Advertisement for orthosearch.org.uk
Results 1 - 20 of 175
Results per page:
Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 318 - 318
1 Dec 2013
Gardner A Angibaud L Stroud N
Full Access

Introduction. Ideally, a patient receiving a unicondylar knee replacement will have fully functional anterior and posterior cruciate ligaments. When at least one of the cruciate ligaments is not fully functional, femoral and tibial implant contact position can potentially increase along the anterior-posterior (AP) axis. Where unicondylar implant wear testing typically uses AP resistance assuming fully functional cruciate ligaments, the authors used reduced AP resistance intended to simulate deficient cruciate ligaments. Methods. Optetrak Logic® Uni (Exactech Inc, Gainesville, FL USA) unicondylar test specimens featuring an all-UHMWPE tibial component and a cobalt chromium femoral component were used in this study. The system has a semi-constrained articular geometry. Testing was conducted at an independent testing facility (EndoLab GMBH, Thansau, Rosenheim, Germany). A four-station knee simulator was used (EndoLab knee simulator) with two unicondylar knee implants per station, giving a total of eight test specimens. Two different tibial fixation designs (keeled and peg) with identical articulating surfaces were tested. Tibial test specimens were 6 mm in thickness. Unloaded soak controls were stored in distilled water at 37°C. The test was conducted according to ISO 14243–1: 2009 [1]. Test specimens were immersed in calf serum (PAA GmBH, Cölbe, LOT B00111-5126) with a protein content of 20 g/l. Custom polyurethane molds allowed for individual component measurement. Per the ISO 14243-1, a 7% medial offset was incorporated into the set-up. The unicondylar knee implants were set at neutral position in extension. Tibial rotational restraint was 0.36 Nm/° and zero when the test specimen was within ± 6° of the reference position. This test was conducted with an AP resistance of 9.3N/mm to maximize AP displacement and simulate deficient cruciate ligaments. Typical unicondylar knee wear testing is conducted with an AP resistance of 44N/mm, which assumes functional cruciate ligaments. Results. Wear data was separated by component design (keeled and peg) as well as for medial and lateral placement [Table 1]. There was no significant difference between lateral components but there was for medial components. This difference could be due to the small sample size. Contact area of the UHMWPE tibial components was elliptical, with the longer portion along the AP axis. Mean wear rates were comparable to historical unicondylar knee systems tested at the same laboratory using the standard AP resistance (i.e., 44 N/mm). Discussion/Conclusion. This study demonstrated using an AP resistance 9.3 N/mm to simulate the presence of deficient cruciate ligaments in a unicondylar knee wear test produced similar wear rates and greater AP displacement when compared to testing using an AP resistance of 44 N/mm, which assumes functioning ligaments. This being said, design and material information about historical unicondylar knee systems tested are not known, so a direct comparison cannot be made. Performing unicondylar knee wear tests with reduced AP resistance could provide realistic wear information for devices implanted in patients without fully functioning cruciate ligaments


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_15 | Pages 87 - 87
7 Nov 2023
Arakkal A Bonner B Scheepers W Van Bornmann R Held M De Villiers R
Full Access

Poor availability of allografts in South Africa has led to an increased use of synthetic augmentation to stabilize knee joints in the treatment of knee dislocations. This study aims to evaluate multiligament knee injuries treated with a posterior cruciate ligament internal brace. The study included patients with knee dislocations who were treated with a PCL internal brace. The internal brace involved the insertion of a synthetic suture tape, which was drilled into the femoral and tibial footprint. Chronic injuries were excluded. Patient-reported outcome scores (PROMs), range of motion, stress X-Rays, and MRI scans were reviewed to assess outcomes. Acceptable outcomes were defined as a Lysholm score of 84 or more, with grade II laxity in no more than one ligament and a range of motion from full extension to 90° or more. The study included eight patients, with a median age of 42, of which five were female. None of the patients had knee flexion less than 90° or an extension deficit of more than 20°. PROMs indicated acceptable outcomes (EQ5D, Tegner Lysholm). Stress radiographs showed less than 7mm (Grade I) of posterior translation laxity in all patients. Four patients underwent MRI scans 1–2 years after the initial surgery, which revealed healing of the PCL in all patients. However, increased signal in a continuous ligament suggested only partial healing in two patients. Tunnel widening of 200% and 250% was noted around the tibial and femoral PCL footprints, respectively. All patients demonstrated stable knees and acceptable PROMs. Tunnel widening was observed in all patients who had MRI scans. Factors such as suspensory fixation, anisometric tunnel position, and the absence of PCL tear repair may have contributed to the tunnel widening


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_5 | Pages 32 - 32
1 Feb 2016
Asseln M Hanisch C Al Hares G Quack V Radermacher K
Full Access

The consideration of the individual knee ligament attachments is crucial for the application of patient specific musculoskeletal models in the clinical routine, e.g. in knee arthroplasty. Commonly, the pre-operative planning is based on CT images, where no soft tissue information is available. The goal of this study was to evaluate the accuracy of a full automatic and robust mesh morphing method that estimates locations of cruciate ligament attachments on the basis of training data. The cruciate ligament attachments from 6 (n=6) different healthy male subjects (BH 184±6cm, BW 90±10kg) were identified in MRI-datasets by a clinical expert. The insertion areas were exported as point clouds and the centres of gravitation served as approximations of the attachments. These insertion points were used to annotate mean shapes of femur and tibia. The mean shapes were built up from 332 training data sets each. The surface data were obtained from CT scans by performing an automatic segmentation followed by manual cleaning steps. The mean shapes were computed by selecting a data set randomly and aligning this reference rigidly to each of the remaining data sets. The data were fitted using the non-rigid ICP variant (N-ICP-A). Due to this morphing step, point correspondences were established. By morphing a mean shape to the target geometries, including the cruciate ligament attachments, the distribution of the insertions on the original mean shape was obtained. Subsequently, a statistical mean was computed (annotated mean). The annotated mean shape was again morphed to the target data sets and the deviations of the respective predicted insertion points from the measured insertion points were computed. The training data was successfully morphed to all 6 subjects in an automatic manner with virtually no distance error (10-5 mm). The mean distance between the measured and morphed ligament attachments was highest for the ACL in the femur (4.26±1.48 mm) and lowest for PCL in the tibia (1.63±0.36 mm). The highest deviation was observed for femoral ACL (6.93 mm). In this study, a morphing based approach was presented to predict origins and insertions of the knee ligaments on the basis of CT-data, exemplarily shown for the cruciate ligaments. It has been demonstrated, that the N-ICP-A is applicable to predict the attachments automatic and robust with a high accuracy. This might help to improve patient-specific biomechanical models and their integration in the clinical routine


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_19 | Pages 33 - 33
1 Dec 2014
van der Merwe W
Full Access

Background:. For the past four decades controversy surrounds the decision to retain or sacrifice the posterior cruciate ligament during a total knee arthroplasty. To our knowledge no study has been done to describe the effect of releasing the PCL on the range of motion of the knee. Study design:. Case series. Methods:. Computer navigation data (Brainlab) was obtained intra-operatively from thirty patients at total knee arthroplasty. Coronal alignment, maximal passive knee extension and maximal passive flexion was captured before and after release of the PCL. Results:. Releasing the posterior cruciate ligament led to an increase in maximal extension in all patients (av 3,6°) and a decrease in coronal deformity in 63%. The surprising finding was an increase in maximal knee flexion (av 5°, range 0 to 10°.) The increase in maximal flexion was statistically significant. Conclusion:. Sacrificing the posterior cruciate ligament alters the kinematics of the knee and the resultant increase in knee flexion might explain why cruciate sacrificing total knee arthroplasty has superior flexion compared to cruciate retaining designs


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 395 - 395
1 Dec 2013
Lee J Yoon J Lee J
Full Access

To investigate the effectiveness of avulsion fracture of tibial insertion of posterior cruciate ligament using the safe postero-medial approach through analyzing the clinical and radiographic outcomes. We treated 14 cases of acute PCL tibial avulsion fracture with “safe postero-medial approach”. The PCL and avulsion bony fragment was fixed with 1 cannuated screw and washer. The patients were assessed clinically and radiographically at 3 months, 6 months, and 12 months. Clinical examination for each visit included assessment of the knee range of motion, using goniometer and the posterior drawer test. The patients were evaluated according to the Lysholm and Tegner rating scales. Patients were followed-up for 12 to 16 months. X-ray showed that satisfactory reducdtion and bony healing was achieved in all cases. There was no neurovascular complication. All patients had negative posterior drawer tests. Excellent outcomes were reported by all patients with the Lysholm score system. And there was no signicant difference between the Tegner scores before injury and last follow-up. Surgical treatment of acute tibial avulsion fracture of the PCL with this approach can restore the stability and fuction of the joint safely in most patients without neurovascular complication. Therefore “safe postero-medial approach” may be suitable for the treatment of isolated tibial avulsion fracture of the PCL


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 39 - 39
1 Dec 2013
D'Alessio J Eckhoff D Kester M
Full Access

Computational modeling has been used to simulate the natural and prosthetic kinematic and kinetic function in an attempt to compare designs and/or predict a desired motion path from a design. The levels of soft tissue can range from basic ligaments (MCL, LCL, and ACL & PCL) to more complex models. The goal of this study was to evaluate the sensitivity of the Posterior Cruciate ligament in a virtual model and its effects on the kinematic outcome in a commercially available and validated kinematics package (KneeSim, LifeModeler San Clemente, CA). Methods:. KneeSIM is a musculoskeletal modeling environment that is built on the foundation of the ADAMS (MSC Software, Santa Ana CA), a rigid body dynamics solver to compute knee kinematics and forces during a deep knee bend. All parameters are customizable and can be altered by the user. Generic three dimensional models of cruciate retaining components of the femoral, tibial, and patellar are available with the software and were used to provide a common reference for the study. The following parameters were modified for each simulation to evaluate the sensitivity of the PCL in the model: 1) Model without PCL, 2) PCL with default properties, 3) PCL Shifted at femoral origin, 7 mm anterior, 7 mm inferior; tibial origin maintained; 4) PCL with increased stiffness properties (2x default), 5) position in the femur and tibia remained default position and 6) PCL with default properties and location, joint line shifted 4 mm superior. The standard output of tracking the flexion facet center (FFC) motion of the medial and lateral condyles was utilized (Figure 1). Results:. Figure 2 and 3 displays the output of the six conditions tested above. Comparing the curves for the medial and lateral motion show different patterns with the lateral point having more posterior translation than the medial. After approximately 95° of flexion, all cases exhibit an anterior translation in the model. This motion was consistent for all test cases. The model showed no difference with motion either with or without the PCL and with changing the stiffness. Altering the location of the PCL on the femoral insertion had the greatest effect on motion, while shifting the joint line superior was second. The shift of the ligament insertion and changing of the joint line results in the ligament being more parallel to the tibial surface which provides resistance to anterior motion or posterior translation. Discussion:. Although the model was able to discern differences, the inability to highlight variation in motion with and without the PCL suggests that the default parameters are not representative of an experimental or clinical setup. Although it is apparent that KneeSim can be used for comparative assessments between designs, simulations should be designed so as to provide similar boundary conditions. Publications (Colwell et al 2011) did successfully use KneeSim to provide a comparative assessment of CR components; however, only after the default model was altered to match the outcome of the experimental rig. Further analysis of the complexities in soft tissue modeling is warranted


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_III | Pages 156 - 156
1 Feb 2012
Khanduja V Somayaji S Utukuri M Dowd G
Full Access

Objective. The aim of this study was to assess the results of combined arthroscopically assisted posterior cruciate ligament reconstruction and open reconstruction of the posterolateral corner in patients with chronic (3 months or more) symptomatic instability and pain. Patients & methods. A retrospective analysis of all the patients who had a combined reconstruction of the posterior cruciate ligament and the posterolateral corner between 1996 and 2003 was carried out. Nineteen patients who had the combined reconstruction were identified from the database. All the patients were assessed pre- and post-operatively by physical examination and three different ligament rating scores. All the patients also had weight bearing radiographs, MRI scans and an examination under anaesthesia and arthroscopy pre-operatively. The PCL reconstruction was performed using an arthroscopically assisted single anterolateral bundle technique and the posterolateral corner structures were reconstructed using an open Larson type of tenodesis. Results. Pre-operatively all the patients had a grade III posterior sag and demonstrated more than 20 degrees of external rotation as compared to the opposite normal knee on the Dial test. The average follow-up was 66.8 months (range 24 -108). Post-operatively 7 patients had no residual posterior sag, 11 patients had a grade I posterior sag and 1 patient had a grade II posterior sag. Five of the 19 patients demonstrated minimal residual posterolateral laxity. The Lysholm score improved from a mean of 41.2 to 76.5 (P=0.0001) and the Tegner score from a mean of 2.6 to 6.4 (p=0.0001). Conclusions. We conclude that while a combined reconstruction of chronic posterior cruciate ligament and posterolateral corner instability does not restore complete anatomical stability, improvement in symptoms and function demonstrate its value in these difficult injuries


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 36 - 36
1 Mar 2017
Takagi T Maeda T Kabata T Kajino Y Yamamoto T Ohmori T
Full Access

Introduction. Compared with the cruciate-retaining (CR) insert for total knee arthroplasty (TKA), the cruciate-substituting (CS) insert has a raised anterior lip, providing greater anterior constraint, and thus, can be used in cases of posterior cruciate ligament (PCL) sacrifice. However, studies have shown that the PCL maintains femoral rollback during flexion, acts as a stabilizer against distal traction force and aids knee joint proprioception; therefore, the argument for PCL excision in CS TKA remains controversial. The purpose of this study was to analyze CS TKA kinematics and identify the role of the PCL. Methods. Seven fresh-frozen lower-extremity cadaver specimens were analyzed using Orthomap. ®. Precision Knee Navigation software (Stryker Orthopaedics, Mahwah, NJ, USA). They were surgically implanted with Triathlon. ®. components (Stryker Orthopaedics). The CS insert has a raised anterior lip, and the posterior geometry shares the same profile as the CR, so we can choose retaining or sacrificing the PCL. Six patterns were analyzed: (1) natural knee; (2) only anterior cruciate ligament excision; (3) CS TKA, PCL retention, and bony island preservation; (4) CS TKA, PCL retention, and bony island resection; (5) CS TKA and PCL excision; and (6) CR TKA and PCL excision. Center of the knee and center of the proximal tibia were registered using navigation system, and the magnitudes of the condylar translation were evaluated. And then, using trigonometric function, the magnitude of anterior-posterior translation of the femur was calculated. Results. PCL excision patterns showed that the magnitude of anterior-posterior (AP) translation was higher in mid-flexion and lower in deep flexion than in other patterns (Fig. 1). Comparing two PCL excision patterns, in CS insert, the anterior translation magnitude was a little lower in extension and 30° flexion. Comparing two PCL retention patterns, the both posterior translation magnitudes in deep flexion were comparable to that of the natural knee. Discussion. Very few studies have reported about comparison of PCL retention with PCL excision in CS TKA. Omori et al. evaluated the medial pivot type TKA, and found that the design showed no femoral rollback under the PCL-sacrificing condition. In our study, increased anterior translation magnitudes in mid-flexion indicated paradoxical roll-forward, and decreased posterior translation magnitudes in deep flexion indicated decreased rollback. In other words, PCL excision in CS TKA caused mid-flexion instability and decreased the femoral rollback, so raised anterior lip was not likely to contribute to TKA kinematics. Another research is necessary to evaluate the effects of the raised anterior lip. On the other hand, PCL retention in CS TKA maintained physiological femoral rollback. The AP translation magnitude was not dependents on the bony island. Conclusions. We had better retain the PCL in raised anterior lip type CS TKA to ensure physiological knee kinematics. For any figures or tables, please contact authors directly (see Info & Metrics tab above).


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 582 - 582
1 Dec 2013
Weijia C Nagamine R Kondo K Osano K
Full Access

INTRODUCTION:. In varus knee, posterior cruciate ligament (PCL) release has been reported to result in the increase of the flexion gap without significant effect on the extension gap. However, the effect of release on gap angle is still obscure. On the other hand, gap angle and distance measured with the tension devices may vary due to different distraction forces. In this study, difference of gap angle and distance before and after PCL resection in knee extension and 90° flexion was inspected. Effect of different distraction force on gap was also assessed. OBJECTIVES:. Fifty cases with medial osteoarthritis undergoing PS-TKA were included in the study. PCL of all the cases were identified intact before resection. METHODS:. After distal femoral and proximal tibial cuts were initially performed and anterior cruciate ligament (ACL) was excised, joint gap angle and distance in full extension and at 90° flexion were obtained by means of a tensioning device, Balancer under 10, 20 and 30 inch-pounds distraction. The gap angle and distance were measured 3 times at each step. Then PCL was excised and the same measurement was performed. RESULTS:. The gap distances significantly enlarged following the increased distraction forces in both knee extension and 90° flexion, before and after the PCL resection (p < 0.0001). The distances were the same before and after the PCL resection in extension under each force. However, after the PCL was resected, the joint distance at 90° flexion significantly enlarged at 10, 20 and 30 inch-pounds for 1.2, 1.6 and 1.8 mm, respectively (p < 0.001). All measured joint gap angles showed varus results. The gap angles significantly increased following the increased distraction forces in both knee extension and 90° flexion, before and after the PCL resection (p < 0.0001). The mean gap angles in extension were 1.9°, 2.9° and 3.6° for 10, 20 and 30 inch-pounds, respectively. The angles significantly decreased after PCL resection in extension at different forces for 0.6°, 0.4° and 0.3°, respectively (p < 0.05). In 90° flexion, the mean angles were 2.9°, 5.5° and 7.3°, respectively, and were decreased after the resection for 0.3° (p = 0.9), 0.6° (p = 0.01) and 1.1° (p = 0.00024). CONCLUSION:. In this study, all the distances and angles had a correlation with the distraction powers of the tensioner. For the joint gap distance, the distance in extension was not influenced by the release of the PCL, but was enlarged at 90° flexion after the release. We found that the varus gap angles were decreased at both extension and flexion. Thirty inch-pounds seem to have more effect on the gap angle than 20 inch-pounds at 90° flexion after the release. It indicated that the evaluation of the joint gap might not be accurate if the distraction power of the tension device is not appropriate. Our study indicated that in varus deformity osteoarthritic knees, PCL resection may have influence on the correction of deformity. The means of modified gap control method and measured bone resection method, and the character of joint gap in CR, CS or PS TKA should be considered independently


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 358 - 358
1 Dec 2013
Jonishi K Kaneyama R Shiratsuchi H Oinuma K Miura Y Higashi H Tamaki T
Full Access

Introduction. In posterior cruciate ligament (PCL)-preserving total knee arthroplasty (TKA), it is important to determine whether the PCL is properly functioning after surgery. As the PCL is partly damaged during the operation, we cannot rule out the possibility that excessive tension further damages the remaining PCL resulting in dysfunction or that initial functioning of the PCL is lost due to excessively low tension. However, it is normally difficult to examine whether the PCL has remained intact and is still functional after TKA. The objective of this study was to visualize knee joint flexion after TKA by MRI and evaluate the PCL based on these images. Method. PCL-preserving TKA was performed in 41 knees using the Fine Total Knee System® (Nakashima Medical, Okayama, Japan) where a titanium component can be selected for both the femur and the tibia. We visualized knee flexion positions by MRI at 6 months after surgery and evaluated visualization or non-visualization of the PCL, the relationship between knee flexion angle and PCL elevation angle against the plane of the tibial joint vertical to the tibial axis, and the forms of PCL based on the MRI data. Results. The PCL was visualized in 40 of the 41 knees. These 40 knees showed a strong positive correlation (correlation coefficient 0.85) between the knee flexion angle (mean 95.8 degrees, 59 to 129 degrees) and the PCL elevation angle (mean 60.4 degrees, 38 to 79 degrees) by MRI. As the PCL was visualized as a straight line in 6 of 13 knees at a knee joint flexion angle of less than 90 degrees, sufficient tension was considered to be transmitted; however, 7 knees showed superior protrusion or S-shaped forms, indicating that the tension in the PCL was not strong. No superior protrusion of the PCL was observed in 27 knees at the flexion angle of 90 degrees or more; 19 knees showed straight-line forms and 7 knees showed inferior protrusion due to posterior pressure from the femur, and the flexion angle was 105 degrees or greater in all knees with inferior protrusion. At the knee flexion angle of 90 degrees or greater, the tension in the PCL was confirmed in 26 knees (96%) by MRI. Conclusion. To date, there have been no morphological evaluations of postoperative PCL in PCL-preserving TKA. While tension in the PCL was determined to be insufficient in some knees at the knee flexion angle of less than 90 degrees, the elevation angle of the PCL against the tibia increased with tension as the knee flexion angle increased. Postoperative MRI indicated that the PCL functions as a stabilizer between the femur and the tibia in knees that have undergone PCL-preserving TKA, especially at the knee flexion angle of 90 degrees or greater


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XLIV | Pages 102 - 102
1 Oct 2012
Petrigliano F Suero E Lane C Voos J Citak M Allen A Wickiewicz T Pearle A
Full Access

Injuries to the posterior cruciate ligament (PCL) and the posterolateral corner (PLC) of the knee remain a challenging orthopaedic problem. Studies evaluating PCL and PLC reconstruction have failed to demonstrate a strong correlation between the degree of knee laxity as measured by uniplanar testing and subjective outcome or patient satisfaction. The effect that changing the magnitude of posterior tibial slope has on multiplanar, rotational stability of the PCL-deficient knee has yet to be determined. We aimed to evaluate the effect that changes in posterior tibial slope would have on static and dynamic stability of the PCL-PLC deficient knee. Ten knees were used for this study. Navigated posterior drawer and standardised reverse mechanised pivot shift maneuvers were performed in the intact knee and after sectioning the PCL, the lateral collateral ligament (LCL), the popliteofibular ligament (PFL) and the popliteus muscle tendon (POP). Navigated high tibial osteotomy (HTO) was performed to obtain the desired change in tibial plateau slope (+5® or −5® from native slope). We then repeated the posterior drawer and the reverse mechanised pivot shift test for each of the two altered slope conditions. Mean posterior tibial translation during the posterior drawer in the intact knee was 1.4 mm (SD = 0.48 mm). In the PCL-PLC deficient knee, posterior tibial translation increased to 18 mm (SD = 5.7 mm) (P < 0.001). Increasing the amount of posterior tibial slope by 5® reduced posterior tibial translation to 12 mm (SD = 4.7 mm) (P < 0.01). Decreasing the amount of posterior slope by 5® compared to the native knee, increased posterior tibial translation to 21 mm (SD = 6.8 mm) (P < 0.01). There was a significant negative correlation between the magnitude of tibial plateau slope and the magnitude of the reverse pivot shift (R2 = 0.71; P < 0.0001). Mean posterior tibial translation during the reverse mechanised pivot shift test in the intact knee was 7.8 mm (SD = 2.8 mm). In the PCL-PLC deficient knee, posterior tibial translation increased to 26 mm (SD = 5.6 mm) (P < 0.001). Increasing the amount of posterior tibial slope by 5® reduced posterior tibial translation to 21 mm (SD = 6.7 mm) (P < 0.01). Decreasing the amount of posterior slope by 5® compared to the native knee, increased posterior tibial translation to 34 mm (SD = 8.2 mm) (P < 0.01). There was a significant negative correlation between the magnitude of tibial plateau slope and the magnitude of the reverse pivot shift (R2 = 0.72; P < 0.0001). Decreasing the magnitude of posterior slope of the tibial plateau resulted in an increase in the magnitude of posterior tibial translation during the posterior drawer and the reverse mechanised pivot shift test in the PCL-PLC deficient knee. Conversely, increasing the slope of the tibial plateau reduced the amount of posterior tibial translation during the posterior drawer and the reverse mechanised pivot shift test. However, the effect of the increase in slope was not sufficient to reduce posterior tibial translation to levels similar to those of the intact knee


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_3 | Pages 108 - 108
23 Feb 2023
Lee W Foong C Kunnasegaran R
Full Access

Most studies comparing medial pivot to the posterior stabilised (PS) systems sacrifice the PCL. It is unknown whether retaining the PCL in the Medial Congruent (MC) system may provide further benefit compared to the more commonly used PS system.

A retrospective review of a single-surgeon's registry data comparing 44 PS and 25 MC with PCL retained (MC-PCLR) TKAs was performed.

Both groups had similar baseline demographics in terms of age, gender, body mass index, and American Society for Anaesthesiology score. There was no significant difference in their preoperative range of motion (ROM) (104º±20º vs. 102º±20º,p=0.80), Oxford Knee Score (OKS) (27±6 vs. 26±7,p=0.72), and Knee Society Scoring System (KS) Function Score (KS-FS) (52±24 vs. 56±24,p=0.62). The preoperative KS Knee Score (KS-KS) was significantly lower in the PS group (44±14 vs. 54 ± 18,p<0.05). At 3-months postoperation, the PS group had significantly better OKS (38±6 vs. 36±6,p=0.02) but similar ROM (111º±14º vs. 108º±12º,p=0.25), KS-FS (73±20 vs. 68±23, p=0.32) and KS-KS (87±10 vs. 86±9,p=0.26). At 12-months postoperation, both groups had similar ROM (115º±13º vs. 115º±11º,p=0.99), OKS (41±5 vs. 40±5,p=0.45), KS-FS (74±22vs.78±17,p=0.80), and KS-KS (89±10vs.89±11,p=0.75). There was statistically significant improvement in all parameters at 1-year postoperation (p<0.05). The PS group had significant improvement in all parameters from preoperation to 3-month postoperation (p<0.05), but not from 3-month to 1-year postoperation (p≥0.05). The MC-PCLR group continued to have significant improvement from 3-month to 1-year postoperation (p<0.05).

The MC provides stability in the medial compartment while allowing a degree of freedom in the lateral compartment. Preserving the PCL when using MC may paradoxically cause an undesired additional restrain that slows the recovery process of the patients after TKA.

In conclusion, compared to MC-PCLR, a PS TKA may expect significantly faster improvement at 3 months post operation, although they will achieve similar outcomes at 1-year post operation.


Purpose. To report clinical results and demonstrate any posterior femoral translation (PFT) in medial rotation total knee arthroplasty (TKA) of posterior cruciate ligament (PCL) retaining type. Materials and Methods. A prospective study was performed upon thirty consecutive subjects who were operated on with medial rotation TKA of PCL retaining type (Advance® Medial Pivot prosthesis with ‘Double High’ insert; Wright Medical Technology, Arlington, TN, USA) (Fig. 1). between March 2009 and March 2010 and had been followed up for a least 2 years. Inclusion criteria were age between 60 and 75 years and primary degenerative joint disease of knee graded as Kellgren Lawrence grade III or higher. Exclusion criteria were age under 60 years, any inflammatory joint disease including rheumatoid arthritis, early stage of primary degenerative joint disease of knee or any history of previous osteotomy around knee. Clinically, the knee society knee score and function score were used to evaluate pain and function. At last follow-up, all subjects performed full extension, thirty degree flexion and full active flexion sequentially under fluoroscopic surveillance. In each of these lateral radiographs, anteroposterior(AP) condylar position was pinpointed and the magnitude of PFT was determined by degree of transition of AP condylar position from full extension to full active flexion radiograph (Fig. 2 A–B). Statistical methods used were paired t-test, Pearson correlation, Steadman rank correlation and regression analysis. Component migration and radiolucent line were also observed. Results. At last follow-up, the mean knee society knee score and the mean function score improved significantly compared to preoperative scores (from 61.5 to 90.4 and from 57.8 to 84.7 respectively). The mean maximum flexion of knee increased postoperatively compared to preoperative one without any significant difference (105.5Ëš±11.2Ëšvs 109.3Ëš±9.8Ëš, p=0.051, β=0.387). Neverthless, regression analysis showed a good linear association (r = 0.53, p=0.0027) between the pre- and post-operative maximum flexions of knee. The AP condylar positions were consistently posterior to midline throughout the entire range of flexion. The mean maximum PFT was 10.5 mm (± 4.3 mm) and the magnitude of maximum PFT was greater in higher flexion cases (r = 0.57, p = 0.0009) (Fig. 3). There were no cases having either component migration or radiolucent line except for one case showing instability related to trauma. Conclusions. In medial rotation total knee arthroplasty of PCL retaining type, clinical outcomes were satisfactory and the maximum obtainable flexions tended to be in narrower ranges than those of preoperative ones and smaller than those of other TKA prostheses. Nonetheless, reliable posterior femoral translations were observed during progressive flexions of knees, which was considered to be one of important kinematic factors in increasing the level of knee flexion of medial-rotation TKA in longer follow-ups by providing greater posterior clearance and reduced femoro-tibial impingement


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 99 - 99
1 Sep 2012
Dwyer T Wasserstein D Gandhi R Mahomed N Ogilvie-Harris D
Full Access

Purpose

Factors that contribute to early and late re-operation after cruciate reconstruction (CR) have not been evaluated on a population level in a public health system. After surgery patients are at risk for knee stiffness, infection or early graft failure prompting revision. Long-term, ipsilateral revision CR, contralateral CR and potentially even joint replacement may occur. Population research in total joint replacement surgery has demonstrated an inverse relationship between complication/failure rates and surgeon procedural volume. We hypothesized that in Ontario, younger patient age and lower surgeon volume would increase the risk of short and long-term re-operation after CR.

Method

Billing, procedural and diagnostic coding from administrative databases (Ontario Health Insurance Plan, Canadian Institutes of Health Research) were accessed through the Institute for Clinical Evaluative Sciences to develop the cohort of all Ontario residents aged 14 to 60 who underwent anterior or posterior CR from July 1992 to April 2008. Logistic regression analysis was used to calculate the odds ratio for patient (age, gender, comorbidity, income, concurrent knee surgery) and provider (surgeon volume, teaching hospital status) factors for having a surgical washout of the knee, manipulation for stiffness or repeat of the index event within six months. A cox proportional hazards survivorship model was used to calculate the hazard ratio of the same covariates for repeat CR and partial/total knee arthroplasty from inception until end of 2009.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 604 - 604
1 Dec 2013
Zumbrunn T Varadarajan KM Rubash HE Li G Muratoglu O
Full Access

INTRODUCTION

Contemporary PCL sacrificing Total Knee Arthroplasty (TKA) implants (CS) consist of symmetric medial and lateral tibial articular surfaces with high anterior lips designed to substitute for the stability of the native PCL. However, designs vary significantly across implant systems in the level of anteroposterior constraint provided. Therefore, the goal of this study was to investigate kinematics of two CS designs with substantially different constraint levels. The hypothesis was that dynamic knee simulations could show the effect of implant constraint on kinematics of CS implants.

METHODS

LifeModeler KneeSIM software was used to analyze contemporary CS TKA (X) with a symmetric and highly dished tibia and contemporary CS TKA (Y) with a symmetric tibia having flat sections bounded by high anterior and posterior lips, during simulated deep knee bend and chair sit. The flat sections of CS-Y implant are designed to allow freedom prior to motion restriction by the implant lips. Components were mounted on an average knee model created from Magnetic Resonance Imaging (MRI) data of 40 normal knees. Relevant ligament/tendon insertions were obtained from the MRI based 3D models and tissue properties were based on literature values. The condyle center motions relative to the tibia were used to compare the different implant designs. In vivo knee kinematics of healthy subjects from published literature was used for reference.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 14 - 14
1 Sep 2012
Han Y Sardar Z McGrail S Steffen T Martineau P
Full Access

Purpose

Twelve case reports of distal femur fractures as post-operative complications after anterior cruciate ligament (ACL) reconstruction have been described in the literature. The femoral tunnel has been suggested as a potential stress riser for fracture formation. The recent increase in double bundle ACL reconstructions may compound this risk. This is the first biomechanical study to examine the stress riser effect of the femoral tunnel(s) after ACL reconstruction. The hypotheses tested in this study are that the femoral tunnel acts as a stress riser to fracture and that this effect increases with the size of the tunnel (8mm versus 10mm) and with the number of tunnels (one versus two).

Method

Femoral tunnels simulating single bundle (SB) hamstring graft (8 mm), bone-patellar tendon-bone graft (10 mm), and double bundle (DB) ACL reconstruction (7mm, 6 mm) were drilled in fourth generation saw bones. These three experimental groups and a control group consisting of native saw bones without tunnels, were loaded to failure.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 76 - 76
1 May 2016
Tei K Kihara S Shimizu T Matsumoto T Kurosaka M Kuroda R
Full Access

Introduction

Recently, tibial insert design of cruciate-substituting (CS) polyethylene insert is employed. However, in vivo kinematics of using CS polyethylene insert is still unclear. In this study, it is hypothesized that CS polyethylene insert leads to stability of femolo-tibial joint as well as posterior-stabilized (PS) polyethylene insert, even if PCL is sacrificed after TKA. The purpose of this study is an investigation of in vivo kinematics of femolo-tibial joint with use of CS polyethylene insert before and after PCL resction using computer assisted navigation system intra-operatively in TKA.

Materials and Methods

Twenty-four consecutive patients who had knees of osteoarthritis with varus deformity were investigated in this study. All TKAs (Triathlon, Stryker) were performed using computer assisted navigation system. In all patients, difference between extension and flexion gap was under 3mm after bony cut of femur and tibia. During surgery, CS polyethylene tibial trial insert were inserted after trial implantation of femoral and tibial components, before and after resection of PCL, respectively. The kinematic parameters of the soft-tissue balance, and amount of coronal (valgus/varus), sagittal (anterior/posterior) and rotational relative movement between femur and tibia were obtained by interpreting kinematics, which display tables throughout the range of motion (ROM) (Figure1). During record of kinematics, the surgeon gently lifted the experimental thigh three times, flexing the hip and knee. In each ROM (30, 45, 60, 90, max degrees), the data were analyzed with paired t-test, and an ANOVA test, and mean values were compared by the multiple comparison test (Turkey HSD test) (p < 0.05).


The posterior drawer is a commonly used test to diagnose an isolated PCL injury and combined PCL and PLC injury. Our aim was to analyse the effect of tibial internal and external rotation during the posterior drawer in isolated PCL and combined PCL and PLC deficient cadaver knee.

Ten fresh frozen and overnight-thawed cadaver knees with an average age of 76 years and without any signs of previous knee injury were used. A custom made wooden rig with electromagnetic tracking system was used to measure the knee kinematics. Each knee was tested with posterior and anterior drawer forces of 80N and posterior drawer with simultaneous external or internal rotational torque of 5Nm. Each knee was tested in intact condition, after PCL resection and after PLC (lateral collateral ligament and popliteus tendon) resection. Intact condition of each knees served as its own control. One-tailed paired student's t test with Bonferroni correction was used.

The posterior tibial displacement in a PCL deficient knee when a simultaneous external rotation torque was applied during posterior drawer at 90° flexion was not significantly different from the posterior tibial displacement with 80N posterior drawer in intact knee (p=0.22). In a PCL deficient knee posterior tibial displacement with simultaneous internal rotation torque and posterior drawer at 90° flexion was not significantly different from tibial displacement with isolated posterior drawer. In PCL and PLC deficient knee at extension with simultaneous internal rotational torque and posterior drawer force the posterior tibial displacement was not significantly different from an isolated PCL deficient condition (p=0.54).

We conclude that posterior drawer in an isolated PCL deficient knee could result in negative test if tibia is held in external rotation. During a recurvatum test for PCL and PLC deficient knee, tibial internal rotation in extension results in reduced posterior laxity.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_15 | Pages 88 - 88
7 Nov 2023
Greenwood K Molepo M Mogale N Keough N Hohmann E
Full Access

Knee arthroscopy is typically approached from the anterior, posteromedial and posterolateral portals. Access to the posterior compartments through these portals can cause iatrogenic cartilage damage and create difficulties in viewing the structures of the posterior compartments. The purpose of this study was to assess the feasibility of needle arthroscopy using direct posterior portals as both working and visualising portals. For workability, the needle scope was inserted advanced from anterior between the cruciate ligament bundle and the lateral wall of the medial femoral condyle until the posterior compartments were visualised. For visualisation, direct postero-lateral and -medial portals were established. The technique was performed in 9 knees by two experienced researchers. Workability and instrumentation of the posteromedial compartment and meniscus was achieved in 56%. The posterior horns could not be visualised in four specimens as the straight lens could not provide a more medial field of view. Visualisation from the direct medial posterior portal allowed a clear view of the medial meniscus, femoral condyle and posterior cruciate ligament in all specimens. Workability and instrumentation of the posterolateral compartment was not possible with the needle scope. Direct posterior approaches for the posteromedial compartment access are challenging with the current needle scope options and could only be achieved in over 50%. The postero-lateral compartment was not accessible. An angled lens or a flexible Needle scope would be better suited for developing this technique further


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_1 | Pages 51 - 51
1 Feb 2021
Smith L Cates H Freeman M Nachtrab J Komistek R
Full Access

Background. While posterior cruciate retaining (PCR) implants are a more common total knee arthroplasty (TKA) design, newer bi-cruciate retaining (BCR) TKAs are now being considered as an option for many patients, especially those that are younger. While PCR TKAs remove the ACL, the BCR TKA designs keep both cruciate ligaments intact, as it is believed that the resection of the ACL greatly affects the overall kinematic patterns of TKA designs. Various fluoroscopic studies have focused on determination of kinematics but haven't defined differentiators that affect motion patterns. This research study assesses the importance of the cruciate ligaments and femoral geometry for Bi-Cruciate Retaining (BCR) and Posterior Cruciate Retaining (PCR) TKAs having the same femoral component, compared to the normal knee. Methods. The in vivo 3D kinematics were determined for 40 subjects having a PCR TKA, 10 having a BCR TKA, and 10 having a normal knee, in a retrospective study. All TKA subjects had the same femoral component. All subjects performed a deep knee bend under fluoroscopic surveillance. The kinematics were determined during early flexion (ACL dominant), mid flexion (ACL/PCL transition) and deep flexion (PCL dominant). Results. During the first 30 degrees of flexion, the ACL played an important role, as subjects having a BCR TKA experienced kinematic patterns more similar to the normal knee. During mid flexion, both TKAs experienced random kinematic patterns, which could be due to the ACL and PCL being less active or resected in PCR TKA. In deeper flexion, both TKAs experienced kinematic patterns similar to the normal knee, thus supporting the assumption that the PCL played a dominant role [Fig. 1, Fig. 2]. All three groups generally experienced progressive axial rotation throughout flexion [Fig. 3]. On average, subjects having a PCR TKA experienced 112.3° of flexion, which was greater than the BCR subjects. Conclusions. Both the BCR TKA and normal groups experienced similar kinematic patterns, but the femoral geometrical differences from the anatomical condition may have influenced decreased motion compared to the normal knee. Both TKAs experienced similar kinematic patterns in deeper flexion, with the PCR TKA experiencing excellent weight-bearing flexion. Results from this study suggest that the cruciate ligaments can play a role in kinematics, but femoral geometry working with the ligaments may be an option to consider