We report a multi-centre prospective clinical trial. 231 hips in 223 patients have been entered into the study. 151 of these have reached 3 years follow-up.
In particular, the RSA subsidence characteristics, cement pressurisation and rotational stability already associated with this implant in-vitro have been supported by excellent survivorship analysis, and the authors believe that increasing familiarity with the concepts raised by this implant will result in clinical benefits in relation to polished taper cemented stem longevity.
Data on the CPS-Plus stem has been obtained from a multi-centre prospective clinical trial. 231 hips in 223 patients have been entered into the study. 151 of these have reached 3 years follow-up.
In particular, the RSA subsidence characteristics, cement pressurisation and rotational stability already associated with this implant in-vitro have been supported by excellent survivorship analysis, and the authors believe that increasing familiarity with the concepts raised by this implant will result in clinical benefits in relation to polished taper cemented stem longevity.
The static properties of bone cements have been widely reported in the literature (Lewis, 1997, Khun, 2000, Armstrong 2002). Commercial bone cements are expected to perform above the minimum values in static tests specified by ISO 5833: 2002. It has been suggested that the viscoelastic properties of bone cement, such as creep and stress relaxation, might bear more relevance to the in-vivo behaviour of the cement-implant construct (Lee 2002). This study aimed to compare numerous properties of Simplex P, Simplex Antibiotic and Simplex Tobramycin and identify those properties most sensitive to subtle changes in cement composition. The three cements were chosen on the basis that they are characterised by the same liquid and powder compositions, the only difference being represented by the type and amount of added antibiotics. In Simplex Antibiotic the additives are 0.5g Erythromycin and 3 million I.U. Colistin, while in Antibiotic Simplex with Tobramycin the only additive is 0.5g of Tobramycin. The static properties of the cements were assessed following protocols described in ISO 5833: 2002, while the viscoelastic properties of the cement were measured with in-house developed apparatus in quasi-static conditions. Creep and stress relaxation tests were performed in four point bending configuration. Porosity was measured on the mid cross section of the creep samples using a digital image technique. All cements exhibited properties compatible with the ISO standard, but in plain Simplex the ISO minimum for bending and compressive strength was within the variation of the batches tested. Bending strength measurements were the least sensitive to differences in the cements. Plain Simplex displayed lower bending and compressive strength but higher bending modulus than the antibiotic laden options. The bending modulus could only discriminate between Simplex P and Simplex Antibiotic (p=0.02). Differences in the compressive strength of the three cements were significant, with the plain option being the weakest. Stress relaxation only discriminated between plain and Tobramycin loaded cement (p=0.028), while creep was more sensitive to differences and allowed distinction between plain and antibiotic loaded bone cements. The creep behaviour correlated with the cross sectional porosity measurements. This study demonstrated that the static tests specified by the current international standard are not as sensitive to subtle changes in the composition of the material as the time temperature dependent parameters characteristic of creep and stress relaxation. The authors advocate the evaluation of time and temperature dependent characteristics as a complement to the current standard.
Revision hip surgery is becoming increasingly common, 300 procedures being performed in 2001 at our institution. In order to achieve a good outcome bone stock needs to be of good quantity frequently necessitating the use of impaction bone grafting using allograft bone. Donor bone may frequently take three months before it becomes available for use due to the stringent screening procedure. Donor patients must have a clean bill of health, swabs taken at the time of surgery must obviously demonstrate no growth and blood samples taken at donation and an interval of three months, free from viral infectious diseases. It is thus easy to see the lag from the time of donation to availability and why, with increasing demand, need for allograft bone is rapidly exceeding supply. We need to look for an alternative supply of human bone allograft. We have compared the harvest of bone at the time of primary total knee replacement with that of the femoral head by both mass and volume. Sixty consecutive patients undergoing primary hip or knee arthroplasty were included in the study, and the masses and volume of the femoral heads compared with that of the total bone cuts in knee arthroplasty. The type of knee replacement used was documented as was whether the femoral head had had a bone block removed. It was found that the mass of femoral heads was 81g, that of knee cuts 95g this is a statistically significant difference; the volume of femoral heads 66ml and that of knee cuts 75ml. The volumes of bone available from knee arthroplasty cuts are at least comparable femoral heads obtained using hip replacement and could, perhaps, provide a realistic source of bone allograft.
We undertook a radiological evaluation of this technique. We assessed fracture union and strut allograft incorporation using the radiological criteria of Emerson et al. The procedure was deemed a success if the fracture had united, with evidence of graft incorporation with a stable implant. We also undertook a notes review identifying any risk factors and any previous surgery.
Retrospective review of 25 patients over 65 years of age with unilateral acetabular fractures managed conservatively showed that seven of the 23 survivors (30%) had an unacceptable functional result. Poor results were associated with: displaced posterior column fractures, osteoporosis, femoral head fracture, delayed diagnosis, inadequate radiographs, inappropriate or too brief traction, and early weight-bearing. In this age group acetabular fractures tend to be low-velocity injuries of osteoporotic bone and are not comparable with those in younger patients.
Using a simple method of quantifying fracture healing, 53 patients who had limb fractures and also severe head injuries were studied; they were compared with 30 patients who had limb fractures but no head injury. Those with head injuries had a greater healing response and united more rapidly. Radiological and histological analysis revealed that the terms "myositis ossificans" and "heterotopic bone" may be more appropriate than "fracture callus" to describe the healing response in these patients.
We report a series of 12 Silastic interposition arthroplasties of the shoulder. There was a very high incidence of dislocation or fragmentation (50%) leading to early failure of the device in seven patients. Our analysis suggests that the operation should be restricted to rheumatoid arthritic shoulders in which there is no loss of bone or distortion of the humeral head.