Advertisement for orthosearch.org.uk
Results 1 - 8 of 8
Results per page:
Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_6 | Pages 133 - 133
1 Jul 2020
Hardesty C Janas M
Full Access

A growing number of children are being implanted with magnetically controlled growing rods. Some also have baclofen pumps, a device that delivers a neuromodulating medication using a magnet. There has been no previous testing to determine if the External Remote Controller (ERC) would stall the baclofen pump during magnetic rod lengthening, a potentially life-threatening event. Two baclofen pumps, 20 mL and 40 mL, were placed in varying proximity to the ERC while the magnetic field was generated. The speed (dose) of the baclofen pump, the distance from the ERC, the length of time the ERC was run, the position of the pump in relationship to the magnetic field, and the presence or absence of a stall were recorded. To stall a baclofen pump, the ERC needed to be running continuously for at least 30 mm of lengthening (about 3 minutes). The pump would not stall if it was more than 1 cm away from the ERC, even if the ERC generated a field for 3 minutes. The speed of the pump did not affect the instance of a motor stall. The pump was more likely to stall if it was positioned in front of or behind the magnetic field, as opposed to the center of the field. Baclofen pumps are always more than 1 cm away from the magnetic rods when both are implanted in a child. The typical lengthening interval is under 10 mm, much less than what was required to stall a baclofen pump. The ERC is highly unlikely to stall a baclofen pump during a magnetic rod lengthening, making a life-threatening event almost impossible


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_3 | Pages 31 - 31
1 Mar 2021
Sun M Buckler N AlNouri M Vaughan M Hilaire TS Sponseller P Smith J Thompson G Howard J El-Hawary R
Full Access

Scoliosis is estimated to occur between 21–64% of patients with cerebral palsy (CP), where a subset of patients develops early onset scoliosis (EOS) before the age of ten. Traditional growth friendly (TGF) surgeries in the context of traditional growing rods have been shown to be effective in treating scoliosis in this population, however significant complication rates are reported. Currently, no studies have been done to examine the effect of novel growth friendly surgeries such as magnetically controlled growing rods (MCGR) on EOS in CP patients. The objective of this study is to compare MCGR with TGF surgeries in this patient population, specifically by evaluating radiographic measurements and risk of unplanned reoperations (UPRORs). Patients with EOS secondary to CP were prospectively identified from an international database, with data retrospectively analyzed. Scoliosis (primary curve), maximum kyphosis, T1-S1 and T1-T12 height were measured pre-operation, immediate post-operation, and at two-years follow-up. The risk and etiology of UPRORs were compared between MCGR and TGF. P < 0.05 was considered statistically significant for all analyses. Of the 120 patients that met inclusion criteria, 86 received TGF (age 7.5 ± 1.8 years; follow-up 7.0 ± 2.9 years) and 34 received MCGR (age 7.1 ± 2.2 years, follow-up 2.8 ± 0.5 years). Compared to TGF, MCGR resulted in significant improvements in maintenance of scoliosis correction (p=0.04). At final follow-up, UPRORs were 24% for MCGR (8/34 patients) and 43% (37/86 patients) for TGF (p=0.05). To minimize the influence of follow-up period, UPRORs within the first two years post-operation were evaluated: MCGR (21%, 7/34 patients) vs. TGF (14%, 12/86 patients; p=0.37). Within the first two years, etiology of UPROR as a percentage of all patients per group were deep infection (5% TGF, 6% MCGR), implant failure/migration (5% TGF, 9% MCGR), dehiscence (2% TGF, 3% MCGR), and superficial infection (1% TGF, 3% MCGR). The most common etiology of UPROR for TGF was deep infection and implant failure/migration and for MCGR was implant failure/migration. For patients with CP, at final follow-up, MCGR had superior maintenance of scoliosis correction; however, there was no difference in risk of UPROR within the first two years post-operatively (21% MCGR, 14% TGF)


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_15 | Pages 10 - 10
1 Oct 2014
Cheung KMC Cheung JPY Kwan K Ferguson J Nnadi C Alanay A Yazici M Demirkiran G Helenius IJ Akbarnia B
Full Access

The magnetically controlled growing rod (MCGR) system allows growth maintenance without the risk of anaesthesia, implant and wound complications associated with repeated surgeries. This is a medium-term report of the complications of MCGR from a multicentre study. Twenty-six patients from 6 spine institutes that are part of a multicentre study with prospectively collected data of minimum 24 months follow-up were assessed. Pre-operative, immediate post-operative and most recent spine radiographs were reviewed to measure the Cobb angle and the rod lengthening distance. The causes and any associated risk factors for re-operations were examined. Eleven patients required re-operation within the follow-up period, with a mean time to re-operation of 17 months after the initial surgery. Five were due to failure of rod distractions; 3 were due to failure of proximal foundation implants; 2 were due to rod breakage; and one case of superficial wound infection with failure of proximal fixation. Proximal junctional kyphosis occurred in 5 patients. Three had proximal anchor dislodgement and all five constructs were revised. This is the largest series with the longest follow-up to date. Our series show that the perception that using MCGR may reduce the frequency of re-operations may not be entirely true. This is the first report to examine the need for re-operation after MCGR implantation, and highlights the inherent risks of any surgical treatment in this group of patients despite the advantages of this new implant. Longer-term studies and comparisons with traditional growing rods are required


Abstract

MAGnetic Expansion Control (MAGEC) rods are used in the surgical treatment of children with early onset scoliosis. The magnetically controlled lengthening mechanism enables rod distractions without the need for repeated invasive surgery. The CE certification of these devices was suspended in March 2021 due, primarily, to performance evidence gaps in the documents provided by the manufacturer to regulators and notified bodies. MAGEC rods are therefore not permitted for use in countries requiring CE marking. This was a survey of 18 MAGEC rod surgeons in the UK about their perception of the impact of the CE suspension on the clinical management of their patients. Unsurprisingly, virtually all perceived a negative impact, reflecting the complexity of this patient group. Reassuringly, these surgeons are highly experienced in alternative treatment methods.

Cite this article: Bone Jt Open 2022;3(2):155–157.


Bone & Joint Open
Vol. 3, Issue 1 | Pages 85 - 92
27 Jan 2022
Loughenbury PR Tsirikos AI

The development of spinal deformity in children with underlying neurodisability can affect their ability to function and impact on their quality of life, as well as compromise provision of nursing care. Patients with neuromuscular spinal deformity are among the most challenging due to the number and complexity of medical comorbidities that increase the risk for severe intraoperative or postoperative complications. A multidisciplinary approach is mandatory at every stage to ensure that all nonoperative measures have been applied, and that the treatment goals have been clearly defined and agreed with the family. This will involve input from multiple specialities, including allied healthcare professionals, such as physiotherapists and wheelchair services. Surgery should be considered when there is significant impact on the patients’ quality of life, which is usually due to poor sitting balance, back or costo-pelvic pain, respiratory complications, or problems with self-care and feeding. Meticulous preoperative assessment is required, along with careful consideration of the nature of the deformity and the problems that it is causing. Surgery can achieve good curve correction and results in high levels of satisfaction from the patients and their caregivers. Modern modular posterior instrumentation systems allow an effective deformity correction. However, the risks of surgery remain high, and involvement of the family at all stages of decision-making is required in order to balance the risks and anticipated gains of the procedure, and to select those patients who can mostly benefit from spinal correction.


Bone & Joint Open
Vol. 1, Issue 7 | Pages 405 - 414
15 Jul 2020
Abdelaal A Munigangaiah S Trivedi J Davidson N

Aims

Magnetically controlled growing rods (MCGR) have been gaining popularity in the management of early-onset scoliosis (EOS) over the past decade. We present our experience with the first 44 MCGR consecutive cases treated at our institution.

Methods

This is a retrospective review of consecutive cases of MCGR performed in our institution between 2012 and 2018. This cohort consisted of 44 children (25 females and 19 males), with a mean age of 7.9 years (3.7 to 13.6). There were 41 primary cases and three revisions from other rod systems. The majority (38 children) had dual rods. The group represents a mixed aetiology including idiopathic (20), neuromuscular (13), syndromic (9), and congenital (2). The mean follow-up was 4.1 years, with a minimum of two years. Nine children graduated to definitive fusion. We evaluated radiological parameters of deformity correction (Cobb angle), and spinal growth (T1-T12 and T1-S1 heights), as well as complications during the course of treatment.


Bone & Joint 360
Vol. 7, Issue 1 | Pages 32 - 35
1 Feb 2018


Bone & Joint 360
Vol. 1, Issue 5 | Pages 28 - 30
1 Oct 2012

The October 2012 Children’s orthopaedics Roundup360 looks at: magnetic growing rods and scoliosis correction; maintaining alignment after manipulation of a radial shaft fracture; Glaswegian children and swellings of obscure origin; long-term outcome of femoral derotation osteotomy in cerebral palsy; lower-leg fractures and compartment syndrome in children; fractures of the radial neck in children; management of the paediatric Monteggia fracture; and missing the dislocated hip in Western Australia.