Advertisement for orthosearch.org.uk
Results 1 - 20 of 20
Results per page:
Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 45 - 45
23 Jun 2023
Lieberman JR
Full Access

Modular dual mobility (DM) articulations are increasingly utilized during total hip arthroplasty (THA). However, concerns remain regarding the metal liner modularity. This study aims to correlate metal artifact reduction sequence (MARS) magnetic resonance imaging (MRI) abnormalities with serum metal ion levels in patients with DM articulations.

All patients with an asymptomatic, primary THA and DM articulation with >2-year follow-up underwent MARS-MRI of the operative hip. Each patient had serum cobalt, chromium, and titanium levels drawn. Patient satisfaction, Oxford Hip Score, and Forgotten Joint Score-12 (FJS-12) were collected. Each MARS-MRI was independently reviewed by fellowship-trained musculoskeletal radiologists blinded to serum ion levels.

Forty-five patients (50 hips) with a modular DM articulation were included with average follow-up of 3.7±1.2 years. Two patients (4.4%) had abnormal periprosthetic fluid collections on MARS-MRI with cobalt levels >3.0 μg/L. Four patients (8.9%) had MARS-MRI findings consistent with greater trochanteric bursitis, all with cobalt levels < 1.0 μg/L. A seventh patient had a periprosthetic fluid collection with normal ion levels. Of the 38 patients without MARS-MRI abnormalities, 37 (97.4%) had cobalt levels <1.0 μg/L, while one (2.6%) had a cobalt level of 1.4 μg/L. One patient (2.2%) had a chromium level >3.0 μg/L and a periprosthetic fluid collection. Of the 41 patients with titanium levels, five (12.2%) had titanium levels >5.0 μg/L without associated MARS-MRI abnormalities.

Periprosthetic fluid collections associated with elevated serum cobalt levels in patients with asymptomatic dual mobility articulations occur infrequently (4.4%), but further assessment of these patients is necessary.

Level of Evidence: Level IV


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_5 | Pages 37 - 37
1 Jul 2020
Poznalska G Auckland C Heddon S Phillips J Wilson M
Full Access

Joint replacement is a life-enhancing, cost-effective surgical intervention widely used to treat disabling joint pain mainly caused by osteoarthritis. Hip and knee joint replacements are common, highly successful operations bringing many patients relief from pain, and improve mobility.

Prosthetic joint infections (PJI) are often difficult to diagnose; diagnosis often relying on a combination of clinical findings, microbiological data and histological evaluation of periprosthetic tissue. The majority of recent studies demonstrate a higher sensitivity for the culture of sonication fluid (62–94%) than periprosthetic tissue (55–88%).

The Royal Devon & Exeter NHS Foundation Trust provided a specialist multidisciplinary team. We evaluated the use of sonication for identifying pathogens during revision hip and knee replacement in our unit with the intention of applying for further grants to study this technology in more detail; it was a diagnostic feasibility study. The target recruitment was 50 patients; 25 consecutive patients undergoing revision hip arthroplasty for any reason, and 25 undergoing revision knee arthroplasty for any reason. The majority of patients were identified in outpatient clinics.

Our results show that tissue culture remains more sensitive than sonication. The sensitivity for direct sonication was 75% compared with 80% for tissue culture; sensitivity of enriched sonication was 80 %, compared to the unit results for tissue culture of 83%. The combined sensitivity of tissue culture was better as a higher number of tissue specimens were obtained.

The different methodology might explain reported differences between this and other studies; however, our study does not support the use of sonication as the only tool in the diagnosis of PJI.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_12 | Pages 26 - 26
1 Jun 2017
Woodnutt D Hickey B Mullins M Dodd M Davies A Mohammed A
Full Access

The ODEP (Orthopaedic Data Evaluation Panel) rating system should offer a surgeon and patient extra information when making a choice on which implant to use. However, in the current economic environment, ratings may also influence implant choice by contracting bodies. Our aim was to determine the performance of commonly used Acetabular and femoral components in our unit and compare these to their published ODEP ratings (or absence of rating).

We analysed all of the following primary THR components (12,792) for revision for any reason, using same date ranges as ODEP where more than 100 implantations had occurred. Hip components: Trinity (3A in 2013), Trilogy (10A* in 2016), Atlas (10A in 2013), Trilogy TMT (10A 2010) Durom (not rated), BHR (10A, 2010), ACCIS (not rated); Femoral components: Taperfit (10A in 2013), Taperloc (10A* in 2016), Metafix (3A in 2013), CPT (10A in 2012), Ecofit (not Rated), ESOP (not rated), Minihip (3A 2013), Durom (not rated), BHR (10A 2010), ACCIS (not rated).

Analysis of Kaplan Meier survival curves was undertaken for all components. The rated components and non-rated components were compared using HR and logrank tests for all time groups when ratings were introduced. No statistical difference was observed in any group except for the Trinity cup which had a 98.2% (1344 cups) survival at 6 years.

Component survival in our unit was better than ODEP suggested failure for A category of not more than 1% per year, for all components.

Whilst we applaud the intention to improve data available for prostheses, the present ODEP system does not distinguish between performances of different implants in our unit. We therefore recommend care when relying upon ODEP ratings to make clinical or contracting decisions.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_11 | Pages 19 - 19
1 Aug 2018
van der Jagt DR Brekon A Mokete L Pietrzak J Nortje M Schepers A
Full Access

A longitudinal study was done assessing the latest radiographs available in a series of collarless Corail uncemented stems which have reported survival rates of 95% at 20 years. Parameters scored included the degree of “Ghosting” or delamination, the Gruen, the stem fit in the femoral canal and the degree of calcar resorption. Patient and implant demographics were noted.

At 3 years the loosening rate was 23% reaching 64% at 6 years after the index procedure. It was present in all age groups, with a peak in the 24 to 49 year age group. Males and females had the same occurrence. When it was present this was always in zone 1. It was present in 6% of patients in zone 7, but then always associated with zone 1 changes. High and standard off-set stems had the same loosening rates. The looser fit of the stem the higher the incidence of loosening. There was no correlation to the type of bearing surface or the degree of calcar resorption. Those patients with a BMI of 25 – 35 had lower loosening rates compared to those with higher or lower BMI's.

We postulate that cancellous bone on-growth onto the hydroxyapatite coating associated with loading flexural micro-motion leads to hydroxyapatite being pulled off the smooth stem substrate in zone 1. Progressive delamination of the hydroxyapatite then occurs. The triple-tapered design though imparts continued stability.

We report high loosening rates in the Corail stem and suggest a mechanism for its development.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_1 | Pages 52 - 52
1 Jan 2018
Devane P
Full Access

Total hip joint replacement (THJR) for high riding congenital hip dislocation (CDH) is often performed in young patients, and presents unique problems with acetabular cup placement and leg length inequality.

A database and the NZ Joint Registry were used to identify 76 hips in 57 patients with a diagnosis of CDH who underwent THJR in the Wellington region between 1994 and 2015. Records and radiographs of 46 hips in 36 patients classified pre-operatively as Crowe II, III or IV were reviewed. Surgical technique used a direct lateral approach, the uncemented acetabular component was located in the anatomic hip center and a primary femoral stem was used in all but one hip. Whether a step-cut sub-trochanteric femoral osteotomy was performed depended on degree of correction, tension on the sciatic nerve, and restoration of leg length.

For the 36 patients classified as Crowe II or higher, the average age at operation was 44 years (26 – 66), female:male ratio was 4.5:1 and follow-up averaged 10 years (2 – 22.3). Of the 15 hips classified as Crowe IV, 10 required a step-cut sub-trochanteric femoral osteotomy to shorten the femur, but 5 were lengthened without undo tension on the sciatic nerve. Nine Crowe IV hips received a conventional proximally coated tapered primary femoral component. Oxford hip scores for 76% of patients was excellent (> 41/48), and 24% had good scores (34 – 41). All femoral osteotomies healed. Five hips have been revised, one at 2 years for femoral loosening, one at 5 years for dislocation, two at 12 years for liner exchanges, and one at 21 years for femoral loosening.

THJR using primary prostheses for CDH can provide durable long-term results.


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 4 | Pages 583 - 589
1 Jul 1997
Kobayashi A Donnelly WJ Scott G Freeman MAR

We reviewed a consecutive series of 527 uninfected hip replacements in patients resident in the UK which had been implanted from 1981 to 1993. All had the same basic design of femoral prosthesis, but four fixation techniques had been used: two press-fit, one HA-coated and one cemented. Review and radiography were planned prospectively. For assessment the components were retrospectively placed into two groups: those which had failed from two years onwards by aseptic femoral loosening and those in which the femoral component had survived without revision or recommendation for revision.

All available radiographs in both groups were measured to determine vertical migration and examined by two observers to agree the presence of radiolucent lines (RLLs), lytic lesions, resorption of the neck, proximal osteopenia and distal intramedullary and distal subperiosteal formation of new bone. We then related the presence or absence of these features and the rate of migration at two years to the outcome with regard to aseptic loosening and determined the predictive value of each of these variables.

Migration of ≥2 mm at two years, the presence of an RLL of 2 mm occupying one-third of any one zone, and subperiosteal formation of new bone at the tip of the stem were predictors of aseptic loosening after two years. There were too few lytic lesions to assess at two years, but at five years a lytic lesion ≥2 mm also predicted failure. We discuss the use of these variables as predictors of femoral aseptic loosening for groups of hips and for individual hips.

We conclude that if a group of about 50 total hip replacements, perhaps with a new design of femoral stem, were studied in this way at two years, a mean migration of < 0.4 mm and an incidence of < 10% of RLLs of 2 mm in any one zone would predict 95% survival at ten years.

For an individual prosthesis, migration of < 2 mm and the absence of an RLL of ≤2 mm at two years predict a 6% chance of revision over approximately ten years. If either 2 mm of migration or an RLL of 2 mm is present, the chances of revision rise to 27%, and if both radiological signs are present they are 50%. If at five years a lytic lesion has developed, whatever the situation at two years, there is approximately a 50% chance of failure in the following five years.

Our findings suggest that replacements using a limited number of any new design of femoral prosthesis should be screened radiologically at two years before they are generally introduced. We also suggest that radiographs of individual patients at two years and perhaps at five years should be studied to help to decide whether or not the patient should remain under close review or be discharged from specialist follow-up.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 2 | Pages 196 - 201
1 Feb 2010
Hallan G Dybvik E Furnes O Havelin LI

The Norwegian Arthroplasty Register has shown that several designs of uncemented femoral stems give good or excellent survivorship. The overall findings for uncemented total hip replacement however, have been disappointing because of poor results with the use of metal-backed acetabular components. In this study, we exclusively investigated the medium-to long-term performance of primary uncemented metal-backed acetabular components. A total of 9113 primary uncemented acetabular components were implanted in 7937 patients between 1987 and 2007. These were included in a prospective, population-based observational study. All the implants were modular and metal-backed with ultra-high-molecular-weight polyethylene liners. The femoral heads were made of stainless steel, cobalt-chrome (CoCr) alloy or alumina ceramic. In all, seven different designs of acetabular component were evaluated by the Kaplan-Meier survivorship method and Cox regression analysis. Most acetabular components performed well up to seven years. When the endpoint was revision of the acetabular component because of aseptic loosening, the survival ranged between 87% and 100% at ten years. However, when the endpoint was revision for any reason, the survival estimates were 81% to 92% for the same implants at ten years. Aseptic loosening, wear, osteolysis and dislocation were the main reasons for the relatively poor overall performance of the acetabular components. Prostheses with alumina heads performed slightly better than those with stainless steel or CoCr alloy in subgroups. Whereas most acetabular components performed well at seven years, the survivorship declined with longer follow-up. Fixation was generally good. None of the metal-backed uncemented acetabular components with ultra-high-molecular-weight polyethylene liners in our study had satisfactory long-term results because of high rates of wear, osteolysis, aseptic loosening and dislocation


The Bone & Joint Journal
Vol. 102-B, Issue 8 | Pages 1003 - 1009
1 Aug 2020
Mononen H Sund R Halme J Kröger H Sirola J

Aims

There is evidence that prior lumbar fusion increases the risk of dislocation and revision after total hip arthroplasty (THA). The relationship between prior lumbar fusion and the effect of femoral head diameter on THA dislocation has not been investigated. We examined the relationship between prior lumbar fusion or discectomy and the risk of dislocation or revision after THA. We also examined the effect of femoral head component diameter on the risk of dislocation or revision.

Methods

Data used in this study were compiled from several Finnish national health registers, including the Finnish Arthroplasty Register (FAR) which was the primary source for prosthesis-related data. Other registers used in this study included the Finnish Health Care Register (HILMO), the Social Insurance Institutions (SII) registers, and Statistics Finland. The study was conducted as a prospective retrospective cohort study. Cox proportional hazards regression and Kaplan-Meier survival analysis were used for analysis.


The Bone & Joint Journal
Vol. 99-B, Issue 6 | Pages 766 - 773
1 Jun 2017
Graves SE de Steiger R Davidson D Donnelly W Rainbird S Lorimer MF Cashman KS Vial RJ

Aims

Femoral stems with exchangeable (modular) necks were introduced to offer surgeons an increased choice when determining the version, offset and length of the femoral neck during total hip arthroplasty (THA). It was hoped that this would improve outcomes and reduce complications, particularly dislocation. In 2010, the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) first reported an increased rate of revision after primary THA using femoral stems with an exchangeable neck. The aim of this study was to provide a more comprehensive up-to-date analysis of primary THA using femoral stems with exchangeable and fixed necks.

Materials and Methods

The data included all primary THA procedures performed for osteoarthritis (OA), reported to the AOANJRR between 01 September 1999 and 31 December 2014. There were 9289 femoral stems with an exchangeable neck and 253 165 femoral stems with a fixed neck. The characteristics of the patients and prostheses including the bearing surface and stem/neck metal combinations were examined using Cox proportional hazard ratios (HRs) and Kaplan-Meier estimates of survivorship.


The Bone & Joint Journal
Vol. 97-B, Issue 9 | Pages 1192 - 1196
1 Sep 2015
Amanatullah DF Siman H Pallante GD Haber DB Sierra RJ Trousdale RT

When fracture of an extensively porous-coated femoral component occurs, its removal at revision total hip arthroplasty (THA) may require a femoral osteotomy and the use of a trephine. The remaining cortical bone after using the trephine may develop thermally induced necrosis. A retrospective review identified 11 fractured, well-fixed, uncemented, extensively porous-coated femoral components requiring removal using a trephine with a minimum of two years of follow-up.

The mean time to failure was 4.6 years (1.7 to 9.1, standard deviation (sd) 2.3). These were revised using a larger extensively porous coated component, fluted tapered modular component, a proximally coated modular component, or a proximal femoral replacement. The mean clinical follow-up after revision THA was 4.9 years (2 to 22, sd 3.1). The mean diameter of the femoral component increased from 12.7 mm (sd 1.9) to 16.2 mm (sd 3.4; p >  0.001). Two revision components had radiographic evidence of subsidence that remained radiographically stable at final follow-up. The most common post-operative complication was instability affecting six patients (54.5%) on at least one occasion.

A total of four patients (36.4%) required further revision: three for instability and one for fracture of the revision component. There was no statistically significant difference in the mean Harris hip score before implant fracture (82.4; sd 18.3) and after trephine removal and revision THA (81.2; sd 14.8, p = 0.918).

These findings suggest that removal of a fractured, well-fixed, uncemented, extensively porous-coated femoral component using a trephine does not compromise subsequent fixation at revision THA and the patient’s pre-operative level of function can be restored. However, the loss of proximal bone stock before revision may be associated with a high rate of dislocation post-operatively.

Cite this article: Bone Joint J 2015;97-B:1192–6.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 12 | Pages 1574 - 1580
1 Dec 2007
Hallan G Lie SA Furnes O Engesaeter LB Vollset SE Havelin LI

Primary uncemented femoral stems reported to the Norwegian arthroplasty register between 1987 and 2005 were included in this prospective observational study. There were 11 516 hips (9679 patients) and 14 different designs of stem. Kaplan-Meier survival probabilities and Cox regression were used to analyse the data.

With aseptic loosening as the end-point, all currently used designs performed excellently with survival of 96% to 100% at ten years. With the end-point as stem revision for any cause, the long-term results of the different designs varied from poor to excellent, with survival at 15 years ranging between 29% and 97%. Follow-up for longer than seven years was needed to identify some of the poorly-performing designs. There were differences between the stems; the Corail, used in 5456 hips, was the most frequently used stem with a survival of 97% at 15 years. Male gender was associated with an increased risk of revision of × 1.3 (95% confidence interval 1.05 to 1.52), but age and diagnosis had no influence on the results. Overall, modern uncemented femoral stems performed well.

Moderate differences in survival between well-performing stems should be interpreted with caution since the differences may be caused by factors other than the stem itself.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1011 - 1021
1 Aug 2013
Krishnan H Krishnan SP Blunn G Skinner JA Hart AJ

Following the recall of modular neck hip stems in July 2012, research into femoral modularity will intensify over the next few years. This review aims to provide surgeons with an up-to-date summary of the clinically relevant evidence. The development of femoral modularity, and a classification system, is described. The theoretical rationale for modularity is summarised and the clinical outcomes are explored. The review also examines the clinically relevant problems reported following the use of femoral stems with a modular neck.

Joint replacement registries in the United Kingdom and Australia have provided data on the failure rates of modular devices but cannot identify the mechanism of failure. This information is needed to determine whether modular neck femoral stems will be used in the future, and how we should monitor patients who already have them implanted.

Cite this article: Bone Joint J 2013;95-B:1011–21.


Bone & Joint Research
Vol. 1, Issue 9 | Pages 225 - 233
1 Sep 2012
Paulsen A Odgaard A Overgaard S

Objectives

The Oxford hip score (OHS) is a 12-item questionnaire designed and developed to assess function and pain from the perspective of patients who are undergoing total hip replacement (THR). The OHS has been shown to be consistent, reliable, valid and sensitive to clinical change following THR. It has been translated into different languages, but no adequately translated, adapted and validated Danish language version exists.

Methods

The OHS was translated and cross-culturally adapted into Danish from the original English version, using methods based on best-practice guidelines. The translation was tested for psychometric quality in patients drawn from a cohort from the Danish Hip Arthroplasty Register (DHR).


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1363 - 1369
1 Oct 2010
Garcia-Cimbrelo E Garcia-Rey E Cruz-Pardos A Madero R

Revision surgery of the hip was performed on 114 hips using an extensively porous-coated femoral component. Of these, 95 hips (94 patients) had a mean follow-up of 10.2 years (5 to 17). No cortical struts were used and the cortical index and the femoral cortical width were measured at different levels.

There were two revisions for aseptic loosening. Survivorship at 12 years for all causes of failure was 96.9% (95% confidence interval 93.5 to 100) in the best-case scenario. Fibrous or unstable fixation was associated with major bone defects. The cortical index (p = 0.045) and the lateral cortical thickness (p = 0.008) decreased at the proximal level over time while the medial cortex increased (p = 0.001) at the proximal and distal levels. An increase in the proximal medial cortex was found in patients with an extended transtrochanteric osteotomy (p = 0.026) and in those with components shorter than 25 cm (p = 0.008).

The use of the extensively porous-coated femoral component can provide a solution for difficult cases in revision surgery. Radiological bony ingrowth is common. Although without clinical relevance at the end of follow-up, the thickness of the medial femoral cortex often increased while that of the lateral cortex decreased. In cases in which a shorter component was used and in those undertaken using an extended trochanteric osteotomy, there was a greater increase in thickness of the femoral cortex over time.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 298 - 306
1 Mar 2011
van der Weegen W Hoekstra HJ Sijbesma T Bos E Schemitsch EH Poolman RW

We systematically reviewed the peer-reviewed literature to relate the survival of hybrid metal-on-metal hip resurfacing arthroplasty devices to a National Institute of Clinical Excellence (NICE) benchmark for choosing a primary total hip replacement, which is a survival rate of 90% at a follow-up of ten years.

A total of 29 articles (10 621 resurfaced hips) met the inclusion criteria. The mean follow-up ranged from 0.6 to 10.5 years and the survival of the implant ranged from 84% to 100%. Of the 10 621 hips, 370 were revised (3.5%), with aseptic loosening as the most frequent mode of failure.

None of the hip resurfacing arthroplasty implants used to date met the full ten-year NICE benchmark of survival. A total of 13 studies showed satisfactory survival compared with the three-year NICE benchmark.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 5 | Pages 586 - 590
1 May 2007
Sauvé P Mountney J Khan T De Beer J Higgins B Grover M

Metal-on-metal bearings for total hip replacement (THR) are becoming increasingly popular. Improved wear characteristics mean that these articulations are being inserted into younger patients in the form of THR and resurfacing procedures. This has led to concerns regarding potential carcinogenicity because of the increased exposure to metal ions that the procedure brings.

We have studied the serum cobalt and chromium concentrations in patients who had primary, well-fixed Ring metal-on-metal THRs for more than 30 years. The levels of cobalt and chromium were elevated by five and three times, respectively compared with those in our reference groups. Metal-on-metal articulations appear to be the source of metal ions throughout the life of the prosthesis. In three patients who had undergone revision of a previous metal-on-metal THR to a metal-on-polyethylene replacement the levels of metal ions were within the normal range. The elevations of cobalt and chromium ions seen in our study were comparable with those in patients with modern metal-on-metal THRs.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 4 | Pages 458 - 462
1 Apr 2005
Walton NP Darrah C Shepstone L Donell ST Phillips H

We prospectively studied 217 patients who underwent 234 Elite Plus total hip arthroplasties. At a mean of 6.4 (SD 0.7) years post-operatively, 39 patients had died and 22 were either lost to follow-up or had no radiographs available. Clinical (Oxford hip score) and radiological assessments were performed on 156 patients (168 hip arthroplasties) who had a mean age of 67.7 (SD 9.7) years at operation.

In the assessed group, 26 of 159 (16.4%) of femoral stems which had not already been revised and 19 of 159 (11.9%) of acetabular cups were definitely loose. In total, 52 of 168 (31%) of hips had either been revised or had definite evidence of loosening of a component. We could not establish any relationship between clinical and radiological outcomes.

Despite the fact that the clinical outcome and rate of revision for the Elite Plus appeared to meet international standards, our findings give us cause for concern. We believe that joint registries should include radiological surveillance in order to provide reliable information about medium-term outcomes for hip prostheses.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 628 - 631
1 May 2005
Dunstan E Sanghrajka AP Tilley S Unwin P Blunn G Cannon SR Briggs TWR

Metal-on-metal hip bearings are being implanted into younger patients. The consequence of elevated levels of potentially carcinogenic metal ions is therefore a cause for concern. We have determined the levels of cobalt (Co), chromium (Cr), titanium (Ti) and vanadium (Va) in the urine and whole blood of patients who had had metal-on-metal and metal-on-polyethylene articulations in situ for more than 30 years. We compared these with each other and with the levels for a control group of subjects.

We found significantly elevated levels of whole blood Ti, Va and urinary Cr in all arthroplasty groups. The whole blood and urine levels of Co were grossly elevated, by a factor of 50 and 300 times respectively in patients with loose metal-on-metal articulations when compared with the control group. Stable metal-on-metal articulations showed much lower levels. Elevated levels of whole blood or urinary Co may be useful in identifying metal-on-metal articulations which are loose.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 9 | Pages 1158 - 1164
1 Sep 2007
Lusty PJ Watson A Tuke MA Walter WL Walter WK Zicat B

We studied 33 third generation, alumina ceramic-on-ceramic bearings retrieved from cementless total hip replacements after more than six months in situ. Wear volume was measured with a Roundtest machine, and acetabular orientation from the anteroposterior pelvic radiograph. The overall median early wear rate was 0.1 mm3/yr for the femoral heads, and 0.04 mm3/yr for the acetabular liners. We then excluded hips where the components had migrated. In this stable subgroup of 22 bearings, those with an acetabular anteversion of < 15° (seven femoral heads) had a median femoral head wear rate of 1.2 mm3/yr, compared with 0 mm3/yr for those with an anteversion of ≥15° (15 femoral heads, p < 0.001). Even under edge loading, wear volumes with ceramic-on-ceramic bearings are small in comparison to other bearing materials. Low acetabular anteversion is associated with greater wear.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 12 | Pages 1631 - 1635
1 Dec 2005
von Schewelov T Sanzén L Önsten I Carlsson Å Besjakov J

We investigated the wear characteristics and clinical performance of four different total hip joint articulations in 114 patients. Wear and migration was measured by roentgenstereophotogrammetric analysis at five years or at the last follow-up. The mean annual wear was 0.11 mm for a stainless steel/Enduron articulation, 0.34 mm for stainless steel/Hylamer cup, 0.17 mm for zirconium oxide ceramic/Enduron and 0.40 mm for zirconium oxide ceramic/Hylamer. The difference between the groups was significant (p < 0.008) except for stainless steel/Hylamer vs zirconium oxide ceramic/Hylamer (p = 0.26).

At present, 12 patients have undergone a revision procedure, four at five years and eight thereafter. No patient who received a stainless steel/Enduron articulation at their primary replacement required revision.

Conflicting results have been reported about the performance of the zirconium oxide ceramic femoral head, but our findings suggest that it should not be used with a polymethylmethacrylate acetabular component. Hylamer has already been withdrawn from the market.