Bone & Joint Open
Vol. 2, Issue 8 | Pages 599 - 610
1 Aug 2021
Hothi H Bergiers S Henckel J Iliadis AD Goodier WD Wright J Skinner J Calder P Hart AJ

Aims. The aim of this study was to present the first retrieval analysis findings of PRECICE STRYDE intermedullary nails removed from patients, providing useful information in the post-market surveillance of these recently introduced devices. Methods. We collected ten nails removed from six patients, together with patient clinical data and plain radiograph imaging. We performed macro- and microscopic analysis of all surfaces and graded the presence of corrosion using validated semiquantitative scoring methods. We determined the elemental composition of surface debris using energy dispersive x-ray spectroscopy (EDS) and used metrology analysis to characterize the surface adjacent to the extendable junctions. Results. All nails were removed at the end of treatment, having achieved their intended lengthening (20 mm to 65 mm) and after regenerate consolidation. All nails had evidence of corrosion localized to the screw holes and the extendable junctions; corrosion was graded as moderate at the junction of one nail and severe at the junctions of five nails. EDS analysis showed surface deposits to be chromium rich. Plain radiographs showed cortical thickening and osteolysis around the junction of six nails, corresponding to the same nails with moderate – severe junction corrosion. Conclusion. We found, in fully united bones, evidence of cortical thickening and osteolysis that appeared to be associated with corrosion at the extendable junction; when corrosion was present, cortical thickening was adjacent to this junction. Further work, with greater numbers of retrievals, is required to fully understand this association between corrosion and bony changes, and the influencing surgeon, implant, and patient factors involved. Cite this article: Bone Jt Open 2021;2(8):599–610


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_10 | Pages 87 - 87
1 Oct 2022
Puetzler J Hasselmann J Gosheger G Niemann S Fobker M Hillebrand J Schwarze J Theil C Schulze M
Full Access

Aim. A novel anti-infective biopolymer implant coating was developed to prevent bacterial biofilm formation and allow on-demand burst release of anti-infective silver (Ag) into the surrounding of the implant at any time after surgery via focused high-energy extracorporeal shock waves (fhESW). Method. A semi-crystalline Poly-L-lactic acid (PLLA) was loaded with homogeneously dissolved silver (Ag) applied onto Ti6Al4V discs. A fibroblast WST-1 assay was performed to ensure adequate biocompatibility of the Ag concentration at 6%. The prevention of early biofilm formation was investigated in a biofilm model with Staphylococcus epidermidis RP62A after incubation for 24 hours via quantitative bacteriology. In addition, the effect of released Ag after fhESW (Storz DUOLITH SD1: 4000 impulses, 1,24 mJ/mm. 2. , 3Hz, 162J) was assessed via optical density of bacterial cultures (Escherichia coli TG1, Staphylococcus epidermidis RP62A, Staphylococcus aureus 6850) and compared to an established electroplated silver coating. The amount of released Ag after the application of different intensities of fhESW was measured and compared to a control group without fhESW via graphite furnace atomic absorption spectrometry (GF-AAS), scanning electron microscopy (SEM) and energy dispersive X-ray spectroscopy (EDS). Results. The coating with 6% Ag reduced Staphylococcus epidermidis biofilm formation by 99.7% (mean±SD: 2.1×10^5 ± 3,9×10^5 CFU/µL) compared to uncoated controls (6.8×10^7 ± 4.9×10^7 CFU/µL); (p=0.0001). After applying fhESW the commercially available electroplated silver coating did not prevent the growth of all tested bacterial strains. Bacterial growth is delayed with 4% Ag and completely inhibited with 6% Ag in the novel coating, except for a small increase of S. aureus after 17 hours. SEM and EDS confirmed a local disruption of the coating after fhESW. Conclusions. This novel anti-infective implant coating has the potential to prevent bacterial biofilm formation. The on-demand burst release of silver via fhESW could be an adjunctive in the treatment of implant related infection and is of particular interest in the concept of single stage revision surgery


Bone & Joint Research
Vol. 10, Issue 7 | Pages 425 - 436
16 Jul 2021
Frommer A Roedl R Gosheger G Hasselmann J Fuest C Toporowski G Laufer A Tretow H Schulze M Vogt B

Aims. This study aims to enhance understanding of clinical and radiological consequences and involved mechanisms that led to corrosion of the Precice Stryde (Stryde) intramedullary lengthening nail in the post market surveillance era of the device. Between 2018 and 2021 more than 2,000 Stryde nails have been implanted worldwide. However, the outcome of treatment with the Stryde system is insufficiently reported. Methods. This is a retrospective single-centre study analyzing outcome of 57 consecutive lengthening procedures performed with the Stryde nail at the authors’ institution from February 2019 until November 2020. Macro- and microscopic metallographic analysis of four retrieved nails was conducted. To investigate observed corrosion at telescoping junction, scanning electron microscopy (SEM) and energy dispersive x-ray spectroscopy (EDX) were performed. Results. Adjacent to the nail’s telescoping junction, osteolytic changes were observed in bi-planar radiographs of 20/57 segments (35%) after a mean of 9.5 months (95% confidence interval 7.2 to 11.9) after surgery. A total of 8/20 patients with osseous alterations (40%) reported rest and ambulation pain of the lengthened segment during consolidation. So far, 24 Stryde nails were retrieved and in 20 (83%) macroscopic corrosion was observed at the nail’s telescoping junction. Before implant removal 11/20 radiographs (55%) of lengthened segments with these 20 nails revealed osteolysis. Implant retrieval analysis by means of SEM showed pitting and crevice corrosion. EDX detected chromium as the main metallic element of corrosion. Conclusion. Patients are exposed to the risk of implant-related osteolysis of unclear short- and long-term clinical consequences. The authors advocate in favour of an early implant removal after osseous consolidation. Cite this article: Bone Joint Res 2021;10(7):425–436


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_6 | Pages 123 - 123
1 Jul 2020
J.Dixon S Beaucage K Nagao M Lajoie G Veras M Fournier D Holdsworth D Bailey C Hammond J Séguin C
Full Access

Equilibrative nucleoside transporter 1 (ENT1) transfers nucleosides, such as adenosine, across plasma membranes. We reported previously that mice lacking ENT1 (ENT1-KO) exhibit progressive ectopic calcification of spinal tissues, including the annulus fibrosus (AF) of intervertebral discs (J Bone Miner Res 28:1135–49, 2013, Bone 90:37–49, 2016). Our purpose was twofold: (1) to compare ectopic calcifications in ENT1-KO mice with those in human DISH, and (2) to investigate the molecular pathways underlying pathological calcification in ENT1-KO mice. Studies were performed with age-matched wild-type (WT) and ENT1-KO mice, as well as human cadaveric vertebral columns meeting radiographic criteria for DISH. Mouse and human specimens were scanned using high-resolution, micro-computed tomography (micro-CT). As well, some samples were decalcified and processed for histological assessment. Calcified lesions in selected specimens were examined using energy dispersive X-ray spectroscopy (EDX) and X-ray diffraction (XRD). To investigate molecular changes associated with ectopic calcification, we isolated AF tissue from thoracic intervertebral discs of WT and ENT1-KO mice. Tissues were then subjected to transcriptomic and proteomic analyses. Micro-CT of ENT1-KO mice revealed ectopic calcification of spinal tissues, first appearing in the cervical-thoracic region and extending caudally with advancing age. Histological examination of calcified lesions in mice revealed accumulations of amorphous, eosinophilic, acellular material in paraspinal ligaments and entheses, intervertebral discs, mandibular symphysis, and sternocostal articulations. There was no evidence of inflammation associated with these lesions. EDX of calcified lesions revealed a high content of calcium and phosphorus in a molar ratio of ∼1.6, with hydroxyapatite detected by micro-XRD. Ten human cadaveric spines (three females and seven males, mean age 81 years) that met radiographic criteria for DISH were analysed in detail by micro-CT. Remarkable heterogeneity in the density and morphology of ectopic calcifications was observed. Analyses of calcifications by EDX and XRD again yielded a calcium/phosphorus ratio of ∼1.6 and a crystalline diffraction pattern matching hydroxyapatite. Histological examination of human lesions revealed regions of mature ossification and other areas of irregular amorphous calcification that resembled lesions in ENT1-KO mice. Microarray analysis of AF tissue from WT and ENT1-KO mice showed extensive dysregulation of transcription in affected tissues. Cell cycle-associated transcripts were the most affected, including the E2f family of transcription factors and proliferating cell nuclear antigen. In addition, expression of genes involved in the regulation of mineralization and bone development were dysregulated. Proteomic analyses confirmed transcriptomic changes and revealed alterations in known modulators of biomineralization such as matrix Gla-protein. Many of the characteristics of ectopic calcification in ENT1-KO mice resemble those of DISH in humans. Human lesions were found to be heterogeneous with regions of pathological ossification and amorphous calcification, the latter resembling lesions in the mouse model. Our studies of the molecular events associated with ectopic calcification in ENT1-KO mice may provide insights into the pathogenesis of DISH in humans. ENT1-KO mice may also be useful for evaluating therapeutics for the prevention of ectopic calcification in DISH and related disorders


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 115 - 115
1 Apr 2019
Bock R Cullen D McEntire B More K Leonard D Bal BS
Full Access

Introduction. The osteogenic capability of any biomaterial is governed by a number of critical surface properties such as surface energy, surface potential, and topography. Prior work suggested that the Si-Y-O-N phase(s) present in the form of a thin (<150 nm), interrupted film at the surface of an annealed silicon nitride bioceramic may be responsible for an observed upregulation of osteoblastic activity due to passive surface properties and dissolution of chemical species. In this study high- resolution analytical electron microscopy was utilized to identify the Si-Y-O-N phase present on the annealed silicon nitride surface, and dissolution studies were employed to elucidate mechanisms of the material's favorable cell interactions. Materials and Methods. Si. 3. N. 4. discs (12.7 mm diameter × 1 mm thick) containing Y. 2. O. 3. and Al. 2. O. 3. sintering aids were processed using conventional techniques and subsequently subjected to annealing in a nitrogen atmosphere. Pre-cultured SaOS-2 osteosarcoma cells at a concentration of 5 × 10. 5. cells/ml were seeded onto sterile polished nitrogen-annealed Si. 3. N. 4. discs in an osteogenic medium consisting of DMEM supplemented with about 50 µg/mL ascorbic acid, 10 mM β-glycerol phosphate, 100 mM hydrocortisone, and 10% fetal bovine calf serum. The samples were incubated for up to 7 days at 37°C with two medium replenishments. Transmission electron microscopy (TEM) images were acquired from focused ion beam (FIB)-prepared samples using a Hitachi HF-3300 TEM (300 kV). Scanning transmission electron microscopy (STEM) images were recorded using a Nion UltraSTEM 100 (60 kV). STEM high-angle annular dark-field (HAADF) imaging and energy dispersive X-ray spectroscopy (EDS) analyses were performed on a JEOL JEM2200FS (200 kV) equipped with a third-order CEOS aberration corrector and a Bruker XFlash silicon drift detector. Results. A cross-section of the of the Si. 3. N. 4. /extracellular polymer (ECP) interface is illustrated in Fig. 1(a)∼(b) as a high- angle annular dark field (HAADF) STEM image (a) with and EDS map overlay (b) highlighting locations of Ca, Y, and Si. The underlying Si. 3. N. 4. microstructure is covered by a yttrium-rich intergranular phase (IGP) film. Deposition of cell-derived hydroxyapatite (HAp) occurred directly onto this IGP film. In Fig. 2, a bright field TEM image (electron diffraction pattern inset) shows the interface between the partially-crystalline HAp and the Y-Si-O-N phase, identified as monoclinic yttrium disilicate (i.e., m-Y. 2. Si. 2. O. 7. ) with a 2 atomic% N impurity, at teh atomic scale. Although rapid electron damage of the mineralized ECP was observed, EDS analyses suggested a Ca/P ratio of ∼1.43, along with the incorporation of Si. Conclusions. The osteogenic Si-Y-O-N phase was successfully identified as a minority concentration of Si. 3. N. 4. dissolved into a m-Y. 2. Si. 2. O. 7. matrix. Evidence of the release of (SiO. 4. ). 4−. tetrahedra from this phase into the local biological microenvironment and their incorporation into the cell-derived HAp layer was also observed. Identification of this phase paves the way for ongoing work to understand and optimize this novel biomaterial. For any figures or tables, please contact the authors directly