Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_4 | Pages 144 - 144
1 Feb 2017
Matz J Lanting B Teeter M Howard J
Full Access

Introduction

Complications related to the patellofemoral joint continue to be a substantial source of patient morbidity, causing anterior knee pain, instability, and dysfunction following total knee arthroplasty. One of the principle factors affecting patellofemoral outcomes may be trochlear design. The optimal design is currently unknown. The purpose of the present study was to study patellofemoral joint contact by analysing areas of wear in retrieved femoral components of three modern designs.

Materials and Methods

Eighteen retrieved femoral components featuring three different designs (constant radius of rotation, multiple radii of rotation, and multiple radii of rotation with built-in external rotation design) were matched on the basis of time-in-vivo, age, BMI and gender. All implants were cobalt chrome, posterior stabilized, cemented components with fixed bearing design with a resurfaced patella. Trochlear wear and surface damage were assessed using visual inspection, low-magnification light microscopy, and light profilometry.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_21 | Pages 62 - 62
1 Dec 2016
Matlovich N Lanting B MacDonald S Teeter M Howard J
Full Access

The concept of constitutional varus and controversy regarding placing the total knee arthroplasty (TKA) in a neutral versus physiologic alignment in varus osteoarthritic (OA) patients is an important current discussion. However, the physiologic mechanical alignment of a varus OA knee is unknown and the relative contribution of the femur and tibia to the mechanical axis is unknown. The goal of this study was to determine and analyse the physiologic mechanical axis of medial OA knees.

Plain radiographs of the knee and full-leg standing radiographs of 1558 patients were reviewed for inclusion criteria; 313 patients with a non-arthritic knee and a contralateral varus end-stage OA knee were analysed in the coronal plane. The Hip-Knee-Ankle (HKA), Condylar-Hip (CH)(femoral), Condylar-Plateau (CP) (intra-articular) and Plateau-Ankle (PA)(tibial) angles were measured for both the arthritic and non-arthritic/physiologic knee. The relationship and contribution of all angles was analysed for every 2° degrees of progressive varus: from 4° valgus to 8° varus. The proportion of patients with constitutional varus was also determined for the sample population and correlated with increasing HKA.

The mean CH (femoral) angle was valgus in all groups and decreased with progressive varus alignment (p< 0.0001), ranging from 3.8° ± 1.0° with HKA of 2–4° valgus, to 0.1° ± 1.5° with HKA of 6–8° varus. The mean PA (tibial) angle was varus in all groups and decreased from valgus to progressively varus alignment (p p<0.0001), ranging from 0.78° ± 1.4° with HKA 2–4° valgus, to 5.6° ± 1.9° with HKA 6–8° varus. The CP angle showed no difference between all groups (p=0.3). Forty five percent of males and 22% of females with arthritic HKA in varus alignment were found to have constitutional varus.

Correlation of unilateral arthritic knees to the unaffected, physiologic aligned knee using full-leg radiographs indicates that it may be possible to understand the patient's physiologic, pre-arthritic coronal plane alignment. The mechanical axis of physiologic knees in a unilateral varus OA population demonstrates a variable contribution of the femur (CH) and tibia (PA) from overall valgus to varus alignment. In addition, a significant proportion of the sample population possessed constitutional varus. This may provide important information regarding the placement of physiologic TKA's and direct future research questions.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_21 | Pages 55 - 55
1 Dec 2016
Lanting B Thoren J Yuan X McCalden R McAuley J MacDonald S Vasarhelyi E Howard J Naudie D Teeter M
Full Access

Adequate fixation of implant components is an important goal for all arthroplasty procedures. Aseptic loosening is one of the leading causes of revision surgery in total knee arthroplasty. Radiostereometric analysis (RSA) is an imaging technique to measure implant migration, with established migration thresholds for well-fixed, at risk, and unacceptably migrating components. The purpose of the present study was to examine the long-term fixation of a cemented titanium fixed bearing polished tibial baseplate.

Patients enrolled in a previous two-year prospective trial were recalled at ten years. All patients received a cemented, posterior-stabilised total knee replacement of the same design implanted by one of three surgeons. Of the original 35 patients, 16 were available for long-term follow-up, with one patient lost to follow-up, nine patients deceased, and a further nine patients unwilling to return to the clinic. Each patient underwent RSA imaging in a supine position using a conventional RSA protocol. Migration of the tibial component in all planes as well as maximum total point motion (MTPM) was compared between all time points (baseline, six weeks, three months, six months, one year, two years) up to the ten year follow-up visits. Outcome scores including the Knee Society Score (KSS), WOMAC, SF-12, and UCLA Activity Score were recorded.

At ten years, the mean migrations of the tibial component were less than 0.1 mm and 0.1 degree in all planes relative to the post-operative RSA exam. There was no significant difference in tibial component migration between time points. However, MTPM increased significantly over time (p = 0.002), from 0.23 ± 0.18 mm at six weeks to 0.42 ± 0.20 mm at ten years. At one year, 13 patients had an acceptable MTPM level, three patients had an ‘at risk’ level, and no patient had an ‘unacceptable’ level. No patients were revised at ten years. WOMAC and KSS were significantly improved (p < 0.0001) at the latest follow-up compared to pre-operatively, but there was no difference in SF-12. The median UCLA Activity Score at latest follow-up was six (range, two to eight).

The tibial baseplate demonstrated solid fixation at ten years. No patients had an unacceptable MTPM level at one year and no patients were revised at ten years, supporting the use of RSA to predict long-term loosening risk. The low level of tibial baseplate migration found in the present study correlates to the low rate of revision for this implant as reported in individual studies and in joint replacement registries.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_21 | Pages 59 - 59
1 Dec 2016
Sisko Z Teeter M Lanting B Howard J McCalden R Naudie D MacDonald S Vasarhelyi E
Full Access

Previous retrieval studies demonstrate increased tibial baseplate roughness leads to higher polyethylene backside wear in total knee arthroplasty (TKA). Micromotion between the polyethylene backside and tibial baseplate is affected by the locking mechanism design and can further increase backside wear. The purpose of this study was to examine modern locking mechanisms, in the setting of both roughened and polished tibial baseplates, on backside tibial polyethylene wear.

Five TKA models were selected, all with different tibial baseplate and/or locking mechanism designs. Six retrieval tibial polyethylenes from each TKA model were matched based on time in vivo (TIV), age at TKA revision, BMI, gender, number of times revised, and revision reason. Two observers scored each polyethylene backside according to a visual damage score and individual damage modes. Primary outcomes were mean damage score and individual damage modes. Demographics were compared by one-way ANOVA. Damage scores and modes were analysed by the Kruskal-Wallis test and Dunn's multiple comparisons test.

There were no differences among the groups based on TIV (p=0.962), age (p=0.651), BMI (p=0.951), gender, revision number, or reason for revision. There was a significant difference across groups for mean total damage score (p=0.029). The polished tibial design with a partial peripheral capture locking mechanism and anterior constraint demonstrated a significantly lower score compared to one of the roughened tibial designs with a complete peripheral-rim locking mechanism (13.0 vs. 22.1, p=0.018). Otherwise, mean total damage scores were not significant between groups. As far as modes of wear, there were identifiable differences among the groups based on abrasions (p=0.005). The polished design with a tongue-in-groove locking mechanism demonstrated a significantly higher score compared to both groups with roughened tibial baseplates (5.83 vs. 0.83, p=0.024 and 5.83 vs. 0.92, p=0.033). Only the two designs with roughened tibial baseplates demonstrated dimpling (5.67 and 8.67) which was significant when compared against all other groups (p0.99). No other significant differences were identified when examining burnishing, cold flow, scratching, or pitting. No polyethylene components exhibited embedded debris or delamination.

Total damage scores were similar between all groups except when comparing one of the polished TKA design to one of the roughened designs. The other TKA model with a roughened tibial baseplate had similar damage scores to the polished designs, likely due to its updated locking mechanism. Dimpling wear patterns were specific for roughened tibial baseplates while abrasive wear patterns were identified in the design with a tongue-in-groove locking mechanism. Our study showed even in the setting of a roughened tibial baseplate, modern locking mechanisms decrease backside wear similar to that of other current generation TKA designs.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_20 | Pages 59 - 59
1 Nov 2016
Goyal P Yuan X Teeter M McCalden R MacDonald S Vasarhelyi E McAuley J Naudie D Lanting B Howard J
Full Access

Studies that have previously examined the relationship between inclination angle and polyethylene wear have shown increased wear of conventional polyethylene with high inclination angles. To date, there have been no long term in vivo studies examining the correlation between cup position and polyethylene wear with highly crosslinked polyethylene.

An institutional arthroplasty database was used to identify patients who had metal-on-highly crosslinked polyethylene primary total hip arthroplasty (THA) using the same component design with a minimum follow up of 10 years ago. A modified RSA examination setup was utilised, recreating standard anteroposterior (AP) and cross-table lateral exams in a single biplane RSA acquisition. Three dimensional head penetration was measured using the centre index method. The same radiographs were used to measure inclination angle and anteversion. Spearman correlation was used to show an association between the parameters of acetabular position and wear rate.

A total of 43 hips were included for analysis in this study. Average follow-up was 12.3 ± 1.2 years. The average linear wear rate was calculated to be 0.066 ± 0.066 mm/year. Inclination angle was not correlated with polyethylene wear rate (p=0.82). Anteversion was also not correlated with polyethylene wear rate (p=0.11). There was no statistical difference between wear rates of hips within Lewinnek's “safe zone” and those outside this “safe zone” (p=0.11). Males had a higher wear rate of 0.094 ± 0.089 mm/year compared to females with a wear rate of 0.046 ± 0.032 mm/year (p=0.045).

At long term follow up of greater than 10 years, highly cross linked polyethylene has very low wear rates. This excellent tribology is independent of acetabular position, but gender did impact wear rates. Due to the low wear rates, follow-up of even longer term is suggested to examine variables affecting wear.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_20 | Pages 87 - 87
1 Nov 2016
Matz J Morden D Teeter M McCalden R MacDonald S Vasarhelyi E McAuley J Naudie D Howard J Lanting B
Full Access

Complications involving the patellofemoral joint are a source of anterior knee pain, instability, and dysfunction following total knee arthroplasty. “Overstuffing” the patello-femoral joint refers to an increase in the thickness of the patellofemoral joint after a total knee replacement compared to the preoperative thickness. While biomechanical studies have indicated that overstuffing the patellofemoral joint may lead to adverse clinical outcomes, limited clinical evidence exists to support this notion. The purpose of this study is to evaluate the effect of changing the thickness of the patellafemoral joint on functional outcomes following total knee arthroplasty.

Our institutional arthroplasty database was used to identify 1347 patients who underwent a primary total knee arthroplasty between 2006 and 2012 with the same component design. Standard preoperative and postoperative anteroposterior, lateral, and skyline radiographs were collected and measured for patello-femoral overstuffing. These measurements included anterior patellar displacement, anterior femoral offset, and anteroposterior femoral size. These measurements were correlated with patient outcome data using WOMAC, KSS scores, and postoperative range of motion. Multiple linear regression analysis was used to assess the association between stuffing and functional outcomes.

A total of 1031 patients who underwent total knee arthroplasty were included. Increased anterior patellar displacement, a measure of patellofemoral joint thickness, was associated with decreased WOMAC scores (p=0.02). Anterior femoral offset (p=0.210) and anteroposterior femoral size (p=0.091) were not significantly associated with patient functional outcomes. Postoperative range of motion (ROM) was not associated with patellofemoral stuffing (p=0.190).

The current study demonstrated that functional outcomes are adversely affected by patellofemoral overstuffing. Based on these results, caution is encouraged against increasing the thickness of the patellofemoral joint, particularly on the patellar side of the joint.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_20 | Pages 64 - 64
1 Nov 2016
Teeter M Naudie D McCalden R Yuan X MacDonald S
Full Access

The philosophy of cemented total hip arthroplasty (THA) femoral components has become polarised. At one extreme are polished, collarless, tapered devices that are expected to subside; at the other extreme are roughened, non-tapered implants with a collar designed not to subside. Radiostereometric analysis (RSA) allows the accurate measurement of implant movement and has been extensively used for measurement of the in vivo migration of implants. The degree of migration as measured by RSA during the first years after surgery has been shown to correlate with the long-term performance of cemented femoral implants. The purpose of this study was to review the two-year RSA results of two different designs of primary cemented THA stems.

Data from two previous prospective RSA trials with two-year follow-up were pooled. The first group included 36 patients who received a Spectron (Smith & Nephew, Memphis, USA) cemented stem. The second group included 13 patients who received an Exeter (Stryker, Mahwah, USA) cemented stem, and 15 patients who received a CPCS (Smith & Nephew, Memphis, USA) cemented stem. All patients underwent RSA examinations shortly post-operation, at 6 weeks, 3 months, 6 months, 1 years, and 2 years. Migration and rotation of the femoral stems was measured at each time point relative to the post-operative exam, and compared between the two groups.

There was no difference in age at surgery (Spectron 78 ± 6 years, Exeter/CPCS 77 ± 5 years, p = 0.43), BMI (Spectron and Exeter/CPCS 28 ± 5 kg/m2, p = 0.92), or percentage of male patients (Spectron 23% male, Exeter/CPCS 21% male) between the implant groups. Subsidence was significantly greater (p < 0.0001) at all time points from three months to two years for the Exeter and CPCS stems (0.94 ± 0.39 mm at two years) compared to the Spectron stem (0.05 ± 0.16 mm at two years). There was no significant difference between the stem types for medial-lateral translation (p = 0.07) or anterior-posterior translation (p = 0.49), or for anterior-posterior tilt (p = 0.15), internal-external rotation (p = 0.89), or varus-valgus rotation (p = 0.05).

Implant material, design, and surface finish are all factors in the long-term performance of cemented femoral hip implants. In this study, both femoral stem designs had a magnitude of sub-sidence that was within the limits of what is considered to be safe with respect to long-term performance. The continuous subsidence of the Exeter and CPCS stems is consistent with previous reports in the literature.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_20 | Pages 57 - 57
1 Nov 2016
Lanting B Tan S Lau A Teeter M Del Balso C McCalden R MacDonald S Vasarhelyi E McAuley J Naudie D Howard J
Full Access

Trunnionosis in modular hip arthroplasty has recently been recognised to be clinically important. Gaining an understanding of how the material interface at the head-trunnion affects the tribology at the modular junctions has current clinical implications as well as an implication on future implant selection and material choice. This matched-cohort study aims to compare tribocorrosion between ceramic and cobalt-chromium trunnions and to investigate other factors that contribute to the difference in tribocorrosion if present.

All hip prostheses retrieved between 1999 and 2015 at one centre were reviewed. Fifty two ceramic heads were retrieved, and these were matched to a cobalt-chromium cohort according to taper design, head size, neck length and implantation time in that order. The two cohorts were similar in male:female ratio (p=0.32) and body mass index (p=0.15) though the ceramic group was younger than the cobalt-chromium group (56.6 (+/−)13.5 years for ceramic group vs 66.3 (+/−14.4) years for cobalt-chromium group; p=0.001). There was no significant difference in the reasons for revision between the two groups (p=0.42). The femoral head trunnions were examined by two independent observers using a previously published 4-point scoring technique. The trunnions were divided into three zones: apex, middle and base. The observers were blinded to clinical and manufacturing data where possible.

Ceramic head trunnions demonstrated a lower median fretting and corrosion score at the base zone (p<0.001), middle zone (p<0.001) and in the combined score (p<0.001). In a subgroup analysis by head size, ceramic heads had a lower fretting and corrosion score at 28mm head diameter (p<0.001). Within the ceramic group, taper design had a significant effect on fretting and corrosion in the apex zone (p=0.04). Taper design also had a similar effect in the cobalt-chromium group in the apex zone (p=0.03). For the ceramic trunnions, the largest effect was contributed by the difference between the 11/13 taper and the 12/14 taper. For the cobalt-chromium trunnions, the largest effect was contributed by the difference between the 5 degree 38′ 37″ taper and type 1 taper.

Ceramic head trunnions showed a significantly lower fretting and corrosion score as compared to cobalt-chromium trunnions. Ceramic heads had a lower score than cobalt-chromium heads at 28mm head diameter. Taper design had an effect on fretting and corrosion within each cohort.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_20 | Pages 62 - 62
1 Nov 2016
MacLean C Vasarhelyi E Lanting B Naudie D Somerville L McCalden R McAuley J MacDonald S Howard J Yuan X Teeter M
Full Access

The advent of highly cross-linked polyethylene has resulted in improved wear rates and reduced osteolysis with at least intermediate follow-up when compared to conventional polyethylene. However, the role of alternative femoral head bearing materials in decreasing wear is less clear. The purpose of this study was to determine in-vivo polyethylene wear rates across ceramic, Oxinium, and cobalt chrome femoral head articulations.

A review of our institutional database was performed to identify patients who underwent a total hip arthroplasty using either ceramic or oxidised zirconium (Oxinium) femoral head components on highly cross-linked polyethylene between 2008 and 2011. These patients were then matched on implant type, age, sex and BMI with patients who had a cobalt chrome bearing implant during the same time period. RSA analysis was performed using the centre index method to measure femoral head penetration (polyethylene wear). Secondary quality of life outcomes were collected using WOMAC and HHS Scores. Paired analyses were performed to detect differences in wear rate (mm/year) between the cobalt chrome cohorts and their matched ceramic and Oxinium cohorts. Additional independent group comparisons were performed by analysis of variance with the control groups collapsed to determine wear rate differences between all three cohorts.

A total of 68 patients underwent RSA analysis. Fifteen patients with a ceramic femoral head component and 14 patients with an Oxinium femoral head component along with the same number of matched patients with cobalt chrome femoral head component were included in the analysis. The time in vivo for the Oxinium (5.17 +/− 0.96 years), Oxinium matched cohort (5.13 +/− 0.72 years), ceramic (5.15 +/− 0.76 years) and ceramic matched cohort (5.36 +/− 0.63 years) were comparable. The demographics of all bearing surface cohorts were similar. The paired comparison between the Oxinium and cobalt chrome cohorts (0.33 vs. 0.29 mm/year, p=0.284) and ceramic vs cobalt chrome cohorts (0.26 vs. 0.20 mm/year, p=0.137) did not demonstrate a significant difference in wear rate. The independent groups analysis revealed a significantly higher wear rate of Oxinium (0.33 mm/year) compared to cobalt chrome (0.24 mm/year) (p = 0. 038). There were no differences in HHS and WOMAC scores between the Oxinium and cobalt chrome cohorts (HHS: p = 0.71, WOMAC: p=0.08) or the ceramic and cobalt chrome cohorts (HHS: p=0.15, WOMAC: p=023).

This study presents evidence of a greater wear rate (mm/year) of the Oxinium femoral head component compared to a cobalt chrome femoral head component. This difference was not demonstrated in the ceramic femoral head component. Despite this difference, there were no clinical differences as measured by the HHS and WOMAC. Future research should focus on factors that may contribute to the higher wear rate seen in the Oxinium cohort.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_20 | Pages 5 - 5
1 Nov 2016
Teeter M Lam K Howard J Lanting B Yuan X
Full Access

Radiostereometric analysis (RSA) has become the gold standard technique for measuring implant migration and wear following joint replacement due to its high measurement precision and accuracy. However, RSA is conventionally performed using two oblique radiographic views with the presence of a calibration cage. Thus, a second set of radiographs must be acquired for clinical interpretation, for example anterior-posterior and cross-table lateral views following total hip arthroplasty (THA). We propose a modification to the RSA setup for examining THA, in which RSA measurements are performed from anterior-posterior and lateral views, with the calibration cage images acquired separately from the patient images. The objective of the current study was to compare the accuracy and precision of the novel technique to the conventional technique using a phantom.

X-ray cassette holders were developed to enable simultaneous acquisition of anterior-posterior and cross-table lateral radiographs with the patient in a supine position in the RSA suite. A Sawbones phantom with total hip implant components was attached to a micrometer-driven stage. The femoral component was translated known distances relative to the acetabular cup in all planes, mimicking head penetration due to wear. Double RSA examinations were acquired for each increment using the traditional and novel radiograph orientations. Translations were measured from the radiographic images using RSA software. For both techniques, accuracy was calculated by comparing the measured translations to the known translation from the micrometer, and reported as the 95% confidence interval. Precision was measured by comparing the measured translations between the double exams, and reported as the standard deviation.

Accuracy was greater for the conventional technique in the inferior-superior axis (p = 0.03), greater for the novel technique in the anterior-posterior axis (p = 0.01), and equivalent in the medial-lateral axis (p = 0.06). Overall accuracy for both the conventional and novel techniques was identical at ±0.022 mm. Precision was equivalent between both techniques for the medial-lateral (p = 0.68), inferior-superior (p = 0.14), and anterior-posterior axes (p = 0.86). Overall precision for the conventional technique was ±0.127 mm and for the novel technique was ±0.095 mm.

Utilising standard clinical radiograph view angles within an RSA exam had no detrimental effect on wear measurement accuracy or precision. This reduces the barriers to implementing RSA imaging in routine follow-up of arthroplasty patients, potentially greatly increasing the numbers of patients that can have quantitative data on implant performance. Future applications can involve applying more clinically relevant radiograph view angles to RSA exams of the knee and shoulder.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 92 - 92
1 Jan 2016
Vandekerckhove P Teeter M Naudie D Howard J MacDonald S Lanting B
Full Access

Summary

The effect of polyethylene wear and lift-off between the tibial and femoral components on the mechanical axis was assessed in primary TKA (Total Knee Arthroplasty) based on retrieval data and full leg radiographs.

Introduction

Controversy exist regarding performing a TKA with component placement in physiologic versus neutral alignment. Some literature indicates good survivorship and superior clinical outcome in undercorrected TKA's for varus osteoarthritic knees. However, other literature indicates decreased survivorship and coronal plane alignment is still one of the contributing factors to wear in total knee arthroplasty (TKA). The two determinants of the intra-articular deformity in TKA arepolyethylene wear and lift-off between the tibial and femoral compartment. The goal of this study was to evaluate the impact of wear and lift-off on the mechanical axis in neutral and varus aligned TKA's.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 94 - 94
1 Jan 2016
Teeter M Lanting B Vasarhelyi E Ivanov T Vandekerckhove P Howard J Naudie D
Full Access

Increased modularity of total hip arthroplasty components has occurred, with theoretical advantages and disadvantages. Recent literature indicates the potential for elevated revision rates of modular neck systems and the potential for metallosis and ALVAL (Aseptic Lymphocyte dominated Vasculitis Associated Lesion) formation at the modular neck/stem site. Retrieval analysis of one modular neck implant design including SEM (Scanning Electron Microscopy) assessment was done and correlated to FEA (Finite Element Analysis) as well as clinical features of patient demographics, implant and laboratory analysis. Correlation of the consistent corrosion locations to FEA indicates that the material and design features of this system may result in a biomechanical reason for failure. The stem aspect of the modular neck/stem junction may be at particular risk.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 93 - 93
1 Jan 2016
Vandekerckhove P Teeter M Naudie D Howard J MacDonald S Lanting B
Full Access

Introduction

Coronal plane alignment is one of the contributing factors to polyethylene wear in total knee arthroplasty (TKA). The goal of this study was to evaluate the wear and damage patterns of retrieved tibial polyethylene inserts in relationship to the overall mechanical alignment and to the position of the tibial component.

Materials and methods

Based on full-length radiographs, ninety-five polyethylene inserts retrieved from primary TKA's with a minimum time in-vivo of five years were analysed for wear and damage. Four alignment groups were compared: valgus, neutral, mild varus and moderate varus. Varus and valgus positioning of the tibial component was analysed for damage score for the neutral and varus aligned groups.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 95 - 95
1 Jan 2016
Teeter M Parikh A Taylor M Vandekerckhove P Sprague J Naudie D
Full Access

We sought to determine what dimensional changes occurred from wear testing of a total knee implant, as well as whether any changes developed within the polyethylene subsurface. Three fixed bearing implants underwent wear simulator testing to 6.1 million cycles. Gravimetric analysis and micro-CT scans were performed pre-test, mid-test, and post-test. Wear volume and surface deviations were greater during 0–3.2 million cycles (91±12 mm3) than from 3.2–6.1 million cycles (52±18 mm3). Deviations (wear and creep) occurred across all surfaces of the tibial inserts, including the articular surface, backside surface, sides, and locking mechanism. No subsurface changes were found. The micro-CT results were a useful adjunct to gravimetric analysis, better defining the dimensional changes that occurred with testing and ruling out subsurface fatigue.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_1 | Pages 14 - 14
1 Jan 2016
Al-Khateeb H Teeter M Naudie D McCalden RW
Full Access

Introduction

The development of new bearing surfaces for total joint replacement is constantly evolving. Oxidized zirconium (Oxinium) has been introduced for use in total hip arthroplasty (THA) and total knee arthroplasty (TKA). One of the most common causes of failure of THA is aseptic loosening secondary to polyethylene wear debris. The aetiology of wear is multifactorial and includes adhesive, abrasive, third-body and fatigue wear mechanisms. Oxidized zirconium is a relatively new material that features an oxidized ceramic surface chemically bonded to a hard metallic substrate. This material possesses the reduced polyethylene wear characteristics of a ceramic, without the increased risk of implant fracture While short-term results of oxidized zirconium in THA have been reported, there have been no reports on retrieved highly cross linked PE articulating with Oxinium headsObjectives:

Objectives

The purpose of this study was to compare matched pairs of retrieved highly cross-linked polyethylene (XLPE) acetabular liners with OxZr and CoCr articulation. The liners were examined for evidence of wear damage, including articular surface damage, impingement, screw-hole creep, and rim cracks


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 554 - 554
1 Dec 2013
Teeter M Pang H McCalden RW Naudie D MacDonald S
Full Access

Background:

Varus-valgus constrained (VVC) implants are used in cases of severe valgus deformity, attenuated medial collateral ligament and difficulty in balancing the medial and lateral gaps of the knee. The increased constraint has been postulated to lead to more stress at the bone-implant interface and early loosening. The objective of this study was to compare the wear characteristics of the polyethylene liner in VVC prosthesis with the posterior-stabilized (PS) prosthesis and identify the factors leading to more wear in the VVC tibial inserts.

Methods:

This was a retrieval analysis of all VVC liners collected from patients who underwent revision surgery from 1999 to 2011. These patients were matched to another group with posterior-stabilized inserts who underwent revision in the same time period. These two groups of patients were similar in terms of their demographic data and implant dimensions. Inserts were divided into 16 zones and a microscopic analysis of surface damage was carried out. We determined overall damage with a scoring system. Pre-revisions radiographs were reviewed and analyzed for correlation with the wear profile.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 182 - 182
1 Dec 2013
Teeter M Pang H McCalden RW Naudie D MacDonald S
Full Access

Background:

Polyethylene wear in total knee arthroplasty (TKA) is influenced by patient, surgeon and implant factors. The objective of this study is to assess the effect of limb alignment, implant position and joint line position on the pattern of wear in posterior stabilized (PS) tibial inserts.

Methods:

This was a retrieval analysis of 83 PS liners collected from patients who underwent revision surgery from 1999 to 2011. Inserts were divided into 16 zones and a microscopic analysis of surface damage was carried out. We determined overall damage with a scoring system. Pre-revisions radiographs were reviewed and analyzed for correlation with the wear profile.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 555 - 555
1 Dec 2013
Teeter M Pang H Naudie D McCalden RW MacDonald S
Full Access

Purpose

The objective of this study was to compare the wear characteristics and damage scores in highly crosslinked (XLPE) and conventional polyethylene (CPE) acetabular liners.

Methods

This was a retrieval analysis of 13 XLPE liners obtained from patients who underwent revision surgery from 1999 to 2011. These patients were matched on patient demographics (age, BMI, side, sex, and length of implantation) and implant characteristics (inner diameter, outer diameter, and lip angle) to another group with CPE who underwent revision in the same time period. The only difference between implants was the use of XLPE. Wear analysis was performed with micro-computed tomography (micro-CT), provided thickness measurements across four quadrants of the bearing surface. Surface damage was scored and the pattern documented. The mean duration of implantation was 5.00 ± 3.36 years in the XLPE group and 5.19 ± 3.69 years in the CPE group (p = 0.12), with the longest duration exceeding 10 years.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 183 - 183
1 Dec 2013
Teeter M Parikh A Taylor M Sprague J Naudie D
Full Access

BACKGROUND:

Implant wear continues to be a limitation of total knee replacement (TKR). Wear simulator studies are a valuable screening tool in new implant development. The purpose of this study was to determine the ability of micro-CT to prospectively measure wear in TKR implants during a wear simulator trial.

METHODS:

Three identical cruciate-retaining, fixed bearing cobalt-chromium-molybdenum (CoCrMo) on conventional EtO-sterilized polyethylene TKA implants underwent wear simulator testing up to 3.2 million cycles using gait inputs; loaded-soaks were used to correct for fluid absorption. The implants were weighed and scanned with micro-CT (at 50 micron resolution) before and after testing. The gravimetric mass was converted to volume based on the density of polyethylene. Volume change due to wear was calculated from both the gravimetric and micro-CT methods. The pre- and post-wear test micro-CT geometries were co-registered and the deviations between the two were measured.