Advertisement for orthosearch.org.uk
Results 1 - 20 of 45
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 1 | Pages 37 - 42
1 Jan 2012
Affatato S Traina F De Fine M Carmignato S Toni A

Alumina–alumina bearings are among the most resistant to wear in total hip replacement. Examination of their surfaces is one way of comparing damage caused by wear of hip joints simulated in vitro to that seen in explanted bearings. The aim of this study was to determine whether second-generation ceramic bearings exhibited a better pattern of wear than those reported in the literature for first-generation bearings. We considered both macro- and microscopic findings.

We found that long-term alumina wear in association with a loose acetabular component could be categorised into three groups. Of 20 specimens, four had ‘low wear’, eight ‘crescent wear’ and eight ‘severe wear’, which was characterised by a change in the physical shape of the bearing and a loss of volume. This suggests that the wear in alumina–alumina bearings in association with a loose acetabular component may be variable in pattern, and may explain, in part, why the wear of a ceramic head in vivo may be greater than that seen after in vitro testing.


The Bone & Joint Journal
Vol. 103-B, Issue 8 | Pages 1345 - 1350
1 Aug 2021
Czubak-Wrzosek M Nitek Z Sztwiertnia P Czubak J Grzelecki D Kowalczewski J Tyrakowski M

Aims. The aim of the study was to compare two methods of calculating pelvic incidence (PI) and pelvic tilt (PT), either by using the femoral heads or acetabular domes to determine the bicoxofemoral axis, in patients with unilateral or bilateral primary hip osteoarthritis (OA). Methods. PI and PT were measured on standing lateral radiographs of the spine in two groups: 50 patients with unilateral (Group I) and 50 patients with bilateral hip OA (Group II), using the femoral heads or acetabular domes to define the bicoxofemoral axis. Agreement between the methods was determined by intraclass correlation coefficient (ICC) and the standard error of measurement (SEm). The intraobserver reproducibility and interobserver reliability of the two methods were analyzed on 31 radiographs in both groups to calculate ICC and SEm. Results. In both groups, excellent agreement between the two methods was obtained, with ICC of 0.99 and SEm 0.3° for Group I, and ICC 0.99 and SEm 0.4° for Group II. The intraobserver reproducibility was excellent for both methods in both groups, with an ICC of at least 0.97 and SEm not exceeding 0.8°. The study also revealed excellent interobserver reliability for both methods in both groups, with ICC 0.99 and SEm 0.5° or less. Conclusion. Either the femoral heads or acetabular domes can be used to define the bicoxofemoral axis on the lateral standing radiographs of the spine for measuring PI and PT in patients with idiopathic unilateral or bilateral hip OA. Cite this article: Bone Joint J 2021;103-B(8):1345–1350


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_12 | Pages 58 - 58
1 Oct 2019
Mirick GM Sabin A Hansen G Lindgren B Aparicio C Carlson CS Bue M Larsen O Schmidt AH Kyle R Gustilo RB Tsukayama D Bechtold JE
Full Access

Introduction. We studied free (= local powder) tobramycin and doxycycline, and controlled release (= local lipid bilayer) doxycycline formulations in a rat model representing a generic joint infection. We . hypothesized. that evidence of infection (quantitative colony forming units (CFU), qualitative SEM, histopathology) (1a) would be reduced with local vs. systemic antibiotic, (1b) any antibiotic would be superior to control (2) there would be a difference among antibiotics, and (3) antibiotic would not be detectable in serum at 4-week euthanasia. Methods. Study groups. included infected and non-infected (1) control, (2) systemic ceftriaxone (daily), (3) local tobramycin, (4) local doxycycline and (5) controlled release doxycycline. With IACUC approval, (10 rats/group; power =0.8), 50-μl, 10×4 CFU Staphylococcus aureus, slowly injecting into distal femoral medullary canal, reliably created joint infection. Antibiotic formulation was introduced locally into cavity and joint, pin was inserted, and tissues closed. After 4-weeks, serum, pin, bone and synovium were obtained. CFU/ml of bone and synovium were quantified using macrotiter method. SEM imaged biofilm on surface of pin, histopathology identified tissue response, liquid chromatography/mass spectrometry measured plasma antibiotic. Kruskal-Wallis one-way ANOVA compared groups. Results. Groups receiving antibiotic reported lower CFU/ml in synovium compared with control (no treatment) group (1b), but there was no difference between systemic, free or controlled antibiotics (1a). Different results with different antibiotics were shown, with free tobramycin reducing CFU/ml to a greater extent than free doxycycline in the synovium (2) (p<0.05). Antibiotic in plasma was nondetectable all groups (3). SEM revealed some biofilm on pin in all groups. . Limitations. include inoculation method, single observation period, administration of only one bacterial and antibiotic dose, and not including pairing local and systemic antibiotic. Conclusion. There was no difference in infection reduction nor detectable antibiotic in serum for any antibiotic formulation, but CFU's in synovium differed based on antibiotic formulation. For any tables or figures, please contact the authors directly


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_12 | Pages 57 - 57
1 Oct 2019
Gil D Grindy S Hugard S Muratoglu OK Oral E
Full Access

Introduction. Ultra-high molecular weight polyethylene (UHMWPE) can provide local sustained delivery of therapeutics. 1,2. For example, it can deliver analgesics to address post-arthroplasty pain. 2. Given that several analgesics, such as bupivacaine (anesthetic) and tolfenamic acid (NSAID), were shown to possess antibacterial activity against Staphylococci, we hypothesize that analgesic-loaded UHMWPE can also yield antimicrobial effects, preventing the development of periprosthetic joint infections. Methods. Bupivacaine and tolfenamic acid were incorporated into UHMWPE via phase-separated compression molding. Drug release from the prepared samples was measured using high-performance liquid chromatography. Antibacterial studies of the obtained materials were conducted against methicillin-sensitive, and methicillin-resistant S. aureus, as well as S. epidermidis. Time-kill curves were obtained to characterize antimicrobial activity against planktonic bacteria. The dynamics of bacterial adhesion were assessed to characterize antibiofilm activity. Scanning electron microscopy (SEM) was used to visualize adherent bacteria. Anticolonizing activity of the tested materials was characterized using the “daughter cell” method as outlined elsewhere. 3. Cytotoxicity profile of drug-loaded UHMWPEs was evaluated using MG-63 osteoblast cell line. Results. The bupivacaine release rate generally increased with increasing drug loading (e.g. a model knee implant loaded with bupivacaine would release ca. 15–500 mg over 24 hours). While also proportional, drug release from UHMWPE loaded with tolfenamic acid was much lower. The bacterial viability curves showed that bupivacaine-loaded UHMWPE possessed moderate antibacterial activity against planktonic MSSA, MRSA, and S. epidermidis, slowing bacteria proliferation by up to 70%. Bupivacaine-loaded UHMWPE also mitigated biofilm formation and development during the initial culture period. SEM images confirmed the observed antibiofilm effect (Fig. 1). Tolfenamic acid-loaded UHMWPE allowed proliferation of planktonic bacteria. At the same time, these materials showed pronounced dose-dependent anticolonizing activity against tested strains, providing 3-log reduction of “daughter” cells. Bupivacaine- and tolfenamic acid-loaded UHMWPEs showed little-to-no cytotoxicity against osteoblasts. Discussion & Conclusions. We demonstrated for the first time that bupivacaine-loaded UHMWPE possesses dose-dependent antibacterial properties against planktonic and adherent MSSA, MRSA, and S. epidermidis – pathogens commonly associated with periprosthetic joint infections. Pronounced anticolonizing activity was evident for tolfenamic acid-loaded UHMWPE. Due to the low solubility of tolfenamic acid, the material's antibacterial effect against planktonic bacteria was lower. These results demonstrate that analgesic-loaded UHMWPE, used as a tool in multimodal pain management, can also yield antibacterial effects, opening an entirely new avenue for providing post-arthroplasty antibacterial prophylaxis. This pioneering approach has a potential to reduce patients' morbidity and mortality after arthroplasty. For any tables or figures, please contact the authors directly


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 760 - 768
1 Jun 2011
ten Broeke RHM Alves A Baumann A Arts JJC Geesink RGT

Four uncemented Symax hip stems were extracted at three weeks and nine, 13 and 32 months, respectively, for reasons other than loosening. The reasons for implant removal were infection in two cases, recurrent dislocation in one and acetabular fracture in one. They were analysed to assess the effect and behaviour of an electrochemically deposited, completely resorbable biomimetic BONIT-hydroxyapatite (HA) coating (proximal part) and a DOTIZE surface treatment (distal part) using qualitative histology, quantitative histomorphometry and scanning electron microscopy (SEM). Early and direct bone-implant bonding with signs of active remodelling of bone and the HA coating were demonstrated by histology and SEM. No loose BONIT-HA particles or delamination of the coating were observed, and there was no inflammation or fibrous interposition at the interface. Histomorphometry showed bone-implant contact varying between 26.5% at three weeks and 83.5% at 13 months at the HA-coated implant surface. The bone density in the area of investigation was between 24.6% at three weeks and 41.1% at 32 months. The DOTIZE surface treatment of the distal part of the stem completely prevented tissue and bone apposition in all cases, thereby optimising proximal stress transfer. The overall features of this implant, in terms of geometry and surface texture, suggest a mechanically stable design with a highly active biomimetic coating, resulting in rapid and extensive osseo-integration, exclusively in the metaphyseal part of the stem. Early remodelling of the HA coating does not seem to have a detrimental effect on short-term bone-implant coupling. There were no adverse effects identified from either the BONIT-HA coating or the DOTIZE surface treatment


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_12 | Pages 64 - 64
1 Oct 2019
McCalden RW Decker MM Klassen RJ Lanting B
Full Access

Introduction. HXLPE acetabular liners were introduced to reduce wear-related complications in THA. However, post-irradiation thermal free radical stabilization can compromise mechanical properties, leave oxidation-prone residual free radicals, or both. Reports of mechanical failure of HXLPE acetabular liner rims raise concerns about thermal free radical stabilization and in vivo oxidization on implant properties. The purpose of this study is to explore the differences in the mechanical, physical and chemical properties of HXLPE acetabular liner rims after extended time in vivo between liners manufactured with different thermal free radical stabilization techniques. Material and methods. Remelted, single annealed and sequentially annealed retrieved HXLPE acetabular liners with in vivo times greater than 4.5 years were obtained from our implant retrieval laboratory. All retrieved liners underwent an identical sanitation and storage protocol. For mechanical testing, a total of 55 explants and 13 control liners were tested. Explant in vivo time ranged from 4.6 – 14.0 years and ex vivo time ranged from 0 – 11.6 years. Rim mechanical properties were tested by microindentation hardness testing using a Micromet II Vickers microhardness tester following ASTM standards. A subset of 16 explants with ex vivo time under one year along with five control liners were assessed for oxidation by FTIR, crystallinity by Raman spectroscopy, and evidence of microcracking by SEM. Results. No significant difference in in vivo or ex vivo was found between thermal stabilization groups in either set of explants studied. In the mechanically tested explants, there was no significant correlation between in vivo time and Vickers hardness in any thermal stabilization group. A significant correlation was found between ex vivo time and hardness in remelted liners (Δ=.520, p=.011), but not in either annealed cohort. ANCOVA with ex vivo time as a covariate found a significant difference in hardness between the thermal free radical stabilization groups (p<.0005, η. 2. = 0.322). Post hoc analysis revealed hardness was significantly lower in the retrieved remelted group compared to both the single annealed (p=.001) and sequentially annealed (p<.0005) cohorts. Hardness was significantly higher in the retrieved remelted liners compared to controls (p=.007), with no different in either annealed cohort (figure 1). Detectable subsurface oxidation (OI > 0.1) was found in retrieved remelted (25%), single annealed (100%) and sequentially annealed (75%) liner rims (figure 2). Crystallinity was increased in the subsurface region relative to control liners for both annealed, but not remelted, liner rims. Hardness was increased in oxidized rims for both annealed cohorts but not in the remelted cohort. Microcracking was only found along the surface of one unoxidized remelted liner rim. Conclusion. Mechanical properties were reduced at baseline and worsened after in vivo time for remelted HXLPE liner rims. Rim oxidation was detected in all groups. Oxidation was associated with increased crystallinity and hardness in annealed cohorts, but not remelted liners. Increased crystallinity and oxidation do not appear to be directly causing the worsened mechanical behavior of remelted HXLPE liner rims after extended in vivo time. For any tables or figures, please contact the authors directly


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1238 - 1246
1 Jul 2021
Hemmerling KJ Weitzler L Bauer TW Padgett DE Wright TM

Aims

Dual mobility implants in total hip arthroplasty are designed to increase the functional head size, thus decreasing the potential for dislocation. Modular dual mobility (MDM) implants incorporate a metal liner (e.g. cobalt-chromium alloy) in a metal shell (e.g. titanium alloy), raising concern for mechanically assisted crevice corrosion at the modular liner-shell connection. We sought to examine fretting and corrosion on MDM liners, to analyze the corrosion products, and to examine histologically the periprosthetic tissues.

Methods

A total of 60 retrieved liners were subjectively scored for fretting and corrosion. The corrosion products from the three most severely corroded implants were removed from the implant surface, imaged using scanning electron microscopy, and analyzed using Fourier-transform infrared spectroscopy.


Bone & Joint Open
Vol. 2, Issue 9 | Pages 765 - 772
14 Sep 2021
Silitonga J Djaja YP Dilogo IH Pontoh LAP

Aims

The aim of this study was to perform a cross-cultural adaptation of Oxford Hip Score (OHS) to Indonesian, and to evaluate its psychometric properties.

Methods

We performed a cross-cultural adaptation of Oxford Hip Score into Indonesian language (OHS-ID) and determined its internal consistency, test-retest reliability, measurement error, floor-ceiling effect, responsiveness, and construct validity by hypotheses testing of its correlation with Harris Hip Score (HHS), vsual analogue scale (VAS), and Short Form-36 (SF-36). Adults (> 17 years old) with chronic hip pain (osteoarthritis or osteonecrosis) were included.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_13 | Pages 7 - 7
1 Oct 2018
Tanzer M
Full Access

Introduction. Porous surfaces developed over the past decades have been shown to promote tissue ingrowth. Hydroxyapatite (HA) coatings have been added to these porous coatings in an attempt to further augment bone ingrowth. The development of additive manufacturing techniques has allowed for precision in building these complex porous structures. The effect of supplemental HA coatings on these new surfaces is unclear. The purpose of this study is to evaluate the biological fixation of a novel 3D printed porous implant in a canine model. In addition, we evaluated the effect of different HA coatings on this 3D printed implant. Methods. A canine transcortical model was used to evaluate the performance of three different laser rapid manufacturing (LRM) Ti6Al4V cylindrical implants (5.2 mm diameter, 10mm length): LRM with precipitated hydroxyapatite (P-HA), LRM with plasma sprayed hydroxyapatite (PS-HA), and a hydroxyapatite-free control (No-HA). The implants were 50–60% porous with a mean pore size of 450 μm and have a random interconnected architecture with irregular pore sizes and shapes that are designed based on the structure of cancellous bone. A lateral approach to the femoral diaphysis was used to prepare 5 mm unicortical, perpendicular drill holes in 12 canines. One of each implant type was press-fit into each femur. The femora were harvested at both 4 and 12 weeks post implantation, radiographed and prepared for either mechanical push-out testing to assess the shear strength of the bone-implant interface (left femora, N=6) or for histological processing (right femora, N=6). An un-paired Student's t-test was used to compare statistical significance between the 4 and 12-week results, as well as differences due to implant type; p<0.05 was considered significant. Results. The post-mortem contact radiographs demonstrated substantial condensation of bone around the implants at both 4 and 12 weeks. Bone ingrowth in the canine femora was observed in all implants, with and without HA, at both time periods under backscattered SEM. The mean extent of bone ingrowth at 4 weeks for no-HA, P-HA, and PS-HA implants was 41.5% (95% CI 32.5 to 50.6), 51.0% (95% CI 45.2 to 56.8) and 53.2% (95% CI 41.6 to 64.7), respectively. The mean extent of bone ingrowth at 12 weeks for no-HA, P-HA, and PS-HA implants was 64.4% (95% CI 61.5 to 67.3), 59.9% (95% CI 51.9 to 67.8) and 64.9% (95% CI 58.2 to 71.6), respectively. There was no significant difference in the amount of bone ingrowth between the HA and non-HA coated implants at any of the time points. All the implants were successfully pushed out after 4 weeks of implantation. The mean shear strength from the push-out test at 4 weeks for the no-HA, P-HA, and PS-HA implants was calculated to be 21.6 MPa (95% CI 17.2 to 26.0), 20.7 MPa (95% CI 18.9 to 22.4), and 20.2 MPa (95% CI 16.3 to 24.2), respectively. At week 12, in two femora all three implant types had compressive failure before rupture of the bone-implant interface with a load of over 2000N. This suggests that the values of shear strength were higher than those calculated from the successful tests at 12 weeks. The mean shear strength for the remaining no-HA, P-HA and PS-HA implants at 12 weeks was calculated to be 39.9 MPa (95% CI 29.8 to 50.9), 33.7 MPa (95% CI 26.3 to 41.2), and 36.0 MPa (95% CI 29.53 to 42.4), respectively. For all implants, the mean shear strength at 12 weeks was statistically significantly greater than at 4 weeks (p<0.05). There was no significant difference in the shear strength between HA coated and non-HA coated implants at 4 or 12 weeks. Conclusion. At 4 and 12 weeks, all non-HA coated LRM Ti6Al4V implants consistently exhibited very high bone ingrowth and mechanical shear strength in the canine model. These results demonstrate that this novel additive manufactured porous implant promoted biological fixation in a canine model. There was no significant improvement in the extent of bone ingrowth with the addition of HA. This is in agreement with the literature indicating that topography is the dominant factor governing bone apposition to hydroxyapatite-coated implants. It is likely that in this model, the morphologic features and roughness of the surface of the LRM implants stimulated osteoblastic activity, so that the addition of HA had a non-significant effect


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_12 | Pages 49 - 49
1 Nov 2015
Karlakki S Graham N Banergee R Hamad A Budhithi C Whittall C
Full Access

Introduction. Hip and knee replacements are predictable orthopaedic procedure with excellent clinical outcomes. Discharging or leaking wounds affect length of hospital stay, affect bed planning and predispose to superficial and potentially deep wound infection. Predictable wound healing therefore remains the first hurdle. This trial aims to study the effectiveness of portable disposable incisional negative pressure wound therapy (NPWT) dressings in hip and knee replacements. This trial aims to study the effectiveness of portable disposable incisional negative pressure wound dressings in hip and knee replacements and the impact on wound healing, length of stay and wound complications. Patients/Materials & Methods. Following ethical approval 110 patients each were randomised to ‘Control group’ and ‘Study group’. Patients in control group received traditional dressings and those in study group received an incisional NPWT (PICO) manufactured by Smith & Nephew. Post operatively, state of the wound, level of wound exudate, length of hospital stay and complications were documented. Results. The average length of stay for control group was 4.72 ± 0.69 (Mean ± SEM) and for the study group 3.79±0.19 (mean ± SEM. Eight percent of the control group patients had wound related complications as opposed to 2% complications in the study group. The mechanical and device related issues in the study group accounted to 10%. Discussion. There is little evidence as to whether wound dressing choices affect wound healing in incisional wounds. There is good evidence for negative pressure wound therapy (NPWT) in complex, open and chronic wounds; however the concept of application of NPWT to routine incisional wounds is novel. Advances such as disposable and portable negative devices have made this option attractive in elective surgery. Conclusion. Our study shows that its application in routine primary hip and knee replacement can improve wound healing, facilitate early and predictable discharge from the hospital and reduce post discharge complications. This is particularly true for high risk wounds such as patients with BMI of >35 and is cost effective


Bone & Joint Open
Vol. 2, Issue 6 | Pages 443 - 456
28 Jun 2021
Thompson JW Corbett J Bye D Jones A Tissingh EK Nolan J

Aims

The Exeter V40 cemented polished tapered stem system has demonstrated excellent long-term outcomes. This paper presents a systematic review of the existing literature and reports on a large case series comparing implant fractures between the Exeter V40 series; 125 mm and conventional length stem systems.

Methods

A systematic literature search was performed adhering to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) criteria. In parallel, we performed a retrospective single centre study of Exeter V40 femoral stem prosthetic fractures between April 2003 and June 2020.


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 7 | Pages 996 - 1000
1 Sep 2001
Haraguchi K Sugano N Nishii T Miki H Oka K Yoshikawa H

We report two cases of surface deterioration of a zirconia ceramic femoral head associated with phase transformation after total hip arthroplasty. One head was retrieved at revision due to recurrent dislocation after six years and the other because of failure of the locking mechanism of the polyethylene liner after three years. The monoclinic content of the zirconia ceramics rose from 1% to about 30% on the surface of the heads. SEM revealed numerous craters indicating extraction of the zirconia ceramics at the surface. Surface roughness increased from an initial value of 0.006 3m up to 0.12 3m. This is the first report to show that phase transformation of zirconia ceramics causes deterioration of the surface roughness of the head in vivo after total hip arthroplasty


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 1 | Pages 48 - 53
1 Jan 1998
Ioannidis TT Zacharakis N Magnissalis EA Eliades G Hartofilakidis G

We report the long-term radiological results of 58 total hip arthroplasties (THA) using the Charnley offset-bore acetabular socket. Wear was measured at four sites and radiolucent lines and possible migration were recorded. Four cups were retrieved at revision and were examined using light microscopy, SEM and X-ray microanalysis. At a mean follow-up of seven years the mean wear in the DeLee and Charnley zone I was 0.4 mm and in zone II 0.26 mm. The wear rate was 0.06 and 0.04 mm/year, respectively. Progression of radiolucent lines was seen in five cases (8.6%). Three sockets (5.2%) were revised because of aseptic migration at a mean follow-up of 9.8 years and one socket for infection at two years. The offset-bore acetabular cup had excellent wear behaviour and a low migration and revision rate. We recommend that it should be considered in THA since the use of small cups is increasing, particularly in revision cases


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 5 | Pages 641 - 646
1 Jul 2002
Lai K Shen W Chen C Yang C Hu W Chang G

Between March 1990 and May 1991 we performed 85 primary total hip replacements in 74 patients using the Landos Atoll hydroxyapatite (HA)-coated cup and the Corail HA-coated stem. The patients were followed up for a mean of ten years. Of the 85 cups, 26 (31%) have already been revised and a further six are radiologically unstable and awaiting revision. Two femoral stems have been revised for infection without loosening. The retrieved acetabular cups were studied by SEM and image-processing techniques to quantify the amount of residual HA on the cup. This was correlated with the clinical variables and modes of failure. The residual HA (as a percentage of the surface) on the loose cups correlated negatively with the duration of implantation (r = −0.732, p < 0.001). Six cups were stable at revision and had more residual HA coating than those which were loose (p < 0.01). The rate of failure of the Landos Atoll HA-coated, smooth hemispherical cup with screw fixation is unacceptably high. Resorption of the HA coating is markedly increased in loose cups compared with stable cups. HA coating cannot substitute for stable mechanical fixation


Bone & Joint Research
Vol. 8, Issue 6 | Pages 246 - 252
1 Jun 2019
Liddle A Webb M Clement N Green S Liddle J German M Holland J

Objectives

Previous studies have evidenced cement-in-cement techniques as reliable in revision arthroplasty. Commonly, the original cement mantle is reshaped, aiding accurate placement of the new stem. Ultrasonic devices selectively remove cement, preserve host bone, and have lower cortical perforation rates than other techniques. As far as the authors are aware, the impact of ultrasonic devices on final cement-in-cement bonds has not been investigated. This study assessed the impact of cement removal using the Orthosonics System for Cemented Arthroplasty Revision (OSCAR; Orthosonics) on final cement-in-cement bonds.

Methods

A total of 24 specimens were manufactured by pouring cement (Simplex P Bone Cement; Stryker) into stainless steel moulds, with a central rod polished to Stryker Exeter V40 specifications. After cement curing, the rods were removed and eight specimens were allocated to each of three internal surface preparation groups: 1) burr; 2) OSCAR; and 3) no treatment. Internal holes were recemented, and each specimen was cut into 5 mm discs. Shear testing of discs was completed by a technician blinded to the original grouping, recording ultimate shear strengths. Scanning electron microscopy (SEM) was completed, inspecting surfaces of shear-tested specimens.


Bone & Joint Research
Vol. 9, Issue 8 | Pages 484 - 492
1 Aug 2020
Zhang W Fang X Shi T Cai Y Huang Z Zhang C Lin J Li W

Aims

To explore the effect of different types of articulating antibiotic-loaded cement spacers in two-stage revision for chronic hip prosthetic joint infection (PJI).

Methods

A retrospective cohort study was performed involving 36 chronic PJI patients treated with different types of articulating antibiotic-loaded cement spacers between January 2014 and December 2017. The incidence of complications and the therapeutic effects of different types of antibiotic-loaded articulating cement spacers were compared.


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 116 - 121
1 Jul 2020
Heise G Black CM Smith R Morrow BR Mihalko WM

Aims

This study aimed to determine if macrophages can attach and directly affect the oxide layers of 316L stainless steel, titanium alloy (Ti6Al4V), and cobalt-chromium-molybdenum alloy (CoCrMo) by releasing components of these alloys.

Methods

Murine peritoneal macrophages were cultured and placed on stainless steel, CoCrMo, and Ti6Al4V discs into a 96-well plate. Cells were activated with interferon gamma and lipopolysaccharide. Macrophages on stainless steel discs produced significantly more nitric oxide (NO) compared to their control counterparts after eight to ten days and remained elevated for the duration of the experiment.


The Bone & Joint Journal
Vol. 100-B, Issue 7 | Pages 882 - 890
1 Jul 2018
Bertrand J Delfosse D Mai V Awiszus F Harnisch K Lohmann CH

Aims

Early evidence has emerged suggesting that ceramic-on-ceramic articulations induce a different tissue reaction to ceramic-on-polyethylene and metal-on-metal bearings. Therefore, the aim of this study was to investigate the tissue reaction and cellular response to ceramic total hip arthroplasty (THA) materials in vitro, as well as the tissue reaction in capsular tissue after revision surgery of ceramic-on-ceramic THAs.

Patients and Methods

We investigated tissue collected at revision surgery from nine ceramic-on-ceramic articulations. we compared our findings with tissue obtained from five metal-on-metal THA revisions, four ceramic-on-polyethylene THAs, and four primary osteoarthritis synovial membranes. The latter were analyzed to assess the amount of tissue fibrosis that might have been present at the time of implantation to enable evaluation, in relation to implantation time, of any subsequent response in the tissues.


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 62 - 67
1 Jun 2019
Tanzer M Chuang PJ Ngo CG Song L TenHuisen KS

Aims

The purpose of this study was to evaluate the biological fixation of a 3D printed porous implant, with and without different hydroxyapatite (HA) coatings, in a canine model.

Materials and Methods

A canine transcortical model was used to evaluate the characteristics of bone ingrowth of Ti6Al4V cylindrical implants fabricated using laser rapid manufacturing (LRM). At four and 12 weeks post-implantation, we performed histological analysis and mechanical push-out testing on three groups of implants: a HA-free control (LRM), LRM with precipitated HA (LRM-PA), and LRM with plasma-sprayed HA (LRM-PSHA).


The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1304 - 1312
1 Oct 2017
Langton DJ Sidaginamale RP Joyce TJ Meek RD Bowsher JG Deehan D Nargol AVF Holland JP

Aims

We sought to determine whether cobalt-chromium alloy (CoCr) femoral stem tapers (trunnions) wear more than titanium (Ti) alloy stem tapers (trunnions) when used in a large diameter (LD) metal-on-metal (MoM) hip arthroplasty system.

Patients and Methods

We performed explant analysis using validated methodology to determine the volumetric material loss at the taper surfaces of explanted LD CoCr MoM hip arthroplasties used with either a Ti alloy (n = 28) or CoCr femoral stem (n = 21). Only 12/14 taper constructs with a rough male taper surface and a nominal included angle close to 5.666° were included. Multiple regression modelling was undertaken using taper angle, taper roughness, bearing diameter (horizontal lever arm) as independent variables. Material loss was mapped using a coordinate measuring machine, profilometry and scanning electron microscopy.