Aims. The
The Exeter totally collarless, double-tapered femoral component was developed in 1969 jointly at the School of Engineering at the University of Exeter and the old Princess Elizabeth Orthopaedic Hospital. At the time, in common with a number of implants in contemporary use, the new
Introduction: The design philosophy of polished tapered THR stems, such as the Exeter, intend for them to migrate distally within the cement mantle. In addition it is likely that micromotion occurs as a result of functional activity. The pattern of induced stresses will be a function of stem geometry &
surface finish, as well as applied loading. Aim: To investigate the stresses induced in the cement mantle of a polished tapered THR stem during functional activity. Method: Using Roentgen Stereophotogrammetric Analysis (RSA) dynamically induced micro-motion (DIMM) was measured in 21 patients implanted with
The
Purpose.
To determine if any notable differences between a cemented Thompson stem hemiarthroplasty and a cemented
Introduction. Despite being from different manufacturers,
Introduction. Varus alignment of the femoral component in total hip arthroplasty (THA) is thought to be a risk factor for implant loosening and early revision surgery. The purpose of this study was to evaluate whether the
Femoral stem varus has been associated with poorer results. We report the incidence of varus/valgus malalignment of the Exeter polished, double taper design in a multicentre prospective study. The surgical outcomes at a minimum of five year and complication rates are also reported. A multicentre prospective study of 987 total hip replacements was undertaken to investigate whether there is an association between surgical outcome and femoral stem malalignment. The primary outcome measure was the change in the Oxford hip score (OHS) at five years. Secondary outcomes included the rate of dislocation and revision. The incidence of varus and valgus malignment were 7.1% and 2.6% respectively. There was no significant difference in OHS between neutral and malaligned femoral stems at 5 years (neutral, mean = 40.2; varus, mean 39.3, p = 0.465; valgus, mean = 40.9, p = 0.605). There was no significant difference in dislocation rate between the groups (p = 0.66). There was also no significant difference in revision rate (p = 0.34). This study provides evidence that the
The
Bone impaction grafting of the femur is associated with more complications when segmental defects are present. The effect of segmental defect repair on initial stem stability was studied in an in vitro study with fresh frozen goat femora. A standardized medial segmental defect was reconstructed using a cortical strut or a metal mesh. As controls we used intact femora and femora with a non-reconstructed defect. In all four groups impacted bone grafting was performed in combination with a cemented
Introduction: Implantation of a THA changes the strain distribution pattern in the proximal femur, with a massive loss of stress in the proximal part of the femur and an increase in stress at the distal part of the femoral component. Aim: The purpose of this study was to quantify the changes in BMD during long-term follow-up, i.e. five years, after insertion of the collarless, two-side conical, cemented
Recent NICE guidance recommends use of a well proven cemented femoral stem for hip hemiarthroplasty in management of fractured neck of femur. The
Roentgen Stereophotogrammetric Analysis (RSA) can predict long-term outcome of prostheses by measuring migration over time. The
Introduction: We describe our experience with the Exeter femoral component in a District General Hospital. We implanted 230
Background.
Introduction: Between 1988 and 1993 we performed 154 primary hip arthroplasties with the Exeter Universal femoral stem and a variety of acetabular components and describe our experience after a minimum follow up of 10.5 years (mean 12.6 years). 65 hips were assessed clinically and radiologically. Our aim was to establish whether results from the inventing centre for the
The
Implant choice was changed from cemented Thompson to
Introduction and aims: To study the survivorship and subsidence patterns of the first 433