Cemented total hip arthroplasty yields reliable results in short to medium term studies, but aseptic loosening remains a problem in long-term follow up, especially in young and active patients. Aseptic loosening has been related to wear, and in order to minimize wear various alternatives to the traditional metal on polyethylene have been proposed. Both ceramic on polyethylene (COP) and metal on metal (MOM) have been shown to produce less wear than metal on polyethylene (MOP). In order to study the effect of the bearing, we have utilized identical stems and cups while comparing the different bearings. Methods and material. 396 hips were randomized to MOP, COP or MOM using a cemented triple tapered polished stem (MS-30; Sulzer Orthopedics) with a cemented polyethylene cup (Weber; Sulzer, Orthopedics) and a 28 millimeter head. For the MOP and COP articulations, a all-polyethylene cup was used with a Protasul™ metal head or a Sulox™ alumina head, whereas a polyethylene cup with metal insert was used for the MOM articulations (Weber Polyethylene Cup with Metasul™ Insert) with a Metasul™ metal head. Harris Hip Score (HHS) and radiological evaluation was performed after two, five and seven years. Ethical approval was obtained. Results. HHS was available for 338 hips after seven years. The HHS in the MOP group (116 hips) was 93.7 (SD 9.0), 93.5 in the COP group (112 hips) (SD 8.8), and 91.0 (SD 13.4) in the MOM group (110 hips). Radiographic evaluation was available for 335 hips.
Introduction.