Patients suffering from finger joint pain or dysfunction due to arthritis and traumatic injury may require arthroplasty and joint replacement. Single-part silicone-based implants remain the material of choice and most widely used option, although reports on their long-term clinical performance are variable. For trauma indications, patients have a high expectation of functionality necessitating the use of materials with high wear resistance and mechanical performance. A new proximal inter phalangeal (PIP) joint designed by Zrinski AG (Wurmlingen, Germany), comprising a self-mating carbon fibre reinforced polyetheretherketone (CFR-PEEK) coupling, may provide a suitable alternative. Here we describe the wear performance of the CFR-PEEK components in a PIP joint wear simulator and subsequent characterisation of the wear particles. Four proximal and distal PIP components were milled (Zrinski AG) from CFR-PEEK (Invibio Ltd, UK) and subjected to wear testing (Endo Lab ® GmbH, Germany). The test was conducted at 37°C over 5 million cycles in 25% bovine serum (refreshed every 0.5 million cycles). The load was a static force of 63N applied at a frequency of 1Hz with a flexion/extension angle of ±40°. Wear rate was determined by mass loss from each component. Pooled serum samples from the wear simulator were subjected to protein digest and the remaining particulate debris isolated by serial filtration through 10μm, 1μm and 0.1μm filters. Particle size and morphology was subsequently determined by scanning electron microscopy (SEM) (Continuum Blue, UK).Introduction
Methods
A total of 179 adult patients with displaced intra-articular fractures of the distal radius was randomised to receive indirect percutaneous reduction and external fixation (n = 88) or open reduction and internal fixation (n = 91). Patients were followed up for two years. During the first year the upper limb musculoskeletal function assessment score, the SF-36 bodily pain sub-scale score, the overall Jebsen score, pinch strength and grip strength improved significantly in all patients. There was no statistically significant difference in the radiological restoration of anatomical features or the range of movement between the groups. During the period of two years, patients who underwent indirect reduction and percutaneous fixation had a more rapid return of function and a better functional outcome than those who underwent open reduction and internal fixation, provided that the intra-articular step and gap deformity were minimised.