A recent review of the literature on metal-on-metal total hip arthroplasties (THA) revealed the lack of comparative clinical studies with a sufficient sample size and the inclusion of patient-reported outcomes as well as patient activity levels. We conducted a prospective cohort study including all metal-on-metal and conventional polyethylene (PE)-ceramic THAs with an uncemented cup (Morscher press-fit cup), a 28mm head and operated upon via a lateral approach at our University hospital between 1/1999 and 12/2008. Only THAs for primary osteoarthritis were included. The study population is part of the Geneva Hip Arthroplasty Registry, a prospective cohort followed since 1996. The following outcomes were compared between the two groups (metal-on-metal=group 1 vs. PE-ceramic bearing=group 2): (1) Complication rates with respect to infection, dislocation and revision, (2) Radiographic outcomes (presence of linear or focal femoral osteolysis, loosening), and (3) Clinical outcomes (Harris Hip score increase, SF-12, activity and patient satisfaction evaluation, presence of groin pain). Patients operated between 1/1999 and 12/2004 were evaluated five years postoperatively by an independent assessor. Cox regression analysis was used to compare incidence rates while adjusting for differences in baseline characteristics.Introduction
Methods
None of the patients required cup revision for aseptic loosening. Complications included 12 dislocations and 3 deep infections resulting in 2 total revisions. In 3 patients the stem was revised for aseptic loosening at a mean of 63 months. At 10-years the survivorship was 98.6% (95% CI 96.7; 99.4) with endpoint revision for any cause. Mean total wear was 0.89 mm (±0.5). 32 cups (18.8%) with a cup inclination >
45° had a mean wear of 1.06 mm (±0.5), whereas 138 cups (81.2%) with inclination <
45° had a mean wear of 0.86 mm (±0.5), p=0.036. In 16 cases osteolytic defects around the stem were present. The outcome scores at 10 years were: HHS 85.9 (±14.1), WOMAC pain 70.7 (±24.7), WOMAC function 68.8 (±24.5), SF-12 physical score 40.3 (±9.2) and mental score 47.0 (±10.4). Ninety-four percent of the patients were satisfied or very satisfied.