or run-out fatigue protocol, each applying
cyclic physiological multiaxial loads. All specimens in the stepwise group failed at the 770 N load
level. The mean number of cycles to failure for the stepwise specimen
was 25 500 cycles (. sd. 1500), 28 800 cycles (. sd. 6300),
and 26 400 cycles (. sd. 2300) cycles for the unplugged, screw
only and fully plugged configurations, respectively (p = 0.16).
The mean number of cycles to failure for the run-out specimens was
42 800 cycles (. sd. 10 700), 36 000 cycles (. sd. 7200),
and 36 600 cycles (. sd. 10 000) for the unplugged, screw
only and fully plugged configurations, respectively (p = 0.50).
There were also no differences in axial or torsional stiffness between
the constructs. The failures were through the
Objectives. Small animal models of fracture repair primarily investigate
indirect fracture healing via external callus formation. We present
the first described rat model of direct fracture healing. Methods. A rat tibial osteotomy was created and fixed with compression
plating similar to that used in patients. The procedure was evaluated
in 15 cadaver rats and then in vivo in ten Sprague-Dawley
rats. Controls had osteotomies stabilised with a uniaxial external
fixator that used the same surgical approach and relied on the same
number and diameter of
The augmentation of fixation with bone cement
is increasingly being used in the treatment of severe osteoporotic fractures.
We investigated the influence of bone quality on the mechanics of
augmentation of plate fixation in a distal femoral fracture model
(AO 33 A3 type). Eight osteoporotic and eight non-osteoporotic femoral
models were randomly assigned to either an augmented or a non-augmented
group. Fixation was performed using a locking compression plate.
In the augmented group additionally 1 ml of bone cement was injected
into the
We investigated the static and cyclical strength of parallel and angulated locking plate screws using rigid polyurethane foam (0.32 g/cm. 3. ) and bovine cancellous bone blocks. Custom-made stainless steel plates with two conically threaded
The aim of this study was to assess the safety and clinical outcome of patients with a femoral shaft fracture and a previous complex post-traumatic femoral malunion who were treated with a clamshell osteotomy and fixation with an intramedullary nail (IMN). The study involved a retrospective analysis of 23 patients. All had a previous, operatively managed, femoral shaft fracture with malunion due to hardware failure. They were treated with a clamshell osteotomy between May 2015 and March 2020. The mean age was 42.6 years (26 to 62) and 15 (65.2%) were male. The mean follow-up was 2.3 years (1 to 5). Details from their medical records were analyzed. Clinical outcomes were assessed using the quality of correction of the deformity, functional recovery, the healing time of the fracture, and complications.Aims
Methods
Background. Locking internal fixation through a relatively small surgical dissection presents an innovative technique for managing distal tibial extra-articular fractures. The aim of this study is to evaluate the biomechanical properties of one locking internal fixation plate used to treat these injuries. Method. An AO/OTA43-A3 fracture was created in synthetic composite tibiae. Locking internal fixation was achieved with an anatomically pre-contoured medial distal tibial locking plate. Comparisons were made between different screw configurations in holes proximal to the fracture and monocortical versus bicortical fixation. Axial stiffness was measured using a universal materials testing machine. Finite element analysis (FEA) was used to model the elastic deformation of the constructs. Outcome measures were axial stiffness under physiological loading conditions and compression load to failure. Results. A trend towards reduced mean axial stiffness from the bicortical to the monocortical fixation constructs was observed. The physical model demonstrated no difference in measured mean axial stiffness between constructs with all
Posterior column plating through the single anterior approach reduces the morbidity in acetabular fractures that require stabilization of both the columns. The aim of this study is to assess the effectiveness of posterior column plating through the anterior intrapelvic approach (AIP) in the management of acetabular fractures. We retrospectively reviewed the data from R G Kar Medical College, Kolkata, India, from June 2018 to April 2023. Overall, there were 34 acetabulum fractures involving both columns managed by medial buttress plating of posterior column. The posterior column of the acetabular fracture was fixed through the AIP approach with buttress plate on medial surface of posterior column. Mean follow-up was 25 months (13 to 58). Accuracy of reduction and effectiveness of this technique were measured by assessing the Merle d’Aubigné score and Matta’s radiological grading at one year and at latest follow-up.Aims
Methods
Introduction. Periprosthetic femur fractures are a serious complication after hip replacement surgery. In an aging population these fractures are becoming more and more common. Open reduction and plate osteosynthesis is one of the available treatment options. Objective. To investigate hip stem stability and cement mantle integrity under cyclic loading conditions after plate fixation with screws perforating the cement in the proximal fragment. Methods. Polished tapered hip stems were implanted in 16 biomechanical testing femora with Palacos cement (3rd generation technique) according to the manufacturer's recommendations. 8 testing bones were osteotomised distal to the stem representing the fracture group (Vancouver Type C). The osteotomy was fixed with a polyaxial locking plate, the other 8 specimens served as a control group. The specimens were tested in a biaxial material testing machine under axial compression (including adduction and torsion moments) for 100.000 cycles at physiological loads. Stem subsidence was measured in 3 planes with a stereoscopic image correlation system during the tests. Subsequently the sliced and crack dyed specimens were investigated microscopically for cement cracks. Results. In the control group no specimen failed during testing. There were no statistically significant differences in stem subsidence along the longitudinal axis (control group mean ± SD −15.4 ± 12.2 μm, fracture group −14.1 ± 13.1 μm). In the fracture group two specimens fractured through the most proximal
Osteopetrosis (OP) is a rare hereditary disease that causes reduced bone resorption and increased bone density as a result of osteoclastic function defect. Our aim is to review the difficulties, mid-term follow-up results, and literature encountered during the treatment of OP. This is a retrospective and observational study containing data from nine patients with a mean age of 14.1 years (9 to 25; three female, six male) with OP who were treated in our hospital between April 2008 and October 2018 with 20 surgical procedures due to 17 different fractures. Patient data included age, sex, operating time, length of stay, genetic type of the disease, previous surgery, fractures, complications, and comorbidity.Aims
Methods
Recent studies of nonoperatively treated displaced midshaft clavicular fractures have shown a high incidence of nonunion and unsatisfactory functional outcome. Some studies have shown superior functional results and higher rates of healing following operative treatment. The aim of this study was to compare the outcome in these patients after nonoperative management with those treated with fixation. In a multicentre, parallel randomized controlled trial, 146 adult patients with an acute displaced fracture of the midthird of the clavicle were randomized to either nonoperative treatment with a sling (71, 55 men and 16 women with a mean age of 39 years, 18 to 60) or fixation with a pre-contoured plate and locking screws (75, 64 men and 11 women with a mean age of 40 years, 18 to 60). Outcome was assessed using the Disabilities of the Arm, Shoulder and Hand (DASH) Score, the Constant Score, and radiographical evidence of union. Patients were followed for one year.Aims
Patients and Methods
External fixators are the traditional fixation method of choice for contaminated open fractures. However, patient acceptance is low due to the high profile and therefore physical burden of the constructs. An externalised locking compression plate is a low profile alternative. However, the biomechanical differences have not been assessed. The objective of this study was to evaluate the axial and torsional stiffness of the externalised titanium locking compression plate (ET-LCP), the externalised stainless steel locking compression plate (ESS-LCP) and the unilateral external fixator (UEF). A fracture gap model was created to simulate comminuted mid-shaft tibia fractures using synthetic composite bones. Fifteen constructs were stabilised with ET-LCP, ESS-LCP or UEF (five constructs each). The constructs were loaded under both axial and torsional directions to determine construct stiffness.Objectives
Methods
Fractures of the distal femur can be challenging to manage and
are on the increase in the elderly osteoporotic population. Management
with casting or bracing can unacceptably limit a patient’s ability
to bear weight, but historically, operative fixation has been associated
with a high rate of re-operation. In this study, we describe the outcomes
of fixation using modern implants within a strategy of early return
to function. All patients treated at our centre with lateral distal femoral
locking plates (LDFLP) between 2009 and 2014 were identified. Fracture
classification and operative information including weight-bearing
status, rates of union, re-operation, failure of implants and mortality
rate, were recorded.Aims
Patients and Methods
The use of two implants to manage concomitant ipsilateral femoral
shaft and proximal femoral fractures has been indicated, but no
studies address the relationship of dynamic hip screw (DHS) side
plate screws and the intramedullary nail where failure might occur
after union. This study compares different implant configurations
in order to investigate bridging the gap between the distal DHS
and tip of the intramedullary nail. A total of 29 left synthetic femora were tested in three groups:
1) gapped short nail (GSN); 2) unicortical short nail (USN), differing
from GSN by the use of two unicortical bridging screws; and 3) bicortical
long nail (BLN), with two angled bicortical and one unicortical bridging
screws. With these findings, five matched-pairs of cadaveric femora
were tested in two groups: 1) unicortical long nail (ULN), with
a longer nail than USN and three bridging unicortical screws; and
2) BLN. Specimens were axially loaded to 22.7 kg (50 lb), and internally
rotated 90°/sec until failure.Objectives
Methods
A consecutive series of 320 patients with an
intracapsular fracture of the hip treated with a dynamic locking
plate (Targon Femoral Neck (TFN)) were reviewed. All surviving patients
were followed for a minimum of two years. During the follow-up period
109 patients died. There were 112 undisplaced fractures, of which three (2.7%) developed
nonunion or re-displacement and five (4.5%) developed avascular
necrosis of the femoral head. Revision to an arthroplasty was required
for five patients (4.5%). A further six patients (5.4%) had elective
removal of the plate and screws. There were 208 displaced fractures, of which 32 (15.4%) developed
nonunion or re-displacement and 23 (11.1%) developed avascular necrosis.
A further four patients (1.9%) developed a secondary fracture around
the TFN. Revision to a hip replacement was required for 43 patients
(20.7%) patients and a further seven (3.3%) had elective removal
of the plate and screws. It is suggested that the stronger distal fixation combined with
rotational stability may lead to a reduced incidence of complications
related to the healing of the fracture when compared with other
contemporary fixation devices but this needs to be confirmed in
further studies. Cite this article:
We investigated a new intramedullary locking
nail that allows the distal interlocking screws to be locked to
the nail. We compared fixation using this new implant with fixation
using either a conventional nail or a locking plate in a laboratory
simulation of an osteoporotic fracture of the distal femur. A total
of 15 human cadaver femora were used to simulate an AO 33-A3 fracture
pattern. Paired specimens compared fixation using either a locking
or non-locking retrograde nail, and using either a locking retrograde
nail or a locking plate. The constructs underwent cyclical loading
to simulate single-leg stance up to 125 000 cycles. Axial and torsional
stiffness and displacement, cycles to failure and modes of failure
were recorded for each specimen. When compared with locking plate
constructs, locking nail constructs had significantly longer mean
fatigue life (75 800 cycles ( The new locking retrograde femoral nail showed better stiffness
and fatigue life than locking plates, and superior fatigue life
to non-locking nails, which may be advantageous in elderly patients. Cite this article:
Intramedullary tibial nailing was performed in ten paired cadavers and the insertion of a medial-to-lateral proximal oblique locking screw was simulated in each specimen. Anatomical dissection was undertaken to determine the relationship of the common peroneal nerve to the cross-screw. The common peroneal nerve was contacted directly in four tibiae and the cross-screw was a mean of 2.6 mm (1.0 to 10.7) away from the nerve in the remaining 16. Iatrogenic injury to the common peroneal nerve by medial-to-lateral proximal oblique locking screws is therefore a significant risk during tibial nailing.
In an interdisciplinary project involving electronic
engineers and clinicians, a telemetric system was developed to measure
the bending load in a titanium internal femoral fixator. As this
was a new device, the main question posed was: what clinically relevant
information could be drawn from its application? As a first clinical
investigation, 27 patients (24 men, three women) with a mean age
of 38.4 years (19 to 66) with femoral nonunions were treated using the
system. The mean duration of the nonunion was 15.4 months (5 to
69). The elasticity of the plate-callus system was measured telemetrically
until union. Conventional radiographs and a CT scan at 12 weeks
were performed routinely, and healing was staged according to the
CT scans. All nonunions healed at a mean of 21.5 weeks (13 to 37).
Well before any radiological signs of healing could be detected,
a substantial decrease in elasticity was recorded. The relative
elasticity decreased to 50% at a mean of 7.8 weeks (3.5 to 13) and
to 10% at a mean of 19.3 weeks (4.5 to 37). At 12 weeks the mean
relative elasticity was 28.1% (0% to 56%). The relative elasticity
was significantly different between the different healing stages
as determined by the CT scans. Incorporating load measuring electronics into implants is a promising
option for the assessment of bone healing. Future application might
lead to a reduction in the need for exposure to ionising radiation
to monitor fracture healing.
Percutaneous stabilisation of tibial fractures by locking plates has become an accepted form of osteosynthesis. A potential disadvantage of this technique is the risk of damage to the neurovascular bundles in the anterior and peroneal compartments. Our aim in this anatomical study was to examine the relationship of the deep peroneal nerve to a percutaneously-inserted Less Invasive Stabilisation System tibial plate in the lower limbs of 18 cadavers. Screws were inserted through stab incisions. The neurovascular bundle was dissected to reveal its relationship to the plate and screws. In all cases, the deep peroneal nerve was in direct contact with the plate between the 11th and the 13th holes. In ten specimens the nerve crossed superficial to the plate, in six it was interposed between the plate and the bone and in the remaining two specimens it coursed at the edge of the plate. Percutaneous insertion of plates with more than ten holes is not recommended because of the risk of injury to the neurovascular structures. When longer plates are required we suggest distal exposure so that the neurovascular bundle may be displayed and protected.
This randomised study compared outcomes in patients with displaced fractures of the clavicle treated by open reduction and fixation by a reconstruction plate which was placed either superiorly or three-dimensionally. Between 2003 and 2006, 133 consecutive patients with a mean age of 44.2 years (18 to 60) with displaced midshaft fractures of the clavicle were allocated randomly to a three-dimensional (3D) (67 patients) or superior group (66). Outcome measures included the peri-operative outcome index, delayed union, revision surgery and symptoms beyond 16 weeks. CT was used to reconstruct an image of each affected clavicle and Photoshop 7.0 software employed to calculate the percentage of the clavicular cortical area in the sagittal plane. The patients were reviewed clinically and radiographically at four and 12 months after the operation. The superior plate group had a higher rate of delayed union and had more symptomatic patients than the 3D group (p <
0.05). The percentage comparisons of cortical bone area showed that cortical bone in the superior distal segment is thicker than in the inferior segment, it is also thicker in the anterior mid-section than in the posterior (p <
0.05). If fixation of midshaft fractures of the clavicle with a plate is indicated, a 3D reconstruction plate is better than one placed superiorly, because it is consistent with the stress distribution and shape of the clavicle.
Secure fixation of displaced proximal fractures of the humerus is a challenging problem. A total of 32 patients with acutely displaced three- or four-part proximal fractures of the humerus were treated by open reduction and internal fixation using the proximal humeral internal locking system (PHILOS) plate. There were 23 women and nine men with a mean age of 59.9 years (18 to 87). Data were collected prospectively and the outcomes were assessed using the Constant score. The mean follow-up was for 11 months (3 to 24). In 31 patients (97%) the fracture united clinically and radiologically at a mean of 10 weeks (8 to 24). The mean Constant score at final review was 66.5 (30 to 92). There was no significant difference in outcome when comparing patients aged more than 60 years (18 patients) with those aged less than 60 years (14 patients) ( This plate provides an alternative method of fixation for fractures of the proximal humerus. It provides a stable fixation in young patients with good-quality bone sufficient to permit early mobilisation. Failure of the screws to maintain fixation in the elderly remains a problem.