Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_2 | Pages 32 - 32
2 Jan 2024
Depboylu F Yasa E Poyraz Ö Korkusuz F
Full Access

Decreasing the bulk weight without losing the biomechanical properties of commercial pure titanium (Cp-Ti) medical implants is now possible by using Laser Powder Bed Fusion (L-PBF) technology. Gyroid lattice structures that have precious mechanical and biological advantages because of similarity to trabecular bone. The aim of the study was to design and develop L-PBF process parameter optimization for manufacturing gyroid lattice Cp-Ti structures. The cleaning process was then optimized to remove the non-melted powder from the gyroid surface without mechanical loss. Gyroid cubic designs were created with various relative densities by nTopology. L-PBF process parameter optimization was progressed using with Cp-Ti (EOS TiCP Grade2) powder in the EOS M290 machine to achieve parts that have almost full dense and dimensional accuracy. The metallography method was made for density. Dimensional accuracy at gyroid wall thicknesses was investigated between designed and manufactured via stereomicroscope, also mechanical tests were applied with real time experiment and numerical analysis (ANSYS). Mass loss and strut thickness loss were investigated for chemical etching cleaning process. Gyroid parts had 99,5% density. High dimensional accuracy was achieved during L-PBF process parameters optimization. Final L-PBF parameters gave the highest 19% elongation and 427 MPa yield strength values at tensile test. Mechanical properties of gyroid were controlled with changing relative density. A minute chemical etching provided to remove non-melted powders. Compression test results of gyroids at numerical and real-time analysis gave unrelated while deformation behaviors were compatible with each other. Gyroid Cp-Ti osteosynthesis mini plates will be produced with final L-PBF process parameters. MTT cytotoxicity test will be characterized for cell viability. Acknowledgements This project is granted by TUBITAK (120N943). Feza Korkusuz MD is a member of the Turkish Academy of Sciences (TÜBA)


Aims. There are concerns regarding nail/medullary canal mismatch and initial stability after cephalomedullary nailing in unstable pertrochanteric fractures. This study aimed to investigate the effect of an additional anteroposterior blocking screw on fixation stability in unstable pertrochanteric fracture models with a nail/medullary canal mismatch after short cephalomedullary nail (CMN) fixation. Methods. Eight finite element models (FEMs), comprising four different femoral diameters, with and without blocking screws, were constructed, and unstable intertrochanteric fractures fixed with short CMNs were reproduced in all FEMs. Micromotions of distal shaft fragment related to proximal fragment, and stress concentrations at the nail construct were measured. Results. Micromotions in FEMs without a blocking screw significantly increased as nail/medullary canal mismatch increased, but were similar between FEMs with a blocking screw regardless of mismatch. Stress concentration at the nail construct was observed at the junction of the nail body and lag screw in all FEMs, and increased as nail/medullary canal mismatch increased, regardless of blocking screws. Mean stresses over regions of interest in FEMs with a blocking screw were much lower than regions of interest in those without. Mean stresses in FEMs with a blocking screw were lower than the yield strength, yet mean stresses in FEMs without blocking screws having 8 mm and 10 mm mismatch exceeded the yield strength. All mean stresses at distal locking screws were less than the yield strength. Conclusion. Using an additional anteroposterior blocking screw may be a simple and effective method to enhance fixation stability in unstable pertrochanteric fractures with a large nail/medullary canal mismatch due to osteoporosis. Cite this article: Bone Joint Res 2022;11(3):152–161


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_16 | Pages 61 - 61
1 Dec 2021
Naghavi SA Hua J Moazen M Taylor S Liu C
Full Access

Abstract. Objectives. Currently, total hip replacement surgery is an effective treatment for osteoarthritis, where the damaged hip joint is replaced with an artificial joint. Stress shielding is a mechanical phenomenon that refers to the reduction of bone density as a result of altered stresses acting on the host bone. Due to solid metallic nature and high stiffness of the current orthopaedic prostheses, surrounding bones undergo too much bone resorption secondary to stress shielding. With the use of 3D printing technology such as selective laser melting (SLM), it is now possible to produce porous graded microstructure hip stems to mimics the surrounding bone tissue properties. Method. In this study we have compared the physical and mechanical properties of two triply periodic minimal surface (TPMS) lattice structure namely gyroid and diamond TPMS. Based on initial investigations, it was decided to design, and 3D print the gyroid and diamond scaffolds having pore size of 800 and 1100 um respectively. Scaffold of each type of structure were manufactured and were tested mechanically in compression (n=8), tension (n=5) and bending (n=1). Results. Upon FEA validation of the scaffold in Abaqus, the desired scaffold for hip implant application was evaluated to have a young's modules of 12.15 GPa, yield strength of 242 MPa and porosity of 55%. Topology and lattice optimization were performed using nTopology to design an optimised graded porous hip implant based on stress shielding reduction. It was understood that the designed optimised hip implant can reduce the stress shielding effect by more than 65% when compared to the conventional generic implant. Conclusions. The designed hip implant presented in this work shows clinical promise in reducing bone loss while having enhanced osseointegration with the surrounding cortical bones. Hence, this will help reduce the risk of periprosthetic fracture and the probability of revision surgery


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_13 | Pages 118 - 118
1 Nov 2021
Pareatumbee P Yew A Koh JSB Howe TS Abidin SZ Tan MH
Full Access

Introduction and Objective. Curative resection of proximal humerus tumours is now possible in this era of limb salvage with endoprosthetic replacement considered as the preferred reconstructive option. However, it has also been linked with mechanical and non-mechanical failures such as stem fracture and aseptic loosening. One of the challenges is to ensure that implants will endure the mechanical strain under physiological loading conditions, especially crucial in long surviving patients. The objective is to investigate the effect of varying prosthesis length on the bone and implant stresses in a reconstructed humerus-prosthesis assembly after tumour resection using finite element (FE) modelling. Methods. Computed tomography (CT) scans of 10 humeri were processed in Mimics 17 to create three-dimensional (3D) cortical and cancellous solid bone models. Endoprostheses of different lengths manufactured by Stryker were modelled using Solidworks 2020. The FE models were divided into four groups namely group A consisting of the intact humerus and groups B, C and D composed of humerus-prosthesis assemblies with a body length of 40, 100 and 120 mm respectively and were meshed using linear 4-noded tetrahedral elements in 3matic 13. The models were then imported into Abaqus CAE 6.14. Isotropic linear elastic behaviour with an elastic modulus of 13400, 2000 and 208 000 MPa were assigned to the cortical bone, cancellous bone and prosthesis respectively and a Poisson's ratio of 0.3 was assumed for each material. To represent the lifting of heavy objects and twisting motion, a tensile load of 200 N for axial loading and a 5 Nm torsional load for torsional loading was applied separately to the elbow joint surface with the glenohumeral joint fixed and with all contact interfaces defined as fully bonded. A comparative analysis against literature was performed to validate the intact model. Statistical analysis of the peak von Mises stress values collected from predicted stress contour plots was performed using a one-way repeated measure of analysis of variance (with a Bonferroni post hoc test) using SPSS Statistics 26. The average change in stress of the resected models from the intact state were then determined. Results. The validation of the intact humerus displayed a good agreement with literature values. The peak bone stress occurred distally above the coronoid and olecranon fossa closer to the load application region in the intact and resected bone models with a significant amount of loading borne by the cortical bone, while the peak implant stress occurred at the bone-prosthesis contact interface under both loading conditions. Based on the results obtained, a statistically significant difference (p =.013) in implant stress was only seen to occur between groups B and C under tension. Results illustrate initiation of stress shielding with the bone bearing lesser stress with increasing resection length which may eventually lead to implant failure by causing bone resorption according to Wolff's law. The peak implant stress under torsion was 3–5 times the stress under tension. The best biomechanical behaviour was exhibited in Group D, having the least average change in stress from the intact model, 5% and 3.8% under tension and torsion respectively. It can be deduced that the shorter the prosthesis length, the more pronounced the effect on cortical bone remodelling. With the maximum bone and implant stresses obtained being less than their yield strength, it can be concluded that the bone-implant construct is safe from failure. Conclusions. The developed FE models verified the influence of varying the prosthesis length on the bone and implant stresses and predicted signs of stress shielding in longer endoprostheses. By allowing for 2 cm shortening in the upper extremity and post-surgical scarring, it is beneficial to err towards a shorter endoprosthesis


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_13 | Pages 11 - 11
1 Nov 2021
Imwinkelried T Heuberger R Eggli S
Full Access

Introduction and Objective. Local cartilage defects in the knee are painful and mostly followed by arthritis. In order to avoid impaired mobility, the osteochondral defect might be bridged by a synthetic compound material: An osteoconductive titanium foam as an anchoring material in the subchondral bone and an infiltrated polymer as gliding material in contact with the surrounding natural cartilage. Materials and Methods. Titanium foam cylinders (Ø38 mm) with porosities ranging from 57% to 77% were produced by powder metallurgy with two different grain sizes of the space holder (fine: 340 ± 110 μm, coarse: 530 ± 160 μm). The sintered titanium foam cylinders were infiltrated with UHMWPE powder on one end and UHMWPE bulk at the other end, at two different temperatures (160 °C, 200 °C), using a pressure of 20 MPa for 15 minutes. Smaller cylinders (Ø16 mm) were retrieved from the compound material by water jet cutting. The infiltration depths were determined by optical microscopy. The anchoring of the UHMWPE was measured by a shear test and the mechanical properties of the titanium foam were verified by a subsequent compression test. The tribological behaviour was investigated in protein containing liquid using fresh cartilage pins (Ø5 mm) sliding against a UHMWPE disc with or without a notch to simulate the gap between the implant and the surrounding cartilage. Friction coefficients were determined in a rotation tribometer and the cartilage wear in a multidirectional six-station tribometer from AMTI (load 10 – 50 N, sliding speed 20 mm/s, 37 °C). Results. UHMWPE could be infiltrated into titanium foam by 1.1 – 1.3 mm with fine pores and by 1.5 – 1.8 mm with coarse pores. The infiltration was neither dependent on the type of UHMWPE (powder or bulk) nor on the temperature. The polymer was so well anchored inside the titanium foam pores that the shear forces for the compounds exceeded the shear strength obtained for a UHMWPE-cylinder. This effect was due to the increased stiffness of the compound plug. Uniaxial compression of the titanium foams after the shear-off of the polymer revealed yield strengths ranging from 50 – 88 MPa for porosities of 62 – 73%. The Ø16 mm samples yielded beyond physiological loads in the knee (≥ 10x body weight) and behaved in a strain hardening and fully ductile manner, reaching deformations of at least 50 % of their initial height without the appearance of macroscopically visible cracks. For smaller plug diameters down to Ø8 mm, however, the lower porosity / higher strength foam should be used to limit elastic deformation of the compound to < 0.1 mm. Pore size did not significantly influence the strength and stiffness values. The elevated coefficient of friction between cartilage and UHMWPE of about 1 was not negatively affected by the presence of the gap. The height loss of the cartilage pin after 1 hour (respectively after 3600 reciproque wear cycles) was 0.2 ± 0.1 mm using a flat disc. For discs with a 1 mm wide V-notch, the wear increased to 0.9 ± 0.3 mm. Conclusions. The tested titanium foams are well suited to act as an anchoring material in the subchondral bone as mechanical properties can be tailored by choosing the adequate porosity and as bone ingrowth has previously been demonstrated for the used pore sizes. UHMWPE is not an ideal gliding partner against cartilage because the friction coefficients of frictions were high. The presence of a V-notched gap was detrimental for cartilage wear. More hydrophilic polymers like PCU should be tested as potential gliding materials


The Bone & Joint Journal
Vol. 103-B, Issue 5 | Pages 939 - 945
1 May 2021
Kakar S Logli AL Ramazanian T Gaston RG Fowler JR

Aims

The purpose was to evaluate early clinical, patient-reported, and radiological outcomes of the scapholunate ligament 360° tenodesis (SL 360) technique for treatment of scapholunate (SL) instability.

Methods

We studied the results of nine patients (eight males and one female with a mean age of 44.7 years (26 to 55)) who underwent the SL 360 procedure for reducible SL instability between January 2016 and June 2019, and who were identified from retrospective review of electronic medical records. Final follow-up of any kind was a mean of 33.7 months (12.0 to 51.3). Clinical, radiological, and patient-reported outcome data included visual analogue scale (VAS) for pain, Quick Disabilities of the Arm, Shoulder and Hand (QuickDASH), Mayo Wrist Score (MWS), and Patient-Rated Wrist Examination (PRWE). Means were analyzed using paired t-test.


Bone & Joint Research
Vol. 10, Issue 4 | Pages 250 - 258
1 Apr 2021
Kwak D Bang S Lee S Park J Yoo J

Aims

There are concerns regarding initial stability and early periprosthetic fractures in cementless hip arthroplasty using short stems. This study aimed to investigate stress on the cortical bone around the stem and micromotions between the stem and cortical bone according to femoral stem length and positioning.

Methods

In total, 12 femoral finite element models (FEMs) were constructed and tested in walking and stair-climbing. Femoral stems of three different lengths and two different positions were simulated, assuming press-fit fixation within each FEM. Stress on the cortical bone and micromotions between the stem and bone were measured in each condition.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_4 | Pages 110 - 110
1 Mar 2021
Pavanram P Li Y Zhou J Kubo Y Lietaert K Leeflang M Fockaert L Pouran B Mol J Weinans H Zadpoor A Jahr H
Full Access

As compared to magnesium (Mg) and iron (Fe), solid zinc (Zn)-based absorbable implants show better degradation rates. An ideal bone substitute should provide sufficient mechanical support, but pure Zn itself is not strong enough for load-bearing medical applications. Modern processing techniques, like additive manufacturing (AM), can improve mechanical strength of Zn. To better mimic the in vivo situation in the human body, we evaluated the degradation behavior of porous Zn implants in vitro under dynamic conditions. Our study applied selective laser melting (SLM) to build topographically ordered absorbable Zn implants with superior mechanical properties. Specimens were fabricated from pure Zn powder using SLM and diamond unit cell topological design. In vitro degradation was performed under both static and dynamic conditions in a custom-built set-up under cell culture conditions (37 °C, 20% O2 and 5% CO2) for up to 28 days. Mechanical properties of the porous structures were determined according to ISO 13314: 2011 at different immersion time points. Modified ISO 10993 standards were used to evaluate biocompatibility through direct cell seeding and indirect extract-based cytotoxicity tests (MTS assay, Promega) against identically designed porous titanium (Ti-6Al-4V) specimens as reference material. Twenty-four hours after cell seeding, its efficacy was evaluated by Live-Dead staining (Abcam) and further analyzed using dual channel fluorescent optical imaging (FOI) and subsequent flow cytometric quantification. Porous Zn implants were successfully produced by means of SLM with a yield strength and Young's modulus in the range of 3.9–9.6 MPa and 265–570 MPa, respectively. Dynamic flow significantly increased the degradation rate of AM porous Zn after 28 days. Results from Zn extracts were similar to Ti-6Al-4V with >95% of cellular activity at all tested time points, confirming level 0 cytotoxicity (i.e., This study clearly shows the great potential of AM porous Zn as a bone substituting material. Moreover, we demonstrate that complex topological design permits control of mechanical properties and degradation behavior


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_2 | Pages 60 - 60
1 Mar 2021
Munford M Ng G Jeffers J
Full Access

Abstract. Objectives. This study aids the control of remodelling and strain response in bone; providing a quantified map of apparent modulus and strength in the proximal tibia in 3 anatomically relevant directions in terms of apparent density and factor groups. Methods. 7 fresh-frozen cadaveric specimens were quantified computed tomography (qCT) scanned, segmented and packed with 3 layers of 9mm side length cubic cores aligned to anatomical mechanical axes. Cores were removed with printed custom cutting and their densities found from qCT. Cores (n = 195) were quasi-statically compression tested. Modulus was estimated from a load cycle hysteresis loop, between 40% and 20% of yield stress. Sequential testing order in 3 orthogonal directions was randomised. Group differences were identified via an analysis of variance for the factors density, age, gender, testing order, subchondral depth, condyle and sub-meniscal location. Regression models were fit for significant factor sub-groups, predicting properties from density. Results. Axial modulus was 1.5 times greater than the two transverse directions (p<0.001), between which no difference was found. For all test directions, differences were quantified for density and modulus across all subchondral depths (p<0.001). 60% of transverse modulus variation was explained by density within subgroups for each subchondral depth. Medial axial modulus was 1.3 times greater than the lateral side (p = 0.011). Lateral axial modulus halved over a 25mm depth whilst remaining constant for the medial side. Density explained 75% of variation when grouped by subchondral depth and condyle. Yield strength was well predicted across all test directions, with density explaining 81% of axial strength variation and no differences over subchondral depth. Conclusions. The quantification of bone multiaxial modulus based on condyle and subchondral depth has been shown for the first time in a clinically viable protocol using conventional CT. Accounting for spatial variation improves upon literature property prediction models. Declaration of Interest. (b) declare that there is no conflict of interest that could be perceived as prejudicing the impartiality of the research reported:I declare that there is no conflict of interest that could be perceived as prejudicing the impartiality of the research project


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_1 | Pages 36 - 36
1 Feb 2021
Spece H Kurtz S Yu T Marcolongo M Law A
Full Access

Introduction. The ability to create patient-specific implants (PSI) at the point-of-care has become a desire for clinicians wanting to provide affordable and customized treatment. While some hospitals have already adopted extrusion-based 3D printing (fused filament fabrication; FFF) for creating non-implantable instruments, recent innovations have allowed for the printing of high-temperature implantable polymers including polyetheretherketone (PEEK). With interest in FFF PEEK implants growing, it is important to identify methods for printing favorable implant characteristics such as porosity for osseointegration. In this study, we assess the effect of porous geometry on the cell response and mechanical properties for FFF-printed porous PEEK. We also demonstrate the ability to design and print customized porous implants, specifically for a sheep tibial segmental defect model, based on CT images and using the geometry of triply periodic minimal surfaces (TPMS). Methods. Three porous constructs – a rectilinear pattern and gyroid/diamond TPMSs – were designed to mimic trabecular bone morphology and manufactured via PEEK FFF. TPMSs were designed by altering their respective equation approximations to achieve desired porous characteristics, and the meshes were solidified and shaped using a CAD workflow. Printed samples were mCT scanned to determine the resulting pore size and porosity, then seeded with pre-osteoblast cells for 7 and 14 days. Cell proliferation and alkaline phosphatase activity (ALP) were evaluated, and the samples were imaged via SEM. The structures were tested in compression, and stiffness and yield strength values were determined from resulting stress-strain plots. Roughness was determined using optical profilometry. Finally, our process of porous structure design/creation was modified to establish a proof-of-concept workflow for creating PSIs using geometry established from segmented sheep tibia CT images. Results. ALP activity measurements of the porous PEEK samples at 7 and 14 days were significantly greater than for solid controls (p < 0.001 for all three designs, 14 days). No difference between the porous geometries was found. SEM imaging revealed cells with flat, elongated morphology attached to the surface of the PEEK and into the pore openings, with filopodia and lamellipodia extensions apparent. mCT imaging showed average pore size to be 545 ± 43 µm (porosity 70%), 708 ± 64 µm (porosity 68%), and 596 ± 94 µm (porosity 69%) for the rectilinear, gyroid, and diamond structures, respectively. The average error between the theoretical and actual values was −16.3 µm (pore size) and −3.3 % (porosity). Compression testing revealed elastic moduli ranging from 210 to 268 MPa for the porous samples. Yield strengths were 6.6 ± 1.2 MPa for lattice, 14.8 ± 0.7 MPa for gyroid, and 17.1 ± 0.6 for diamond. Average roughness ranged from 0.8 to 3 µm. Finally, we demonstrated the ability to design and print a fully porous implant with the geometry of a sheep tibia segment. Assessments of implant geometrical accuracy and mechanical performance are ongoing. Discussion. We created porous PEEK with TPMS geometries via FFF and demonstrated a positive cellular response and mechanical characteristics similar to trabecular bone. Our work offers an innovative approach for advancing point-of-care 3D printing and PSI creation


Bone & Joint 360
Vol. 9, Issue 6 | Pages 18 - 21
1 Dec 2020


Bone & Joint Research
Vol. 9, Issue 10 | Pages 645 - 652
5 Oct 2020
Chao C Chen Y Lin J

Aims

To determine whether half-threaded screw holes in a new titanium locking plate design can substantially decrease the notch effects of the threads and increase the plate fatigue life.

Methods

Three types (I to III) of titanium locking plates were fabricated to simulate plates used in the femur, tibia, and forearm. Two copies of each were fabricated using full- and half-threaded screw holes (called A and B, respectively). The mechanical strengths of the plates were evaluated according to the American Society for Testing and Materials (ASTM) F382-14, and the screw stability was assessed by measuring the screw removal torque and bending strength.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_6 | Pages 84 - 84
1 Jul 2020
Chow D Qin L Wang J Yang K Wan P
Full Access

Patellar fractures account for approximately 1% of all fractures. Open reduction and internal fixation is recommended to restore extensor continuity and articular congruity. However, complications such as nonunion and symptomatic hardware, still exist. Furthermore, there is a risk of re-fracturing of the healed bone during the removal of the implants. Magnesium (Mg), a biodegradable metal, has elastic moduli and compressive yield strength that are comparable to those of natural bone. Our previous study showed that released Mg ions enhanced fracture healing. However, Mg-based implants degrade rapidly after implantation and lead to insufficient mechanical strength to support the fracture. Microarc oxidation (MAO) is a metal surface coating that reduces corrosion. We hypothesized that Mg pins, with or without MAO, would enhance fracture healing radiologically, mechanically, and histologically, while MAO would decrease degradation of Mg pins. Patellar fracture was performed on forty-eight 18-week-old female New Zealand White rabbits according to established protocol. Briefly, the patella is osteotomized transversely and a tunnel (1.1mm) was drilled longitudinally through the two bone fragments. A pin (1 mm, stainless steel, Mg, or MAO-Mg) was inserted into the tunnel. The reduced construct was stabilized with a figure-of-eight band wire (⊘ 0.6 mm stainless steel wire). Cast immobilization was applied for 6 weeks. The rabbits were euthanized at week 8 and 12 post-operation. Microarchitecture and mechanical properties of the repaired patella were analyzed with microCT and tensile testing respectively. Histological sections of the repaired patella were stained. To evaluate the effect of the MAO treatment on degradation rate of Mg pin, the volume of the Mg pins in the patella was measured with microCT. At week 8, both Mg and Mg-MAO showed higher ratio of bone volume to tissue volume (BV/TV) than the control while there was no significant different between Mg and Mg-MAO. At week 12, Control, Mg, and Mg-MAO groups showed enlarged patella when compared to the normal patella. Tissue volume (TV) and bone volume (BV) of the patella in Mg and Mg-MAO were larger than those in the Control group. However, the Control had higher ratio of bone volume to tissue volume (BV/TV), TV density, and BV density than Mg and Mg-MAO. Tensile testing showed that the mechanical properties of the repaired patella (failure load, stiffness, ultimate strength, and energy-to-failure) of Mg and Mg-MAO were higher than that of the control at both week 8 and week 12. Histological analysis showed that there was significant new bone formation in the Mg and Mg-MAO group compared with the Control group at week 8 and 12. The degradation rate of the MAO-coated Mg pins was significantly slower than those without MAO at week 8 but no significant difference was detected at week 12. Mechanical, microarchitectural, and histological assessments showed that Mg pins, with or without MAO, enhanced fracture healing of the repaired patella compared to the Control. MAO treatment enhanced the corrosion resistance of the Mg pins at the early time point


Bone & Joint Research
Vol. 9, Issue 7 | Pages 386 - 393
1 Jul 2020
Doyle R van Arkel RJ Muirhead-Allwood S Jeffers JRT

Aims

Cementless acetabular components rely on press-fit fixation for initial stability. In certain cases, initial stability is more difficult to obtain (such as during revision). No current study evaluates how a surgeon’s impaction technique (mallet mass, mallet velocity, and number of strikes) may affect component fixation. This study seeks to answer the following research questions: 1) how does impaction technique affect a) bone strain generation and deterioration (and hence implant stability) and b) seating in different density bones?; and 2) can an impaction technique be recommended to minimize risk of implant loosening while ensuring seating of the acetabular component?

Methods

A custom drop tower was used to simulate surgical strikes seating acetabular components into synthetic bone. Strike velocity and drop mass were varied. Synthetic bone strain was measured using strain gauges and stability was assessed via push-out tests. Polar gap was measured using optical trackers.


Introduction. The ability to manufacture implants at the point-of-care has become a desire for clinicians wanting to provide efficient patient-specific treatment. While some hospitals have adopted extrusion-based 3D printing (fused filament fabrication; FFF) for creating non-implantable instruments with low-temperature plastics, recent innovations have allowed for the printing of high-temperature polymers such as polyetheretherketone (PEEK). Due to its low modulus of elasticity, high yield strength, and radiolucency, PEEK is an attractive biomaterial for implantable devices. Though concerns exist regarding PEEK for orthopaedic implants due to its bioinertness, the creation of porous networks has shown promising results for bone ingrowth. In this study, we endeavor to manufacture porous PEEK constructs via clinically-used FFF. We assess the effect of porous geometry on cell response and hypothesize that porous PEEK will exhibit greater preosteoblast viability and activity compared to solid PEEK. The work represents an innovative approach to advancing point-of-care 3D printing, cementless fixation for total joint arthroplasty, and additional applications typically reserved for porous metal. Methods. Three porous constructs – a rectilinear pattern and two triply period minimal surface (TPMSs) - were designed to mimic the morphology of trabecular bone. The structures, along with solid PEEK samples for use as a control, were manufactured via FFF using PEEK. The samples were mCT scanned to determine the resulting pore size and porosity. The PEEK constructs were then seeded with pre-osteoblast cells for 7 and 14 days. Cell proliferation and alkaline phosphatase activity (ALP) were evaluated at each time point, and the samples were imaged via SEM. Results. mCT imaging showed the pores in the PEEK constructs to be open and interconnected. The average pore size was 535 ± 92 µm for the rectilinear, 484 ± 237 µm for the diamond, and 669 ± 216 µm for the gyroid. Porosity was 71% for the rectilinear, 76% for the diamond, and 68% for the gyroid. The average error between the theoretical and actual values was −37.3 µm for pore size and −2.3 % for porosity. Normalized ALP activity of the three porous PEEK samples at 7 days were found to be significantly greater than the solid sample (p < 0.05 rectilinear, p < 0.005 gyroid, p < 0.001 diamond). At 14 days, the same relationships were observed (p < 0.001 for all three designs). No difference between the three geometries was found. SEM imaging revealed cells with flat, elongated morphology attached to the surface of the PEEK. The 14-day samples appeared to have proliferated well and spread along the PEEK pores. Extensions of filopodia and lamellipodia were observed along with large blankets of cells covering the PEEK surface. Discussion. We demonstrated the ability of FFF printed porous PEEK surfaces to promote cellular processes necessary for bone-implant fixation. While all porous structures showed promising results, more investigation into their material characteristics and osteogenic potential are necessary to determine which geometry may be suitable for orthopaedic use. Our work offers an innovative approach to advancing point-of-care 3D printing, cementless implant fixation, and additional applications typically reserved for porous metal


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 90 - 90
1 Feb 2020
Haeussler K Pandorf T
Full Access

Introduction. Lipped liners have the potential to decrease the rate of revision for instability after total hip replacement since they increase the jumping distance in the direction of the lip. However, the elevated lip also may reduce the Range of Motion and may lead to early impingement of the femoral stem on the liner. It is unclear whether the use of a lipped liner has an impact on the level of lever-out moments or the contact stresses. Therefore, the aim of the current study was to calculate these values for lipped liners and compare these results to a conventional liner geometry. Materials and Methods. 3D Finite Element studies were conducted comparing a ceramic lipped liner prototype and a ceramic conventional liner both made from BIOLOX. ®. delta. The bearing diameter was 36 mm. To apply loading, a test taper made of titanium alloy was bonded to a femoral head, also made from BIOLOX. ®. delta. Titanium was modeled with a bilinear isotropic hardening law. For the bearing contact a coefficient of friction of both 0.09 or 0.3 was assumed to model a well and poorly lubricated system. Frictionless contact was modeled between taper and liner. Pre-load was varied between 500 N and 1500 N and applied along the taper axis. While keeping pre-load constant, lever-out force was applied perpendicular to the taper axis until subluxation occurred. Liners were fixed at the taper region. Lever-out moment, equivalent plastic strain and von Mises stress of the taper, bearing contact area and contact area between taper and liner was evaluated. Results. With increasing pre-load, larger lever-out moment, equivalent plastic strain, contact area between taper and liner and bearing contact area was found for both liner designs. However, von Mises stresses were nearly constant but slightly exceeded yield strength of titanium. For all evaluated parameters almost no differences were found between the liner designs. Lever-out moments were comparable for both designs ranging from 4.5–10.5 Nm for the lipped liner and 4.4–10.2 Nm for the conventional liner. The increase of the coefficient of friction strongly affected lever-out moments, equivalent plastic strain and contact area between taper and liner. The other parameters were not affected by varying the coefficient of friction. Discussion. This study did not find significant differences in the lever-out behavior of the lipped acetabular liner compared to the conventional liner design. The inner geometry of the lipped liner is comparable to the conventional liner inner geometry. Therefore, contact area showed no significant differences and contact mechanics are identical in the current setup leading to similar results of both liner designs. For both designs small plastic deformations in the contact point of the taper were found at the contact region between liner and taper. However, the investigated mechanical parameters did not differ between the two investigated liner types. For any figures or tables, please contact authors directly


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 80 - 80
1 Feb 2020
Robotti P Luchin V Galeotti F Molinari A
Full Access

AM Open Cell porous Ti Structures were investigated for compressive strength, morphology (i.e. pore size, struts size and porosity), and wear resistance with the aim to improve design capability at support of implant manufacturing. Specimens were manufactured in Ti6Al4V using a SLM machine. Struts sizes had nominal diameters of 200µm or 100µm, pores had nominal diameters of 700µm, 1000µm or 1500µm. These dimensions were applied to three different open-cell geometrical configurations: one with unit-cells based on a regular cubic arrangement (Regular), one with a deformed cubic arrangement (Irregular), and one based on a fully random arrangement (Fully Random). Morphological analysis was performed by image analysis applied onto optical and SEM acquired pictures. The analyses estimated the maximum and minimum Feret pores diameter, and the latter was used as one of the key parameters to describe the interconnected network of pores intended for bone colonization. Outcome revealed the systematic oversizing of the actual struts diameter Vs designed diameter; by opposite min. Feret diameters of the pores resulted significantly smaller than nominal pore diameters, thus better fitting within the range of pores dimension acknowledged to favor the osseointegration. Consequently, the actual total porosity is also reduced. Many technologic factors are responsible for the morphologic differences design vs actual, among these the influence of melting pool dimension, the struts orientation during building and the layer thickness have a significant impact. Mechanical compression was performed on porous cylinder samples. Test revealed the Yield Strength and Stiffness are highly sensitive to the actual porosity. Deformation behavior follows densification phenomenon at lower porosity, whereas at higher porosity the Gibson-Ashby model fits for most of the structure tested. The relationship among load direction, struts alignment and the collapse behavior of the unit cell geometries are discussed. Stiffness of the porous structure is evaluated in both quasistatic and cyclic compression. Wear was investigated according to Taber test method. The abrasion resistance is measured by scratching a ceramic wheel against the different AM porous structures along a circular path. Metal debris eventually loss were quantified by gravimetric analysis at different number of cycles. Correlation among AM porous structure geometry, porosity and wear loss is discussed. All the tested structures showed a debris loss within the limit suggested by FDA for the porous coating in contact with the bone tissue. The actual AM porous Titanium unit cell geometry and features are a key design input. In combination with all the other design factors of a device they may result helpful in address the stress shielding and prevent metal debris release issues. The study underlines the importance of the research activity in AM to support Design for Additive Manufacturing (DFAM) capability. For any figures or tables, please contact authors directly


Bone & Joint Research
Vol. 7, Issue 12 | Pages 639 - 649
1 Dec 2018
MacLeod AR Serrancoli G Fregly BJ Toms AD Gill HS

Objectives

Opening wedge high tibial osteotomy (HTO) is an established surgical procedure for the treatment of early-stage knee arthritis. Other than infection, the majority of complications are related to mechanical factors – in particular, stimulation of healing at the osteotomy site. This study used finite element (FE) analysis to investigate the effect of plate design and bridging span on interfragmentary movement (IFM) and the influence of fracture healing on plate stress and potential failure.

Materials and Methods

A 10° opening wedge HTO was created in a composite tibia. Imaging and strain gauge data were used to create and validate FE models. Models of an intact tibia and a tibia implanted with a custom HTO plate using two different bridging spans were validated against experimental data. Physiological muscle forces and different stages of osteotomy gap healing simulating up to six weeks postoperatively were then incorporated. Predictions of plate stress and IFM for the custom plate were compared against predictions for an industry standard plate (TomoFix).


Bone & Joint Research
Vol. 7, Issue 12 | Pages 629 - 635
1 Dec 2018
Hung L Chao C Huang J Lin J

Objectives. Screw plugs have been reported to increase the fatigue strength of stainless steel locking plates. The objective of this study was to examine and compare this effect between stainless steel and titanium locking plates. Methods. Custom-designed locking plates with identical structures were fabricated from stainless steel and a titanium alloy. Three types of plates were compared: type I unplugged plates; type II plugged plates with a 4 Nm torque; and type III plugged plates with a 12 Nm torque. The stiffness, yield strength, and fatigue strength of the plates were investigated through a four-point bending test. Failure analyses were performed subsequently. Results. For stainless steel, type II and type III plates had significantly higher fatigue strength than type I plates. For titanium, there were no significant differences between the fatigue strengths of the three types of plates. Failure analyses showed local plastic deformations at the threads of screw plugs in type II and type III stainless steel plates but not in titanium plates. Conclusion. The screw plugs could increase the fatigue strength of stainless steel plates but not of titanium plates. Therefore, leaving screw holes open around fracture sites is recommended in titanium plates. Cite this article: L-W. Hung, C-K. Chao, J-R. Huang, J. Lin. Screw head plugs increase the fatigue strength of stainless steel, but not of titanium, locking plates. Bone Joint Res 2018;7:629–635. DOI: 10.1302/2046-3758.712.BJR-2018-0083.R1


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_13 | Pages 8 - 8
1 Oct 2018
Du JY Flanagan CD Bensusan JS Knusel KD Akkus O Rimnac CM
Full Access

Background. Structural bone allografts are an established treatment method for long-bone structural defects arising from such conditions as trauma, sarcoma, and osteolysis following total joint replacement. However, the quality of structural bone allografts is difficult to non-destructively assess prior to use. The functional lifetime of structural allografts depend on their ability to resist cyclic loading, which can lead to fracture even at stress levels well below the yield strength. Because allograft bone has limited capacity for remodeling, optimizing allograft selection for bone quality could decrease long-term fracture risk. Raman spectroscopy biomarkers can non-destructively assess the three primary components of bone (collagen, mineral, and water), and may predict the resistance of donor bone allografts to fracture from cyclic loads. The purpose of this study was to prospectively assess the ability of Raman biomarkers to predict number of cycles to fracture (“cyclic fatigue life”) of human allograft cortical bone. Methods. Twenty-one cortical bone specimens were from the mid-diaphysis of human donor bone tissue (bilateral femurs from 4 donors: 63M, 61M, 51F, 48F) obtained from the Musculoskeletal Transplant Foundation. Six Raman biomarkers were analyzed: collagen disorganization, type B carbonate substitution (a surrogate for mineral maturation), matrix mineralization, and 3 water compartments. Specimens underwent cyclic fatigue testing under fully reversed conditions at 35 and 45MPa (physiologically relevant stress levels for structural allografts). Specimens were tested to fracture or to 30 million cycles (“run-out”), simulating 15 years of moderate activity (i.e., 6000 steps per day). Multivariate regression analysis was performed using a tobit model (censored linear regression) for prediction of cyclic fatigue life. Specimens were right-censored at 30 million cycles. Results. All of the 6 biomarkers that were evaluated were independently associated with cyclic fatigue life (p < 0.05). The multivariate model explained 70% of the variance in cyclic fatigue life (R2=0.695, p<0.001,). Increasing disordered collagen (p<0.001) and loosely collagen-bound water compartments (p<0.001) were associated with decreased cyclic fatigue life. Increasing type B carbonate substitution (p<0.001), matrix mineralization (p<0.001), tightly collagen-bound water (p<0.001), and mineral-bound water (p=0.002) were associated with increased cyclic fatigue life. In the predictive model, 42% of variance in cyclic fatigue life was attributable to degree of collagen disorder, all bound water compartments accounted for 6%, and age and sex accounted for 17%. Conclusions. Raman biomarkers of three bone components (collagen, mineral, and water) predict cyclic fatigue life of human cortical bone. Increased baseline collagen disorder was associated with decreased cyclic fatigue life, and was the strongest determinant of cyclic fatigue life. Increased matrix mineralization and mineral maturation were associated with increased cyclic fatigue life. Bound-water compartments of bone contributed minimally to cyclic fatigue life. These results are complementary with prior Raman studies of monotonic testing of bone that reported decreased toughness and strength with increased collagen disorder and increased stiffness with increased bone mineralization and mineral maturation. This model should be prospectively validated. Raman analysis is a promising tool for the non-destructive evaluation of structural bone allograft quality and may be useful as a screening tool for selection of allograft bone. Acknowledgements. Supported by a grant from the Musculoskeletal Transplant Foundation. The Dudley P. Allen Fellowship (JYD), Wilbert J. Austin Professor of Engineering Chair (CMR) and the Leonard Case Jr. Professor of Engineering Chair (OA) are gratefully acknowledged