Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_16 | Pages 38 - 38
17 Nov 2023
Al-Namnam NM Luczak AT Collishaw S Li X Lucas M Simpson AHRW
Full Access

Abstract

Introduction

Ultrasonic cutting in surgery has great potential. However, a key limitation is heat created by friction between the bone and the blade. Bone has poor thermal conductivity which hinders the dissipation of heat, causing cell death near the cut site In addition, ultrasonic vibration may create microcracks. It was hypothesised that these effects on bone would vary with the frequency and displacement of the ultrasonically powered blade. Therefore varying frequencies and displacements of the tip of the blade were studied to find the combination with fewest microcracks and lowest temperature rise at the bone-tool interface.

Aim

To explore the effect of different frequencies and tip displacements of ultrasonic cutting devices on the amount of thermal and mechanical damage.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_16 | Pages 59 - 59
17 Nov 2023
Yang I Buchanan R Al-Namnam N Li X Lucas M Simpson AH
Full Access

Abstract

Background

Ultrasonic cutting of bone boasts many advantages over alternatively powered surgical instruments, including but not limited to: elimination of swarf, reduced reaction forces, increased precision in cutting and reduced adjacent soft tissue damage, reduced post-operative complications such as bleeding and bone fracture, reduced healing time, reduced intra-operative noise and ease of handling. Despite ultrasonic cutting devices being well established in oral and maxillofacial surgery, applications in orthopaedic surgery are more niche and are not as well understood. The aim of this study was to investigate the cutting speed (mm/s) and cutting forces (N) of orthopaedic surgeons using a custom-designed state of the art ultrasonic cutting tool to cut fresh human bone samples.

Methods

A setup based on the Robot Operating System (ROS) and AprilTag was designed to track and to record the real time position of the ultrasonic cutting tool in space. Synchronised load cell axial force readings of three separate orthopaedic surgeons during ultrasonic cutting were recorded. Each surgeon was asked to find a comfortable position that reflects as close as possible their clinical handling of a cutting instrument used in surgery, and to perform two cuts in each of three samples of human cortical bone. Bone samples were obtained following ethical approval from an institutional review board (ethics approval number: SR1342) and prior informed consent was obtained from all patients. Bone samples were extracted from the femoral neck region of three hip osteoarthritis patients. During cutting, surgeons were allowed a total cutting time of one minute and cutting was conducted using an ultrasonic tool with frequency of a 35kHz (35.7 µm peak to peak displacement amplitude) under constant irrigation using a MINIPULS® 3 Peristaltic pump (38 revolutions per minute) using Phosphate-Buffered Saline (PBS) at 25°C. From the recorded data, the average instantaneous cutting velocity was calculated and the maximum cutting force was identified.


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 37 - 37
1 Jul 2014
Molina-Manso D Del-Prado G Lucas-Diaz M Gómez-Barrena E Cordero-Ampuero J Esteban J
Full Access

Summary Statement

Combination of antibiotics with N-acetylcisteine and sub-MIC concentration of erythromycin was evaluated in two collection and 16 clinical strains of staphylococci isolated from PJI. The results were strain-dependent, so it evidences the necessity of perform individual studies of biofilm susceptibility.

Objectives

Staphylococci are the most common cause of prosthetic joint infections (PJI) (1), making the treatment of this disease difficult due to the increased resistance to antibiotics of biofilms. Combination between antibiotics and other compounds could be a good alternative. The aim of this study was to evaluate the effect of the combination of two compounds with nine antibiotics in biofilms formed by staphylococcal strains isolated from PJI.