Advertisement for orthosearch.org.uk
Results 1 - 20 of 86
Results per page:
The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 779 - 786
1 Jul 2019
Lamb JN Baetz J Messer-Hannemann P Adekanmbi I van Duren BH Redmond A West RM Morlock MM Pandit HG

Aims. The aim of this study was to estimate the 90-day risk of revision for periprosthetic femoral fracture associated with design features of cementless femoral stems, and to investigate the effect of a collar on this risk using a biomechanical in vitro model. Materials and Methods. A total of 337 647 primary total hip arthroplasties (THAs) from the United Kingdom National Joint Registry (NJR) were included in a multivariable survival and regression analysis to identify the adjusted hazard of revision for periprosthetic fracture following primary THA using a cementless stem. The effect of a collar in cementless THA on this risk was evaluated in an in vitro model using paired fresh frozen cadaveric femora. Results. The prevalence of early revision for periprosthetic fracture was 0.34% (1180/337 647) and 44.0% (520/1180) occurred within 90 days of surgery. Implant risk factors included: collarless stem, non-grit-blasted finish, and triple-tapered design. In the in vitro model, a medial calcar collar consistently improved the stability and resistance to fracture. Conclusion. Analysis of features of stem design in registry data is a useful method of identifying implant characteristics that affect the risk of early periprosthetic fracture around a cementless femoral stem. A collar on the calcar reduced the risk of an early periprosthetic fracture and this was confirmed by biomechanical testing. This approach may be useful in the analysis of other uncommon modes of failure after THA. Cite this article: Bone Joint J 2019;101-B:779–786


The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 144 - 150
1 Feb 2024
Lynch Wong M Robinson M Bryce L Cassidy R Lamb JN Diamond O Beverland D

Aims. The aim of this study was to determine both the incidence of, and the reoperation rate for, postoperative periprosthetic femoral fracture (POPFF) after total hip arthroplasty (THA) with either a collared cementless (CC) femoral component or a cemented polished taper-slip (PTS) femoral component. Methods. We performed a retrospective review of a consecutive series of 11,018 THAs over a ten-year period. All POPFFs were identified using regional radiograph archiving and electronic care systems. Results. A total of 11,018 THAs were implanted: 4,952 CC femoral components and 6,066 cemented PTS femoral components. Between groups, age, sex, and BMI did not differ. Overall, 91 patients (0.8%) sustained a POPFF. For all patients with a POPFF, 16.5% (15/91) were managed conservatively, 67.0% (61/91) underwent open reduction and internal fixation (ORIF), and 16.5% (15/91) underwent revision. The CC group had a lower POPFF rate compared to the PTS group (0.7% (36/4,952) vs 0.9% (55/6,066); p = 0.345). Fewer POPFFs in the CC group required surgery (0.4% (22/4,952) vs 0.9% (54/6,066); p = 0.005). Fewer POPFFs required surgery in males with a CC than males with a PTS (0.3% (7/2,121) vs 1.3% (36/2,674); p < 0.001). Conclusion. Male patients with a PTS femoral component were five times more likely to have a reoperation for POPFF. Female patients had the same incidence of reoperation with either component type. Of those having a reoperation, 80.3% (61/76) had an ORIF, which could greatly mask the size of this problem in many registries. Cite this article: Bone Joint J 2024;106-B(2):144–150


The Bone & Joint Journal
Vol. 102-B, Issue 12 | Pages 1654 - 1661
1 Dec 2020
Perelgut ME Polus JS Lanting BA Teeter MG

Aims. The direct anterior (DA) approach has been associated with rapid patient recovery after total hip arthroplasty (THA) but may be associated with more frequent femoral complications including implant loosening. The objective of this study was to determine whether the addition of a collar to the femoral stem affects implant migration, patient activity, and patient function following primary THA using the DA approach. Methods. Patients were randomized to either a collared (n = 23) or collarless (n = 26) cementless femoral stem implanted using the DA approach. Canal fill ratio (CFR) was measured on the first postoperative radiographs. Patients underwent a supine radiostereometric analysis (RSA) exam postoperatively on the day of surgery and at two, four, six, 12, 26, and 52 weeks postoperatively. Patient-reported outcome measures (Western Ontario and McMaster Universities Osteoarthritis (WOMAC) Index, the 12-item Short Form Health Survey Mental and Physical Score, and University of California, Los Angeles (UCLA) Activity Score) were measured preoperatively and at each post-surgery clinic visit. Activity and function were also measured as the weekly average step count recorded by an activity tracker, and an instrumented timed up-and-go (TUG) test in clinic, respectively. Results. Comparing the RSA between the day of surgery baseline exam to two weeks postoperatively, subsidence was significantly lower (mean difference 2.23 mm (SD 0.71), p = 0.023) with collared stems, though these patients had a greater CFR (p = 0.048). There was no difference (p = 0.426) in subsidence between stems from a two-week baseline through to one year postoperatively. There were no clinically relevant differences in PROMs; and there was no difference in the change in activity (p = 0.078) or the change in functional capacity (p = 0.664) between the collared stem group and the collarless stem group at any timepoint. Conclusion. Presence of a collar on the femoral stem resulted in reduced subsidence during the first two postoperative weeks following primary THA using the DA approach. However, the clinical implications are unclear, and larger studies examining patient activity and outcomes are required. Cite this article: Bone Joint J 2020;102-B(12):1654–1661


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_6 | Pages 30 - 30
1 May 2019
Lamb J Baetz J Messer-Hannemann P Redmond A West R Morlock M Pandit H
Full Access

Background. Post-operative periprosthetic femoral fractures (PFF) are a devastating complication associated with high mortality and are costly. Few risk factors are modifiable apart from implant choice. The design features governing risk of PFF are unknown. We estimated the 90-day risk of revision for PFF associated with design features of cementless femoral stems and to investigate the effect of a collar on early PFF risk using a biomechanical in-vitro model. Patients, materials and methods. 337 647 primary THAs from the National Joint Registry (UK) were included in a multivariable survival and regression analysis to identify the adjusted hazard of PFF revision following primary THA using cementless stems. The effect of a collar in cementless THA on early PFF was evaluated in an in-vitro model using paired fresh frozen cadaveric femora. Results. Prevalence of PFF revision was 0.34% (1180/337647) and 44.0% occurred (520/1180) within 90 days of surgery. Implant risk factors included: collarless stem, non grit-blasted finish and triple tapered design. In the in-vitro PFF model a medial calcar collar consistently improved construct stability and fracture resistance. Discussion. During rotational injury the collar can load the calcar in compression increasing the force required for a fracture. This increases the force required to cause a PFF around a collared implant versus collarless implants. The calcar possibly acts as a check-rein which prevents excessive peri-prosthetic trabecular deformation in rotational injuries and may improve the resistance to loosening after high energy injuries which do not cause cortical fracture. Conclusion. Analysis of stem design features in registry data is a useful method to identify implant characteristics which affect the risk of early PFF around cementless femoral stems. Calcar collar reduced early PFF risk and this was confirmed by biomechanical testing. This approach may be useful in the analysis of other uncommon arthroplasty failure modes


The Bone & Joint Journal
Vol. 103-B, Issue 4 | Pages 650 - 658
1 Apr 2021
Konow T Baetz J Melsheimer O Grimberg A Morlock M

Aims. Periprosthetic femoral fractures (PPF) are a serious complication of total hip arthroplasty (THA) and are becoming an increasingly common indication for revision arthroplasty with the ageing population. This study aimed to identify potential risk factors for PPF based on an analysis of registry data. Methods. Cases recorded with PPF as the primary indication for revision arthroplasty in the German Arthroplasty Registry (Endoprothesenregister Deutschland (EPRD)), as well as those classified as having a PPF according to the International Classification of Diseases (ICD) codes in patients’ insurance records were identified from the complete datasets of 249,639 registered primary hip arthroplasties in the EPRD and included in the analysis. Results. The incidence of PPFs was higher (24.6%; 1,483) than reported in EPRD annual reports listing PPF as the main reason for revision (10.9%; 654). The majority of fractures occurred intraoperatively and were directly related to the implantation process. Patients who were elderly, female, or had comorbidities were at higher risk of PPFs (p < 0.001). German hospitals with a surgical volume of < 300 primary procedures per year had a higher rate of PPFs (p < 0.001). The use of cemented and collared prostheses had a lower fracture risk PPF compared to uncemented and collarless components, respectively (both p < 0.001). Collared prostheses reduced the risk of PPF irrespective of the fixation method and hospital’s surgical volume. Conclusion. The high proportion of intraoperative fractures emphasises the need to improve surgeon training and surgical technique. Registry data should be interpreted with caution because of potential differences in coding standards between institutions. Cite this article: Bone Joint J 2021;103-B(4):650–658


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 609 - 614
1 May 2012
Flecher X Blanc G Sainsous B Parratte S Argenson J

We describe the results of 81 consecutive revision total hip replacements with impaction grafting in 79 patients using a collared polished chrome–cobalt stem, customised in length according to the extent of distal bone loss. Our hypothesis was that the features of this stem would reduce the rate of femoral fracture and subsidence of the stem. The mean follow-up was 12 years (8 to 15). No intra-operative fracture or significant subsidence occurred. Only one patient suffered a post-operative diaphyseal fracture, which was associated with a fall. All but one femur showed incorporation of the graft. No revision for aseptic loosening was recorded. The rate of survival of the femoral component at 12 years, using further femoral revision as the endpoint, was 100% (95% confidence interval (CI) 95.9 to 100), and at nine years using re-operation for any reason as the endpoint, was 94.6% (95% CI 92.0 to 97.2). These results suggest that a customised cemented polished stem individually adapted to the extent of bone loss and with a collar may reduce subsidence and the rate of fracture while maintaining the durability of the fixation


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 43 - 43
19 Aug 2024
Lustig S Batailler C
Full Access

The stem design in total hip arthroplasty (THA) is constantly evolving. The impact of the collar on the risk of periprosthetic fracture remains controversial. This study aimed to determine whether adding a collar to the femoral stem impacts the early periprosthetic fracture risk within 90 days of surgery. This retrospective study included 1,623 patients who underwent primary total hip arthroplasty in a single orthopedic department between January 2010 and December 2020. The inclusion criteria were uncemented stem with or without a collar, in a primary intention, without previous hip surgery with a similar “corail like” design. The assessed data were demographic characteristics (age, gender, number of obese (BMI > 30)), single or dual mobility, the surgical approach, the early complication, particularly the periprosthetic femoral fractures. Of the 1,623 patients, 1,380 received a collared stem (85%), and 243 received a collarless stem (15%). A multivariate analysis was performed to determine the collar's effect on the risk of early periprosthetic fracture (<90 days). Nine (0.55%) early periprosthetic fractures were identified in the whole cohort. There were four fractures (1.65%) in the collarless stem group and five fractures (0.36%) in the collared stem group (p=0,005). One patient required revision surgery in the collared stem group, while two patients required revision surgery in the collarless stem group. The multinomial logistic regression model indicated a statistically significant lower (p<0.05) risk of early periprosthetic fracture within 90 days of surgery in the collared stems group. No other risk factor for early periprosthetic fractures has been identified. Using collared stems in cementless THA protects early periprosthetic femoral fractures within 90 days of surgery


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 15 - 15
23 Jun 2023
Ricotti RG Flevas D Sokrab R Vigdorchik JM Sculco TP Sculco PK
Full Access

Periprosthetic femur fracture (PFF) is a major complication following total hip arthroplasty (THA) that carries significant morbidity, mortality, and economic burden. Currently, uncemented stems are highly preferred in primary THA, but have been associated with higher risk of PFF compared to cemented stems. The use of collared stems in uncemented primary THA has shown promise in reducing PFF rates postoperatively. This retrospective study included 2,294 uncemented primary THAs using the posterior approach performed by two attending surgeons from January 2016 to December 2022. Both surgeons switched from a collarless femoral stem design to a collared design in May 2020. Data was collected regarding stem design, frequency of PFF, and requirement for revision surgery. Periprosthetic fractures were identified and confirmed using medical records and/or radiographic imaging. Fracture rates and percentages between collared and collarless stems were then analyzed. A Fisher's Exact Test was performed to determine if there was a significant association between collared and collarless stem use on PFF rates. A total of 2,294 uncemented primary THAs performed by 2 surgeons were eligible for analysis. 903 (39.4%) patients received a collared stem, and 1,391 (60.6%) patients received a collarless stem. In total, 14 (0.6%) PFFs occurred over the study period. There was 1 fracture (0.1%) out of 903 collared stems, and 13 fractures (0.9%) out of 1,391 collarless stems (p = 0.012). Collared stems were associated with a significant decrease in PFF rate when compared to collarless stems in uncemented primary THA. Future studies are encouraged to continue to investigate PFF and other complication rates with the use of a collared stem design


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 18 - 18
23 Jun 2023
Morlock M Melsheimer O
Full Access

The early revision rate in elective Total Hip Arthroplasty (THA) three years after surgery in elderly patients over 80 years is significantly lower for cemented stems in the German Arthroplasty Register (EPRD): cemented 3,1% (3.0 – 3.2) vs. uncemented 4.2% (4.1 – 4.3; p < 0.001). However, the mortality rate in elderly patients is elevated for cemented fixation. This study presents a detailed analysis of the influence of stem type and fixation on revision and mortality rate in this patient cohort. Elective primary THA cases for primary Coxarthrosis using uncemented cups from the EPRD data base were analysed (n. 0. = 37,183). Four stem type groups were compared: cementless, cementless with collar, cementless short, and cemented. Stems with at least 300 cases at risk three years after surgery were analysed individually. The reference stem was determined as the stem with the lowest revision rate and at least 1000 cases under surveillance 3 years after surgery (n. 3. = 28,637). The revision rate for cemented stems (2.5% [2.2–1.81] was lower than for uncemented (4.5% [4.2–4.9]; p<0.001) and uncemented short stems (4.2% [3.1–5.7]; p=0.002). The revision rate of uncemented collared stems (2.3% [1.5–3.6]) was similar to cemented stems (p=0.89) and lower than for uncemented stems (p=0.02). One year mortality showed no sig. differences between the groups (p>0.17): cemented 3.2% [2.9–3.6], uncemented 3.4% [3.1–3.7], uncemented short 3.5% [2.5–4.9], uncemented collar 2.0% [1.2–3.2]. “Cementless” and “cementless short” stems should not be used in patients over 80 years due to the higher revision risk. If cementing should be avoided, “cementless collared” stems seem to be a good alternative combined with a tendency for a lower one year mortality rate


There is a strong association between the presence of a calcar collar on a cementless stem and a reduced risk of revision surgery for periprosthetic fracture of the femur (PFF). A medial calcar collar may act to reduce relative movement between the implant and femur during PFF, through calcar-collar contact (CCC). The aims were:. Estimate the effect of CCC on periprosthetic fracture mechanics. Estimate the effect of initial calcar-collar separation on the likelihood of CCC. Three groups of six composite femurs were implanted with a fully coated collared cementless femoral stem. Neck resection differed between groups (group 1 = no additional resection, group 2 = 3mm additional resection, group 3 = 6mm additional resection). PFF were simulated using a previously published technique. Fracture torque and rotational displacement were measured and torsional stiffness and rotational work prior to fracture were estimated. Results between trials where CCC did and did not occur where compared using Mann-Whitney U tests. Logistic regression estimated the odds (OR) of failing with 95% confidence interval (CI) to achieve CCC for a given initial separation. Where CCC occurred fracture torque was greater (47.33 [41.03 to 50.45] Nm versus 38.26 [33.70 to 43.60] Nm, p= 0.05) and torsional stiffness was greater (151.38 [123.04 to 160.42] rad.Nm. −1. versus 96.86 [84.65 to 112.98] rad.Nm. −1. , p <0.01). CCC was occurred in all cases in group one, 50% in group two and 0% in group three. OR of failure to obtain CCC increased 3.8 fold (95% CI 1.6 to 30.2, p <0.05) for each millimetre of separation. Resistance to fracture and construct stiffness increased when a the collar made contact with the calcar prior to fracture and the chances of contact decrease with increasing initial separation at the time of implantation. Surgeons should aim to achieve a calcar-collar distance of less than 1mm following implantation to ensure CCC and to reduce the risk of fracture


Bone & Joint Research
Vol. 11, Issue 5 | Pages 270 - 277
6 May 2022
Takegami Y Seki T Osawa Y Imagama S

Aims. Periprosthetic hip fractures (PPFs) after total hip arthroplasty are difficult to treat. Therefore, it is important to identify modifiable risk factors such as stem selection to reduce the occurrence of PPFs. This study aimed to clarify differences in fracture torque, surface strain, and fracture type analysis between three different types of cemented stems. Methods. We conducted biomechanical testing of bone analogues using six cemented stems of three different types: collarless polished tapered (CPT) stem, Versys Advocate (Versys) stem, and Charnley-Marcel-Kerboull (CMK) stem. Experienced surgeons implanted each of these types of stems into six bone analogues, and the analogues were compressed and internally rotated until failure. Torque to fracture and fracture type were recorded. We also measured surface strain distribution using triaxial rosettes. Results. There was a significant difference in fracture torque between the three stem types (p = 0.036). Particularly, the median fracture torque for the CPT stem was significantly lower than that for the CMK stem (CPT vs CMK: 164.5 Nm vs 200.5 Nm; p = 0.046). The strain values for the CPT stem were higher than those for the other two stems at the most proximal site. The fracture pattern of the CPT and Versys stems was Vancouver type B, whereas that of the CMK stem was type C. Conclusion. Our study suggested that the cobalt-chromium alloy material, polished surface finish, acute-square proximal form, and the absence of a collar may be associated with lower fracture torque, which may be related to PPF. Cite this article: Bone Joint Res 2022;11(5):270–277


Aims. The aim of this study was to compare the mid-term patient-reported outcome, bone remodelling, and migration of a short stem (Collum Femoris Preserving; CFP) with a conventional uncemented stem (Corail). Methods. Of 81 patients who were initially enrolled, 71 were available at five years’ follow-up. The outcomes at two years have previously been reported. The primary outcome measure was the clinical result assessed using the Oxford Hip Score (OHS). Secondary outcomes were the migration of the stem, measured using radiostereometric analysis (RSA), change of bone mineral density (BMD) around the stem, the development of radiolucent lines, and additional patient-reported outcome measures (PROMs). Results. There were no statistically significant differences between the groups regarding PROMs (median OHS (CFP 45 (interquartile range (IQR) 35 to 48); Corail 45 (IQR 40 to 48); p = 0.568). RSA showed stable stems in both groups, with little or no further subsidence between two and five years. Resorption of the femoral neck was evident in nine patients in the CFP group and in none of the 15 Corail stems with a collar that could be studied. Dual X-ray absorbiometry showed a significantly higher loss of BMD in the proximal Gruen zones in the CFP group (mean changes in BMD: Gruen zone 1, CFP -9.5 (95% confidence interval (CI) -14.8 to -4.2), Corail 1.0 (95% CI 3.4 to 5.4); Gruen zone 7, CFP -23.0 (95% CI -29.4 to -16.6), Corail -7.2 (95% CI -15.9 to 1.4). Two CFP stems were revised before two years’ follow-up due to loosening, and one Corail stem was revised after two years due to chronic infection. Conclusion. The CFP stem has a similar clinical outcome and subsidence pattern when compared with the Corail stem. More pronounced proximal stress-shielding was seen with the CFP stem, suggesting diaphyseal fixation, and questioning its femoral neck-sparing properties in the long term. Cite this article: Bone Joint J 2022;104-B(5):581–588


Current advice regarding implant choice is based on estimates of cost-benefit derived from implant survival to an endpoint of revision. Current estimates do not account for many implant failures which are treated with non-revision surgery and may not be accurate. The aim of this study was to estimate survival of major stem implant design groups to an endpoint of reoperation. Primary total hip replacement and linked revision form the National Joint Registry (NJR) and Hospital Episode Statistics (HES) data linked by unique identifier were used. Survival of femoral implant groups (cemented stainless steel polished taper [PTSS], cemented cobalt chrome polished taper [PTCC], cemented composite beam [CB], collarless cementless [NCOL] and collared cementless [COL]) was estimated using Kaplan-Meier method. 809,832 patients with valid NJR and HES data from England, were included. Cumulative failure at ten years for PTSS increased overall from 2.9% (95%CI 2.8–2.9) to 3.6% (95%CI 3.6–3.7) after inclusion of reoperations. Cumulative failure at ten years for PTSS increased from 2.5% (95%CI 2.5–2.6) to 3.3% (95%CI 3.2–3.4), for PTCC increased from 3.8% (95%CI 3.5–4.0) to 5.4% (95%CI 5.1–5.6), for CB increased from 3.1% (95%CI 2.9–3.3) to 4.1% (95%CI 3.8–4.3), for NCOL increased from 3.4% (95%CI 3.3–3.5) to 3.9% (95%CI 3.8–4.0), and for COL increased from 2.5% (95%CI 2.4–2.6) to 3.1% (95%CI 2.9–3.2), after inclusion of reoperations. Re-operation for internal fixation is as significant life event for the patient as revision. When a more inclusive metric is used, the patient and clinician's perspective on what constitutes a GIRFT implant may not be the same. Further work is required to update implant selection guidance in view of the change in implant performance


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_6 | Pages 35 - 35
2 May 2024
Robinson M Wong ML Cassidy R Bryce L Lamb J Diamond O Beverland D
Full Access

The significance of periprosthetic fractures about a total hip arthroplasty (THA) is becoming increasingly important. Recent studies have demonstrated post-operative periprosthetic fracture rates are higher amongst cemented polished taper slip (PTS) stem designs compared to collared cementless (CC) designs. However, in the National Joint Registry, the rate of intra-operative periprosthetic femoral fractures (IOPFF) with cementless implant systems remains higher (0.87% vs 0.42%. p <0.001) potentially leading to more post-operative complications. This study identifies the incidence of IOPFF, the fracture subtype and compares functional outcomes and revision rates of CC femoral implants with an IOPFF to CC stems and PTS stems without a fracture. 5376 consecutive CC stem THA, carried out through a posterior approach were reviewed for IOPFF. Each fracture was subdivided into calcar fracture, greater trochanter (GT) fracture or shaft fracture. 1:1:1 matched analysis was carried out to compare Oxford scores at one year. Matching criteria included; sex (exact), age (± 1 year), American Society of Anaesthesiologists (ASA) grade (exact), and date of surgery (± 6 months). Electronic records were used to review revision rates. Following review of the CC stems, 44 (0.8%) were identified as having an IOPFF. Of these 30 (0.6%) were calcar fractures, 11 (0.2%) GT fractures and 3 (0.06%) were shaft fractures. There were no shaft penetrations. Overall, no significant difference in Oxford scores at one year were observed when comparing the CC IOPFF, CC non-IOPFF and PTS groups. There were no CC stems revised for any reason with either a calcar fracture or trochanteric fracture within the period of 8 years follow-up. IOPFF do occur more frequently in cementless systems than cemented. The majority are calcar and GT fractures. These fractures, when identified and managed intra-operatively, do not have worse functional outcomes or revision rates compared to matched non-IOPFF cases


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_6 | Pages 36 - 36
1 May 2019
Syed F Magra M Saunders P Ugwuoke A Young S
Full Access

Introduction. Primary fixation in uncemented Total Hip Replacement (THR) is achieved by implanting an appropriately sized stem. Last year, in our pilot study, we classified undersizing & reported the incidence of subsidence & radiolucent lines (RLL) in undersized uncemented stems. Now we report on a larger cohort of patients, reporting undersizing and its association with radiolucent lines and subsidence. Secondarily we report on the effect of using a collar, patients' BMI, and in a smaller group the five-year failure rate. Methods. A retrospective review of electronic records of 1337 cases who underwent THR using the Corail stem was carried out. 1204 (Collared 1042, Uncollared 162) had a minimum follow up of 1 year, 693 of those had a follow up of at least 5 years. Radiological review was carried out by 2 independent investigators. Results. 19.77% stems were undersized. There was significantly higher incidence of subsidence & RLL in undersized stems with more than 1 year follow up. There no significant increase in incidence of RLL in those with BMI greater than 30 (16.88% vs 18.81%). Analysis of undersized collared stems demonstrated that even in this sub group undersizing is associated with increased incidence of RLL (P 0.00001) & subsidence (P 0.0003). A five year follow up, the revision rate for aseptic loosening of stem was 3.09% in undersized stems & 0.29% in appropriately sized stems (P 0.06). Conclusions. Our study indicates undersizing of uncemented stems is associated with poorer radiological outcomes & there's a trend of higher revision rate at 5 year follow up. This reinforces the importance of implanting appropriately sized stem, as undersizing can potentially compromise stem survival in uncemented total hip arthroplasty


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_6 | Pages 28 - 28
1 May 2019
Pryce G Al-Hajjar M Wilcox R Thompson J Board T Williams S
Full Access

Impingement of total hip replacements (THRs) can cause rim damage of polyethylene liners, and lead to dislocation and/or mechanical failure of liner locking mechanisms[1]. Previous work has focussed on the influence of femoral neck profile on impingement without consideration of neck-shaft angle. This study assessed the occurrence of impingement with two different stem designs (Corail standard [135°] and coxa vara [125°]) under different activities with varying acetabular cup orientation (30° to 70° inclination; 0° to 50° anteversion) using a geometric modelling tool. The tool was created in a computer aided design software programme, and incorporated an individual's hemi-pelvis and femur geometry[3] with a THR (DePuy Synthes Pinnacle. ®. shell and neutral liner; size 12 Corail. ®. standard or coxa vara and 32mm head). Kinematic data of activities associated with dislocation[2], such as stooping to pick an object from the floor was applied and incidences of impingement were recorded. Predicted implant impingement was influenced by stem design. The coxa vara stem was predicted to cause implant impingement less frequently across the range of activities and cup orientations investigated, compared to the standard stem [Fig. 1]. The cup orientations predicted to cause impingement the least frequently were at lower inclination and anteversion angles, relative to the standard stem [Fig. 1]. The coxa vara stem included a collar, while the standard stem was collarless; additional analysis indicated that differences were due to neck angle and not the presence of a collar. This study demonstrated that stem neck-shaft angle is an important variable in prosthetic impingement in THR and surgeons should be aware of this when choosing implants. Future work will consider further implant design and bone geometry variables. This tool has the potential for use in optimising stem design and position and could assist with patient specific stem selection based on an individual's activity profile. For any figures or tables, please contact the authors directly


The Bone & Joint Journal
Vol. 98-B, Issue 12 | Pages 1589 - 1596
1 Dec 2016
Magill P Blaney J Hill JC Bonnin MP Beverland DE

Aims. Our aim was to report survivorship data and lessons learned with the Corail/Pinnacle cementless total hip arthroplasty (THA) system. Patients and Methods. Between August 2005 and March 2015, a total of 4802 primary cementless Corail/Pinnacle THAs were performed in 4309 patients. In March 2016, we reviewed these hips from a prospectively maintained database. Results . A total of 80 hips (1.67%) have been revised which is equivalent to a cumulative risk of revision of 2.5% at ten years. The rate of revision was not significantly higher in patients aged ≥ 70 years (p = 0.93). The leading indications for revision were instability (n = 22, 0.46%), infection (n = 20, 0.42%), aseptic femoral loosening (n = 15, 0.31%) and femoral fracture (n = 6, 0.12%). There were changes in the surgical technique with respect to the Corail femoral component during the ten-year period involving a change to collared components and a trend towards larger size. These resulted in a decrease in the rate of iatrogenic femoral fracture and a decrease in the rate of aseptic loosening. Conclusion. The rate of revision in this series is comparable with the best performing THAs in registry data. Most revisions were not directly related to the implants. Despite extensive previous experience with cemented femoral components, the senior author noted a learning curve requiring increased focus on primary stability. The number of revisions related to the femoral component is reducing. Any new technology has a learning curve that may be independent of surgical experience. Cite this article: Bone Joint J 2016;98-B:1589–96


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_13 | Pages 6 - 6
1 Oct 2018
Naudie DD Paish AD Nikolov HN Chmiel T El-Warrak AO Welch ID Teeter MG Holdsworth DW
Full Access

Introduction. As new innovations are developed to improve the longevity of joint replacement components, preclinical testing is necessary in the early stages of research into areas such as osseointegration, metal-cartilage wear and periprosthetic joint infection (PJI). Large-animal studies that test load-bearing components are expensive, however, requiring that animals be housed in special facilities that are not available at all institutions. Comparably, small animal models, such as the rat, offer several advantages including lower cost. Load-bearing implants remain difficult to manufacture via traditional methods in the sizes required for small-animal testing. Recent advances in additive manufacturing (3D metal-printing) have allowed for the creation of miniature joint replacement components in a variety of medical-grade metal alloys. The objective of this work is to create and optimize an image-based 3D-printed rat hip implant system that will allow in vivo testing of functional implant properties in a rat model. Methods. A database of n=25 previously-acquired, 154μm micro-CT volumes (eXplore Locus Ultra, GE Medical) of male Sprague-Dawley rats (390–610g) were analyzed to obtain spatial and angular relationships between several anatomical features of the proximal rat femora. Mean measurements were used to guide the creation of a femoral implant template in computer-aided design software (Solidworks, Dassault Systemes). Several different variations were created, including collarless and collared designs, in a range of sizes to accommodate rats of various weights. Initial prototypes were 3D-printed 316L stainless steel with subsequent iterations printed in Ti6Al4V titanium and F75 cobalt-chrome. Implants were post-processed via sandblasting, hand-polishing, ultrasonic bath, and sterilization in an autoclave. Innate surface texturing was left on manufactured stems to promote osseointegration. Surgical implantation was performed in three live Sprague-Dawley rats (900g, 500g, 750g) with preservation of muscle attachments to the greater trochanter. Micro-CT imaging and X-ray fluoroscopy were performed post-operatively on each animal at 1 day, and 1, 3, 9 and 12 weeks to evaluate gait and component positioning. Results. Implantation of components was successful and each animal was observed to ambulate on its affected limb immediately following recovery from surgery. The 900g rat, given a collarless 316L stainless steel component, was kept for 11 months post-implantation before succumbing to old age. Micro-CT and fluoroscopic findings revealed no evidence of implant subsidence. The 500g animal, given a collarless 316L stainless steel implant, showed evidence of implant subsidence at 3 weeks, with full subsidence and hip dislocation at 12 weeks. The 750g rat, given a collared F75 cobalt-chrome implant, was observed ambulating on its affected limb, but experienced implant rotation and failure at 9 weeks. Conclusions. We report the first hip hemi-arthroplasty in a rat using a 3D-printed metal implant. This model aims to provide a low-cost platform for studying osseointegration, metal-cartilage interactions, and PJI using a functional, loaded implant. Efforts to further optimize the surgical approach will be made to reduce early implant loosening. A study with larger sample sizes is needed to determine if implants can be installed repeatedly, without complications, before the utility of this approach can be validated. Future work will include surface preparations on implant stems, with micro-CT to longitudinally track changes at the bone-metal interface, and gait analysis on a radiolucent treadmill to quantify post-operative kinematics


Bone & Joint Open
Vol. 5, Issue 11 | Pages 971 - 976
5 Nov 2024
Baker G Hill J O'Neill F McChesney J Stevenson M Beverland D

Aims

In 2015, we published the results of our ceramic-on-metal (CoM) total hip arthroplasties (THAs) performed between October 2007 and July 2009 with a mean follow-up of 34 months (23 to 45) and a revision rate of 3.1%. The aim of this paper is to present the longer-term outcomes.

Methods

A total of 264 patients were reviewed at a mean of 5.8 years (4.6 to 7.2) and 10.1 years (9.2 to 10.6) to determine revision rate, pain, outcome scores, radiological analysis, and blood ion levels. Those who were unwilling or unable to travel were contacted by telephone.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_11 | Pages 34 - 34
1 Jun 2016
Magill P Blaney J Hill J Beverland D
Full Access

Introduction. The results of cementless total hip arthroplasty (THA) vary with data from the UK national Joint Registry being less favourable than that from the Australian registry. The senior author started using a fully cementless THA in 2005 and we aimed to gauge the performance of the implants based on their revision data. Patients and methods. Between August 2005 and March 2015, 4,802 primary THA (4,309 patients) were performed with a cementless Corail. ®. stem and a cementless Pinnacle. ®. cup. There were 2,086 (43.4%) males and 2,716 (56.6%) females with a median age of 70 years (IQR 13, Range 16–95). There were a number of changes to the surgical technique with respect to the Corail. ®. stem during the ten-year period, which we have categorised as phase 1 and phase 2. We compared the data in the two phases. Data were extracted from a prospectively maintained patient information database. Results. A total of 80 (1.67%) revisions have been performed to date (median follow-up 65.9 months, IQR 46.8, Range 0 to 121), which is equivalent to a cumulative revision risk of 2.5% at ten years. Revision rate was not significantly different in those less than 70 years old (1.63%) compared to those greater than or equal to 70 years old (1.76%, P=0.81). The leading causes of revision were instability (n=22, 0.46%), infection (n=20, 0.42%) and aseptic loosening of the stem (n=15, 0.31%). More collarless stems than collared stems have been revised. Phase 2 changes in surgical technique resulted in cessation of collarless stem use, a small but significant increase in mean stem size, and a paradoxical decrease in iatrogenic femoral fracture. Conclusions. The overall revision rates of the Corail. ®. stem and Pinnacle. ®. cup in this series are comparable to the best performing THA in equivalent registry data. Instability was the leading cause of revision but these data did not identify a causative factor. The changes in infection rate in this series are possibly influenced by changes in local antibiotic prophylaxis policy. There is a learning curve for the cementless stem as seen here by a reduction in revisions for aseptic loosening and iatrogenic femoral fracture during the ten-year period. We believe that collared stems, avoidance of undersizing and surgical technique focusing on primary stability are the key aspects