Bone & Joint Open
Vol. 5, Issue 10 | Pages 818 - 824
2 Oct 2024
Moroder P Herbst E Pawelke J Lappen S Schulz E

Aims

The liner design is a key determinant of the constraint of a reverse total shoulder arthroplasty (rTSA). The aim of this study was to compare the degree of constraint of rTSA liners between different implant systems.

Methods

An implant company’s independent 3D shoulder arthroplasty planning software (mediCAD 3D shoulder v. 7.0, module v. 2.1.84.173.43) was used to determine the jump height of standard and constrained liners of different sizes (radius of curvature) of all available companies. The obtained parameters were used to calculate the stability ratio (degree of constraint) and angle of coverage (degree of glenosphere coverage by liner) of the different systems. Measurements were independently performed by two raters, and intraclass correlation coefficients were calculated to perform a reliability analysis. Additionally, measurements were compared with parameters provided by the companies themselves, when available, to ensure validity of the software-derived measurements.


The Bone & Joint Journal
Vol. 105-B, Issue 12 | Pages 1314 - 1320
1 Dec 2023
Broida SE Sullivan MH Barlow JD Morrey M Scorianz M Wagner ER Sanchez-Sotelo J Rose PS Houdek MT

Aims

The scapula is a rare site for a primary bone tumour. Only a small number of series have studied patient outcomes after treatment. Previous studies have shown a high rate of recurrence, with functional outcomes determined by the preservation of the glenohumeral joint and deltoid. The purpose of the current study was to report the outcome of patients who had undergone tumour resection that included the scapula.

Methods

We reviewed 61 patients (37 male, 24 female; mean age 42 years (SD 19)) who had undergone resection of the scapula. The most common resection was type 2 (n = 34) according to the Tikhoff-Linberg classification, or type S1A (n = 35) on the Enneking classification.


The Bone & Joint Journal
Vol. 105-B, Issue 11 | Pages 1189 - 1195
1 Nov 2023
Kim JS Kim SH Kim SC Park JH Kim HG Lee SM Yoo JC

Aims. The aim of this study was to compare the clinical and radiological outcomes of reverse shoulder arthroplasty (RSA) using small and standard baseplates in Asian patients, and to investigate the impact of a mismatch in the sizes of the glenoid and the baseplate on the outcomes. Methods. This was retrospective analysis of 50 and 33 RSAs using a standard (33.8 mm, ST group) and a small (29.5 mm, SM group) baseplate of the Equinoxe reverse shoulder system, which were undertaken between January 2017 and March 2021. Radiological evaluations included the size of the glenoid, the β-angle, the inclination of the glenoid component, inferior overhang, scapular notching, the location of the central cage in the baseplate within the vault and the mismatch in size between the glenoid and baseplate. Clinical evaluations included the range of motion (ROM) and functional scores. In subgroup analysis, comparisons were performed between those in whom the vault of the glenoid was perforated (VP group) and those in whom it was not perforated (VNP group). Results. Perforation of the vault of the glenoid (p = 0.018) and size mismatch in height (p < 0.001) and width (p = 0.013) were significantly more frequent in the ST group than in the SM group. There was no significant difference in the clinical scores and ROM in the two groups, two years postoperatively (all p > 0.05). In subgroup analysis, the VP group had significantly less inferior overhang (p = 0.009), more scapular notching (p = 0.018), and more size mismatch in height (p < 0.001) and width (p = 0.025) than the VNP group. Conclusion. In Asian patients with a small glenoid, using a 29.5 mm small baseplate at the time of RSA was more effective in reducing size mismatch between the glenoid and the baseplate, decreasing the incidence of perforation of the glenoid vault, and achieving optimal positioning of the baseplate compared with the use of a 33.8 mm standard baseplate. However, longer follow-up is required to assess the impact of these findings on the clinical outcomes. Cite this article: Bone Joint J 2023;105-B(11):1189–1195


Bone & Joint Open
Vol. 4, Issue 7 | Pages 490 - 495
4 Jul 2023
Robinson PG Creighton AP Cheng J Dines JS Su EP Gulotta LV Padgett D Demetracopoulos C Hawkes R Prather H Press JM Clement ND

Aims

The primary aim of this prospective, multicentre study is to describe the rates of returning to golf following hip, knee, ankle, and shoulder arthroplasty in an active golfing population. Secondary aims will include determining the timing of return to golf, changes in ability, handicap, and mobility, and assessing joint-specific and health-related outcomes following surgery.

Methods

This is a multicentre, prospective, longitudinal study between the Hospital for Special Surgery, (New York City, New York, USA) and Edinburgh Orthopaedics, Royal Infirmary of Edinburgh, (Edinburgh, UK). Both centres are high-volume arthroplasty centres, specializing in upper and lower limb arthroplasty. Patients undergoing hip, knee, ankle, or shoulder arthroplasty at either centre, and who report being golfers prior to arthroplasty, will be included. Patient-reported outcome measures will be obtained at six weeks, three months, six months, and 12 months. A two-year period of recruitment will be undertaken of arthroplasty patients at both sites.


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_10 | Pages 89 - 89
1 Oct 2022
Alier A Gasol B Pérez-Prieto D Santana F Torrens C
Full Access

Aim. A significant number of patients undergoing shoulder arthroplasty surgery have C acnes contamination at the end of the primary surgery. The objective of this study is to determine whether patients with C acnes contamination at the end of their primary shoulder surgery have a worse prognosis than those who end up without C. acnes contamination. Method. Prospective study including all patients who underwent a reverse shoulder prosthesis from January 2015 to December 2018. In all of them, 5 to 12 cultures were performed during primary surgery. The patients underwent surgery for shoulder arthritis secondary to rotator cuff tears, acute fracture of the proximal humerus, and sequelae of fracture of the proximal humerus. Exclusion criteria included the existence of previous surgeries on the affected shoulder, the presence of signs of infection, having received infiltrations and / or complementary invasive examinations (Arthro-MRI and Arthro-CT). Follow-up from 2 to 5 years. Functional assessment according to the Constant Functional Scale. All complications were also recorded. Results. 162 patients were included. Of these, 25 had positive cultures for C. acnes at the end of primary shoulder surgery. Average age of 74.8 years. 136 women and 26 men. 75.9% Shoulder arthritis secondary to rotator cuff tears, 13.6% acute fractures and 10.5% sequelae of fractures. There were no differences between patients with C. acnes and those without C. acnes regarding age and indication for surgery. Predominance of men in the group with positive C. acnes (p <0.001). No differences at 2 and 5 years in the Constant functional scale between the two groups (2 years, 59.6 vs 59.2 p 0.870) (5 years, 62.4 vs 59.5 p 0.360). Significant differences regarding the number of complications (p 0.001). Patients without C. acnes had 1 aseptic loosening of the metaglene and patients with C. acnes had 2 infections, 1 dislocation, and 1 revision surgery. Patients with contamination by C. acnes had more comorbidities (p 0.035) than patients without contamination. Conclusions. Patients with C acnes contamination at the end of primary surgery do not have functional differences when compared with patients without contamination at 2 and 5 years, but they have a higher number of complications in the medium term


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_9 | Pages 8 - 8
1 Jun 2021
Giorgini A Tarallo L Porcellini G Micheloni G
Full Access

Introduction. Reverse shoulder Arthroplasty is a successful treatment for gleno-humeral osteoarthritis. However, components loosening and painful prostheses, related to components wrong positioning, are still a problem for those patients who underwent this kind of surgery. Several new technology has been developed the improve the implant positioning. CT-based intraoperative navigation system is a suitable technology that allow the surgeon to prepare the implant site exactly as planned with preoperative software. Method. Thirty reverse shoulder prostheses were performed at Modena Polyclinic using GPS CT-based intraoperative navigation system (Exactech, Gainsville, Florida). Walch classification was used to assess glenoid type. Planned version and inclination of the glenoid component, planned seating, final version and inclination of the reamer were recorded. Intraoperative and perioperative complication were recorded. Planned positioning was conducted aiming to the maximum seating, avoiding retroversion >10° and superior inclination. Results. Eight patients were male, 22 were female. Mean age was 75 years old (range 58–87). 4 glenoid were type B3, four were B2, 10 cases were B1, 12 case were A1/A2. Posterior or superior augment was used in 15 cases. Mean planned seating was 93%. Mean preoperative version was -7.5±6.9°; Mean planned version was -2±2.8°; Mean intraoperative measured version was -1.9±2.8°; no statistical difference was found between planned and intraoperative version (p=0.16). Mean preoperative inclination was 1.8±6.°; Mean planned inclination was -2.2±2.4°; Mean intraoperative measured inclination was -2.1.9±2.3°; no statistical difference was found between planned and intraoperative version or inclination (respectively p=0.16 and p=0.32). Mean surgical time was 71 minute (range 51–82). Three cases of coracoid ruptures were reported, 1 failure of the system occurred. Discussion. GPS navigation system allows the surgeon to prepare the implant site as planned on Preoperative software in Reverse shoulder arthroplasty, with no statistical difference between planned orientation and intraoperative measured orientation. That means that even in the most difficult cases the surgeon is able to find a good positioning (93% seating)and to replicate it in the operative room. Only one failure of the system occurred, because too much time was passed between CT scan and surgery (9 months). Three coracoid fractures occurred in the first 10 cases: these could be addressed to a lack of confidence with the double lateralization of this prosthesis which increase tensioning on the coracoid and a lack of confidence in tracker positioning, which should be made as proximal as it is possible. Finally, the system needs several improvements to be considered a breakthrough technology, such as humeral component positioning and final control of the implant, but by now is a useful way to improve our surgery, especially in difficult cases


Bone & Joint Open
Vol. 1, Issue 12 | Pages 731 - 736
1 Dec 2020
Packer TW Sabharwal S Griffiths D Reilly P

Aims

The purpose of this study was to evaluate the cost of reverse shoulder arthroplasty (RSA) for patients with a proximal humerus fracture, using time-driven activity based costing (TDABC), and to compare treatment costs with reimbursement under the Healthcare Resource Groups (HRGs).

Methods

TDABC analysis based on the principles outlined by Kaplan and a clinical pathway that has previously been validated for this institution was used. Staffing cost, consumables, implants, and overheads were updated to reflect 2019/2020 costs. This was compared with the HRG reimbursements.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_8 | Pages 76 - 76
1 Aug 2020
Habis A Bicknell R Mei X
Full Access

Reverse shoulder arthroplasty (RSA) has an increasing effective use in the treatment of patients with a variety of diagnoses, including rotator cuff deficiency, inflammatory arthritis, or failed shoulder prostheses. Glenoid bone loss is not uncommonly encountered in these cases due to the significant wear. Severe bone loss can compromise glenoid baseplate positioning and fixation, consequently increasing the risk for early component loosening, instability, and scapular notching. To manage severe glenoid bone deficiencies, bone grafts are commonly used. Although, many studies report outcome of bone grafting in revision RSA, the literature on humeral head autograft for glenoid bone loss in primary RSA is less robust. The purpose of this study is to evaluate the clinical and radiographic outcomes of primary RSA with humeral head autograft for glenoid bone loss at our institution. Institutional review board approval was obtained to retrospectively review the records of 22 consecutive primary RTSA surgeries in 21 patients with humeral head autograft for glenoid bone loss between January 2008 and December 2016. Five patients died during follow-up, three were unable to be contacted and one refused to participate, leaving a final study cohort of 12 patients with 13 shoulders that underwent RSA. All patients had a clinical evaluation including detailed ROM and clinical evaluation using the American Shoulder and Elbow Surgeons (ASES) Score, Constant Score, Western Ontario Osteoarthritis of the Shoulder Index (WOOS), and Short Form-12 (SF-12) questionnaires. Preoperative and postoperative plain radiographs and CT scans were assessed for component position, loosening, scapular notching, as well as graft incorporation, resorption, or collapse. There were 6 males and 6 females, with an average age of 74 ± 6.8 years. The average BMI was 31.7 ± 5.3, and the median ASA score was 3. Average follow-up was 3.4 ± 1.1 years. The average postoperative range of motion measurements for the operative arm are: flexion = 120 ± 37, abduction = 106 ± 23, external rotation = 14 ± 12, internal rotation at 90 degrees of abduction = 49 ± 7, external rotation at 90 degrees of abduction = 50 ± 28. Average functional scores are: ASES: 76.9 ± 19.2, WOOS: 456 ± 347, SF12 physical: 34.2 ± 8.2, SF12 mental: 54.1 ± 10.2, Constant Score: 64.6 ± 14. No evidence of hardware loosening or evidence of bone graft resorption were encountered. On CT, the average of pre operative B-angle was 79.3 ± 9.3 while the pre operative reverse shoulder angle was 101.4 ± 28. Glenoid retroversion average on CT was 13.3 ± 16.6. Post operative baseplate inclination average was 82 ± 7.4 while the baseplate version 7.8 ±10. The operative technique was able to achieve up to 30 degrees of inclination correction and up to 50 degrees of version correction. In conclusion, primary reverse shoulder arthroplasty with humeral head autograft for glenoid bone loss provides excellent ROM and functional outcomes at mid-term follow-up. This technique has a high rate of bone incorporation and small risk of bone resorption at mid term follow up


Bone & Joint 360
Vol. 9, Issue 3 | Pages 31 - 34
1 Jun 2020


The Bone & Joint Journal
Vol. 102-B, Issue 3 | Pages 365 - 370
1 Mar 2020
Min KS Fox HM Bedi A Walch G Warner JJP

Aims

Patient-specific instrumentation has been shown to increase a surgeon’s precision and accuracy in placing the glenoid component in shoulder arthroplasty. There is, however, little available information about the use of patient-specific planning (PSP) tools for this operation. It is not known how these tools alter the decision-making patterns of shoulder surgeons. The aim of this study was to investigate whether PSP, when compared with the use of plain radiographs or select static CT images, influences the understanding of glenoid pathology and surgical planning.

Methods

A case-based survey presented surgeons with a patient’s history, physical examination, and, sequentially, radiographs, select static CT images, and PSP with a 3D imaging program. For each imaging modality, the surgeons were asked to identify the Walch classification of the glenoid and to propose the surgical treatment. The participating surgeons were grouped according to the annual volume of shoulder arthroplasties that they undertook, and responses were compared with the recommendations of two experts.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 75 - 75
1 Feb 2020
Pitocchi Wirix-Speetjens Lenthe V Perez
Full Access

Introduction

Loosening of the baseplate is one of the most common causes of failure in Reverse Shoulder Arthroplasty. To allow osteo-integration to occur and thus provide long-term stability, initial screws fixation plays a pivotal role. In particular, tightening torque and force of nonlocking screws are two parameters that are considered to have a clear impact on implant stability, yet the relation is not fully understood. For this reason, this study aims to define an experimental set-up, to measure force and torque in artificial bone samples of different quality, in order to estimate ranges of optimal surgical values and give guidelines to maximize screw fixation and therefore initial implant stability.

Methods

A custom-made torque sensor (Figure 1a) was built and calibrated using a lever deadweight system. To measure the compression force generated by the screw head, three thin FlexiForce sensors (Tekscan, South Boston, US) were enclosed between two 3D printed plates with a central hole to allow screw insertion (Figure 1b). The tightening force, represented by the sum of the three sensors, was calibrated using a uniaxial testing machine (Zwick/Roell, Ulm, Germany). Multiple screw lengths (26mm, 32mm and 47mm) were selected in the protocol. Synthetic bone blocks (Sawbones; Malmö, Sweden) of 20 and 30 PCF were used to account for bone quality variation. To evaluate the effect of a cortical bone layer, for each density three blocks were considered with 0 mm (no layer), 1.5 mm and 3 mm of laminate foam of 50 PCF. The holes for the screws were pre-drilled in the same way as in the operation room. For each combination of screw dimensions and bone quality, ten measurements were performed by acquiring the signal of the insertion torque and tightening force until bone breaking.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 102 - 102
1 Feb 2020
DiGeorgio C Yegres J VanDeven J Stroud N Cheung E Grey S Yoo J Deshmukh R Crosby L Roche C
Full Access

Introduction. Little guidance exists regarding the minimum screw length and number necessary to achieve fixation with reverse shoulder arthroplasty (rTSA). The goal of this study is to quantify the pre- and post-cyclic baseplate displacements associated with two baseplate designs of different sizes using multiple screw lengths and numbers in a low density polyurethane bone substitute model. Methods. The test was conducted according to ASTM F 2028–17. The baseplate displacements of standard and small reverse shoulder constructs (Equinoxe, Exactech, Inc.) were quantified in a 15pcf polyurethane block (Pacific Research, Inc.) before and after cyclic testing with an applied load of 750N for 10,000 cycles. Baseplates were constructed using 2 or 4 screws with 3 different poly-axial locking compression screw lengths: 4.5×18mm, 4.5×30mm, and 4.5×46mm. Five of each configuration were tested for a total of 30 specimens for each baseplate. A two-tailed, unpaired student's t-test (p<0.05) compared baseplate displacements before and after cyclic loading in both the superior-inferior (S/I) and anterior-posterior (A/P) directions. The standard and small results were then compared. Results. All standard and small reverse glenoid baseplates remained well-fixed after cyclic loading in the low-density bone substitute model regardless of screw length or number. The average pre- and post-cyclic displacement for baseplates with 2 screws was significantly greater than that of baseplates with 4 screws in both the A/P and S/I directions. The average pre- and post-cyclic displacements for baseplates with 18mm screws were significantly greater than baseplates with 46mm screws in the A/P and S/I directions, post-cyclic displacement with 18mm screws was significantly greater than with 30mm screws in the A/P and S/I directions, and post-cyclic displacement with 30mm screws was significantly greater than with 46mm screws in the S/I direction only. Few differences in fixation were observed between baseplate sizes. Statistically significant difference was reached for post cyclic S/I displacement for 30mm (small baseplate superior) and 46mm screws (standard baseplate superior). Discussion and Conclusions. The results demonstrate that rTSA glenoid displacement is impacted by both the number and length of screws for both standard and small baseplate sizes. Regardless of the number of screws, the use of longer screws was associated with significantly better initial fixation. Additionally, the use of more screws was associated with significantly better fixation irrespective of screw length in the A/P direction. None of the tested devices catastrophically failed, demonstrating that adequate fixation can be achieved with as little as two 18mm screws for the baseplates utilized. However, this screw configuration was associated with the largest pre- and post-cyclic displacements, so it is assumed to be at a greater risk for aseptic loosening. If using 4 screws is not feasible in a given case, the results suggest that using longer screws can be used to improve fixation. The results of the small and standard baseplates were comparable for the given lengths and quantities of screws, suggesting that the reduced surface area of the small baseplate has no detrimental impact on fixation. Care should be made when extrapolating these results to glenoid defects. For any figures or tables, please contact authors directly


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 40 - 40
1 Feb 2020
Tarallo L Porcellini G Giorgini A Pellegrini A Catani F
Full Access

Introduction. Total shoulder replacement is a successful treatment for gleno-humeral osteoarthritis. However, components loosening and painful prostheses, related to components wrong positioning, are still a problem for those patients who underwent this kind of surgery. CT-based intraoperative navigation system is a suitable option to improve accuracy and precision of the implants as previously described in literature for others district. Method. Eleven reverse shoulder prostheses were performed at Modena Polyclinic from October 2018 to April 2019 using GPS CT-based intraoperative navigation system (Exactech, Gainsville, Florida). In the preoperative planning, Walch classification was used to assess glenoid type. The choice of inclination of the glenoid component, the screw length, as well as the inclination of the reamer was study and recorded using specific software using the CT scan of shoulder of each patient (Fig.1, Fig.2). Intraoperative and perioperative complications were recorded. Three patients were male, eight were female. Mean age was 72 years old (range 58=84). Three glenoid were type B2, six cases were B1, two case were type C1. Results. In all cases treated by reverse shoulder prostheses we had obtain good functional results at preliminary follow up. Eight degree posterior augment was used in seven case. Planned version was 0° in eight case, an anti-version of 3° was planned in the other three cases. Final reaming was as preoperatively planned in all cases except one. Mean surgical time was 71 minutes (range 51–82). One case of coracoid rupture has been reported. In all cases the system worked in proper manner without failures, no case of infection was reported. Discussion. It is well known as the more accurate placement of the glenoid led to enhanced long-term survivorship of the implant and decrease complication rates in RSTA. Our first experience with GPS navigation system has been satisfied. Good components’ positioning has been reached in all cases, without deviation from the preoperative planning. Pre-operative preparation using software has been always respected except in one case in which we decided to ream 1mm less to avoid excessive bone loss. In 3 case we decide to increase glenoid anti-version to allow a good cage containment in the scapula. No failure of the system has been recorded, with a little increase in the surgical time respect to traditional surgeries performed in our institute. The first case performed reported coracoid fracture, probably due to lack of experience in coracoid tracker positioning. It is very important to set the surgical theatre and the position of the patient in order to make the coracoid tracker visible for the computer. Screw positioning and length is decisively improved with GPS system compared with traditional implant. The most important advantage is to avoid the malposition of the glenoid component, solving problems like loosening or restriction in shoulder range of motion. We believe that a final cross check between preoperative planning and final control of the prostheses implanted, should be used in the future, but by now the GPS navigation system is a useful way to improve our surgery, especially in difficult cases. For any figures or tables, please contact the authors directly


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 134 - 134
1 Feb 2020
Greene A Parsons I Jones R Youderian A Byram I Papandrea R Cheung E Wright T Zuckerman J Flurin P
Full Access

INTRODUCTION. 3D preoperative planning software for anatomic total shoulder arthroplasty (ATSA) provides surgeons with increased ability to visualize complex joint relationships and deformities. Interestingly, the advent of such software has seemed to create less of a consensus on the optimal way to plan an ATSA rather than more. In this study, a survey of shoulder specialists from the American Shoulder and Elbow Society (ASES) was conducted to examine thought patterns in current ATSA implant selection and placement. METHODS. 172 ASES members completed an 18-question survey on their thought process for how they select and place an ATSA glenoid implant. Data was collected using a custom online Survey Monkey survey. Surgeon answers were split into two cohorts based on number of arthroplasties performed per year: between 0–75 was considered low volume (LV), and between 75–200+ was considered high volume (HV). Data was analyzed for each cohort to examine differences in thought patterns, implant selection, and implant placement. RESULTS. 70 surgeons were grouped into the LV cohort, and 102 surgeons were grouped into the HV cohort. 46.1% of surgeons in the HV cohort reported using a preoperative planning software for the majority of cases vs. 41.4% in the LV cohort, 48% of surgeons in the HV cohort reported seldom use vs. 24.3% in the LV cohort, and 5.9% of surgeons in the HV cohort reported no use vs. 34.3% in the LV cohort (Figure 1). When questioned on what percentage of ATSA cases do surgeons use augmented glenoid implants, 20.6% in the HV cohort responded never using augments vs. 30% in the LV cohort, 39.2% responded using augments <15% of the time in the HV cohort vs. 34.3% in the LV cohort, 26.5% responded using augments between 15–45% of the time in the HV cohort vs. 28.6% in the LV cohort, and 13.7% responded using augments >45% of the time in the HV cohort vs. 7.2% in the LV cohort (Figure 2). When asked what the maximum allowable residual retroversion for an ATSA glenoid implant is, surgeons answered 0–5° 6.9% of the time in the HV cohort vs. 4.3% in the LV cohort, 6–9° 35.6% of the time in the HV cohort vs. 50% in the LV cohort, 10–12° 34.7% of the time in the HV cohort vs. 32.9% in the LV cohort, 13–15° 10.9% of the time in the HV cohort vs. 8.6% in the LV cohort, and lastly >16° 11.9% of the time in the HV cohort vs. 4.3% in the LV cohort (Figure 3). CONCLUSION. Research suggests ATSA glenoid implants may be less forgiving of malalignment than reverse shoulder glenoid implants, but the contrasting survey results in this study reveal that a consensus in optimal placement has yet to be reached. Interestingly, even though HV use more augmented glenoid components than LV surgeons, HV surgeons are more accepting of residual glenoid component retroversion than LV surgeons. Despite these differences, there is no way to prove the optimal implant selection and placement without long-term clinical outcomes. For any figures or tables, please contact the authors directly


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 23 - 23
1 Feb 2020
Van De Kleut M Athwal G Yuan X Teeter M
Full Access

Introduction. Reverse total shoulder arthroplasty (RTSA) is a semi-constrained joint replacement with an articulating cobalt-chromium glenosphere and ultra-high molecular weight polyethylene (PE). Because of its limited load bearing, surgeons and implant manufacturers have not elicited the use of highly cross-linked PE in the shoulder, and to date have not considered excessive PE wear in the reverse shoulder a primary concern. As the number of shoulder procedures is expected to grow exponentially in the next decade, however, it is important to evaluate how new designs and bearing materials interact and to have an understanding of what is normal in well-functioning joint replacements. Currently, no in vivo investigation into RTSA PE wear has been conducted, with limited retrieval and simulation studies. In vitro and in silico studies demonstrate a large range in expected wear rates, from 14.3 mm. 3. /million cycles (MC) to 126 mm. 3. /MC, with no obvious relationship between wear rate and polyethylene diameter. The purpose of this study is to evaluate, for the first time, both volumetric and linear wear rates in reverse shoulder patients, with a minimum six-year follow-up using stereo radiographic techniques. Methods. To date, seven patients with a self-reported well-functioning Aequalis Reversed II (Wright Medical Group, Edina, MN, USA) RTSA implant system have been imaged (mean years from surgery = 7.0, range = 6.2 to 9). Using stereo radiographs, patients were imaged at the extents of their range of motion in internal and external rotation, lateral abduction, forward flexion, and with their arm at the side. Multiple arm positions were used to account for the multiple wear vectors associated with activities of daily living and the shoulder's six degrees of motion. Using proprietary software, the position and orientation of the polyethylene and glenosphere components were identified and their transformation matrices recorded. These transformation matrices were then applied to the CAD models of each component, respectively, and the apparent intersection of the glenosphere into the PE recorded. Using previously validated in-house software, volumetric and maximum linear wear depth measurements were obtained. Linear regression was used to identify wear rates. Results. The volumetric and linear wear rates for the 36 mm PE liners (n = 5) were 39 mm. 3. /y (r. 2. = 0.86, range = 24 to 42 mm. 3. /y) and 0.09 mm/y (r. 2. = 0.96, range = 0.08 to 0.11 mm/y), respectively. Only two patients with 42 mm PE liners were evaluated. For these, volumetric and linear wear rates were 110 mm. 3. /y (r. 2. = 0.81, range = 83 to 145 mm. 3. /y) and 0.17 mm/y (r. 2. = 0.99, range = 1.12 to 1.15 mm/y), respectively. Conclusion. For the first time, PE wear was evaluated in the reverse shoulder in vivo. More patients are required for conclusive statements, but preliminary results suggest first order volumetric and linear wear rates within those predicted by simulation studies. It is interesting to note the increased wear with larger PE size, likely due to the increased contact area between congruent faces and the potential for increased sliding distance during arm motion


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 72 - 72
1 Feb 2020
Hall D Garrigues G Blanchard K Shewman E Nicholson G Pourzal R
Full Access

Introduction

The combined incidence of anatomic (aTSA) and reverse total shoulder arthroplasties (rTSA) in the US is 90,000 per annum and rising. There has been little attention given to potential long-term complications due to periprosthetic tissue reactions to implant debris. The shoulder has been felt to be relatively immune to these complications due to lower acting loads compared to other joint arthroplasties. In this study, retrieved aTSAs and rTSAs were examined to determine the extent of implant damage and to characterize the nature of the corresponding periprosthetic tissue responses.

Methods

TSA components and periprosthetic tissues were retrieved from 23 (eleven aTSA, twelve rTSA). Damage to the implants was characterized using light microscopy. Head/stem taper junction damage was graded 1–4 as minimal, mild, moderate or marked. Damage on polyethylene (PE) and metal bearing surfaces was graded 1–3 (mild, moderate, marked). H&E stained sections of periprosthetic soft tissues were evaluated for the extent and type of cellular response. A semi-quantitative system was used to score (1=rare to 4=marked) the overall number of particle-laden macrophages, foreign body giant cells, lymphocytes, plasma cells, eosinophils, and neutrophils. Implant damage and histopathological patterns were compared between the two TSA groups using the Mann-Whitney and Spearman tests.


The Bone & Joint Journal
Vol. 101-B, Issue 12 | Pages 1520 - 1525
1 Dec 2019
Clark NJ Samuelsen BT Alentorn-Geli E Assenmacher AT Cofield RH Sperling JW Sánchez-Sotelo J

Aims

Reverse shoulder arthroplasty (RSA) reliably improves shoulder pain and function for a variety of indications. However, the safety and efficacy of RSA in elderly patients is largely unknown. The purpose of this study was to report the mortality, morbidity, complications, reoperations, and outcomes of primary RSA in patients aged > 80 years.

Patients and Methods

Between 2004 and 2013, 242 consecutive primary RSAs were performed in patients aged > 80 years (mean 83.3 years (sd 3.1)). Of these, 53 were lost to follow-up before two years and ten had died within two years of surgery, leaving 179 for analysis of survivorship, pain, motion, and strength at a minimum of two years or until revision surgery. All 242 patients were considered for the analysis of 90-day, one-year, and overall mortality, medical complications (90-day and overall), surgical complications, and reoperations. The indications for surgery included rotator cuff arthropathy, osteoarthritis, fracture, the sequela of trauma, avascular necrosis, and rheumatoid arthritis. A retrospective review of the medical records was performed to collect all variables. Survivorship free of revision surgery was calculated at two and five years.


Bone & Joint 360
Vol. 8, Issue 5 | Pages 27 - 30
1 Oct 2019


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1107 - 1114
1 Sep 2019
Uy M Wang J Horner NS Bedi A Leroux T Alolabi B Khan M

Aims

The aim of this study was to evaluate the differences in revision and complication rates, functional outcomes, and radiological outcomes between cemented and press-fit humeral stems in primary anatomical total shoulder arthroplasty (TSA).

Materials and Methods

A comprehensive systematic review and meta-analysis was conducted searching for studies that included patients who underwent primary anatomical TSA for primary osteoarthritis or rheumatoid arthritis.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 92 - 92
1 Apr 2019
Matsuki K Matsuki K Sugaya H Takahashi N Hoshika S Tokai M Ueda Y Hamada H Banks S
Full Access

Background. Scapular notching is a complication after reverse shoulder arthroplasty with a high incidence up to 100%. Its clinical relevance remains uncertain; however, some studies have reported that scapular notching is associated with an inferior clinical outcome. There have been no published articles that studied positional relationship between the scapular neck and polyethylene insert in vivo. The purpose of this study was to measure the distance between the scapular neck and polyethylene insert in shoulders with Grammont type reverse shoulder arthroplasty during active external rotation at the side. Methods. Eighteen shoulders with Grammont type prosthesis (Aequalis Reverse, Tornier) were enrolled in this study. There were 13 males and 5 female, and the mean age at surgery was 74 years (range, 63–91). All shoulders used a glenosphere with 36mm diameter, and retroversion of the humeral implant was 10°in 4 shoulders, 15°in 3 shoulders, and 20°in 11 shoulders. Fluoroscopic images were recorded during active external rotation at the side from maximum internal to external rotation at the mean of 14 months (range, 7–24) after surgery. The patients also underwent CT scans, and three-dimensional glenosphere models with screws and scapula neck models were created from CT images. CT-derived models of the glenosphere and computer-aided design humeral implant models were matched with the silhouette of the implants in the fluoroscopic images using model-image registration techniques (Figure 1). Based on the calculated kinematics of the implants, the closest distance between the scapular neck and polyethylene insert was computed using the scapular model and computer-aided design insert models (Figure 2). The distance was computed at each 5° increment of glenohumeral internal/external rotation, and the data from 20°internal rotation to 40°external rotation were used for analyses. One-way repeated-measures analysis of variance was used to examine the change of the distance during the activity, and the level of significance was set at P < 0.05. Results. The mean glenohumeral abduction during the activity was 17°-22°. The mean distance between the neck and insert was approximately 1mm throughout the activity (Figure 3). The distance tended to become smaller with the arm externally rotated, but the change was not significant. Discussion. The reported incidence of scapular notching after Grammont type reverse shoulder arthroplasty is generally higher than the newer design prosthesis with the lateralized center of rotation. This may be associated with the design of the prosthesis, and the results of this study that the distance between the neck and insert was approximately 1mm throughout active external rotation at the side will support the high incidence of notching. We may need to analyze the distance with the newer design reverse shoulder prosthesis to prove the architectural advantage of the newer systems. Conclusion. The distance between the scapular neck and polyethylene insert was approximately 1mm throughout active external rotation activity in shoulders with Grammont type prosthesis