Bone & Joint Research
Vol. 8, Issue 3 | Pages 146 - 155
1 Mar 2019
Langton DJ Natu S Harrington CF Bowsher JG Nargol AVF

Objectives

We investigated the reliability of the cobalt-chromium (CoCr) synovial joint fluid ratio (JFR) in identifying the presence of a severe aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL) response and/or suboptimal taper performance (SOTP) following metal-on-metal (MoM) hip arthroplasty. We then examined the possibility that the CoCr JFR may influence the serum partitioning of Co and Cr.

Methods

For part A, we included all revision surgeries carried out at our unit with the relevant data, including volumetric wear analysis, joint fluid (JF) Co and Cr concentrations, and ALVAL grade (n = 315). Receiver operating characteristic curves were constructed to assess the reliability of the CoCr JFR in identifying severe ALVAL and/or SOTP. For part B, we included only patients with unilateral prostheses who had given matched serum and whole blood samples for Co and Cr analysis (n = 155). Multiple regression was used to examine the influence of JF concentrations on the serum partitioning of Co and Cr in the blood.


Bone & Joint Research
Vol. 7, Issue 6 | Pages 388 - 396
1 Jun 2018
Langton DJ Sidaginamale RP Joyce TJ Bowsher JG Holland JP Deehan D Nargol AVF Natu S

Objectives

We have encountered patients who developed large joint fluid collections with massive elevations in chromium (Cr) and cobalt (Co) concentrations following metal-on-metal (MoM) hip arthroplasties. In some cases, retrieval analysis determined that these ion concentrations could not be explained simply by the wear rates of the components. We hypothesized that these effects may be associated with aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL).

Patients and Methods

We examined the influence of the ALVAL grade on synovial fluid Co and Cr concentrations following adjustment for patient and device variables, including volumetric wear rates. Initially restricting the analysis to include only patients with one MoM hip resurfacing device, we performed multiple regression analyses of prospectively collected data. We then repeated the same statistical approach using results from a larger cohort with different MoM designs, including total hip arthroplasties.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_1 | Pages 87 - 87
1 Jan 2016
Clarke I Sufficool D Bowsher JG Savisaar C Burgett-Moreno M Donaldson T
Full Access

Introduction

Hip simulators proved to be valuable, pre-clinical tests for assessing wear. Preferred implant positioning has been with cup mounted above head, i.e. ‘Anatomical’ (Figs. 1a-c) 1,2 while the ‘Inverted’ test (cup below head) was typically preferred in debris studies (Figs. 1d-f).3,4 In an Anatomical study, wear patterns on cups and heads averaged 442 and 1668 mm² area, respectively, representing 8% and 30% of available hemi-surface (Table 1), i.e. the head pattern was ×3.8 times larger than cup. This concept of wear patterns is illustrated well in the ‘pin-on-disk’ test (Fig. 1) in which the oscillating pin has the ‘contained’ wear area (CWP) and the large wear track on the disk is the ‘distributed’ pattern (DWP). Hip simulators also create CWP and DWP patterns, site dependant on whether Anatomical (Fig. 1a-c) or ‘Inverted’ (Fig. 1d-f) test. However there is scant foundation as to clinical merits of either test mode. Retrieval studies of MOM bearings have indicated that cups have the larger wear patterns, i.e. contrary to simulator tests running Anatomical cups (Table 1).5 Therefore we compared Anatomical and Inverted cup modes using 38mm and 40mm MOM in two 5-million cycle simulator studies.

Methods

38mm and 40mm MOM bearings were run in Anatomical mode (study-1) and Inverted (study-2) mode, respectively, in a hip simulator. Lubricant was bovine serum diluted to provide protein concentration 17 mg/ml. Wear was measured gravimetrically and wear-rates calculated by linear regression. Wear patterns were assessed by stereomicroscopy and compared to algorithms using standard spherical equations.


Bone & Joint Research
Vol. 4, Issue 3 | Pages 29 - 37
1 Mar 2015
Halim T Clarke IC Burgett-Moreno MD Donaldson TK Savisaar C Bowsher JG

Objectives

Third-body wear is believed to be one trigger for adverse results with metal-on-metal (MOM) bearings. Impingement and subluxation may release metal particles from MOM replacements. We therefore challenged MOM bearings with relevant debris types of cobalt–chrome alloy (CoCr), titanium alloy (Ti6Al4V) and polymethylmethacrylate bone cement (PMMA).

Methods

Cement flakes (PMMA), CoCr and Ti6Al4V particles (size range 5 µm to 400 µm) were run in a MOM wear simulation. Debris allotments (5 mg) were inserted at ten intervals during the five million cycle (5 Mc) test.