Cephalomedullary nails are widely used for fixation of unstable pertrochanteric fractures. In 2018, the Depuy Synthes Trochanteric Fixation Nail - Advanced (TFNA) implant was introduced at a level I academic trauma center. Thereafter, the TFNA swiftly replaced the older implant models used at the time. Subsequently, clinical concerns were raised about the use of the TFNA due to reports of nail breakage. The purpose of this study was to investigate whether the concerns raised about the performance of the TFNA were valid and to assess long-term outcomes. The data consisted of 2397 patients who had undergone a proximal femoral hip fracture procedure between 2014 and 2020. Data were handpicked from patient records. TFNA was compared with TFN, PFNA, Gamma3, and Intertan regarding nail breakage, breakage time and long-term outcomes.Background
Methods
Metal particles detached from metal-on-metal hip prostheses (MoM-THA) have been shown to cause inflammation and destruction of tissues. To further explore this, we investigated the histopathology (aseptic lymphocyte-dominated vasculitis-associated lesions (ALVAL) score) and metal concentrations of the periprosthetic tissues obtained from patients who underwent revision knee arthroplasty. We also aimed to investigate whether accumulated metal debris was associated with ALVAL-type reactions in the synovium. Periprosthetic metal concentrations in the synovia and histopathological samples were analyzed from 230 patients from our institution from October 2016 to December 2019. An ordinal regression model was calculated to investigate the effect of the accumulated metals on the histopathological reaction of the synovia.Aims
Methods
Previous publications have suggested that the incidence of revisions due to infection after THA is increasing. We performed updated time-trend analyses of risk and timing of revision due to infection after primary THAs in the Nordic countries during the period 2004–2018. 569,463 primary THAs reported to the Nordic Arthroplasty Register Association from 2004 through 2018 were studied. We estimated adjusted hazard ratios (aHR) with 95% confidence interval by Cox regression with the first revision due to infection after primary THA as endpoint. The risk of revision was investigated. In addition, we explored changes in the time span from primary THA to revision due to infection.Aim
Methods
Young patients are at increased risk of revision after primary THA (THA). The bearing surface may be of importance for the longevity of the joint. We aimed to compare the risk of revision of primary stemmed cementless THA with MoM and CoC with metal-on-highly-crosslinked-polyethylene (MoXLP) bearings in patients between 20–54 years. From NARA, we included 2,153 MoM, 4,120 CoC and 10,329 MoXLP THA operated between 1995 and 2017. Kaplan-Meier estimator was used for calculation of THA survivorship and Cox regression to estimate the hazard ratio (HR) of revision (95% CI) due to any and specific causes. MoXLP was reference. The median follow-up was 10.3 years for MoM, 6.6 years for CoC and 4.8 years for MoXLP. 15 years postoperatively the Kaplan-Meier survival estimates were 80% (78–83%) for MoM, 92% (91–93%) for CoC and 94% (93–95%) for MoXLP. The 0–2, 2–7 and 7–15 years adjusted HRs of revision by any cause were 1.4 (0.9–2.4), 3.2 (2.1–5.1) and 3.9 (1.9–7.9) for MoM and 1.1 (0.8–1.4), 1.0 (0.7–1.3) and 2.5 (1.3–4.8) for CoC bearings. After 7–15 years follow-up, the unadjusted HR of revision due to aseptic loosening was 5.4 (1.2–24) for MoM and 4.2 (0.9–20) for CoC THA. MoM and CoC had a 7–15 year adjusted HR of revision due to ‘other’ causes of 4.8 (1.6–14) and 2.1 (0.8–5.8). MoXLP bearings were associated with better survival than MoM and CoC bearings, mainly because of lower risk of revision due to aseptic loosening and ‘other’ causes.
The main advantage of 3D-printed, off-the-shelf acetabular implants is the potential to promote enhanced bony fixation due to their controllable porous structure. In this study we investigated the extent of osseointegration in retrieved 3D-printed acetabular implants. We compared two groups, one made via 3D-printing (n = 7) and the other using conventional techniques (n = 7). We collected implant details, type of surgery and removal technique, patient demographics, and clinical history. Bone integration was assessed by macroscopic visual analysis, followed by sectioning to allow undecalcified histology on eight sections (~200 µm) for each implant. The outcome measures considered were area of bone attachment (%), extent of bone ingrowth (%), bone-implant contact (%), and depth of ingrowth (%), and these were quantified using a line-intercept method.Aims
Methods
The optimum clearance between the bearing surfaces of hip arthroplasties is unknown. Theoretically, to minimize wear, it is understood that clearances must be low enough to maintain optimal contact pressure and fluid film lubrication, while being large enough to allow lubricant recovery and reduce contact patch size. This study aimed to identify the relationship between diametrical clearance and volumetric wear, through the analysis of retrieved components. A total of 81 metal-on-metal Pinnacle hips paired with 12/14 stems were included in this study. Geometrical analysis was performed on each component, using coordinate and roundness measuring machines. The relationship between their as-manufactured diametrical clearance and volumetric wear was investigated. The Mann-Whitney U test and unpaired Aims
Methods
The risk for developing a periprosthetic joint infection (PJI) as a consequence of bacteremia is not clear, except for All patients with a primary knee or hip joint replacement performed between September 2002 and December 2013 in a tertiary care hospital (n=14 378) were retrospectively followed up until December 2014. The mean follow-up time was 6.0 years (range 0–12 years). Positive blood culture results of the patients during the study period were obtained. PJIs during the study period were identified from several data sources. PJIs as a consequence of bacteremia were recorded and confirmed from patient records. Primary PJIs resulting in bacteremia were excluded. Binary logistic regression with univariate analysis was used to study potential risk factors for PJI among those with bacteremia.Aim
Method
The Attune total knee arthroplasty (TKA) has been used in over 600 000 patients worldwide. Registry data show good clinical outcome; however, concerns over the cement-tibial interface have been reported. We used retrieval analysis to give further insight into this controversial topic. We examined 12 titanium (Ti) PFC Sigma implants, eight cobalt-chromium (CoCr) PFC Sigma implants, eight cobalt-chromium PFC Sigma rotating platform (RP) implants, and 11 Attune implants. We used a peer-reviewed digital imaging method to quantify the amount of cement attached to the backside of each tibial tray. We then measured: 1) the size of tibial tray thickness, tray projections, peripheral lips, and undercuts; and 2) surface roughness (Ra) on the backside and keel of the trays. Statistical analyses were performed to investigate differences between the two designs.Objectives
Methods
Previous studies have suggested that metal-on-metal (MoM) Pinnacle (DePuy Synthes, Warsaw, Indiana) hip arthroplasties implanted after 2006 exhibit higher failure rates. This was attributed to the production of implants with reduced diametrical clearances between their bearing surfaces, which, it was speculated, were outside manufacturing tolerances. This study aimed to better understand the performance of Pinnacle Systems manufactured before and after this event. A total of 92 retrieved MoM Pinnacle hips were analyzed, of which 45 were implanted before 2007, and 47 from 2007 onwards. The ‘pre-2007’ group contained 45 implants retrieved from 21 male and 24 female patients, with a median age of 61.3 years (interquartile range (IQR) 57.1 to 65.5); the ‘2007 onwards’ group contained 47 implants retrieved from 19 male and 28 female patients, with a median age of 61.8 years (IQR 58.5 to 67.8). The volume of material lost from their bearing and taper surfaces was measured using coordinate and roundness measuring machines. These outcomes were then compared statistically using linear regression models, adjusting for potentially confounding factors.Objectives
Methods
Numerous studies have reported on clinically significant volumes of material loss and corrosion at the head-stem junction of metal-on-metal (MOM) hips; less is understood about metal-on-polyethylene (MOP) hips. We compared the effect of bearing type (MOM vs MOP) on taper material loss for a hip system of a single design (DePuy Pinnacle). We recruited retrieved MOM (n=30) and MOP (n=22) bearing hips that were consecutively received at our centre. We prospectively collected associated clinical and imaging data. We measured the severity of corrosion and volumes of material loss at each head taper surface and used multivariate statistical analysis to investigate differences between the two bearing types.Introduction
Methods
Recent studies have demonstrated that implant-specific blood metal ion thresholds exist in unilateral and bilateral metal-on-metal (MoM) hip arthroplasty patients, with these thresholds being most effective for identifying patients at low-risk of adverse reactions to metal debris (ARMD). We investigated whether these new blood metal ion thresholds could effectively identify patients at risk of ARMD in an external cohort of MoM hip arthroplasty patients. We performed a validation study involving 803 MoM hip arthroplasties implanted in 710 patients at three European centres (323=unilateral Birmingham Hip Resurfacing (BHR); 93=bilateral BHR; 294=unilateral Corail-Pinnacle). All patients underwent whole blood metal ion sampling. Patients were divided into those with ARMD (revised for ARMD or ARMD on imaging; n=75), and those without ARMD (n=635). Previously devised implant-specific blood metal ion thresholds (cobalt=2.15μg/l for unilateral BHR; maximum cobalt or chromium=5.5μg/l for bilateral BHR; cobalt=3.57μg/l for unilateral Corail-Pinnacle) were applied to the validation cohort, with receiver operating characteristic curve analysis used to establish the discriminatory characteristics for each respective threshold. The area under the curve, sensitivity, specificity, positive predictive value and negative predictive value for distinguishing between patients with and without ARMD for each implant-specific threshold were respectively: unilateral BHR=89.4% (95% CI=82.8%-96.0%), 78.9%, 86.7%, 44.1%, 96.9%; bilateral BHR=89.2% (95% CI=81.3%-97.1%), 70.6%, 86.8%, 54.5%, 93.0%; unilateral Corail-Pinnacle=76.9% (95% CI=63.9%-90.0%), 65.0%, 85.4%, 24.5%, 97.1%. The 7μg/l UK MHRA threshold missed significantly more patients with ARMD compared with the implant-specific thresholds (4.9% vs. 2.8%; p=0.0003). This external multi-centre validation study has confirmed that MoM hip arthroplasty patients with blood metal ion levels below newly devised implant-specific thresholds have a low-risk of ARMD. Compared to implant-specific thresholds, the currently proposed fixed MHRA threshold missed more patients with ARMD. We recommend using implant-specific thresholds over fixed thresholds when managing MoM hip arthroplasty patients.