Aims. The aim of this study was to compare a bicruciate-retaining (BCR) total knee arthroplasty (TKA) with a posterior
The posterior tibial slope angle (PTS) in posterior cruciate retaining total knee arthroplasty influences the knee kinematics, knee stability, flexion gap, knee range of motion (ROM) and the tension of the posterior cruciate ligament (PCL). The current technique of using an arbitrary (often 3–5 degrees) PTS in all cases seldom will restore native slope in cruciate retaining TKA. Questions/Purposes: The primary objective was to determine if we could surgically reproduce the native PTS in
Background. To prevent excessive tension on the posterior cruciate ligament (PCL) in
Purpose. This study was to investigate the effect of posterior tibial slope (PTS) on the kinematics in the
The aim of the study was to compare clinical results after
Introduction: In general a loose-balanced total knee arthroplasty (TKA) are believed to gain good postoperative range of motion (ROM), however, too much laxity is thought to be a cause for persistent pain and catastrophic long-term results. Materials and methods: We measured the antero-posterior and medio-lateral laxity to evaluate the influence of stability after
Background. In recent literatures, medial instability after TKA was reported to deteriorate early postoperative pain relief and have negative effects on functional outcome. Furthermore, lateral laxity of the knee is physiological, necessary for medial pivot knee kinematics, and important for postoperative knee flexion angle after
Background. The constraint of total knee replacement (TKR) implants is not simply defined and many of the factors that influence it are not well understood. Variability in the constraint of different TKR implants designed for the same indication (e.g. cruciate-retaining, or posterior-stabilized) have been previously demonstrated, but these differences among implants have yet to be simply quantified. Furthermore, the relative importance of several variables on the implant constraint remains unknown. The purpose of this study was to quantify the differences in constraint that exist between different implant designs, and to examine the effects of axial load and flexion angle on the constraint of current
Introduction. Tibial slope was shown to majorly affect the outcomes of Total Knee Arthroplasty (TKA). More slope of the tibial component could help releasing a too tight flexion gap in
Introduction / Purpose. Many factors can influence postoperative knee flexion angle after total knee arthroplasty (TKA), and range of flexion is one of the most important clinical outcomes. Although many studies have reported that postoperative knee flexion is influenced by preoperative clinical conditions, the factors which affect postoperative knee flexion angle have not been fully elucidated. As appropriate soft-tissue balancing as well as accurate bony cuts and implantation has traditionally been the focus of TKA success, in this study, we tried to investigate the influence of intraoperative soft-tissue balance on postoperative knee flexion angle after
Purpose. Surgeons sometimes encounter moderate or severe varus deformed osteoarthritic cases in which medial substantial release including semimembranosus is compelled to appropriately balance soft tissues in total knee arthroplasty (TKA). However, medial stability after TKA is important for acquisition of proper knee kinematics to lead to medial pivot motion during knee flexion. The purpose of the present study is to prove the hypothesis that step by step medial release, especially semimembranosus release, reduces medial stability in
A bicruciate retaining (BCR) TKA is thought to maintain a closer resemblance to the native knee kinematics compared to a posterior cruciate retaining (CR) TKA. With BCR TKAs retainment of the anterior cruciate ligament (ACL) facilitates proprioception and balance which is thought to lead to more natural knee kinematics and increased functional outcome. The aim of this study was to quantify and compare the kinematics of a BCR and CR TKA during functional tests. In this patient-blinded randomized controlled trial, a total of 40 patients with knee osteoarthritis were included, 18 of them received a BCR TKA (Vanguard XP, Zimmer-Biomet) and 22 received a CR TKA (Vanguard CR, Zimmer-Biomet). Fluoroscopic analysis was done 1 year post-operatively. The main outcome was posterior femoral rollback (i.e. translation of the femorotibial contact point (CP)) of the BCR and CR TKA during a step-up test. Secondary, the kinematics during a lunge test were quantified as anterior-posterior (AP) translation of the femorotibial CP. Independent student t-tests (or non-parametric equivalent) were used to analyze the effect of BCR versus CR TKA on these measures, to correct for the multiple testing problem post-hoc Bonferroni-Holm corrections were applied.Introduction
Materials and Methods
Introduction. Appropriate intraoperative soft tissue balancing is recognized to be essential in total knee arthroplasty (TKA). However, it has been rarely reported whether intraoperative soft tissue balance reflects postoperative outcomes. In this study, we therefore assessed the relationship between the intra-operative soft tissue balance measurements and the post-operative stress radiographs at a minimum 1-year follow-up in
Although the pre- or intraoperative flexion angle in TKA has been commonly considered as a predictor of the postoperative flexion angle, patients with well flexion intraoperatively cannot necessarily obtain deep flexion angle postoperatively. The reason why inconsistencies remains has been unsolved. The intraoperative compressive force between femoral and tibial components has the advantage of the sequential changes during knee motion. However, the relationship between the compressive force and the postoperative ROM has not yet been clarified. We aimed to evaluate the intraoperative femorotibial compressive force during passive knee motion, and determine the relationship between the compressive force and the postoperative flexion angle. A total of 11 knees in 10 patients who underwent primary
Maintaining posterior stability in total knee arthroplasty (TKA) may be achieved by using a posterior stabilized TKA, retaining and balancing the posterior cruciate ligament (PCL) using a traditional cruciate-retaining design (CR), or by increasing the sagittal plane conformity of the tibial insert. In the latter case, stability is achieved by the addition of an anterior buildup on the tibial polyethylene creating the so-called “anterior stabilized” (AS) design. We hypothesized that using an AS tibial insert would provide similar function and survivorship as compared to using a more traditional CR bearing when the PCL is either recessed or balanced. Between 2004 and 2016, 1,731 modular CR TKAs were implanted in 1,509 patients using the same CR TKA design. The diagnosis was osteoarthritis in 98%. 58% of patients were female. Average age of 64.9 years. Within this group, 868 TKAs (50.1%) had a standard CR tibial bearing (3-degree posterior slope and no posterior lip) implanted (CR-S). 480 TKAs (27.8%) had a lipped CR modular tibial bearing (2.5 mm elevated posterior lip) implanted (CR-L). Starting in 2013, 383 TKAs (22.1%) were implanted with an AS modular tibial bearing (9–11 mm anterior lip and a 5 mm posterior lip). If the PCL was considered non-functional or absent, an AS bearing was placed. If the PCL was considered functional, a standard bearing or lipped bearing was used. Clinical and radiographic analysis was performed according to the Knee Society (KS) grading system. The most recent clinical and radiographic evaluation was used for post-operative analysis. The average follow-up in the entire cohort of TKAs was 5.5 years (range 2 to 14.3 years). Kaplan-Meier analysis was used to determine prosthesis survivorship with failure defined as aseptic loosening of the prosthesis (with or without revision) or tibial insert exchange.Introduction
Methods
Introduction. Total knee arthroplasty (TKA) with a computer-assisted navigation system has been developed to improve the accuracy of the alignment of osteotomies and implantations. One of the most important goals of TKA is to improve the flexion angle. Although accurate soft tissue balancing has been recognized as an essential surgical intervention influencing flexion angle, the direct relationship between post-operative flexion angle and intra-operative soft tissue balance during TKA, has little been clarified. In the present study, therefore, we focused on the relationship between them in
There is little information available to surgeons regarding how the lateral soft-tissue structures prevent instability in knees implanted with total knee arthroplasty (TKA). The aim of this study was to quantify the lateral soft-tissue contributions to stability following cruciate retaining (CR) TKA. Nine cadaveric knees with CR TKA implants (PFC Sigma; DePuy Synthes Joint Reconstruction) were tested in a robotic system (Fig. 1) at full extension, 30°, 60°, and 90° flexion angles. ±90 N anterior-posterior force, ±8 Nm varus-valgus and ±5 Nm internal-external torque were applied at each flexion angle. The anterolateral structures (ALS, including the iliotibial band, anterolateral ligament and anterolateral capsule), the lateral collateral ligament (LCL), the popliteus tendon complex (Pop T) and the posterior cruciate ligament (PCL) were then sequentially transected. After each transection the kinematics obtained from the original loads were replayed, and the decrease in force / moment equated to the relative contributions of each soft-tissue to stabilising the applied loads.Introduction
Methods
Posterior cruciate ligament (PCL) preservation in total knee arthroplasty (TKA) is adovocated on the grounds that it provides better restoration of knee joint kinematics as opposed to PCL sacrifice. Mobile-bearing (MB) total knee prostheses have been in the market for a long time, but the PFC-Sigma Rotating Platform (RP) prosthesis (DePuy Orthopaedics, Inc, Warsaw, Ind) has been introduced in the market since 2000. Since, little is known about the in vivo kinematics of MB prostheses especially with cruciate retaining (CR). The objective of this study is to investigate the in vivo kinematics of MB RP-CR total knee arthroplasty during weight-bearing deep knee bending motion. We investigated the in vivo knee kinematics of 20 knees (17 patients) implanted with PFC-Sigma RP-CR. All TKAs were judged clinically successful (Hospital for Special Surgery scores >90), with no ligamentous laxity or pain. Mean patient age at the time of operation was 78.0 ± 6.0 years. Mean period between operation and surveillance was 15.0 ± 9.0 months. Under fluoroscopic surveillance, each patient did a wight-bearing deep knee bending motion. Femorotibial motion was analyzed using 2D/3D registration technique, which uses computer-assisted design (CAD) models to reproduce the spatial position of the femoral, tibial components from single-view fluoroscopic images. We evaluated the range of motion, axial rotation, and antero-posterior (AP) translation of the nearest point between the femoral and tibial component.Introduction
Patients and methods
Bicruciate-retaining (BiCR) total knee replacements (TKRs) were designed to improve implant performance; however, functional advantages during daily activity have yet to be demonstrated. Although level walking is a common way to analyze functionality, it has been shown to be a weak test for identifying gait abnormalities related to ACL pathologies. The goal of this study is to set up a functional motion analysis test that will examine the effects of the ACL in TKR patients by comparing knee kinematics, kinetics, and muscle activation patterns during level and downhill walking for patients with posterior-cruciate retaining (PCR) and BiCR TKRs. Motion and electromyography (EMG) data were collected simultaneously for 12 subjects (4/8 m/f, 64±11 years, 31.3±7.3 BMI, 6/6 right/left) with BiCR TKRs and 15 subjects (6/9 m/f, 67±7 years, 30.5±5.1 BMI, 4/11 right/left) with PCR TKRs during level and downhill walking using the point cluster marker set. Surface electrodes were placed on the vastus medialis obliquus (VMO), rectus femoris (RF), biceps femoris (BF), and semitendinosus (ST) muscles. EMG data are reported as percent relative voluntary contraction (%RVC), normalizing the signal during downhill walking to the mean maximum EMG value during level walking.Introduction
Methods
Wear debris from polyethylene tibial inserts has been associated with limited longevity of total knee replacements (TKRs). While material factors were studied extensively and considerable progress has been made, there is little knowledge about surgical factors, particularly on how the wear rate is related to implant positioning. It was the purpose of this study to determine the combined effect of patient and implant positioning factors on the volumetric wear rate of TKRs. Our hypothesis was that implant alignment has a significant impact on the wear rate when controlled for other patient factors. This study included 59 tibial inserts of a cruciate retaining TKR design (Nexgen, Zimmer Inc.). The patients' age, sex, weight, height, and implant size were obtained. All implants were scanned with a coordinate measuring machine. Volumetric wear was determined using an autonomous mathematical reconstruction method (Figure 1). Radiographs were used to determine the anatomic lateral distal femoral angle (aLDFA), anatomic medial proximal tibial angle (aMPTA), femoral tilt angle (FTA) and posterior tibial slope (PTS). Also, the patella position was assessed using the Blackburne-Peel Index (BPI) and the Insall-Salvati Ratio (Figure 2). General linear modeling (SPSS) was conducted in order to determine the most significant patient and implant positioning factors on wear rate.Introduction
Methods