Background: Polymethylmethacrylate (PMMA) is a potent stimulant of inflammatory response. This study investigated the role of Prostaglandin E2 (PGE2), Platelet activating factor (PAF) and histamine and their specific antagonists in bone changes. Materials: 120 white-male-wistar rats were divided into ten groups. Using sterile technique, a 2mm drill hole was made in the tibia 1cm distal to the knee joint bilaterally. The left tibia was filled with Simplex particulate cement polymer (PMMA) and the right tibia was used as control. The first nine groups respectively received terfenadine 1mg/kg, 10mg/kg and 25mg/kg, alprazolam 0.08mg/kg, 0.32mg/kg and 0.64mg/kg, and naproxen 1mg/kg, 5mg/kg and 25mg/kg; however, the tenth group received no drug and served as control. The animals were killed after 16 weeks and implant areas were harvested aseptically and studied by one pathologist. Results: Our study revealed that the cellular reaction in the left side was statistically more than the right one in all cases (p<
0.05). Also, a significant decrease in
Summary. Particulate wear debris with different chemical composition induced similar periprosthetic tissue reactions in patients with loosened uncemented and cemented titanium hip implants, which suggests that osteolysis can develop independent of particle composition. Introduction. Periprosthetic osteolysis is a serious long-term complication in total hip replacements (THR). Wear debris-induced inflammation is thought to be the main cause for periprosthetic bone loss and implant loosening. The aim of the present study was to compare the tissue reactions and wear debris characteristics in periprosthetic tissues from patients with failed uncemented (UC) and cemented (C) titanium alloy hip prostheses. We hypothesised that implant wear products around two different hip designs induced periprosthetic inflammation leading to osteolysis. Patients & Methods. Thirty THR-patients undergoing revision surgery were included: Fifteen patients had loose cemented titanium stems (Titan. ®. , DePuy) and 15 had well-fixed uncemented titanium stems (Profile, DePuy), but loose or worn uncemented metal-backed cups with conventional UHMWPE liners (Gemini, Tropic and Tri-Lock Plus, DePuy; Harris/Galante and Trilogy, Zimmer). A semi-quantitative histological evaluation was performed in 59 sections of periprosthetic tissues using light microscopy. Wear particles were counted by polarised light and high resolution dark-field microscopy. Additionally, particle composition was determined by SEM-EDXA following particle isolation using an enzymatic digestion method. Blood metal ions were determined with high resolution-ICP-MS. Results. The implants in the uncemented group were revised after a median of 15.7 years (range: 7.25–19.3) due to osteolysis and high wear of the polyethylene liner and metal backing resulting in gross metallosis, and/or cup loosening. The average lifetime of implants in the cemented group was only 6.5 years (range: 1.5–11.75) due to early stem loosening with large osteolysis pockets in the femur close to the cement mantle. Tissue examination revealed similar results for both groups: numerous mononuclear
1. The clinical, radiological and pathological features of 106 cases of eosinophilic granuloma of bone (solitary and multiple) are reported. 2. Our findings support strongly the concept that eosinophilic granuloma, Hand-Schüller-Christian and Letterer-Siwe syndrome are closely related manifestations of a single pathological disorder, characterised fundamentally by the proliferation of non-neoplastic
Total joint arthroplasty is the most significant advance in the treatment of end-stage arthritic disease of major joints. Despite the clinical success of this surgical procedure, however, some total joint prostheses fail, and although a failed prosthesis can be replaced, the results of revision arthroplasty are not as good as the first time. Studying the failed prosthesis and the associated bone and soft tissues provides insight into the causes of failure. Most prosthetic failures are the result of structural limitations of the implant components. Although material failure may be sudden, a much more common cause is gradual aseptic loosening of the prostheses. Aseptic loosening is caused by both mechanical (gradual loss of material by wear) and biological (osteoclastic resorption of adjacent bone) factors. Wear particles induce a foreign body reaction characterized by a pseudomembrane composed of granulomatous tissues including macrophages, fibroblasts, giant cells, and osteoclasts in addition to debris particles. The extent of this response is driven by the number, size, composition, surface area, and types of particles present. Although there are differences in the relative local toxicity of each of these particles, the end result is the same. These mechanical and biological factors are unavoidable, and the success of a total joint prosthesis depends on the rate with which they occur. Polyethylene wear particles (1–200 ?) are the primary cause of loosening. They are strongly birefringent under polarized light microscopy. Smaller particles are phagocytized by
Introduction:. Sinus histiocytosis with massive lymphadenopathy (SHML) also known as Rosai – Dorfman disease is a disease of bone marrow stem cell origin. It affects lymph nodes primarily. Solitary bone lesions are very rare and can cause diagnostic difficulty. Aim:. To increase the awareness of SHML as a cause of cystic bone lesions. Materials and methods:. A 2 year old presented with 4 months history of pain and swelling of the distal forearm. There was no history of tuberculosis or HIV disease. The swelling was 4 × 3 cm firm, non-fluctuant and slightly tender. There were no lymph nodes. Radiographs showed an oval cystic lesion expanding with a well-defined margin. The ulnar cortex was deficient. CT scan confirmed a cystic lesion with contents of granulation tissue. The Hb and WCC were normal, ESR 20 was, CRP<5 and mantoux was negative. At surgery the lesion was curretted. The contents resembled tuberculous granulation but there was no caseation. The borders were well formed, the ulnar cortex was deficient. Results:. The histology revealed granulation tissue with numerous large
Thirteen total hip replacements with titanium alloy femoral components required revision for loosening at an average of two years after implantation. At revision the soft tissues around the implant were darkly stained and a proliferative membrane had invaded the cement-bone interface. The femoral components showed polishing of parts of their shot-blasted surfaces. Histology showed a fibroblastic reaction with abundant titanium lying free and within
We treated 75 patients with benign bone tumours by curettage and filling the defect with calcium hydroxyapatite (HA). There were 28 women and 47 men with a mean age of 27.7 years (3 to 80). The mean follow-up was for 41.3 months. Postoperative radiological assessment revealed that the implanted HA was well incorporated into the surrounding host bone in all patients. Two patients suffered fractures in the postoperative period. Two patients complained of pain associated with HA in the soft tissues, but this diminished within six months. No patient had local pain at the final follow-up. Recurrence of the tumour was seen in three cases. Histopathological study of the implanted area showed removal of the HA by
Introduction. Silicon nitride (SiN) is a recently introduced bearing material for THR that has shown potential in its bulk form and as a coating material on cobalt-chromium (CoCr) substrates. Previous studies have shown that SiN has low friction characteristics, low wear rates and high mechanical strength. Moreover, it has been shown to have osseointegration properties. However, there is limited evidence to support its biocompatibility as an implant material. The aim of this study was to investigate the responses of peripheral blood mononuclear cells (PBMNCs) isolated from healthy human volunteers and U937 human
The aim of this study was to report the long-term prognosis of patients with multiple Langerhans cell histiocytosis (LCH) involving the spine, and to analyze the risk factors for progression-free survival (PFS). We included 28 patients with multiple LCH involving the spine treated between January 2009 and August 2021. Kaplan-Meier methods were applied to estimate overall survival (OS) and PFS. Univariate Cox regression analysis was used to identify variables associated with PFS.Aims
Methods
In our independent centre, from 2002 to 2009, 155 BHRs (mean F/U 60 months) have been implanted as well as 420 ASR resurfacings and 75 THRs using ASR XL heads on SROM stems (mean F/U 35). During this period we have experienced a number of failures with patients complaining of worsening groin pain at varying lengths of time post operatively. Aspiration of the hip joints yielded a large sterile effusion on each occasion. At revision, there were copious amounts of green grey fluid with varying degrees of necrosis. There were 17 failures of this nature in patients with ASR implants (12 females) and 0 in the BHR group. This amounts to a failure of 3.5% in the ASR group. Tissue specimens from revision surgery showed varying degrees of “ALVAL” as well as consistently high numbers of
In our independent centre, in the period from January 2003 to august 2008, over 1100 36mm MoM THRs have been implanted as well as 155 Birmingham Hip Resurfacing procedures, 402 ASR resurfacings and 75 THRs using ASR XL heads on SROM stems. During this period we have experienced a number of failures with patients complaining of worsening groin pain at varying lengths of time post operatively. Aspiration of the hip joints yielded a large sterile effusion on each occasion. At revision, there were copious amounts of green grey fluid with varying degrees of necrosis. There were 11 failures of this nature in patients with ASR implants (10 females) and 2 in the 36 MoM THR group (one male one female). Tissue specimens from revision surgery showed varying degrees of ‘ALVAL’ as well as consistently high numbers of
Langerhans-cell histiocytosis (LCH) is a reactive proliferative disease characterized by the accumulation of abnormal
We analysed revised Mathys isoelastic polyacetal femoral stems with stainless-steel heads and polyethylene acetabular cups from eight patients in order to differentiate various types of particle of wear debris. Loosening of isoelastic femoral stems is associated with the formation of polyacetal wear particles as well as those of polyethylene and metal. All three types of particle were isolated simultaneously by tissue digestion followed by sucrose gradient centrifugation. Polyacetal particles were either elongated, ranging from 10 to 150 μm in size, or shred-like and up to 100 μm in size. Polyethylene particles were elongated or granules, and were typically submicron or micronsized. Polyacetal and polyethylene polymer particles were differentiated by the presence of BaSO. 4. , which is added as a radiopaque agent to polyacetal but not to polyethylene. This was easily detectable by back-scattered SEM analysis and verified by energy dispersive x-ray analysis. Two types of foreign-body giant cell (FBGC) were recognised in the histological specimens. Extremely large FBGCs with irregular polygonal particles showing an uneven, spotty birefringence in polarised light were ascribed to polyacetal debris. Smaller FBGCs with slender elongated particles shining uniformly brightly in polarisation were related to polyethylene. Mononucleated
Aims: Wood is a product of nature, has a structural architecture resembling bone and is chemically polymer-like. Birchwood modified with heat and humid air was selected to study its possibilities for bone reconstruction. Methods: Bulk birchwood was prepared for 2–3 hours at temperature of 220°C in humid air, this modifies the wood chemically and physically. 16 cone shaped implants 7x4 mm in size were carved from the heat treated material (Bioactive Wood Bone, BWB) and implanted by press-fit technique into holes drilled in the distal femurs of rabbits. Untreated cones served as controls. The resected knees were embedded in plastic (Techmont, Kulzer GmBH). For evaluation histology, histomorfometry and scanning electron microscopy (SEM) were carried out. Results: In vitro SEM showed the canal structures of the wood. In vivo no articular hydrops or wound infections were seen. At 4 and 8 weeks an inflammatory cellular reaction of a mild degree with some
Metallosis is a combined chemical and toxic reaction which, if the wear of a metal implant is large, may cause extensive reaction of synovial membrane and thus triggering the loosening. We present a case of a 72 year-old man, who underwent to a cemented unicompartimental porous metal coated knee implant because affected by rheumatoid arthritis complicated by osteonecrosis of medial femoral condyle of the knee. Four years after replacement, the patient presented symptoms included moderate swelling, pain, synovitis inability to bear full weight as well as grinding; plain radiographs shows well fixed implant and not finding of loosening of prosthesis; arthroscopy revealed the diagnosis of metallosis. The specimens of synovial tissue were prepared to observation to light and electron microscopy. Total synovialectomy and revison with total knee replacement were successful in relieving the symptoms. Arthroscopy examination revealed a posterior break of tibial component, source of the release of multiple metal beads; we observe alsogray black discoloration of hypertofic and hyperplastic synovium pannus like; metal beads were detected in the joint space soft tissue and were also embedded in the articulating surface of the tibia component. Microscopic examination shows metal debris as black aggregates and a diffuse sheet like proliferation inside
The pathogenesis of aseptic loosening of total joint prostheses is not clearly understood. Two features are associated with loosened prostheses, namely, particulate debris and movement of the implant. While numerous studies have evaluated the cellular response to particulate biomaterials, few have investigated the influence of movement of the implant on the biological response to particles. Our aim was therefore to test the hypothesis that excessive mechanical stimulation of the periprosthetic tissues induces an inflammatory response and that the addition of particulate biomaterials intensifies this. We allocated 66 adult Beagle dogs to four groups as follows: stable implants with (I) and without (II) particulate polymethylmethacrylate (PMMA) and moving implants with (III) and without (IV) particulate PMMA. They were then evaluated at 2, 4, 6, 12 and 24 weeks. The stable implants were well tolerated and a thin, fibrous membrane of connective tissue was observed. There was evidence of positive staining in some cells for interleukin-6 (IL-6). Addition of particulate PMMA around the stable implants resulted in an increase in the fibroblastic response and positive staining for IL-6 and tumour necrosis factor-alpha (TNF-α). By contrast, movement of the implant resulted in an immediate inflammatory response characterised by large numbers of
Introduction: There is little data available about numerical analysis of polyethylene particles from regions with different degrees of osteolysis in aseptic loosening of total hip replacements. Hence, it was the purpose of the present study to investigate, whether particles from large ostolitic lesions are different in size or shape to particles from regions with little or no osteolysis. Methods: during hip revision surgery tissue samples from regions with maximal and minimal osteolysis at the stem and acetabulum were collected in five patients. The samples were examined histologically and numerically for each region. The polyethylene particles were isolated from one gram soft tissue by papain digestion and analyzed with a scanning electron microscope (SEM). Size, elongation, area, form factor and perimeter were calculated for a total of 6526 particles. Results: The histological examination showed significantly more lymphocytes (p <
0,001),
Early failure associated with adverse reactions to metal debris is an emerging problem after hip resurfacing but the exact mechanism is unclear. We analysed our entire series of 660 metal-on-metal resurfacings (Articular Surface Replacement (ASR) and Birmingham Hip Resurfacing (BHR)) and large-bearing ASR total hip replacements, to establish associations with metal debris-related failures. Clinical and radiological outcomes, metal ion levels, explant studies and lymphocyte transformation tests were performed. A total of 17 patients (3.4%) were identified (all ASR bearings) with adverse reactions to metal debris, for which revision was required. This group had significantly smaller components, significantly higher acetabular component anteversion, and significantly higher whole concentrations of blood and joint chromium and cobalt ions than asymptomatic patients did (all p <
0.001). Post-revision lymphocyte transformation tests on this group showed no reactivity to chromium or cobalt ions. Explants from these revisions had greater surface wear than retrievals for uncomplicated fractures. The absence of adverse reactions to metal debris in patients with well-positioned implants usually implies high component wear. Surgeons must consider implant design, expected component size and acetabular component positioning in order to reduce early failures when performing large-bearing metal-on-metal hip resurfacing and replacement.
Adverse reaction to wear and corrosion debris
is a cause for concern in total hip arthroplasty (THA). Modular junctions
are a potential source of such wear products and are associated
with secondary pseudotumour formation. We present a consecutive series of 17 patients treated at our
unit for this complication following metal-on-highly cross-linked
polyethylene (MoP) THA. We emphasise the risk of misdiagnosis as
infection, and present the aggregate laboratory results and pathological
findings in this series. The clinical presentation was pain, swelling or instability.
Solid, cystic and mixed soft-tissue lesions were noted on imaging
and confirmed intra-operatively. Corrosion at the head–neck junction
was noted in all cases. No bacteria were isolated on multiple pre-
and intra-operative samples yet the mean erythrocyte sedimentation
rate was 49 (9 to 100) and C-reactive protein 32 (0.6 to 106) and
stromal polymorphonuclear cell counts were noted in nine cases. Adverse soft–tissue reactions can occur in MoP THA owing to corrosion
products released from the head–neck junction. The diagnosis should
be carefully considered when investigating pain after THA. This
may avoid the misdiagnosis of periprosthetic infection with an unidentified
organism and mitigate the unnecessary management of these cases
with complete single- or two-stage exchange. Cite this article:
The LockDown device (previously called Surgilig)
is a braided polyester mesh which is mostly used to reconstruct the
dislocated acromioclavicular joint. More than 11 000 have been implanted
worldwide. Little is known about the tissue reaction to the device
nor to its wear products when implanted in an extra-articular site
in humans. This is of importance as an adverse immunological reaction
could result in osteolysis or damage to the local tissues, thereby affecting
the longevity of the implant. We analysed the histology of five LockDown implants retrieved
from five patients over the last seven years by one of the senior
authors. Routine analysis was carried out in all five cases and
immunohistochemistry in one. The LockDown device acts as a scaffold for connective tissue
which forms an investing fibrous pseudoligament. The immunological
response at the histological level seems favourable with a limited
histiocytic and giant cell response to micron-sized wear particles.
The connective tissue envelope around the implant is less organised
than a native ligament. Cite this article: