Healing after bone fracture is assessed by frequent radiographs, which expose patients to radiation and lacks behind biological healing. This study aimed to investigate whether the electrical impedance using electrical impedance spectroscopy correlated to quantitative scores of bone healing obtained from micro-CT and mechanical bending test. Eighteen rabbits were subjected to tibial fracture that was stabilized with external fixator. Two electrodes were positioned, one electrode placed within the medullary cavity and the other on the lateral cortex, both three millimeters from the fracture site. Impedance was measured daily across the fracture site at a frequency range of 5 Hz to 1 MHz. The animals were divided into three groups with different follow-up time: 1, 3 and 6 weeks for micro-CT (Bone volume/tissue volume (BV/TV, %)) and mechanical testing (maximum stress (MPa), failure energy (kJ/cm3), young modulus (Mpa)). There was a statistically significant correlation between last measured impedance at 5 Hz frequency immediately prior to euthanasia and BV/TV of callus (−0.68, 95%CI: (−0.87; −0.31)). Considering the mechanical testing with three-point bending, no significant correlation was found between last measured impedance at 5 Hz frequency immediately prior to euthanasia and maximum stress (−0.35, 95%CI: (−0.70; 0.14)), failure energy (−0.23, 95%CI: (−0.63; 0.26)), or young modulus (−0.28, 95%CI: (−0.66; 0.22)). The significant negative correlation between impedance and BV/TV might indicate that impedances correlate with the relative bone volume in the callus site. The lack of correlation between impedance and mechanical parameters when at the same time observing a correlation between impedance and days since operation (0-42 days), might indicate that the impedance can measure biological changes at an earlier time point than rough mechanical testing.
Rotator cuff muscle atrophy and fatty infiltration affect the clinical outcomes of rotator cuff tear patients. However, there is no effective treatment for fatty infiltration at this time. High-intensity interval training (HIIT) helps to activate beige adipose tissue. The goal of this study was to test the role of HIIT in improving muscle quality in a rotator cuff tear model via the β3 adrenergic receptor (β3AR). Three-month-old C57BL/6 J mice underwent a unilateral rotator cuff injury procedure. Mice were forced to run on a treadmill with the HIIT programme during the first to sixth weeks or seventh to 12th weeks after tendon tear surgery. To study the role of β3AR, SR59230A, a selective β3AR antagonist, was administered to mice ten minutes before each exercise through intraperitoneal injection. Supraspinatus muscle, interscapular brown fat, and inguinal subcutaneous white fat were harvested at the end of the 12th week after tendon tear and analyzed biomechanically, histologically, and biochemically.Aims
Methods
To detect early signs of infection infrared thermography has been suggested to provide quantitative information. Our vision is to invent a pin site infection thermographic surveillance tool for patients at home. A preliminary step to this goal is the aim of this study, to automate the process of locating the pin and detecting the pin sites in thermal images efficiently, exactly, and reliably for extracting pin site temperatures. A total of 1708 pin sites was investigated with Thermography and augmented by 9 different methods in to totally 10.409 images. The dataset was divided into a training set (n=8325), a validation set (n=1040), and a test set (n=1044) of images. The Pin Detection Model (PDM) was developed as follows: A You Only Look Once (YOLOv5) based object detection model with a Complete Detection Intersection over Union (CDIoU), it was pre-trained and finetuned by the through transfer learning. The basic performance of the YOLOv5 with CDIoU model was compared with other conventional models (FCOS and YOLOv4) for deep and transition learning to improve performance and precision. Maximum Temperature Extraction (MTE) Based on Region of Interest (ROI) for all pin sites was generated by the model. Inference of MTE using PDM with infected and un-infected datasets was investigated. An automatic tool that can identify and annotate pin sites on conventional images using bounding boxes was established. The bounding box was transferred to the infrared image. The PMD algorithm was built on YOLOv5 with CDIoU and has a precision of 0.976. The model offers the pin site detection in 1.8 milliseconds. The thermal data from ROI at the pin site was automatically extracted. These results enable automatic pin site annotation on thermography. The model tracks the correlation between temperature and infection from the detected pin sites and demonstrates it is a promising tool for automatic pin site detection and maximum temperature extraction for further infection studies. Our work for automatic pin site annotation on thermography paves the way for future research on infection assessment using thermography.
Home-based monitoring of fracture healing has the potential of reducing routine follow-up and improve personalized fracture care. Implantable sensors measuring electrical impedance might detect changes in the electrical current as the fracture heals. The aim was to investigate whether electrical impedance correlated with radiographic fracture healing. Eighteen rabbits were subjected to a tibial osteotomy that was stabilized with an external fixator. Two electrodes were positioned, one electrode placed within the medullary cavity and the other on the lateral cortex, both three millimeters from the osteotomy site. Transverse electrical impedance was measured daily across the fracture site at a frequency range of 5 Hz to 1 MHz using an Analog Discovery 2 Oscilloscope with Impedance Analyzer. Biweekly x-rays were taken and analyzed blinded using a modified anterior-posterior (AP) radiographic union score of the tibia (RUST). Each animal served as its own control by performing repeated measurements from time zero until the end of follow-up.Introduction and Objective
Materials and Methods
In 2019, Lin et al. published a proof-of-concept study of electrical impedance spectroscopy as a simple and low-cost method to characterize progression of fracture repair (Lin et al., Sci Rep 2019). However, the electrical impedance sensors were placed in the fracture site which may impair the transfer to clinical use. To further explore the concept of monitoring fracture healing by electrical impedance spectroscopy, we established a tibial fracture model in the rabbit where sensors are positioned in proximity to the fracture site but without being placed in the fracture site. The aim of this pilot study was to explore whether distinct patterns of electrical impedance would evolve as tibial fractures in rabbits were evaluated until radiographic signs of healing. Approval was granted from the Inspectorate of the Animal Experimentation under the Danish Ministry of Justice. Four rabbits were anaesthetized, and in each rabbit a tibial osteotomy was made and stabilized by an external fixator. Electrical impedance was measured immediately postoperative and hereafter daily until euthanization after 3 weeks. Recordings were obtained within a wide frequency range (10 Hz to 1 MHz) from an inner electrode placed into the medullary canal and an outer electrode placed extracortical on the lateral with a distance of 3 mm to the defect. A similar pattern of electrical impedance over time was observed in the four rabbits. During the very early stages of fracture healing, an initial fluctuation in electrical impedance occurred. However, after 10 days the curves revealed a steady daily increase in electrical impedance. The first radiological signs of bone healing were detected after 14 days and progressed in all four rabbits in accordance with increments in the electrical impedance until termination of the pilot study after 21 days. Consistent electrical impedance patterns were detected during bone healing in a pilot study of four rabbits. Further research is needed to explore whether the presented method of electrical impedance measurements can be used to monitor bone healing over time.
Malnutrition is a potentially modifiable risk factor that may contribute to complications following geriatric hip fracture surgery. The purpose of this study was to investigate the association between preoperative hypoalbuminemia, a marker for malnutrition, and complications during the thirty days following surgery for geriatric hip fracture. The American College of Surgeons National Surgical Quality Improvement Program was used to conduct a retrospective cohort study of geriatric patients (>65 years) undergoing surgery for hip fracture. Patients without preoperative serum albumin concentration were excluded. Outcomes were compared between patients with and without hypoalbuminemia (defined as serum albumin concentration <3.5g/dL). All comparisons were adjusted for baseline differences between populations. 17,651 Patients were identified. Of these, 8,272 (46.9%) underwent hemiarthroplasty, 759 (4.3%) total joint arthroplasty, 324 (1.9%) percutaneous fixation, 2,445 (13.9%) plate/screw fixation, and 5,833 (33.1%) intramedullary fixation. The prevalence of hypoalbuminemia was 45.9% (Figure 1). The risk for death was strongly associated with serum albumin concentration, with a linear increase in risk observed as albumin fell below 3.5 g/dL (p<0.001; Figure 2). Following adjustment for all demographic, comorbidity, and procedural characteristics, patients with hypoalbuminemia had higher rates of death (9.94% versus 5.53%, adjusted relative risk [RR]=1.54, p<0.001), pneumonia (5.30% versus 3.77%, adjusted RR=1.20, p=0.012), sepsis (1.19% versus 0.53%, adjusted RR=1.90, p<0.001), and hospital readmission (10.91% versus 9.03%, adjusted RR=1.11, p<0.036; Table 1). The present study suggests that hypoalbuminemia is a powerful independent risk factor for death following surgery for geriatric hip fracture. This association persists over-and-above any associations of death with age, sex, body mass index, and comorbidities. Based on these data, we propose that the nutritional status of hip fracture patients should receive greater attention, and that randomized trials testing for efficacy of aggressive postoperative nutritional interventions may be warranted. For any figures or tables, please contact the authors directly by clicking on ‘Info & Metrics’ above to access author contact details.
Bone ingrowth is desired with uncemented hip implants. Infection is clearly undesirable. We have worked on developing a nanofiber coating for implants that would enhance bone formation while inhibiting infection. Few studies have focused on developing an implant surface nanofiber (NF) coating to prevent infection and enhance osseointegration by local drug release. In this study, coaxial doxycycline (Doxy)-doped polycaprolactone/polyvinyl alcohol (PCL/PVA) Nanofibers were directly deposited on the titanium (Ti) implant surface during electrospinning. The interaction of loaded Doxy with both PVA and PCL NFs was characterized by Raman spectroscopy. The bonding strength of Doxy-doped NF coating on Ti implants was confirmed by a stand single-pass scratch test. The improved implant osseointegration by PCL/PVA NF coatings in vivo was confirmed by scanning electron microscopy, histomorphometry and micro computed tomography at 2, 4 and 8 weeks after implantation. The bone contact surface (%) changes of NF coating group (80%) is significantly higher than that of no NF group (< 5%, p<0.05). Finally, we demonstrated that Doxy-doped NF coating effectively inhibited bacterial infection and enhanced osseointegration in an infected (Staphylococcus aureus) tibia implantation rat model. Doxy released from NF coating inhibited bacterial growth up to 8 weeks in vivo. The maximal push-in force of Doxy-NF coating (38 N) is much higher than that of NF coating group (6.5 N) 8 weeks after implantation (p<0.05), which was further confirmed by quantitative histological analysis and micro computed tomography. These findings indicate that coaxial PCL/PVA NF coating doped with Doxy and/or other drugs have great potential in enhancing implant osseointegration and preventing infection.
Although complications associated with patello-femoral (PF) joint account for up to 50% of total knee replacement (TKR) revision procedures (Lee), the PF joint has been overlooked in wear simulations. The goal of this study was to develop an in vitro model to simulate patella wear in TKR’s. This report describes the concepts of an in vitro model for normal gait and the preliminary results of experimental validation. The primary consideration in the development of the current model was modeling of the in vivo kinetics and kinematics. Since the in vivo kinetics are not well documented, the current model adapted a PF joint force pattern of gait measured one year postoperatively in a telemetric distal femoral replacement (Taylor et al). The maximum force was increased from 571N to 1780N (2.5xBody Weight) to compensate for muscle deficiency and to better reflect a maximum load representative of the in vivo situation. In vivo kinematics were adopted from measurements of Lafortune. Only the PF flexion was included in the model as a simplification of the complex patella motions. The phase relationship between the kinematic and kinetic waveforms was adjusted to replicate the in vivo situation. A 6-station knee simulator carried out the experimental validation with a test frequency of 1.5Hz. The test was intended to run for 5 million cycles, with CMM wear measurements (Muratoglu et al.) taken every million cycles. The preliminary measurements showed wear patterns in the tested patellae similar to retrieved patellae. Currently there are no standards for wear testing the PF joint. The current in vitro wear model presents a useful tool to critically assess the PF joint during gait. Future work should incorporate testing for adverse loading conditions, such as PF mal-alignment, rising from a chair or deep knee flexion.
A high proportion of complications following TKR occur at the patellofemoral articulation secondary to delami-nation and adhesive/abrasive wear. Electron beam cross-linking and melting has been shown to substantially reduce delamination and adhesive/abrasive wear in polyethylene tibial inserts. A series of in-vitro patella wear and fatigue tests were developed to explore the benefits of this material at the patellofemoral articulation. Patellae (NKII, Sulzer Orthopedics, Inc., Austin, TX) were tested on an AMTI (Watertown, MA) knee simulator articulating against the trochlear grove of the femoral component. The simulator controlled flexion/ extension and patellofemoral contact force. Each test included patellae manufactured from conventional and electron beam crosslinked and melted polyethylene. Three different simulations were created: i) normal gait (5 million cycles) with optimal component alignment, ii) stair climbing (2 million cycles) with optimal component alignment, iii) stair climbing (2 million cycles) with 4° of femoral component internal rotation to simulate a component malalignment condition. In the last two simulations all patellae were artificially aged for 35 days in 80°C air to simulate one aspect of the long term oxidative state of each material. In normal gait, the unaged conventional and highly cross-linked materials demonstrated similar behaviour. In stair climbing with optimal component alignment, the aged conventional patellae developed cracks by 2 million cycles. In stair climbing with component malalign-ment the aged conventional patellae developed cracks and delamination by 1 million cycles. None of the highly cross-linked components showed cracks or delamination. These results demonstrate the potential advantage of highly cross-linked polyethylene for the patella.