This study compared patient-reported outcomes of three total knee arthroplasty (TKA) designs from one manufacturer: one cruciate-retaining (CR) design, and two cruciate-sacrificing designs, anterior-stabilized (AS) and posterior-stabilized (PS). Patients scheduled for primary TKA were included in a single-centre, prospective, three-armed, blinded randomized trial (n = 216; 72 per group). After intraoperative confirmation of posterior cruciate ligament (PCL) integrity, patients were randomly allocated to receive a CR, AS, or PS design from the same TKA system. Insertion of an AS or PS design required PCL resection. The primary outcome was the mean score of all five subscales of the Knee injury and Osteoarthritis Outcome Score (KOOS) at two-year follow-up. Secondary outcomes included all KOOS subscales, Oxford Knee Score, EuroQol five-dimension health questionnaire, EuroQol visual analogue scale, range of motion (ROM), and willingness to undergo the operation again. Patient satisfaction was also assessed.Aims
Methods
Medial unicompartmental knee arthroplasty (mUKA) is an advised treatment for anteromedial knee osteoarthritis. While long-term survival after mUKA is well described, reported incidences of short-term surgical complications vary and the effect of surgical usage on complications is less established. We aimed to describe the overall occurrence and treatment of surgical complications within 90 days of mUKA, as well as occurrence in high-usage centres compared to low-usage centres. mUKAs performed in eight fast-track centres from February 2010 to June 2018 were included from the Lundbeck Foundation Centre for Fast-track Hip and Knee Replacement Database. All readmissions within 90 days of surgery underwent chart review and readmissions related to the surgical wound or the prosthesis were recorded. Centres were categorized as high-usage centres when using mUKA in ≥ 20% of annual knee arthroplasties. The occurrence of complications between high- and low-usage centres were compared using Fisher’s exact test.Aims
Methods
Limb alignment in total knee arthroplasty (TKA) influences periarticular soft-tissue tension, biomechanics through knee flexion, and implant survival. Despite this, there is no uniform consensus on the optimal alignment technique for TKA. Neutral mechanical alignment facilitates knee flexion and symmetrical component wear but forces the limb into an unnatural position that alters native knee kinematics through the arc of knee flexion. Kinematic alignment aims to restore native limb alignment, but the safe ranges with this technique remain uncertain and the effects of this alignment technique on component survivorship remain unknown. Anatomical alignment aims to restore predisease limb alignment and knee geometry, but existing studies using this technique are based on cadaveric specimens or clinical trials with limited follow-up times. Functional alignment aims to restore the native plane and obliquity of the joint by manipulating implant positioning while limiting soft tissue releases, but the results of high-quality studies with long-term outcomes are still awaited. The drawbacks of existing studies on alignment include the use of surgical techniques with limited accuracy and reproducibility of achieving the planned alignment, poor correlation of intraoperative data to long-term functional outcomes and implant survivorship, and a paucity of studies on the safe ranges of limb alignment. Further studies on alignment in TKA should use surgical adjuncts (e.g. robotic technology) to help execute the planned alignment with improved accuracy, include intraoperative assessments of knee biomechanics and periarticular soft-tissue tension, and correlate alignment to long-term functional outcomes and survivorship.
The interest in unicompartmental knee arthroplasty (UKA) for
medial osteoarthritis has increased rapidly but the long-term follow-up
of the Oxford UKAs has yet to be analysed in non-designer centres.
We have examined our ten- to 15-year clinical and radiological follow-up
data for the Oxford Phase III UKAs. Between January 1999 and January 2005 a total of 138 consecutive
Oxford Phase III arthroplasties were performed by a single surgeon
in 129 patients for medial compartment osteoarthritis (71 right
and 67 left knees, mean age 72.0 years (47 to 91), mean body mass
index 28.2 (20.7 to 52.2)). Both clinical data and radiographs were
prospectively recorded and obtained at intervals. Of the 129 patients,
32 patients (32 knees) died, ten patients (12 knees) were not able
to take part in the final clinical and radiological assessment due
to physical and mental conditions, but via telephone interview it
was confirmed that none of these ten patients (12 knees) had a revision
of the knee arthroplasty. One patient (two knees) was lost to follow-up.Aims
Patients and Methods
The purpose of this study was to develop an accurate, reliable and easily applicable method for determining the anatomical location of the joint line during revision knee arthroplasty. The transepicondylar width (TEW), the perpendicular distance between the medial and lateral epicondyles and the distal articular surfaces (DMAD, DLAD) and the distance between the medial and lateral epicondyles and the posterior articular surfaces (PMAD, DLAD) were measured in 40 knees from 20 formalin-fixed adult cadavers (11 male and nine female; mean age at death 56.9 years, Objectives
Methods