Advertisement for orthosearch.org.uk
Results 1 - 20 of 124
Results per page:
Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_7 | Pages 72 - 72
1 Apr 2017
Brooks P
Full Access

Distal femoral varus osteotomy is a procedure intended to relieve pain, correct valgus deformity, and delay or possibly prevent the progression of lateral compartment osteoarthritis in the knee. It is indicated in patients who are considered too young or are too active to be considered candidates for total knee arthroplasty. It also allows protection of the lateral compartment in cases of meniscal or cartilage allograft. In patients who are a good candidate for total knee replacement, TKR is the procedure of choice. A sloping joint line requires that the correction be performed above the knee. Several methods of distal femoral varus osteotomy have been proposed. These include a medial closing wedge, a lateral opening wedge, and a dome osteotomy. In the author's experience, the medial closing wedge has proven reliable. This technique uses a 90-degree blade plate, and does not require any angle measurements during surgery. Fixation is secure, allowing early motion. Healing proceeds rapidly in the metaphyseal bone, and non-unions have not occurred. The desired final alignment was zero degrees, which was reliably achieved using this method. Medium to long-term results are generally satisfactory. When conversion to total knee replacement is required, standard components may generally be used, and function was not compromised by the prior osteotomy. Distal femoral varus osteotomy is a successful procedure for lateral compartment osteoarthritis in a valgus knee. It is indicated in patients who are too young or active for total knee arthroplasty, and provides an excellent functional and cosmetic result


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 104 - 104
1 Feb 2020
Zarei M Hamlin B Urish K Anderst W
Full Access

INTRODUCTION. Controversy exists regarding the ability of unicompartmental knee arthroplasty (UKA) to restore native knee kinematics, with some studies suggesting native kinematics are restored in most or all patients after UKA. 1–3. , while others indicate UKA fails to restore native knee kinematics. 4,5. Previous analysis of UKA articular contact kinematics focused on the replaced compartment. 2,5. , neglecting to assess the effects of the arthroplasty on the contralateral compartment which may provide insight to future pathology such as accelerated degeneration due to overload. 6. or a change in the location of cartilage contact. 7. The purpose of this study was to assess the ability of medial UKA to restore native knee kinematics, contact patterns, and lateral compartment dynamic joint space. We hypothesized that medial UKA restores knee kinematics, compartmental contact patterns, and lateral compartment dynamic joint space. METHODS. Six patients who received fixed-bearing medial UKA consented to participate in this IRB-approved study. All patients (4 M, 2 F; average age 62 ± 6 years) completed pre-surgical (3 weeks before) and post-surgical (7±2 months) testing. Synchronized biplane radiographs were collected at 100 images per second during three repetitions of a chair rise movement (Figure 1). Motion of the femur, tibia, and implants were tracked using an automated volumetric model-based tracking process that matches subject-specific 3D models of the bones and prostheses to the biplane radiographs with sub-millimeter accuracy. 8. Anatomic coordinate systems were created within the femur and tibia. 9. and used to calculate tibiofemoral kinematics. 10. Additional outcome measures included the center of contact in the medial and lateral compartments, and the lateral compartment dynamic joint space (i.e. the distance between subchondral bone surfaces). 11. The results of the three movement trials were averaged for each knee in each test session. All outcome measures were interpolated at 5° increments of knee extension (Figure 2). The average differences between knees at corresponding flexion angles were analyzed using paired t-tests with significance set at p < 0.05. RESULTS. The UKA knee was in 5.3° more varus than the contralateral knee prior to surgery (p=0.005). After surgery, the UKA knee was in 4.9° more valgus than before surgery (p=0.005). The UKA knee was 4.3° more externally rotated than the contralateral knee post-surgery (p=0.05) (Table 1). No significant differences were observed between knees or pre- to post-surgery in lateral compartment dynamic joint space or the center of contact in the medial and lateral tibia compartments (Table 1). DISCUSSION. These results suggest that medial UKA can restore native knee varus without significantly altering lateral compartment joint space or contact location during the chair rise movement. For any figures or tables, please contact the authors directly


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_22 | Pages 70 - 70
1 Dec 2016
Brooks P
Full Access

Distal femoral varus osteotomy is a procedure intended to relieve pain, correct valgus deformity, and delay or possibly prevent the progression of lateral compartment osteoarthritis in the knee. It is indicated in patients who are considered too young or are too active to be considered candidates for total knee arthroplasty. It also allows protection of the lateral compartment in cases of meniscal or cartilage allograft. In patients who are a good candidate for total knee replacement, TKR is the procedure of choice. A sloping joint line requires that the correction be performed above the knee. Several methods of distal femoral varus osteotomy have been proposed. These include a medial closing wedge, a lateral opening wedge, and a dome osteotomy. In the author's experience, the medial closing wedge has proven reliable. This technique uses a 90-degree blade plate, and does not require any angle measurements during surgery. Fixation is secure, allowing early motion. Healing proceeds rapidly in the metaphyseal bone, and non-unions have not occurred. The desired final alignment was zero degrees, which was reliably achieved using this method. Medium to long-term results are generally satisfactory. When conversion to total knee replacement is required, standard components may generally be used, and function was not compromised by the prior osteotomy. Distal femoral varus osteotomy is a successful procedure for lateral compartment osteoarthritis in a valgus knee. It is indicated in patients who are too young or active for total knee arthroplasty, and provides an excellent functional and cosmetic result


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 372 - 372
1 Dec 2013
Nam D Khamaisy S Zuiderbaan H Pearle A
Full Access

Introduction:. The number of medial unicompartmental knee arthroplasties (UKA) performed over the last decade has increased by 30%, as studies have demonstrated improved knee kinematics, range of motion, and decreased perioperative morbidity versus total knee arthroplasty. However, concerns remain regarding the future risk of revision due to lateral compartment degeneration. In patients with a varus mechanical alignment and tibiofemoral subluxation secondary to medial compartment osteoarthritis, the femoral and tibial articular surfaces of the lateral compartment subsequently become incongruous, potentially increasing the focal contact stresses seen with loading. The purpose of this study is to evaluate whether the tibiofemoral congruence of the lateral compartment of the knee is improved following a medial UKA. Methods:. This study is a retrospective review of 192 consecutive medial UKAs included in an IRB-approved, single-surgeon database. All UKAs were performed using a robot-assisted surgical technique. Preoperative and postoperative standing, anteroposterior hip-to-ankle radiographs controlling for lower extremity rotation were performed from which the congruence of the lateral compartment was measured. The preoperative and postoperative degree of articular congruence (congruence index, CI) was calculated using an iterative closest point (ICP)-based software code (Matlab, MathWorks Inc., Natick, MA), specially developed to evaluate congruence of knee compartments. Following digitization of the articular surfaces of the femur and tibia, the code performs a rigid transformation that best aligns the articular surfaces and evaluates the current degree of articular congruence. A congruence index (CI) is then calculated, with a value of 1 indicating complete congruence, and a value of 0 indicating a 100% dislocation of the articular surfaces. A student's t-test was used to compare the preoperative and postoperative values of lateral compartment congruence. Results:. The mean, preoperative congruence index of the lateral compartment was 0.88 (± 0.1), which was improved to 0.93 (± 0.07), following implantation of a medial UKA (p < 0.001). Congruence of the lateral compartment was improved in 158 of the UKAs (83%), while 34 (17%) demonstrated a decrease in the congruence index postoperatively. Conclusion:. Implantation of a medial unicompartmental knee arthroplasty improves the articular surface congruence of the lateral compartment in the majority of patients with isolated, medial compartment osteoarthritis (Figure 1). We hypothesize that this factor, combined with a controlled undercorrection of the overall mechanical alignment, will improved load distribution across the lateral compartment, reduce the risk of focal contact stress points, and decrease the risk of subsequent osteoarthritic degeneration of the lateral compartment. Medial UKA not only resurfaces the medial compartment, but also may treat potential lateral compartment degeneration by improving congruence and load distribution


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_3 | Pages 102 - 102
23 Feb 2023
Campbell T Hill L Wong H Dow D Stevenson O Tay M Munro JT Young S Monk AP
Full Access

Contemporary indications for unicompartmental knee replacement (UKR) include bone on bone radiographic changes in the medial compartment with relatively preserved lateral and patellofemoral compartments. The role of MRI in identifying candidates for UKR is commonplace. The aim of this study was to assess the relationship between radiographic and MRI pre-operative grade and outcome following UKR. A retrospective analysis of medial UKR patients from 2017 to 2021. Inclusion criteria were medial UKR for osteoarthritis with pre-operative and post-operative Oxford Knee Scores (OKS), pre-operative radiographs and MRI. 89 patients were included. Whilst all patients had grade 4 ICRS scores on MRI, 36/89 patients had grade 3 KL radiographic scores in the medial compartment, 50/89 had grade 4 KL scores on the medial compartment. Grade 3 KL with grade 4 IRCS medial compartment patients had a mean OKS change of 17.22 (Sd 9.190) meanwhile Grade 4 KL had a mean change of 17.54 (SD 9.001), with no statistical difference in the OKS change score following UKR between these two groups (p=0.873). Medial bone oedema was present in all but one patient. Whilst lateral compartment MRI ICRS scores ranged from 1 to 4 there was no association with MRI score of the lateral compartment and subsequent change in oxford score (P value 0.458). Patellofemoral Compartment (PFC) MRI ICRS ranged from 0 to 4. There was no association between PFC ICRS score and subsequent change in oxford knee score (P value .276). Radiographs may under report severity of some medial sided knee osteoarthritis. We conclude that in patients with grade 3 KL score that would normally not be considered for UKR, pre-operative MRI might identify grade 4 ICRS scores and this subset of patients have equivalent outcomes to patients with radiographic Grade 4 KL medial compartment osteoarthritis


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 88 - 88
1 Feb 2020
Choi I Seon J
Full Access

Purpose. The purpose of this study was to elucidate kinematic change according to the implant's specific femoral rotation by using orthosensor (Verasense) implant with three degrees external rotation of femoral rotation rebuilt (Genesis-II) and traditional TKA implant without rebuilt of the femoral rotation (Anthem). Methods. Twenty-eight patients (34 knees) underwent TKA using Anthem (Smith & Nephew, Memphis, TN, USA) and 16 patients (22 knees) underwent TKA using Genesis-II (Smith & Nephew, Memphis, TN, USA). Patients were followed up for at least 1 year. Mean age of patients was 71.1 years (range, 60 to 80 years) at the time of surgery. After implantation of femur and tibial components, we applied Verasense, the orthosensor system, to evaluate femoral rollback of the new artificial joint. Femoral rollback was analyzed using digitized screenshot function of Verasense. Results. Overall femoral tracking proportion regardless of implants was significantly higher on the medial compartment compared to that on the lateral compartment (13.3 ± 8.4% vs. 6.3 ± 5.0%, p < 0.001). Regarding femoral tracking according to each compartment, Genesis-II and Anthem showed 12.1 ± 8.2% and 14.2 ± 8.6% (p = 0.371) on the medial compartment and 8.0 ± 5.8% and 5.2 ± 4.2% (p = 0.059) on the lateral compartment, respectively. Conclusion. Our study showed reverse femoral roll-back movement with higher tracking distance on the lateral compartment during TKA. Genesis-II TKA system with femoral component 3-degree rebuilt showed less roll-back difference between medial and lateral compartments compared to traditional TKA system. Fortunately, both TKA systems had excellent short-term clinical outcomes without having significant difference between the two. With longer follow-up and larger cohort, the advantage and effectiveness of femoral component rotation can be elucidated in the future


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 21 - 21
1 Feb 2020
Park C Kang S Song S
Full Access

Purpose. The purpose of the present study was to evaluate the intercompartmental loads with a sensor placed on implants after conventional gap balancing during total knee arthroplasty (TKA) with a tensiometer. Methods. Fifty sensor-assisted TKA procedures were performed prospectively between August and September 2018 with a cruciate-retaining prosthesis. After applying a modified measured technique, conventional balancing between the resected surfaces was achieved. The equal and rectangular flexion–extension gaps were confirmed using a tensiometer. Then, the load distribution was evaluated with a sensor. Results. The average load of the medial compartment was greater than that of the lateral compartment in both the flexion and extension of the knee. The proportion of medial–tight coronal load imbalance (medial load – lateral load ≥ 15 lb) was 50% in the extension and 28% in the flexion positions, respectively (p = 0.035). The loads in each medial and lateral compartment increased with extension of the knee; of note, the amount of increase was higher in the medial compartment (9.7 lb vs. 4.0 lb; p < 0.001). The proportion of the extension–tight sagittal load imbalance (extension load – flexion load ≥ 15lbs) was 34% in the medial compartment and 4% in the lateral compartment (p < 0.001). Conclusions. Coronal and sagittal load imbalances existed as determined by the sensor even after the achievement of appropriate conventional gap balance. The use of an intraoperative load sensor offers the advantage of being able to directly evaluate the load on TKA implants following surgery


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_3 | Pages 108 - 108
23 Feb 2023
Lee W Foong C Kunnasegaran R
Full Access

Most studies comparing medial pivot to the posterior stabilised (PS) systems sacrifice the PCL. It is unknown whether retaining the PCL in the Medial Congruent (MC) system may provide further benefit compared to the more commonly used PS system. A retrospective review of a single-surgeon's registry data comparing 44 PS and 25 MC with PCL retained (MC-PCLR) TKAs was performed. Both groups had similar baseline demographics in terms of age, gender, body mass index, and American Society for Anaesthesiology score. There was no significant difference in their preoperative range of motion (ROM) (104º±20º vs. 102º±20º,p=0.80), Oxford Knee Score (OKS) (27±6 vs. 26±7,p=0.72), and Knee Society Scoring System (KS) Function Score (KS-FS) (52±24 vs. 56±24,p=0.62). The preoperative KS Knee Score (KS-KS) was significantly lower in the PS group (44±14 vs. 54 ± 18,p<0.05). At 3-months postoperation, the PS group had significantly better OKS (38±6 vs. 36±6,p=0.02) but similar ROM (111º±14º vs. 108º±12º,p=0.25), KS-FS (73±20 vs. 68±23, p=0.32) and KS-KS (87±10 vs. 86±9,p=0.26). At 12-months postoperation, both groups had similar ROM (115º±13º vs. 115º±11º,p=0.99), OKS (41±5 vs. 40±5,p=0.45), KS-FS (74±22vs.78±17,p=0.80), and KS-KS (89±10vs.89±11,p=0.75). There was statistically significant improvement in all parameters at 1-year postoperation (p<0.05). The PS group had significant improvement in all parameters from preoperation to 3-month postoperation (p<0.05), but not from 3-month to 1-year postoperation (p≥0.05). The MC-PCLR group continued to have significant improvement from 3-month to 1-year postoperation (p<0.05). The MC provides stability in the medial compartment while allowing a degree of freedom in the lateral compartment. Preserving the PCL when using MC may paradoxically cause an undesired additional restrain that slows the recovery process of the patients after TKA. In conclusion, compared to MC-PCLR, a PS TKA may expect significantly faster improvement at 3 months post operation, although they will achieve similar outcomes at 1-year post operation


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_10 | Pages 80 - 80
1 Jun 2018
Lombardi A
Full Access

Introduction. Persistent pain after medial unicompartmental knee arthroplasty (UKA) is a prevailing reason for revision to total knee arthroplasty (TKA). Many of these pathologies can be addressed arthroscopically. The purpose of this study is to examine the outcomes of patients who undergo an arthroscopy for any reason after medial UKA. Methods. A query of our practice registry revealed 58 patients who had undergone medial UKA between October 2003 and June 2015 with subsequent arthroscopy. Mean interval from UKA to arthroscopy was 22 months (range, 1–101 months). Indications for arthroscopy were acute anterior cruciate ligament tear (1), arthrofibrosis (7), synovitis (12), recurrent hemarthrosis (2), lateral compartment degeneration including isolated lateral meniscus tears (11), and loose cement fragments (25). Results. Mean follow-up after arthroscopy was 37 months (range, 1–134 months). Twelve patients have been revised from UKA to TKA. Relative risk of revision after arthroscopy for lateral compartment degeneration was 4.27 (6 of 11; 55%; p=0.002) and for retrieval of loose cement fragments was 0.05 (0 of 25; 0%; p=0.03). Relative risk for revision after arthroscopy for anterior cruciate ligament tear, arthrofibrosis, synovitis, or recurrent hemarthrosis did not meet clinical significance secondary to the low number of patients in these categories. Conclusions. The results of this study suggest that arthroscopic retrieval of cement fragments does not compromise UKA longevity. However, arthroscopy for lateral compartment degradation after UKA predicts a high risk of revision to TKA regardless of its relative radiographic insignificance


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_10 | Pages 120 - 120
1 Jun 2018
Berend M
Full Access

Not all degenerative knees need a total knee replacement. Over the last few decades we have shifted our surgical treatment of end-stage osteoarthritis (OA) of the knee to a “compartmental approach” resulting in approximately half of end-stage OA knees receiving a partial knee replacement. Of these an emerging procedure is isolated lateral compartment replacement with the indications being isolated bone-on-bone osteoarthritis or avascular necrosis of the lateral compartment of the knee. Associated significant patellofemoral disease and inflammatory arthritis are contraindications. The purpose of this study is to present the indications, surgical technique, and early outcome of lateral partial knees from our institution. From Aug 2011 until June 2017 we have performed 3,548 knee arthroplasties. Of these 147 were fixed bearing lateral partial knee replacements via a lateral parapatellar approach (4%), 1,481 medial partial knee replacements (42%), and 1,920 total knee replacements (54%). The average age was 66 years old and 76% were female. Average follow-up in the lateral partials was 1.3 years (range 0.5 years to 6 years). Knee Society Scores improved from 41 (pre-op) to 86 points (post-op). Range of motion improved from 6 – 113 degrees (pre-op) to 0 – 123 degrees (post-op). No knees were revised to a TKA. One knee required I&D for traumatic wound dehiscence. This is the largest single center series of lateral partial knee replacements. We have observed this cohort to have more female patients and gain additional range of motion compared to our historic cohorts of TKA's. Longer-term follow-up is needed for determination of implant and unreplaced compartment survivorship. We believe the lateral partial knee replacement to be a viable option for isolated lateral compartment disease in approximately 4% of patients


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 461 - 461
1 Dec 2013
Nochi H Abe S Ruike T Kobayashi H Ito H
Full Access

Introduction:. The assumption that symmetric extension-flexion gaps improve the femoral condyle lift-off phenomenon and the patellofemoral joint congruity in total knee arthroplasty (TKA) is now widely accepted. For tease reasons, the balanced gap technique has been developed. However, the management of soft tissue balancing during surgery remains difficult and much is left to the surgeon's feel and experience. Furthermore, little is known about the differences of the soft-tissue stiffness (STS) of medial and lateral compartment in extension and flexion in the both cruciate ligaments sacrificed knee. It has a deep connection with the achievement of appropriate gaps operated according to the balanced gap technique. Therefore, the purpose of this study was to analyze the STS of individual compartment in vivo. Materials and Methods:. The subjects presented 100 osteoarthritic knees with varus deformity underwent primary posterior stabilized (PS) – TKA (NexGen LPS-flex, Zimmer, Warsaw, USA). All subjects completed written informed consent. The patient population was composed of 14 men and 68 women with a mean age of 74.5 ± 7.5 years. The average height, weight, BMI, weight-bearing femorotibial mechanical angle (FTMA), the patella height (T/P ratio), extension and flexion angle of the knee under anesthesia were 151.9 ± 7.8 cm, 62.1 ± 9.4 kg, 26.9 ± 3.7 kg/m. 2. , 167.7 ± 5.6 °, 0.91 ± 0.15 °, −12.0 ± 6.7° and 129.4 ± 13.8°, respectively. After finishing osteotomy and soft tissue balancing, the femoral trial prosthesis was fitted with patello-femoral joint reduction. Then, the medial and lateral compartment gaps (CG) were measured at various distraction forces (89–178 N) using a newly developed versatile tensor device at full extension and 90° flexion positioning, respectively. (Fig. 1) The STS (N/mm) was calculated from a load displacement curve generated by the intra-operative CG data and joint distraction force. Comparisons were made by Wilcoxon signed-ranks test. Correlations were analyzed with Pearson's correlation coefficient. Predictive variables were analyzed with Stepwise regression. A value of p < 0.05 was considered significant. Results:. The CG discrepancy between the medial and the lateral compartments significantly tended to increase as the force dependent manner in the knee at extension (p < 0.0004) and 90° flexion position (p < 0.0001). (Fig. 2) Significant differences (p < 0.0001) were observed in the STS among all compartments respectively; extension medial (71.0 ± ãζζ33.9), flexion medial (26.1 ± 11.6), extension lateral (60.2 ± 36.4) and flexion lateral (19.4 ± 8.2). The ratio of medial to lateral compartment STS (R = −0.54) and the difference of the STS between the medial and lateral compartments (R = 0.385) were significantly correlated with the flexion CG discrepancy (p < 0.0001). The predict variables of the STS could be acquired in extension medial, extension lateral and the ratio of flexion lateral to flexion medial. (Fig. 3). Discussion:. We should notice the significant difference of the STS between the medial and lateral compartments and the ratio of the medial to lateral compartments STS, especially when the balanced gap technique is used. It suggests the importance of refinement of the joint distraction force for individual patients based on their own characteristics of soft tissue


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 71 - 71
1 Mar 2017
Verstraete M Meere P Salvadore G Victor J Walker P
Full Access

Introduction. A correct balancing of the knee following TKA surgery is believed to minimize instability and improve patient satisfaction. In that respect, trial components containing force sensors can be used. These force sensors provide insight in the medial/lateral force ratio as well as absolute contact forces. Although this method finds clinical application already, the target values for both the force magnitude and ratio under surgical conditions remain uncertain. Methods. A total of eight non-arthritic cadaveric knees have been tested mimicking surgical conditions. Therefore, the specimens are mounted in a custom knee simulator (Verstraete et al., 2015). This simulator allows to test full lower limb specimens, providing kinematic freedom throughout the range of motion. Knee flexion is obtained by lifting the femur (thigh pull). Knee kinematics are simultaneously recorded by means of a navigation system and based on the mechanical axis of the femur and tibia. In addition, the load transferred through the medial and lateral compartment of the knee is monitored. Therefore, a 2.4 mm thick sawing blade is used to machine a slot in the tibia perpendicular to the mechanical axis, at the location of the tibial cut in TKA surgery. A complete disconnection was thereby assured between the tibial plateau and the distal tibia. To fill the created gap, custom 3D printed shims were inserted (Fig. 1). Through their specific geometry, these shims create a load deviation between two pressure pads (Tekscan type 4011 sensor) seated on the medial and lateral side. Following the insertion of the shims, the knee was closed before performing the kinematic and kinetic tests. Results. Seven specimens showed a limited varus throughout the range of motion (ranging from 1° to 7° varus). The other knee was in valgus (4° valgus). Amongst varus knees, the results were very consistent, indicating high loads in full extension that rapidly decrease. Subsequently, the loads on the lateral side vanish. This leads to consistently high compartmental load ratios (medial load / total load) in flexion (Fig. 3). Discussion. In full extension the screw-home mechanism results in increased loads, both medially and laterally. Upon flexion, the lateral loads disappear. This is attributed to slackening of the lateral collateral ligament, in turn linked to the femoral rollback and slope of the lateral compartment. The isometry of the medial collateral ligament contributes on the other hand to the near-constant load in the medial compartment. The above particularly applies for varus knees. The single valgus knee tested indicated a higher load transmission by the lateral compartment, potentially attributed to a contracture of the lateral structures. With respect to TKA surgery, these findings are particularly relevant when considering anatomically designed implants. For those implants, this study concludes that a tighter medial compartment reflects that of healthy varus knees. Be aware however that in full extension, higher and up to equal loads can be acceptable for the medial and lateral compartment. For any figures or tables, please contact authors directly (see Info & Metrics tab above).


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_22 | Pages 72 - 72
1 Dec 2016
Cobb J
Full Access

Lateral meniscal failure and secondary valgus with lateral compartment arthrosis is quite common in the developed world. The varus knee is the common phenotype of the ‘jock’ of both genders, while the valgus knee is a common consequence of lateral meniscal tear, skiing or ‘catwalk’ life. Occurring more commonly in ‘flamingo’ phenotypes, lateral meniscal failure can be disabling, entirely preventing high heels being worn for instance. Indications. Lateral UKA is indicated for most valgus knees, and is substantially safer than TKA. ACL integrity is not essential in older people, as the patello-femoral mechanism is in line with the lateral compartment. Severe valgus with substantial bone loss is not a contraindication, if the deformity is simply angular. As long as there is not marked subluxation, fixed flexion deformity invariably corrects after notch osteophyte removal from femur and tibia. Combinations. Lateral UKA can be combined safely with PFJA: performed through a lateral approach, this is a safe and conservative procedure. ACL integrity is not essential – reconstruction can be undertaken simultaneously, if necessary. Combining lateral UKA with medial UKA is only rarely needed, and sometimes needs ACL reconstruction too. Adding a medial UKA in under 5 years usually results from overcorrection of the valgus. Mid Term Results, at a median of 7 years postop: Between 2005 to 2009, 64 knees in 58 patients had a lateral UKA using a device designed for the lateral compartment. This included 41 females and 17 males with a mean age of 71 years at the time of surgery (range 44–92). Thirty-nine patients underwent surgery on the right knee and 6 underwent bilateral procedures, of which four were performed under a single anesthetic. Primary lateral compartment osteoarthritis was the primary diagnosis in 63 cases with secondary osteoarthritis to a lateral tibial plateau fracture the indication in one patient. At 119 months follow up, the predicted cumulative survival was 0.97. With re-operation as an endpoint, 11% of patients within the study had undergone re-operation with a predicted cumulative survival of 0.81 at 119 months. This compares well with historic fixed bearing series. Preoperative OKS scores were available for 50 knees, scores were available for 63 knees at 9–48 months and 52 knees at 61–119 months post index operation. There was a significant improvement in the OKS between the preoperative scores (median 26 range 9–36) and early postoperative time points of 9–48 months, (median 42 range 23–48) (p<0.001). At the later postoperative time point of 61–119 months the score had been maintained (42 range 10–48). Conclusion. Lateral UKA is a small and safe procedure, with clinical outcomes that are equivalent to a medial UKA and are maintained at a median of 7 years postoperatively


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 251 - 251
1 Dec 2013
Buechel F
Full Access

Introduction:. Isolated lateral compartment osteoarthritis (OA) occurs in 5–10% of knees with unicompartmental OA. Lateral unicompartmental knee arthroplasty has been limited in its prevalence due to challenging surgical technique issues. A robotic-arm assisted surgical technique has emerged as a way to achieve precise implant positioning which can potentially improve surgical outcomes. Methods:. 63 consecutive lateral unicompartmental knee arthroplasties were performed by a single surgeon with the use of a metal backed, cemented prosthesis installed with the three-dimensional intra-operative kinematics and haptic robotic guidance. The average age of the patient was 72.7 years (range: 59–87) and the average BMI was 27.2 (range: 19.0–38.6). The follow-up ranged from 2 months to 30 months. Results:. All patients recovered flexion to an average of 130° at 6 weeks post-operatively, compared to an average of 127° pre-operatively. There was one revision to a total knee at 26 months post-operatively for progression of OA to the patellofemoral compartment. Conclusion:. Early results of robotically guided lateral UKA are encouraging and provide evidence to show that lateral UKA is a viable option for patients with lateral OA disease. Three dimensional planning, intra-operative kinematic analysis and haptic robotic guidance provide a significant advantage over manual installation for lateral compartment arthroplasty of the knee. The significant anterior to posterior translation of the lateral femoral condyle along with the “screw-home” mechanism associated with the lateral compartment makes the tracking of the lateral compartment highly complex when retaining the cruciate ligaments. Intraoperative planning and adjustment of the preoperative plan provided only with this robotic system results in optimized lateral compartment kinematics. Further mid to long term studies are needed to determine survivorship as it compares to medial unicompartmental knee arthroplasty


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_20 | Pages 65 - 65
1 Dec 2017
Macke C Westphal R Citak M Hawi N Liodakis E Krettek C Stuebig T Suero EM
Full Access

Introduction. High tibial osteotomy (HTO) is a commonly used surgical technique for treating moderate osteoarthritis (OA) of the medial compartment of the knee by shifting the center of force towards the lateral compartment. The amount of alignment correction to be performed is usually calculated prior to surgery and it's based on the patient's lower limb alignment using long-leg radiographs. While the procedure is generally effective at relieving symptoms, an accurate estimation of change in intraarticular contact pressures and contact surface area has not been developed. Using electromyography (EMG), Meyer et al. attempted to predict intraarticular contact pressures during gait patterns in a patient who had received a cruciate retaining force-measuring tibial prosthesis. Lundberg et al. used data from the Third Grand Challenge Competition to improve contact force predictions in total knee replacement. Mina et al. performed high tibial osteotomy on eight human cadaveric knees with osteochondral defects in the medial compartment. They determined that complete unloading of the medial compartment occurred at between 6° and 10° of valgus, and that contact pressure was similarly distributed between the medial and lateral compartments at alignments of 0° to 4° of valgus. In the current study, we hypothesised that it would be possible to predict the change in intra-articular pressures based on extra-articular data acquisition. Methods. Seven cadavers underwent an HTO procedure with sequential 5º valgus realignment of the leg up to 15º of correction. A previously developed stainless-steel device with integrated load cell was used to axially load the leg. Pressure-sensitive sensors were used to measure intra-articular contact pressures. Intraoperative changes in alignment were monitored in real time using computer navigation. An axial loading force was applied to the leg in the caudal-craneal direction and gradually ramped up from 0 to 550 N. Intra-articular contact pressure (kg) and contact area (mm2) data were collected. Generalised linear models were constructed to estimate the change in contact pressure based on extra-articular force and alignment data. Results. The application of an axial load results in axial angle changes and load distribution changes inside the knee joint. Preliminary analysis has shown that it is possible to predict lateral and medial compartment pressures using externally acquired data. For lateral compartment pressure estimation, the following equation had an R of 0.86: Lateral compartment pressure = −1.26*axial_force + 37.08*horizontal_force − 2.40*vertical_force − 271.66*axial_torque − 32.64*horizontal_torque + 18.98*vertical_torque − 24.97*varusvalgus_angle_change + 86.68*anterecurvature_angle_change − 17.33*axial_angle_change − 26.14. For medial compartment pressure estimation, the following equation had an R2 of 0.86: Medial compartment pressure = −2.95*axial_force −22.93*horizontal_force − 9.48*vertical_force − 34.53*axial_torque + 6.18*horizontal_torque − 127.00*vertical_torque − 110.10*varusvalgus_angle_change − 15.10*anterecurvature_angle_change + 55.00*axial_angle_change + 193.91. Discussion. The most important finding of this study was that intra-articular pressure changes in the knee could be accurately estimated given a set of extra-articular parameters. The results from this study could be helpful in developing more accurate lower limb realignment procedures. This work complements and expands on previous research by other groups aimed at predicting intra-articular pressures and identifying optimal alignment for unloading arthritic defects. A possible clinical application of these findings may involve the application of a predetermined axial force to the leg intra-operatively. Given the estimated output from the predictive equation, one could then perform the opening wedge until the desired estimated intra-articular pressure is achieved. With this method, an arthrotomy and placement of intra-articular pressure sensors would not be needed. This work is not without its limitations. This experiment was performed on cadaveric specimens. Therefore, we cannot directly predict what the pressures would be in a de-ambulating patient. However, these sort of experiments do help us understand the complex biomechanics of the knee in response to alterations in multi-planar alignment. Further in vivo research would be warranted to validate these results. Additionally, given our current experimental setup, only axial loading could be performed for testing. Further experiments involving dynamic motion of the lower limb under load would further help us understand the changes in pressure at difference flexion angles. Continued experiments would help us gather additional data to better understand the relationship between these variables and to construct a more accurate predictive model. In summary, we have established a framework for estimating the change in intra-articular contact pressures based on extra-articular, computer-navigated measurements. Quantifying the resulting changes in load distribution, alignment changes, torque generation and deflection will be essential for generating appropriate algorithms able to estimate joint alignment changes based on applied loads


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 93 - 93
1 Jan 2016
Vandekerckhove P Teeter M Naudie D Howard J MacDonald S Lanting B
Full Access

Introduction. Coronal plane alignment is one of the contributing factors to polyethylene wear in total knee arthroplasty (TKA). The goal of this study was to evaluate the wear and damage patterns of retrieved tibial polyethylene inserts in relationship to the overall mechanical alignment and to the position of the tibial component. Materials and methods. Based on full-length radiographs, ninety-five polyethylene inserts retrieved from primary TKA's with a minimum time in-vivo of five years were analysed for wear and damage. Four alignment groups were compared: valgus, neutral, mild varus and moderate varus. Varus and valgus positioning of the tibial component was analysed for damage score for the neutral and varus aligned groups. Results. A progression in the angle of wear was observed with progressively mechanical varus alignment. The valgus group was thinner laterally and the neutral, mild varus, and moderate varus groups were progressively thinner medially. The lateral compartment had greater damage in the mild and moderate varus group compared to the valgus group. There was a progression of increased lateral damage with increasingly varus HKA. No difference in damage was seen between groups for tibial component positioning. Conclusion. While greater wear of the lateral compartment in valgus aligned implants and progressively greater medial compartment wear in varus aligned implants was observed, greater damage scores were observed in the lateral compartment in the mild and moderate varus aligned TKAs compared to the valgus group. This observation is unique and might by explained by lateral condylar lift-off inducing impact and shear loading in the varus group


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_21 | Pages 69 - 69
1 Dec 2016
Kopka M Rahnemani-Azar A Abebe E Labrum J Irrgang J Fu F Musahl V
Full Access

Knee laxity following anterior cruciate ligament (ACL) injury is a complex phenomenon influenced by various biomechanical and anatomical factors. The contribution of soft tissue injuries – such as ligaments, menisci, and capsule – has been previously defined, but less is known about the effects of bony morphology. (Tanaka et al, KSSTA 2012) The pivot shift test is frequently employed in the clinical setting to assess the combined rotational and translational laxity of the ACL deficient knee. In order to standardise the maneuver and allow for reproducible interpretation, the quantitative pivot shift test was developed. (Hoshino et al, KSSTA 2013) The aim of this study is to employ the quantitative pivot shift test to determine the effects of bone morphology as determined by magnetic resonance imaging (MRI) on rotatory laxity of the ACL deficient knee. Fifty-three ACL injured patients scheduled for surgical reconstruction (36 males and 17 females; 26±10 years) were prospectively enrolled in the study. Preoperative magnetic resonance imaging (MRI) scans were reviewed by two blinded observers and the following parameters were measured: medial and lateral tibial slope, tibial plateau width, femoral condyle width, bicondylar width, and notch width. (Musahl et al. KSSTA 2012). Preoperatively and under anaesthesia, a quantitative pivot shift test was performed on each patient by a single experienced examiner. An image analysis technique was used to quantify the lateral compartment translation during the maneuver. Subjects were classified as “high laxity” or “low laxity” based upon the median value of lateral compartment translation. (Hoshino et al. KSSTA 2012) Independent t-tests and univariate logistic regression were used to investigate the relationship between the pivot shift grade and various features of bone morphology. Statistical significance was set at p<0.05. A high inter-rater reliability was observed in all MRI measurements of bone morphology (ICC=0.72–0.88). The median lateral compartment translation during quantitative pivot shift testing was 2.8mm. Twenty-nine subjects were classified as “low laxity” (2.8mm). The lateral tibial plateau slope was significantly increased in “high laxity” patients (9.3+/−3.4mm versus 6.1+/−3.7mm; p<0.05). No other significant difference in bone morphology was observed between the groups. This study employed an objective assessment tool – the quantitative pivot shift test – to assess the contribution of various features of bone morphology to rotatory laxity in the ACL deficient knee. Increased lateral tibial plateau slope was shown to be a significant independent predictor of high laxity. These findings could help guide treatment strategies in patients with high grade rotatory laxity. Further research into the role of tibial osteotomies in this sub-group is warranted


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_7 | Pages 10 - 10
1 May 2018
Monk P Boyd R Brown C Gibbons M Alvand A Price A
Full Access

The optimal correction of the weight bearing line during High Tibial Osteotomy has not been determined. We used finite element modelling to simulate the effect that increasing opening wedge HTO has on the distribution of stress and pressure through the knee joint during normal gait. Subject-specific models were developed by combining geometry from 7T MRI scans and applied joint loads from ground reaction forces measured during level walking. Baseline stresses and pressures on the articulating proximal tibial cartilage and menisci were calculated. Progressive osteotomies were then simulated to shift the weight-bearing line from the native alignment towards/into the lateral compartment (between 40 – 80% of medial-lateral tibial width). Changes in calculated stresses and pressures were recorded. Both stress and pressure decreased in the medial compartment and increased in the lateral compartment as increasingly valgus osteotomies were simulated. The models demonstrated a consistent “safe zone” for weight bearing line position at 50%-65% medial-lateral tibial width, outside of which compartment stresses and pressures substantial increased. This study suggests a safe correction zone within which a medial opening wedge HTO can be performed correcting the WBL to 55% medio-lateral width of the tibia


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 25 - 25
1 Sep 2012
Prud'homme-Foster M Louati H Parai M Dervin G
Full Access

Purpose. Unicompartmental knee replacement (UKR) is an established, bone preserving surgical treatment option for medial compartment osteoarthritis (OA). Early revision rates appear consistently higher than those of total knee replacement (TKR) in many case series and consistently in national registry data. Failure with progression of OA in the lateral compartment has been attributed, in part, to surgical technical errors. In this study we used navigation assisted surgery to investigate the effects of improper sizing of the mobile bearing and malrotation of the tibial component on alignment and lateral compartment loading. Method. A total of eight fresh frozen cadaveric lower limbs were used in the study. After thawing overnight, a Brainlab navigation system with an Oxford (Biomet, Inc) medial UKR module was used to capture the native knee anatomy and alignment using a digitizing probe. Following registration, the case was performed with navigation verified neutral cuts and an ideal insert size was selected to serve as a baseline. The bearing thickness was subsequently increased by 2 mm increments to simulate progressive medial joint overstuffing. Excessive tibial internal rotation of 12 was also simulated at each of the intervals. Knee alignment in varus or valgus was recorded in real time for each surgical scenario with the knee in full extension and at 20 of flexion. Lateral compartment peak pressure was measured using a Tekscan pressure map. Results. Incremental overstuffing of the medial compartment with inserts of increasing thickness resulted in a progressive shift to more valgus knee alignment. Internally rotated sagittal cuts at 12 resulted in a further valgus shift for a given insert size. The valgus shift was detectable at full extension however it was more pronounced at 20 of flexion. Conclusion. The intentional technical errors of overstuffing and malrotation in UKR produced coronal valgus knee alignment and a greater load shift to the lateral compartment. These errors can be construed to contribute to the higher early failure rates associated with UKR when compared to TKR. Special care should be taken to ensure a neutral sagittal tibia cut and appropriate bearing selection. The Intra operative verification of knee alignment should be conducted at 20 of flexion where such errors will be easier for the surgeon to detect and rectify


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 113 - 113
1 Apr 2019
Gray H Guan S Young T Dowsey M Choong P Pandy M
Full Access

INTRODUCTION. The medial-stabilised (MS) knee implant, characterised by a spherical medial condyle on the femoral component and a medially congruent tibial bearing, was developed to improve knee kinematics and stability relative to performance obtained in posterior-stabilised (PS) and cruciate-retaining (CR) designs. We aimed to compare in vivo six-degree-of-freedom (6-DOF) kinematics during overground walking for these three knee designs. METHODS. Seventy-five patients (42 males, 33 females, age 68.4±6.6 years) listed for total knee arthroplasty (TKA) surgery were recruited to this study, which was approved by the relevant Human Research Ethics committees. Each patient was randomly- assigned a PS, CR or MS knee (Medacta International AB, Switzerland) resulting in three groups of 23, 26 and 26 patients, respectively. Patients visited the Biomotion Laboratory at the University of Melbourne 6±1.1 months after surgery, where they walked overground at their self-selected speed. A custom Mobile Biplane X-ray (MoBiX) imaging system tracked and imaged the implanted knee at 200 Hz. The MoBiX system measures 6-DOF tibiofemoral kinematics of TKA knees during overground gait with maximum RMS errors of 0.65° and 0.33 mm for rotations and translations, respectively. RESULTS AND DISCUSSION. Mean walking speeds for the three groups were not significantly different (PS, 0.86±0.14 m/s CR, 0.82±0.17 m/s and MS, 0.87±0.14 m/s, p>0.25). While most kinematic parameters were similar for the PS and CR groups, mean peak-to-peak anterior drawer was greater for PS (9.89 mm) than CR (7.75 mm, p=0.004), which in turn was greater than that for MS (4.43 mm, p<0.001). Mean tibial external rotation was greater for MS than PS (by 3.12°, p=0.033) and CR (by 3.34°, p=0.029). Anterior drawer and lateral shift were highly coupled to external rotation for MS but not so for PS and CR. The contact centres on the tibial bearing translated predominantly in the anterior-posterior direction for all three designs. Peak-to-peak anterior-posterior translation of the contact centres in the medial compartment was largest for PS (7.09 mm) followed by CR (5.45 mm, p=0.003) and MS (2.89 mm, p<0.001). The contact centre in the lateral compartment was located 2.5 mm more laterally for MS than PS and CR (p<0.001). The centre of rotation of the knee in the transverse plane was located in the medial compartment for MS and in the lateral compartment for both PS and CR. CONCLUSIONS. We quantitatively compared in vivo 6-DOF joint motion for PS, CR, and MS knees during locomotion. A higher degree of coupling between external rotation and anterior-posterior translation, greater constraint in the anterior-posterior direction, and a more medialised joint centre of rotation observed for the MS knees are explained by the highly congruent medial articulation characterising this design