Introduction. Proper cup positioning is a critical component in the success of total hip arthroplasty surgery. A multicenter study has been initiated to study a new type of highly cross-linked polyethylene. This study provides a unique opportunity to a review the
The MITCH PCR is an anatomic, flexible, horse-shoe shaped acetabular component, with 2 polar fins. The rationale of the PCR cup design is to reproduce a near-physiological stress distribution in the bone adjacent to the prosthesis. The thin composite cup is designed to fuse and flex in harmony with the surrounding bony structure. Only the pathological acetabular cartilage and underlying subchondral bone of the horseshoe-shaped, load-bearing portion of the acetabular socket is replaced, thus preserving viable bone stock. The PCR is manufactured from injection moulded carbon fibre reinforced polyetheretherketone (PEEK), with a two layer outer surface comprising hydroxyapatite and plasma sprayed commercially pure titanium. It is implanted in conjunction with a large diameter low wear femoral head, producing a bearing that will generate minimal wear debris with relatively inert particles. Pre-clinical mechanical testing, finite element analysis and biocompatibility studies have been undertaken. FEA evaluation predicts preservation of host bone density in the load bearing segments. A pilot clinical study was completed on a proto-type version of the PCR cup (the “Cambridge” cup), achieving excellent 5 and 10 year results. We report the three-year results from a two-centre, prospective clinical evaluation study of the MITCH PCR cup. Patient outcome has been assessed using standardised clinical and radiological examinations and validated questionnaires. The change in physical level of activity and quality of life has been assessed using the Oxford Hip Score, Harris Hip score and the EuroQol-5D score, at scheduled time-points. Serial radiographs have been analysed to monitor the fixation and stability of the components.Introduction
Subjects and Methods
Introduction.
Introduction. Alternative bearings – metal-on-metal and ceramic-on-ceramic bearings have been introduced in the last decade with the aim to diminish wear and, subsequently, aseptic loosening and osteolysis. These bearings were aimed for younger, more active patients. Clinical results which would compare the performance of various alternative and traditional bearings are scarce. Methods. Between January 2000 and December 2002, we performed 487 total hip replacements in 474 patients using three types of bearings: metal-on-metal (MOM), metal-on-polyethylene (MOP), and ceramic-on-ceramic (COC). All patients received the same type of
The concept of stainless steel dual mobility cups in total hip arthroplasty has demonstrated very low long-term instability rates and a 98% survival rate after 12 years. We systematically implanted titanium alloy
Introduction. The advantages of metal on metal (MOM) hip replacement are decreased wear rate, preservation of bone stock, anatomical restoration and enhanced stability. Large amounts of metal wear particles and metal ions are released which may induce adverse reactions including local soft tissue toxicity, hypersensitivity reactions, bone loss and risk of carcinogenesis. Aseptic loosening can be the result of a peri prosthetic osteolysis generated as a result of a biological response to particulate wear debris. Our aim in this study was to determine whether a steeply inclined acetabular component would give rise to a higher concentration of metal ions. Patients and methods. Between April 2003 and June 2006, 22 patients had MOM hip replacement for osteoarthritis by a single Surgeon. There were 12 male and 10 female patients. The average age at the time of surgery was 56 years (Range: 44–69 years). We divided the 22 patients into 2 groups, one group (A) of 11 patients with the acetabular inclination angle more than 50 degrees and the other group (B) of 11 patients with the angle less than 50 degrees. The inclination of the
The prevalent cause of implant failure after total joint replacement is aseptic loosening caused by wear debris. Improvement of the wear behaviour of the articulating bearing between the cup and femoral head is essential for increased survival rate of artificial hip joints. Cross-linking of the polyethylene (PE) material is one attempt to reduce wear particle release at the articulating surface. Various cross-linked polyethylenes (X-PE) are used in orthopaedics since several years. In total hip arthroplasty (THA) the use of larger femoral head sizes has specific reasons. Larger heads lead to a decreased risk of total hip dislocation and impingement as well as an improved range of motion in comparison to smaller head sizes like 28mm or less. However, the increasing diameter of femoral head can be associated with lower thickness of the PE liner and increased wear rate. Cross-linking of PE can improve the wear rate of the liner and hence supports the use of larger femoral heads. The aim of this experimental study was to evaluate the wear of standard vs. sequential X-PE (X3-PE) liner in combination with different ceramic femoral head sizes. Wear testing was performed for 5 million load cycles using standard UHMW-PE liners (N2Vac) and X3-PE liners (each Stryker GmbH & Co. KG, Duisburg, Germany) combined with 28mm ceramic ball heads and the Trident PSL
INTRODUCTION. Whilst there is a great deal of research on hip implants, few studies have looked at implant orientation and the subsequent effect upon the wear performance of a hip resurfacing. This study aimed to measure implantation angles through radiographic analysis and linear wear for retrieved
Introduction. Hip wear simulation is a widely used technique for the pre-clinical evaluation of new bearing designs. However, wear rates seen in vitro can often be significantly different to those seen clinically. This can be attributed to the difference between the optimal conditions in a simulator and wide ranging conditions in real patients. This study aimed to develop more clinically relevant simulator tests, looking specifically at the effects of cup inclination angle (in vivo) and stop-dwell-start (SDS) protocols on a clinically available product. Method. Five tests using a Paul type walking cycle (ISO 14242) were carried out on two ProSim hip simulators:
. 28mm MoM, standard walking, cup inclination 45°, (n = 5). 36mm MoM, standard walking, cup 45°, (n = 4). 36mm MOM, SDS: 10 walking cycles and pause of 5s with stance load of 1250N cup 45°, (n = 5). 36mm MOM, SDS: 10 walking cycles and pause of 30s with stance load of 1250N, (n = 5) cup 45°. 36mm MOM, standard walking, cup 55°(n = 5), and 65°(n = 5). All samples had matched clearances, measured using a CMM (Prismo Navigator, Zeiss, Germany). Wear was measured gravimetrically (Sartorius ME235S: 0.01mg). Results and Discussion. Metal-on-metal hip bearings are known to exhibit a bi-phasic wear pattern in-vitro (1), with the majority of wear occurring over the first 1–1.5mc, followed by lower steady state wear as the components conform to each other. Therefore bedding in wear over the first 1 million cycles was compared for each of the sample groups in this study. Bedding in wear for 36mm bearings under standard angle (45 inclination) and walking conditions was 0.16±0.15mm3/mc, while for 28mm bearings, wear was 2.67±2.06mm3/mc. Positioning the
Early total hip replacement (THR) for acetabular
fractures offers accelerated rehabilitation, but a high risk of heterotopic
ossification (HO) has been reported. The purpose of this study was
to evaluate the incidence of HO, its associated risk factors and
functional impact. A total of 40 patients with acetabular fractures
treated with a THR weres retrospectively reviewed. The incidence
and severity of HO were evaluated using the modified Brooker classification,
and the functional outcome assessed. The overall incidence of HO
was 38%
(n = 15), with nine severe grade III cases. Patients who underwent
surgery early after injury had a fourfold increased chance of developing
HO. The mean blood loss and operating time were more than twice
that of those whose surgery was delayed (p = 0.002 and p <
0.001,
respectively). In those undergoing early THR, the incidence of grade
III HO was eight times higher than in those in whom THR was delayed
(p = 0.01). Only three of the seven patients with severe HO showed
good or excellent Harris hip scores compared with eight of nine
with class 0, I or II HO (p = 0.049). Associated musculoskeletal
injuries, high-energy trauma and head injuries were associated with
the development of grade III HO. The incidence of HO was significantly higher in patients with
a displaced acetabular fracture undergoing THR early compared with
those undergoing THR later and this had an adverse effect on the
functional outcome. Cite this article: