Knee

IMPACT OF COMPONENT DESIGN AND FIXATION IN BONE REMODELLING AFTER TOTAL KNEE ARTHROPLASTY

The Knee Society (TKS) 2020 Members Meeting, held online, 10–12 September 2020.



Abstract

Introduction

Bone mineral density (BMD) is correlated with component migration and aseptic loosening after total knee arthroplasty (TKA). Older implant designs have demonstrated BMD loss up to 23% in the first 6 months after TKA, and continued to BMD decline at an average of 5% per year for as long as 2 years after TKA. The impact of component design and fixation method on BMD loss after TKA in modern implant designs has not been fully elucidated. The purpose of this study is to determine the effect of tibial tray thickness and fixation method (cemented versus cementless) on BMD loss patterns of the proximal tibia in two different modern TKA implant systems

Methods

A prospective, nonrandomized, single center study of patients undergoing primary TKA by one of two surgeons was performed with four study cohorts: cemented DePuy Attune, cementless DePuy Attune, cemented Stryker Triathlon, cementless Stryker Triathlon. Target sample size was 80, with 20 per cohort based on adhoc power analysis. Exclusion criteria included: age over 75, BMI >40, inflammatory arthritis, previous knee surgery involving the femur, tibia or tibial bone, and diagnosis of osteopenia/osteoporosis. Implant fixation type was based on surgeon intraoperative assessment of patient bone quality. Demographic data was collected preoperatively. Dual Energy X-ray Absorptiometry (DEXA) Bone Density Monitoring was performed at 6 weeks and one year postoperatively. Bone mineral density was calculated from the DEXA scans for 4 zones for the tibia relative to the keel or central peg: anterior, posterior, medial and lateral. Results were reported as BMD at 1 year postoperatively as a percentage of BMD at 6 weeks postoperatively.

Results

81 knees with complete DEXA studies at 6 weeks and 1 year postoperatively were available for analysis. The mean (SD) age at the time of surgery was 65.4 years (6.1), with 39 men and 42 women and a mean (SD) BMI of 31.2 kg/m2 (4.7). The patients who received cementless implants were significantly younger with a mean (SD) age of 62.1 (6.0) than those who received cemented implants with a mean age of 69.1 (3.7) (p < 0.05). There was no difference in age between implant type (p > 0.05). There was no difference in BMI or gender distribution between implant types or fixation type (p > 0.05). There was no significant difference in mean BMD loss between 6 weeks and 1 year postoperatively between the cemented and cementless Triathlon in any zone, between the cemented Triathlon and Attune in any zone, the cementless Triathlon and Attune in any zone, and the cemented and cementless attune in the lateral, anterior and posterior proximal tibia (p > 0.05). There was significantly more BMD loss seen in the medial tibia in the cemented (BMD at 1 year 88% of BMD at 6 weeks) than cementless (BMD at 1 year 103% of BMD at 6 weeks) Attune (p = 0.043).

Conclusion

BMD loss at 1 year postoperatively versus 6 weeks postoperatively after TKA with two modern implant designs is not significantly affected by tibial tray thickness. BMD loss was unaffected by fixation type for the Stryker Triathlon, but was significant less for the cementless DePuy Attune than the cemented version.