Introduction. Total knee arthroplasty (TKA) femoral components are known to wear and roughen with clinical use, and reaction to metal is a well-documented complication of TKA. Ceramic materials are resistant to wear and corrosion, but the surface wear of ceramic femoral components in TKA has not been reported. This study measured the changes in roughness and mass lost from cobalt-chromium (CoCr) femoral components tested in a knee simulator, and compared them to those observed in ceramic components. Methods. Six cast CoCr femoral components were tested in a knee simulator bearing against UHMWPE, with high-kinematics/high-load waveforms. Roughness and scratches were measured via optical profilometry. Scratch volume was equated to mass lost from abrasive wear, while nano-particulate wear and corrosion were estimated from the change in the distance of the surface to the depth of the scratches after increasing numbers of cycles. Three magnesia-stabilized zirconia femoral components that had been wear tested to 15 million cycles were compared with the CoCr components. Results. After 500,000 cycles, the CoCr femoral components were visibly scratched and measurably roughened (average roughness Sa = 38.4nm, vs. 22.2nm for never-implanted components; p < 0.0001). The ceramic components were not scratched or roughened (Sa = 18.5nm). The CoCr components had an average of 7.5μg lost due to scratches (0.0018mm. 3. per million cycles). The material lost due to nanoparticulate wear and corrosion was estimated at 1.5μg. Discussion. This study demonstrates rapid loss of CoCr material from wear in TKA, and reports absence of scratching and roughening of ceramic TKA femoral components in simulated wear. This observation suggests a mechanism for the lower polyethylene wear rate against ceramic femoral components, and suggests an advantage in avoiding metal sensitivity


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_9 | Pages 51 - 51
1 Oct 2020
Wooster BM Kennedy NI Mallet KE Taunton MJ Abdel MP Trousdale RT
Full Access

Introduction. Mechanical or corrosive failure of total knee arthroplasties (TKAs) is difficult to diagnose with current laboratory and radiographic analyses. As such, the goal of this study was to determine the mean blood concentration of cobalt, chromium, and titanium in a series of revision TKAs with mechanical implant failure and evaluate whether they facilitated identification of the underlying TKA failure mechanism. Methods. Serum cobalt, chromium, and titanium levels and synovial fluid characteristics were evaluated in 12 patients (13 aseptic revision TKAs) who underwent revision TKA between 2000 and 2020 at a single academic institution for mechanical implant failure or corrosion. Seventy-five percent were re-revisions of previously revised TKAs. Mean time to revision was 6 years. Modular metallic junctions were present in 100%. Twenty-five percent did not have another in situ total joint arthroplasty, and the remaining patients did not have a metal-on-metal articulation that could lead to elevation in serum metal ion levels. Mean follow-up after the revision TKA was 8 months. Results. Mean serum cobalt, chromium, and titanium concentrations were 11 ng/mL, 6 ng/mL, and 3 ng/mL, respectively. Serum metal ion levels facilitated failure mechanism identification in 75%, which included modular junction failure (6 cases), constraint locking mechanism failure (3 cases), corrosion of modular metallic interfaces (2 cases), and implant fracture (1 case). Arthrocentesis was performed in 75%. Mean synovial fluid cell count was 950 cells/mcL. Monocytes were the predominant mean cell type (41%), followed by neutrophils (35%), and lymphocytes (22%). Conclusion. Serum metal ion assessment should be considered when the etiology of painful primary or revision TKAs, particularly those with modular metallic junctions, remains elusive after routine evaluation


The Bone & Joint Journal
Vol. 104-B, Issue 3 | Pages 376 - 385
1 Mar 2022
Gramlich Y Hofmann L Kress S Ruckes C Kemmerer M Klug A Hoffmann R Kremer M

Aims

This study compared the cobalt and chromium serum ion concentration of patients treated with two different metal-on-metal (MoM) hinged total knee arthroplasty (TKA) systems, as well as a titanium nitride (TiN)-coated variant.

Methods

A total of 63 patients (65 implants) were treated using either a MoM-coated (n = 29) or TiN-coated (n = 7) hinged TKA (GenuX mobile bearing, MUTARS; Implantcast, Germany) versus the BPKS (Brehm, Germany) hinged TKA (n = 27), in which the weight placed on the MoM hinge is diffused through a polyethylene (PE) inlay, reducing the direct load on the MoM hinge. Serum cobalt and chromium ion concentrations were assessed after minimum follow-up of 12 months, as well as functional outcome and quality of life.


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 138 - 144
1 Jun 2020
Heckmann ND Nahhas CR Yang J Della Valle CJ Yi PH Culvern CN Gerlinger TL Nam D

Aims

In patients with a “dry” aspiration during the investigation of prosthetic joint infection (PJI), saline lavage is commonly used to obtain a sample for analysis. The aim of this study was to investigate prospectively the impact of saline lavage on synovial fluid analysis in revision arthroplasty.

Methods

Patients undergoing revision hip (THA) or knee arthroplasty (TKA) for any septic or aseptic indication were enrolled. Intraoperatively, prior to arthrotomy, the maximum amount of fluid possible was aspirated to simulate a dry tap (pre-lavage) followed by the injection with 20 ml of normal saline and re-aspiration (post-lavage). Pre- and post-lavage synovial white blood cell (WBC) count, percent polymorphonuclear cells (%PMN), and cultures were compared.


The Bone & Joint Journal
Vol. 101-B, Issue 2 | Pages 227 - 232
1 Feb 2019
Walker T Rutkowski L Innmann M Panzram B Herre J Gotterbarm T Aldinger PR Merle C

Aims

The treatment of patients with allergies to metal in total joint arthroplasty is an ongoing debate. Possibilities include the use of hypoallergenic prostheses, as well as the use of standard cobalt-chromium (CoCr) alloy. This non-designer study was performed to evaluate the clinical outcome and survival rates of unicondylar knee arthroplasty (UKA) using a standard CoCr alloy in patients reporting signs of a hypersensitivity to metal.

Patients and Methods

A consecutive series of patients suitable for UKA were screened for symptoms of metal hypersensitivity by use of a questionnaire. A total of 82 patients out of 1737 patients suitable for medial UKA reporting cutaneous metal hypersensitivity to cobalt, chromium, or nickel were included into this study and prospectively evaluated to determine the functional outcome, possible signs of hypersensitivity, and short-term survivorship at a minimum follow-up of 1.5 years.


The Bone & Joint Journal
Vol. 98-B, Issue 4 | Pages 437 - 441
1 Apr 2016
Middleton S Toms A

We explored the literature surrounding whether allergy and hypersensitivity has a clinical basis for implant selection in total knee arthroplasty (TKA). In error, the terms hypersensitivity and allergy are often used synonymously. Although a relationship is present, we could not find any evidence of implant failure due to allergy. There is however increasing basic science that suggests a link between loosening and metal ion production. This is not an allergic response but is a potential problem. With a lack of evidence logically there can be no justification to use ‘hypoallergenic’ implants in patients who have pre-existing skin sensitivity to the metals used in TKA.

Cite this article: Bone Joint J 2016;98-B:437–41.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 5 | Pages 634 - 638
1 May 2010
Savarino L Tigani D Greco M Baldini N Giunti A

We investigated the role of ion release in the assessment of fixation of the implant after total knee replacement and hypothesised that ion monitoring could be a useful parameter in the diagnosis of prosthetic loosening. We enrolled 59 patients with unilateral procedures and measured their serum aluminium, titanium, chromium and cobalt ion levels, blinded to the clinical and radiological outcome which was considered to be the reference standard. The cut-off levels for detection of the ions were obtained by measuring the levels in 41 healthy blood donors who had no implants. Based on the clinical and radiological evaluation the patients were divided into two groups with either stable (n = 24) or loosened (n = 35) implants.

A significant increase in the mean level of Cr ions was seen in the group with failed implants (p = 0.001). The diagnostic accuracy was 71% providing strong evidence of failure when the level of Cr ions exceeded the cut-off value. The possibility of distinguishing loosening from other causes of failure was demonstrated by the higher diagnostic accuracy of 83%, when considering only patients with failure attributable to loosening.

Measurement of the serum level of Cr ions may be of value for detecting failure due to loosening when the diagnosis is in doubt. The other metal ions studies did not have any diagnostic value.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1069 - 1074
1 Aug 2013
Rao BM Kamal TT Vafaye J Moss M

We report the results of revision total knee replacement (TKR) in 26 patients with major metaphyseal osteolytic defects using 29 trabecular metal cones in conjunction with a rotating hinged total knee prosthesis. The osteolytic defects were types II and III (A or B) according to the Anderson Orthopaedic Research Institute (AORI) classification. The mean age of the patients was 72 years (62 to 84) and there were 15 men and 11 women. In this series patients had undergone a mean of 2.34 previous total knee arthroplasties. The main objective was to restore anatomy along with stability and function of the knee joint to allow immediate full weight-bearing and active knee movement. Outcomes were measured using Knee Society scores, Oxford knee scores, range of movement of the knee and serial radiographs. Patients were followed for a mean of 36 months (24 to 49). The mean Oxford knee clinical scores improved from 12.83 (10 to 15) to 35.20 (32 to 38) (p < 0.001) and mean American Knee Society scores improved from 33.24 (13 to 36) to 81.12 (78 to 86) (p < 0.001). No radiolucent lines suggestive of loosening were seen around the trabecular metal cones, and by one year all the radiographs showed good osteo-integration. There was no evidence of any collapse or implant migration. Our early results confirm the findings of others that trabecular metal cones offer a useful way of managing severe bone loss in revision TKR.

Cite this article: Bone Joint J 2013;95-B:1069–74.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 5 | Pages 609 - 614
1 May 2007
Himanen A Belt EA Lehto MUK Hämäläinen MMJ

We evaluated the survival of moulded monoblock and modular tibial components of the AGC total knee replacement in patients with rheumatoid arthritis. Between 1985 and 1995, 751 knees with this diagnosis were replaced at our institution. A total of 256 tibial components were of the moulded design and 495 of the modular design. The mean follow-up of the moulded subgroup was 9.6 years (0.5 to 14.7), and that of the modular group 7.0 years (0.1 to 14.7).

The groups differed significantly from each other in Larsen grade, cementing of components and patellar resurfacing, but no statistically significant difference between the survival of the components was found (Log rank test, p = 0.91). The cumulative success rate of the moulded group was 96.8% (95% confidence interval 93.6% to 98.4%) at five years and 94.4% (95% confidence interval 90.4% to 96.7%) at ten years, and of the modular group 96.2% (95% confidence interval 94% to 97.6%) and 93.6% (95% confidence interval 89.7% to 96%), respectively. Revision was required in 37 total knee replacements, the main causes were infection, pain, loosening of the tibial component and patellar problems. Survival rates for both components were satisfactory.