Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVIII | Pages 4 - 4
1 May 2012
Simpson D Kendrick B Hughes M Rushforth G Gill H Murray D
Full Access

Introduction

Primary mechanical stability is important with uncemented THR because early migration is reduced, leading to more rapid osseointegration between the implant and bone. Such primary mechanical stability is provided by the design features of the device. The aim of this study was to compare the migration patterns of two uncemented hip stems, the Furlong Active and the Furlong HAC stem; the study was designed as a randomised control trial. The implants were the Furlong HAC, which is an established implant with good long term results, and the Furlong Active, which is a modified version of the Furlong HAC designed to minimise stress concentrations between the implant and bone, and thus to improve fixation.

Materials and methods

The migration of 43 uncemented femoral components for total hip replacement was measured in a randomised control trial using Roentgen Stereophotogrammetric Analysis (RSA) over two years. Twenty-three Furlong HAC and twenty Furlong Active stems were implanted into 43 patients. RSA examinations were carried out post-operatively, and at six months, 12 months and 24 months post-operatively. The patients stood in-front of a purpose made calibration frame which contained accurately positioned radio-opaque markers. From the obtained images, the 3-D positions of the prosthesis and the host bone were reconstructed. Geometrical algorithms were used to identify the components of the implant. These algorithms allowed the femoral component to be studied without the need to attach markers to the prosthesis. The migration was calculated relative to the femoral coordinate system representing the anterior-posterior (A-P), medial-lateral (M-L) and proximal-distal (P-D) directions respectively. Distal migration was termed subsidence.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVIII | Pages 59 - 59
1 May 2012
Simpson DJ Kendrick B Thomas G Gill H O'Connor J Murray DW
Full Access

Introduction

The results of the original mobile bearing Oxford unicompartmental knee replacement (UKR) in the lateral compartment have been disappointing because of high dislocation rates (11%). This original implant used a flat bearing articulation on the tibial tray. To address the issue of dislocation a new implant (domed tibia with biconcave bearing to increase entrapment) was introduced with a modified surgical technique. The aim of this study was to compare the risk of dislocation between a domed and flat lateral UKR.

Methods

Separate geometric computer models of an Oxford mobile bearing lateral UKR were generated for the two types of articulation between the tibial component and the meniscal bearing: Flat-on-flat (flat) and Concave-on-convex (domed). Each type of mobile bearing was used to investigate three distinct dislocation modes observed clinically: lateral to medial dislocation, with the bearing resting on the tray wall (L-M-Wall); medial to lateral dislocation, out of the joint space (M-L); anterior to posterior dislocation, out of the joint space (A-P).

A size C tray and a medium femoral component and bearing were used in all models. The femoral component, tibial tray and bearing were first aligned in a neutral position. For each dislocation the tibial tray was restrained in all degrees of freedom. The femoral component was restrained from moving in the anterior-posterior directions and in the medial-lateral directions. The femoral component was also restrained from rotating about the anterior-posterior, medial-lateral and superior-inferior directions. This meant that the femoral component was only able to move in the superior-inferior direction. Different bearing sizes were inserted into the model and the effect that moving the femoral component medially and laterally had on the amount of distraction required to cause bearing dislocation was investigated.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVIII | Pages 47 - 47
1 May 2012
Bottomley N McNally E Jones L Javaid M Arden N Gill H Dodd C Murray D Beard D Price A
Full Access

Introduction

Anteromedial osteoarthritis of the knee (anteromedial gonarthrosis-AMG) is a common form of knee arthritis. In a clinical setting, knee arthritis has always been assessed by plain radiography in conjunction with pain and function assessments. Whilst this is useful for surgical decision making in bone on bone arthritis, plain radiography gives no insight to the earlier stages of disease. In a recent study 82% of patients with painful arthritis had only partial thickness joint space loss on plain radiography. These patients are managed with various surgical treatments; injection, arthroscopy, osteotomy and arthroplasty with varying results. We believe these varying results are in part due to these patients being at different stages of disease, which will respond differently to different treatments. However radiography cannot delineate these stages. We describe the Magnetic Resonance Imaging (MRI) findings of this partial thickness AMG as a way of understanding these earlier stages of the disease.

Method

46 subjects with symptomatic partial thickness AMG underwent MRI assessment with dedicated 3 Tesla sequences. All joint compartments were scored for both partial and full thickness cartilage lesions, osteophytes and bone marrow lesions (BML). Both menisci were assessed for extrusion and tear. Anterior cruciate ligament (ACL) integrity was also assessed. Osteophytes were graded on a four point scale in the intercondylar notch and the lateral margins of the joint compartments. Scoring was performed by a consultant radiologist and clinical research fellow using a validated MRI atlas with consensus reached for disagreements. The results were tabulated and relationships of the interval data assessed with linear by linear Chi2 test and Pearson's Correlation.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVIII | Pages 36 - 36
1 May 2012
Boyd J Zavatsky A Gill H
Full Access

Background

Osteoarthritis (OA) is a degenerative, chronic disease of the articular cartilage that affects more than 150 million people [1]. In the knee, OA can begin as either isolated medial OA or isolated lateral OA. Previous research [2,3] shows medial OA and lateral OA have characteristic cartilage lesion locations and progression patterns as well as flexion angles associated with lesion development, indicating strong involvement of mechanical factors in disease initiation. Therefore, it is important to investigate these mechanical factors. Previous studies combined data sets (geometry, motion, load) from separate sources. The aim of the current work was to use a consistent multi-modal approach.

Method

A finite element (FE) model of a healthy knee in full extension was created using magnetic resonance imaging (MRI) and motion analysis data from the same subject (female, 24 yrs). MRI data was obtained using a 3T MRI scanner (Philips Medical Systems/Achieva). Surface geometries of the tibia, femur, and associated cartilage were then semi-automatically segmented and processed (Mimics 12.5; Geomagic Studio 11; SolidWorks 2009). Motion data was collected at 100 Hz (Vicon 612) during level walking and subsequently applied to a lower limb model (AnyBody Version 3.0) to calculate muscle forces. Both sets of data were then combined to create a subject-specific FE model (ANSYS 11.0) which was solved to determine relative contact areas, pressures, and deformations in the medial and lateral tibiofemoral compartments.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVIII | Pages 12 - 12
1 May 2012
Simpson DJ Kendrick B Price AJ Murray D Gill H
Full Access

Introduction

Unicompartmental Knee Replacement (UKR) is an appealing alternative to Total Knee Replacement (TKR) when the patient has isolated compartment osteoarthritis (OA). A common observation post-operatively is radiolucency between the tibial tray wall and the bone. In addition, some patients complain of persistent pain following implantation with a UKR; this may be related to elevated bone strains in the tibia. The aim of this study was to investigate the mechanical environment of the tibia bone adjacent to the tray wall, following UKR, to determine whether this region of bone resorbs, and how altering the mechanical environment affects tibia strains.

Materials and methods

A finite element (FE) model of a cadaver tibia implanted with an Oxford UKR was used in this study, based on a validated model. A single static load, measured in-vivo during a step-up activity was used. There was a 1 mm layer of cement surrounding the keel in the cemented UKR, and the cement filled the cement pocket. In accordance with the operating procedure, no cement was used between the tray wall and bone. For the cementless UKR a layer of titanium filled the cement pocket. An intact tibia was used to compare to the cemented and cementless UKR implanted tibiae. The tibia was sectioned by the tray wall, defining the radiolucency zone (parallel to the vertical tray wall, 2 mm wide with a volume of 782.5 mm3), corresponding to the region on screened x-rays where radiolucencies are observed. Contact mechanics algorithms were used between all contacting surfaces; bonded contact was also introduced between the tray wall and adjacent bone, simulating a mechanical tie between them. Strain energy density (SED), was compared between the intact and implanted tibia for the radiolucency zone. Equivalent strains were compared on the proximal tibia between the intact and implanted tibia models. Forty patients (20 cemented, 20 cementless) who had undergone UKR were randomly selected from a database, and assessed for radiolucency.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVIII | Pages 24 - 24
1 May 2012
Bottomley N Javaid M Gill H Dodd C Murray D Beard D Price A
Full Access

Introduction

Anteromedial gonarthrosis is a common well described pattern of knee osteoarthritis with cartilage wear beginning in the anteromedial quadrant of the medial tibial plateau in the presence of an intact and functioning ACL. It is well known that mechanical factors such as limb alignment and meniscal integrity affect the progression of arthritis and there is some evidence that the morphology of the tibial plateau may be a risk factor in the development of this disease. The extension facet angle is the angle of the downslope of the anterior portion of the medial tibial plateau joint surface in relation to the middle portion on a sagittal view. If this is an important factor in the development of AMG there may be potential for disease modifying intervention.

This study investigates if there is a significant difference in this angle as measured on MRI between a study cohort with early AMG (partial thickness cartilage damage and intact ACL) and a comparator control cohort of patients (no cartilage damage and ACL rupture).

Methods

3 Tesla MRI scans of 99 patients; 54 with partial thickness cartilage damage and 44 comparitors with no cartilage damage (acute ACL rupture) were assessed. The extension facet angle was measured (Osirix v3.6) using a validated technique on two consecutive MRI T2 sagittal slices orientated at the mid-coronal point of the medial femoral condyle. (InterClass Correlation 0.95, IntraClass Correlation 0.97, within subject variation of 1.1° and coefficient of variation 10.7%). The mean of the two extension angle values was used. The results were tabulated and analysed (R v2.9.1).


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVIII | Pages 13 - 13
1 May 2012
Gray H Zavatsky A Gill H
Full Access

Iterative finite element (FE) models are used to simulate bone remodelling that takes place due to the surgical insertion of an implant or to simulate fracture healing. In such simulations element material properties are calculated after each iteration of solving the model. New material properties are calculated based on the results derived by the model during the last iteration. Once the FE model has gone through a number of such iterations it is often necessary to assess the remodelling that has taken place. The method widely used to do this is to analyse element Young's modulus plots taken at particular sections through the model. Although this method gives relevant information which is often helpful when comparing different implants, the information is rather abstract and is difficult to compare with patient data which is commonly in the form of radiographs.

The authors suggest a simple technique that can be used to generate synthetic radiograph images from FE models. These images allow relatively easy comparisons of FE derived information with patient radiographs. Another clear advantage of this technique is that clinicians (who are familiar with reading radiographs) are able to understand and interpret them readily.

To demonstrate the technique a three dimensional (3D) model of the proximal tibia implanted with an Oxford Unicompartmental Knee replacement was created based on CT data obtained from a cadaveric tibia. The model's initial element material properties were calculated from the same CT data set using a relationship between radiographic density and Young's modulus.

The model was subject to simplified loading conditions and solved over 365 iterations representing one year of in vivo remodelling. After each iteration the element material properties were recalculated based on previously published remodelling rules. Next, synthetic anteroposterior radiographs were generated by back calculating radiographic densities from material properties of the model after 365 iterations. A 3D rectangular grid of sampling points which encapsulated the model was defined. For each of the elements in the FE model radiographic densities were back calculated based on the same relationships used to calculate material properties from radiographic densities. The radiographic density of each element was assigned to all the sampling grid points within the element. The 3D array of radiographic densities was summed in the anteroposterior direction thereby creating a 2D array of radiographic densities. This 2D array was plotted giving an image analogous to anteroposterior patient radiographs. Similar to a patient radiograph denser material appeared lighter while less dense material appeared darker.

The resulting synthetic radiographs were compared to patient radiographs and found to have similar patterns of dark and light regions.

The synthetic radiographs were relatively easy to produce based on the FE model results, represented FE results in a manner easily comparable to patient radiographs, and represented FE results in a clinician friendly manner.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVIII | Pages 96 - 96
1 May 2012
Monk A McKenna D Simpson D Beard D Thomas N Gill H
Full Access

The heat produced by drills, saws and PMMA cement in the handling of bone can cause thermal necrosis. Thermal necrosis could be a factor in the formation of a fibrous tissue membrane and impaired bony ingrowth into porous prostheses. This has been proposed to lead to non-union of osteotomies and fractures, the failure of the bone-cement interface and the failure of resurfacing arthroplasty.

We compared three proprietary blades (the De Soutter, the Stryker Dual Cut and the Stryker Precision) in an in-vitro setting with porcine tibiae, using thermocouples embedded in the bone below the cutting surface to measure the increases in bone temperature.

There was a significant (p=0.001) difference in the change in temperature (δT) between the blade types. The mean increase in temperature was highest for the De Soutter, 2.84°C (SD: 1.83°C, range 0.48°C to 9.30°C); mean δT was 1.81°C (SD: 1.00°C, range 0.18°C to 4.85°C) for the Precision and 1.68°C (SD: 0.95°C, range 0.24°C to 5.67°C). Performing paired tests, there was no significant difference in δT between the Precision and Dual Cut blades (p=0.340), but both these blades had significantly (p=0.003 for Precision vs De Soutter, p<0.001 for Dual Cut vs De Soutter) lower values for δT than the Dual Cut.