Aims. Periprosthetic fracture and implant loosening are two of the major reasons for revision surgery of cementless implants. Optimal implant fixation with minimal bone damage is challenging in this procedure. This pilot study investigates whether vibratory implant
Introduction. The Exeter RimFit™ flanged cemented cup features ultra-highly cross-linked polyethylene and was introduced onto the market in the UK in 2010. We aimed to examine the rates of radiolucent lines observed when the Rimfit™ cup was implanted using a ‘rimcutter’ technique with the cup sitting on a prepared acetabular rim, and a ‘trimmed flange’ technique were the flange is cut so that it sits inside the prepared acetabular rim. Patients/Materials & Methods. The radiographs of 150 (75 ‘rimcutter’, 75 ‘trimmed flange’) Rimfit™ hip replacements were critically evaluated to assess for radiolucency at the cement bone interface. This group was then compared to a historic pre-Rimfit™ cohort of 76 patients. Results. Rimfit™ Cups inserted using the rimcutter technique exhibited significantly higher rates of radiolucency than cups inserted using the trimmed flange technique at 1 year post surgery (1 Delee and Charnley zone: rimcutter=73/77 (95%) versus trimmed flange=14/75 (19%); 2 zones: 43/75 (57%) versus 0/75 (0%); all 3 zones: 13/75 (17%) versus 0/75 (0%):(p<0.01)). There were significantly more radiolucencies for the ‘rimcutter’/Rimfit™ group when compared to a historic group of non-‘rimcutter’/non-Rimfit™ cups implanted by the same surgeon (p<0.01). Discussion. 95% of Rimfit™ cups inserted using the rimcutter technique exhibited radiolucent lines at 12 months. This may be due to increased flange stiffness related to the ultra-highly cross-linked polyethylene and the closer seal due to the narrow circumferential ridge created by the rimcutter.
Introduction. Operative inclination (OI) is defined as the angle between the acetabular axis and the sagittal plane. With the patient in the true lateral decubitus position, this corresponds to the angle formed between the handle of the acetabular component inserter and the theatre floor intra-operatively. Patients/Materials & Methods. The primary study aim was to determine which method of acetabular component
In a prospective randomised clinical study acetabular components were implanted either freehand (n = 30) or using CT-based (n = 30) or imageless navigation (n = 30). The position of the component was determined post-operatively on CT scans of the pelvis. Following conventional freehand placement of the acetabular component, only 14 of the 30 were within the safe zone as defined by Lewinnek et al (40° inclination Imageless navigation proved as reliable as that using CT in positioning the acetabular component.
Aims. Post-traumatic periprosthetic acetabular fractures are rare but serious. Few studies carried out on small cohorts have reported them in the literature. The aim of this work is to describe the specific characteristics of post-traumatic periprosthetic acetabular fractures, and the outcome of their surgical treatment in terms of function and complications. Methods. Patients with this type of fracture were identified retrospectively over a period of six years (January 2016 to December 2021). The following data were collected: demographic characteristics, date of
Securing the osteotomized greater trochanter (GT) during total hip arthroplasty (THA) for dislocated dysplastic hips (DDH) poses a significant challenge. This study evaluates the union rate and effectiveness of a 2-strand transverse wiring technique utilizing the lesser trochanter for wire anchorage and tensioning. A digastric anterior slide trochanteric osteotomy was performed in 106 patients (118 hips) undergoing THA for DDH. Following uncemented stem
In metal-on-polyethylene (MoP) THA large femoral metal heads are designed to increase stability and to reduce dislocation risk. The increased head size could lead to increased taper corrosion with the release of metal ions and adverse reactions. Using blood ion measurements, we aimed to investigate the association between femoral head size and metal-ion release after MoP THA. 96 patients were enrolled at two centers and randomized to receive either a 32-mm metal head or a 36–44 mm metal head (the largest possible fitting the thinnest available polyethylene insert). Blood metal ions and PROMs (OHS, UCLA) were measured at two- and five-year follow-ups. Both 2- and 5-year median chrome, cobalt, and titanium levels were below taper corrosion indicative ion levels. At 5 years, median chrome, cobalt, and titanium levels were 0.5 μg/L (0.50–0.62), 0.24 μg/L (0.18–0.30), and 1.16 μg/L (1.0–1.68) for the 32-mm group, and 0.5 μg/L (0.5–0.54), 0.23 μg/L (0.17–0.39), and 1.30 μg/L (1–2.05) for the 36–44 mm group, with no difference between groups (p=0.825, p=1.000, p=0.558). At 2 years, 7 (32-mm) versus 4 (36–44-mm) patients had elevated ions. At 5 years, 6 (32-mm) versus 7 (36–44-mm) patients had elevated ions. There was no difference in either OHS (p=0.665) or UCLA (p=0.831) between patients with or without elevated blood metal ions. 5 years after the
Aims. Hip arthroplasty does not always restore normal anatomy. This is due to inaccurate surgery or lack of stem sizes. We evaluated the aptitude of four total hip arthroplasty systems to restore an anatomical and medialized hip rotation centre. Methods. Using 3D templating software in 49 CT scans of non-deformed femora, we virtually implanted: 1) small uncemented calcar-guided stems with two offset options (Optimys, Mathys), 2) uncemented straight stems with two offset options (Summit, DePuy Synthes), 3) cemented undersized stems (Exeter philosophy) with three offset options (CPT, ZimmerBiomet), and 4) cemented line-to-line stems (Kerboul philosophy) with proportional offsets (Centris, Mathys). We measured the distance between the templated and the anatomical and 5 mm medialized hip rotation centre. Results. Both rotation centres could be restored within 5 mm in 94% and 92% of cases, respectively. The cemented undersized stem performed best, combining freedom of stem positioning and a large offset range. The uncemented straight stem performed well because of its large and well-chosen offset range, and despite the need for cortical bone contact limiting stem positioning. The cemented line-to-line stem performed less well due to a small range of sizes and offsets. The uncemented calcar-guided stem performed worst, despite 24 sizes and a large and well-chosen offset range. This was attributed to the calcar curvature restricting the stem
Short cementless femoral stems are increasingly popular as they allow for less dissection for
With the approval of our institute, we reviewed all the robot-assisted hip revision during October 2019 and August 2021. MAKO joint arthroplasty system was used to perform the hip revision surgery. Seventy-one robot-assisted hip revision cases were included. Cup revisions were carried out in 68 patients while stem revisions were also carried out in 68 patients. Three types of registration techniques (extra acetabular bone surface based, liner based, metal shell based or cage surface based) on the acetabular side. The extra acetabular bone surface was the commonest used for registration (48/70, 68.6%, mean accuracy 0.37mm), followed by liner surface (11/70, 15.7%, mean accuracy 0.36mm), acetabulum cup (10/70, 14.3%, mean accuracy 0.37mm), and cage surface (1/70, 1.4%, accuracy 0.40mm). We succeeded cup registration and robotic arm guided cup
Thru purpose of this study was to evaluate the clinical outcomes of a consecutive series of conversion total hip arthroplasty (cTHA) following previous hip fractures. A retrospective chart review of patients who underwent cTHAs from 2008–2017 at an urban academic teaching institution was performed. Eighty-eight patients were included in this study. The mean age at the cTHA was 66 years (range 27 to 89). 67% of the patients wre women. The mean BMI was 28 kg/m. 2. (range 17 to 41). The mean Charlson Comorbidity Index was 3 (range 0 to 9). The mean follow-up was 49 months (range 24 to 131). The mean duration from the hip fracture fixation to the cTHA was 51 months (range 10 to 144). The mean operating time was 188 minutes, (range 71 to 423) with a mean estimated blood loss of 780 ml (range 300 to 2500). Revision-type (long-stem) designs were used in 65% of the cases. The mean length of hospital stay was 8 days (range 2 to 61). The readmission rate was 37% within 90 days after the CTHAs. Of these, 57% were due to non-orthopaedic complications. There were 10 orthopaedic complications: 7 PJIs, all of which required I&D and 3 required staged revision. There were 2 dislocations treated with closed reduction and 1 case of intraoperative periprosthetic femur fracture during femoral component
We evaluated an anatomical uncemented stem, SP-CL, (Static Physiologicus – CementLess) designed to facilitate
Aims. Traditionally, acetabular component
Aims. One method of femoral head preservation following avascular necrosis
(AVN) is core decompression and
Aims. Although good clinical outcomes have been reported for monolithic tapered, fluted, titanium stems (TFTS), early results showed high rates of subsidence. Advances in stem design may mitigate these concerns. This study reports on the use of a current monolithic TFTS for a variety of indications. Methods. A multi-institutional retrospective study of all consecutive total hip arthroplasty (THA) and revision total hip arthroplasty (rTHA) patients who received the monolithic TFTS was conducted. Surgery was performed by eight fellowship-trained arthroplasty surgeons at four institutions. A total of 157 hips in 153 patients at a mean follow-up of 11.6 months (SD7.8) were included. Mean patient age at the time of surgery was 67.4 years (SD 13.3) and mean body mass index (BMI) was 28.9 kg/m. 2. (SD 6.5). Outcomes included intraoperative complications, one-year all-cause re-revisions, and subsidence at postoperative time intervals (two weeks, six weeks, six months, nine months, and one year). Results. There were eight intraoperative complications (4.9%), six of which were intraoperative fractures; none occurred during stem
Total hip arthroplasty (THA) is considered the preferred treatment for displaced proximal femoral neck fractures. However, in many countries this option is economically unviable. To improve outcomes in financially disadvantaged populations, we studied the technique of concomitant valgus hip osteotomy and operative fixation (VOOF). This prospective serial study compares two treatment groups: VOOF versus operative fixation alone with cannulated compression screws (CCSs). In the first series, 98 hip fixation procedures were performed using CCS. After fluoroscopic reduction of the fracture, three CCSs were placed. In the second series, 105 VOOF procedures were performed using a closing wedge intertrochanteric osteotomy with a compression lag screw and lateral femoral plate. The alignment goal was to create a modified Pauwel’s fracture angle of 30°. After fluoroscopic reduction of fracture, lag screw was placed to achieve the calculated correction angle, followed by inter-trochanteric osteotomy and placement of barrel plate. Patients were followed for a minimum of two years.Aims
Methods
It is important to analyze objectively the hammering sound in cup press-fit technique in total hip arthroplasty (THA) in order to better understand the change of the sound during impaction. We hypothesized that a specific characteristic would present in a hammering sound with successful fixation. We designed the study to quantitatively investigate the acoustic characteristics during cementless cup impaction in THA. In 52 THAs performed between November 2018 and April 2022, the acoustic parameters of the hammering sound of 224 impacts of successful press-fit fixation, and 55 impacts of unsuccessful press-fit fixation, were analyzed. The successful fixation was defined if the following two criteria were met: 1) intraoperatively, the stability of the cup was retained after manual application of the torque test; and 2) at one month postoperatively, the cup showed no translation on radiograph. Each hammering sound was converted to sound pressures in 24 frequency bands by fast Fourier transform analysis. Basic patient characteristics were assessed as potential contributors to the hammering sound.Aims
Methods
It is important to analyze objectively the hammering sound in cup press-fit technique in total hip arthroplasty (THA) in order to better understand the change of the sound during impaction. We hypothesized that a specific characteristic would present in a hammering sound with successful fixation. We designed the study to quantitatively investigate the acoustic characteristics during cementless cup impaction in THA. In 52 THAs performed between November 2018 and April 2022, the acoustic parameters of the hammering sound of 224 impacts of successful press-fit fixation, and 55 impacts of unsuccessful press-fit fixation, were analyzed. The successful fixation was defined if the following two criteria were met: 1) intraoperatively, the stability of the cup was retained after manual application of the torque test; and 2) at one month postoperatively, the cup showed no translation on radiograph. Each hammering sound was converted to sound pressures in 24 frequency bands by fast Fourier transform analysis. Basic patient characteristics were assessed as potential contributors to the hammering sound.Aims
Methods
The primary objective of this study was to develop a validated classification system for assessing iatrogenic bone trauma and soft-tissue injury during total hip arthroplasty (THA). The secondary objective was to compare macroscopic bone trauma and soft-tissues injury in conventional THA (CO THA) versus robotic arm-assisted THA (RO THA) using this classification system. This study included 30 CO THAs versus 30 RO THAs performed by a single surgeon. Intraoperative photographs of the osseous acetabulum and periacetabular soft-tissues were obtained prior to implantation of the acetabular component, which were used to develop the proposed classification system. Interobserver and intraobserver variabilities of the proposed classification system were assessed.Aims
Methods
Our objective was describing an algorithm to identify and prevent vascular injury in patients with intrapelvic components. Patients were defined as at risk to vascular injuries when components or cement migrated 5 mm or more beyond the ilioischial line in any of the pelvic incidences (anteroposterior and Judet view). In those patients, a serial investigation was initiated by a CT angiography, followed by a vascular surgeon evaluation. The investigation proceeded if necessary. The main goal was to assure a safe tissue plane between the hardware and the vessels.Aims
Methods