Abstract

Background

Cephalomedullary nails are widely used for fixation of unstable pertrochanteric fractures. In 2018, the Depuy Synthes Trochanteric Fixation Nail - Advanced (TFNA) implant was introduced at a level I academic trauma center. Thereafter, the TFNA swiftly replaced the older implant models used at the time. Subsequently, clinical concerns were raised about the use of the TFNA due to reports of nail breakage. The purpose of this study was to investigate whether the concerns raised about the performance of the TFNA were valid and to assess long-term outcomes.

Methods

The data consisted of 2397 patients who had undergone a proximal femoral hip fracture procedure between 2014 and 2020. Data were handpicked from patient records. TFNA was compared with TFN, PFNA, Gamma3, and Intertan regarding nail breakage, breakage time and long-term outcomes.

Results

After exclusion a total of 23/1667 (1.4%) nails broke during the follow-up period. The TFNA broke the most often with 15 cases (2.0%), followed by the Gamma3 with five cases (1.1 %) and the PFNA with three cases (1.3%). Overall, the mean (SD) nail breakage time was 233 (147.8) days. However, for the TFNA, PFNA, and Gamma3, the mean breakage times were 176.8 days (109.9), 419 days (108.6), and 291.8 (153.4), respectively. In cox regression analysis we observed significant reduction in nail breakage when using PFNA with adjusted hazard risk of 0.081 [95% Ci, 0.011-0.576, p=0.011].

Conclusions

In our data, the TFNA had a slightly higher risk for nail breakage when compared to the PFNA and the Gamma3, with a risk difference of 0.7% and 0.9%, respectively. On average, the TFNA broke nearly four months earlier than the Gamma3 and more than eight months earlier than the PFNA. It should be noted, however, that implant breakage is a relatively infrequent complication.


Corresponding author: Leevi Karjalainen