Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_9 | Pages 69 - 69
1 Oct 2020
Lawrie CM Barrack RL Nunley RM
Full Access

Introduction

Bone mineral density (BMD) is correlated with component migration and aseptic loosening after total knee arthroplasty (TKA). Older implant designs have demonstrated BMD loss up to 23% in the first 6 months after TKA, and continued to BMD decline at an average of 5% per year for as long as 2 years after TKA. The impact of component design and fixation method on BMD loss after TKA in modern implant designs has not been fully elucidated. The purpose of this study is to determine the effect of tibial tray thickness and fixation method (cemented versus cementless) on BMD loss patterns of the proximal tibia in two different modern TKA implant systems

Methods

A prospective, nonrandomized, single center study of patients undergoing primary TKA by one of two surgeons was performed with four study cohorts: cemented DePuy Attune, cementless DePuy Attune, cemented Stryker Triathlon, cementless Stryker Triathlon. Target sample size was 80, with 20 per cohort based on adhoc power analysis. Exclusion criteria included: age over 75, BMI >40, inflammatory arthritis, previous knee surgery involving the femur, tibia or tibial bone, and diagnosis of osteopenia/osteoporosis. Implant fixation type was based on surgeon intraoperative assessment of patient bone quality. Demographic data was collected preoperatively. Dual Energy X-ray Absorptiometry (DEXA) Bone Density Monitoring was performed at 6 weeks and one year postoperatively. Bone mineral density was calculated from the DEXA scans for 4 zones for the tibia relative to the keel or central peg: anterior, posterior, medial and lateral. Results were reported as BMD at 1 year postoperatively as a percentage of BMD at 6 weeks postoperatively.