Bone & Joint Open
Vol. 2, Issue 6 | Pages 443 - 456
28 Jun 2021
Thompson JW Corbett J Bye D Jones A Tissingh EK Nolan J

Aims. The Exeter V40 cemented polished tapered stem system has demonstrated excellent long-term outcomes. This paper presents a systematic review of the existing literature and reports on a large case series comparing implant fractures between the Exeter V40 series; 125 mm and conventional length stem systems. Methods. A systematic literature search was performed adhering to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) criteria. In parallel, we performed a retrospective single centre study of Exeter V40 femoral stem prosthetic fractures between April 2003 and June 2020. Results. There are 25 reported cases of such prosthetic fractures confined to small case series and case reports within the literature. We report an additional 19 cases to the literature (mean age 66.3 years (SD 11.7); 12 (63%) females; BMI 32.9 kg/m. 2. (SD 5.9)). The mean time from index procedure to fracture was 7.8 years (SD 3.6; 2.5 to 16.3). Exeter V40 stem fracture incidence was 0.15% and 1.21% for primary and revision arthroplasty, respectively. Incidence was significantly higher in revision arthroplasty (p < 0.001) and 125 mm length stems compared to ≥ 150 mm length stems (1.25% vs 0.13%, respectively; p < 0.001). When comparing different stem length cohorts, 125 mm short-stems were associated with stem-body fractures (92% vs 29%; p = 0.0095), earlier time to fracture (6.2 years vs 11.0 years; p = 0.0018), younger patient age at time of fracture (62.7 years vs 72.6 years; p = 0.037), and female sex (75% vs 43%; p = 0.326). Conclusion. This complication remains rare, although we report a significantly higher incidence at up to 17 years follow-up than in the literature. Short 125 mm length Exeter V40 stems undoubtedly have a role in restoring anatomy and biomechanics in smaller femoral geometries, although the surgeon has to appreciate the higher risk of stem fracture and the associated predisposing factors which may necessitate particular attention to surgical technique and planning. Cite this article: Bone Jt Open 2021;2(6):443–456


Bone & Joint Open
Vol. 1, Issue 12 | Pages 743 - 748
1 Dec 2020
Mahon J McCarthy CJ Sheridan GA Cashman JP O'Byrne JM Kenny P

Aims. The Exeter V40 cemented femoral stem was first introduced in 2000. The largest single-centre analysis of this implant to date was published in 2018 by Westerman et al. Excellent results were reported at a minimum of ten years for the first 540 cases performed at the designer centre in the Exeter NHS Trust, with stem survivorship of 96.8%. The aim of this current study is to report long-term outcomes and survivorship for the Exeter V40 stem in a non-designer centre. Methods. All patients undergoing primary total hip arthroplasty using the Exeter V40 femoral stem between 1 January 2005 and 31 January 2010 were eligible for inclusion. Data were collected prospectively, with routine follow-up at six to 12 months, two years, five years, and ten years. Functional outcomes were assessed using Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) scores. Outcome measures included data on all components in situ beyond ten years, death occurring within ten years with components in situ, and all-cause revision surgery. Results. A total of 829 stems in 745 patients were included in the dataset; 155 patients (20.8%) died within ten years, and of the remaining 664 stems, 648 stems (97.6%) remained in situ beyond ten years. For the 21 patients (2.5%) undergoing revision surgery, 16 femoral stems (1.9%) were revised and 18 acetabular components (2.2%) were revised. Indications for revision in order of decreasing frequency were infection (n = 6), pain (n = 6), aseptic component loosening (n = 3), periprosthetic fracture (n = 3), recurrent dislocation (n = 2), and noise production (ceramic-on-ceramic squeak) (n = 1). One patient was revised for aseptic stem loosening. The mean preoperative WOMAC score was 61 (SD 15.9) with a mean postoperative score of 20.4 (SD 19.3) (n = 732; 88.3%). Conclusion. The Exeter V40 cemented femoral stem demonstrates excellent functional outcomes and survival when used in a high volume non-designer centre. Outcomes are comparable to those of its serially validated predecessor, the Exeter Universal stem. Cite this article: Bone Jt Open 2020;1-12:743–748


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_4 | Pages 25 - 25
1 Apr 2022
Evans J Salar O Whitehouse S Kassam A Howell J Wilson M Timperley J Sayers A Whitehouse M Wilton T Hubble M
Full Access

The Exeter V40 femoral stem is the most implanted stem in the NJR for primary THA. In 2004, the 44/00/125 stem was released for use in “cement-in-cement” revision cases. It has however been used ‘off-label’ as a primary stem when, for example, patient anatomy requires a smaller stem with a 44mm offset. We aimed to investigate survival of this stem in comparison to others in the range when used in primary THAs recorded in the NJR. Analyses were performed using a dataset based on that used for the 2020 NJR annual report. Our exposure was the stem; the outcome was all-cause construct revision. Crude analyses were performed using Kaplan-Meier and adjusted using Cox models. The 44/00/125 stem was directly compared to other stems in the Exeter range. We analysed 330,732 primary THAs using the Exeter V40 stem comprising 34.5% of the 958,869 THAs with complete information from the start of the NJR to 31 December 2018. The 44/00/125 stem was implanted in 2,158 primary THAs with 67.5% in female patients and a mean age of 67.8. The 10-year revision estimate for the 44/00/125 stem was 4.9% (95%CI 3.6, 6.8) and in constructs using an Exeter V40 stem was 2.8% (95%CI 2.7, 2.8). Controlling for age, sex and ASA demonstrated an increased overall hazard of revision for constructs using the 44/00/125 stem compared to constructs using other Exeter V40 femoral stems (HR 1.8 (95%CI 1.4, 2.3)). Although the revision estimate is within the NICE 10-year benchmark, survivorship of constructs using the 44/00/125 stem appears to be lower than the rest of the Exeter V40 range. Attempts to control for age, sex and ASA will not take into account confounding by indication i.e. patients with more complex anatomy who may have a higher risk of revision. Surgeons and patients should be reassured by this but should be aware of the observed increased revision estimate and use the stem according to its indications


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_11 | Pages 31 - 31
1 Jun 2016
Westerman R Whitehouse S Howell J Hubble M Timperley A Wilson M
Full Access

Introduction. The Exeter Hip femoral component remains largely unchanged from the original design, introduced in 1970. It is a highly polished, modular, double tapered stem and has undergone various minor modifications to surface, modularity and most recently the taper; changed to the current V40. TM. design in 2000. The effect of any design modification cannot easily be foreseen and greater emphasis is now placed on ensuring appropriate monitoring for such implants. Methods. We present the results of the first 540 V40. TM. Exeter THAs performed in our Centre between December 2000 and May 2002. All patients were reviewed prospectively at 1, 5 and 10 years following surgery. Results. Ten years after surgery there are no revisions for aseptic loosening and no evidence of aseptic loosening in any hip. The fate of every implant is known and all patients remain under review. One hundred and forty-five patients (26.9%) died before 10 years and of the remaining 395 stems, 374 (94.7%) remain in situ, 11 were revised (2.7%). Three were revised for infection, one for recurrent dislocation and one implant failed at the neck of the prosthesis. Ten were exchanged using cement-in-cement techniques to aid acetabular revision. There were ten periprosthetic fractures: six requiring stem revision, two underwent ORIF and two had already healed by the time the patients presented for follow-up. There was no radiographic evidence of failure preceding the periprosthetic fractures. All radiographs and pain scores were analysed as part of long-term follow-up, and are discussed. Discussion. There were no revisions for aseptic loosening. Previous studies from this centre reviewing the Original and Universal series have demonstrated the stem to provide a predictable and reproducible long term outcome with 100% survival well beyond10 years for aseptic loosening. Conclusion. The polished V40. TM. Exeter stem continues to perform well and survival has remained comparable to the Universal series


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_10 | Pages 5 - 5
1 Aug 2021
Thompson J Corbett J Bye D Jones A Tissingh E Nolan J
Full Access

The Exeter. ™. V40 cemented polished tapered stem system has demonstrated excellent long-term outcomes. This paper presents a systematic review of the existing literature and reports on a large case series comparing implant fractures between the Exeter V40 series; 125 mm and conventional length stem systems. A systematic literature search was performed adhering to PRISMA criteria. In parallel we performed a retrospective single centre study of Exeter. ™. V40 femoral stem prosthetic fractures between April 2003– June 2020. There are 25 reported cases of such prosthetic fractures confined to small case series and case reports within the literature. We report an additional 19 cases to the literature (mean age 66.3 ± 11.7 years; 12 female [63%]; body mass index 32.9 ± 5.9 kg/m. 2. ). The mean time from index procedure to fracture was 7.8 years (2.5–16.3, ±3.6). Exeter V40 stem fracture incidence was 0.27%. Incidence was significantly higher in 125 mm length stems compared to ≥150 mm length stems (1.26% vs 0.13%, respectively, p <0.001) and revision arthroplasty (1.209% vs 0.149%, p <0.001). When comparing different stem length cohorts, 125-mm short-stem were associated with stem body fractures (92% vs 29%, p = 0.0095), earlier time to fracture (6.2 vs 11.0 years, p = 0.0018), younger patient age at time of fracture (62.7 vs 72.6 years old, p = 0.037) and female sex (75% vs 43%, p = 0.33). This case series in conjunction with the systematic review provides evidence stem morphology plays a role in femoral implant fracture. This complication remains rare, although we report a significantly higher incidence at up to 17 years follow-up than in the literature. As femoral geometries remain the same, increasing BMIs in THR patients should raise concern. Short 125 mm length Exeter V40 stems undoubtedly have a role in restoring anatomy and biomechanics in smaller femoral geometries, although the surgeon has to appreciate the higher risk of stem fracture and the associated predisposing factors which may necessitate meticulous surgical technique and planning


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_6 | Pages 37 - 37
1 May 2019
Windell L Haque A Kulkarni A Alabort E Barba D Reed R Singh H
Full Access

Background. Periprosthetic femoral fractures following total hip arthroplasty are relatively uncommon but are associated with significant morbidity. With an increasing number of total hip arthroplasties being carried out in an aging population we need to ensure correct implants are chosen for our patients. A recent review of NJR data suggested a significantly higher revision risk for the Zimmer CPT stems due to periprosthetic fractures when compared to the Stryker Exeter stems. Objectives. Our aim was to compare the biomechanics of periprosthetic fractures around the CPT and Exeter V40 stems in a composite saw bone model to identify if a difference in fracture risk exists between the two stems. We also compared the engineering design of the two implants in order to analyse the possible effect this may have on fracture risk. Study Design & Methods. Fourteen composite femurs were divided into two groups and cemented using Palacos R cement with either the CPT or Exeter V40 stem by a single surgeon. The implanted femurs were then mounted onto an Instron machine and were axially loaded and torqued to fracture with an axial compressive force of 2000N over 10 seconds followed by a rotation of 40 degrees applied over 1 second. A power calculation from a previous composite saw bone model study suggested that a minimum of 6 implanted femurs would be required in each group. Results. The implanted femurs invariably sustained fracture patterns similar to the Vancouver B2 periprosthetic fracture which are commonly seen in clinical practice. Implanted femurs with CPT stems suffered periprosthetic fractures with less rotation when compared to those femurs with the Exeter V40 stem (20.10 versus 33.60, p<0.01). We also found that CPT implanted femurs were fracturing at significantly lower torque values when compared to the Exeter V40 implanted femurs (124Nm Versus 174Nm, p<0.01). The energy release rate (G111) for CPT stems was 21.8Nm compared to 61.2Nm for Exeter V40 stems. The higher energy release with Exeter stems led to more comminuted fractures in Exeter implanted femurs when compared to the CPT femurs, which fractured earlier, but with simpler fracture patterns. Finite element method (FEM) simulation analysis showed that fractures initiated between the prosthesis and cement at the proximal end of the femur. Two dimensional sections at the same height showed a difference in bone-cement-implant geometrics at the critical point of failure suggesting that a design cause may be the reason for the higher risk of periprosthetic fractures in CPT implanted femurs. Conclusions. Our observations may explain the higher revision risk secondary to periprosthetic fractures that has been observed with the CPT stem when compared to the Exeter V40 stem


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_9 | Pages 18 - 18
1 May 2018
Mammoliti L Van Bavel D De Steiger R Rainbird S
Full Access

Introduction/Aims. The Exeter Stem can be used with metal femoral head that are made of either cobalt chrome, or stainless steel. The aim of this study was to compare the rates of revision of these two metal femoral head types when used with this femoral component. Method. Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) Data from September 1999 until December 2015 for all primary THRs using an Exeter or an Exeter v40 stem with the diagnosis of osteoarthritis were analysed. Only bearing couples that used a metal head with polyethylene were included. The cumulative percent revision (CPR) calculated using Kaplan-Meier estimates were compared for the two metal head types. CPR were further analysed by age, polyethylene type and head size. Reasons for revision and types of revision were assessed. Results. There were 51666 THR that used Exeter or Exeter v40 stems of which 12554 had femoral heads made of cobalt chrome and 39112 had heads made of stainless steel. There was no difference in the rate of revision overall when head types were compared. There was also no difference in CPR between the two head types with age. Hips that used cobalt chrome heads had a higher CPR than stainless steel heads when these were used against non-crosslinked polyethylene. When heads with a diameter of 32mm or greater were compared, those made of cobalt chrome had a higher rate of revision than stainless steel HR 1.38 (1.15, 1.66) P<0.001). Conclusion. There was no difference comparing cobalt chrome or stainless-steel heads, except where non-crosslinked polyethylene was used, or where the heads size was 32mm or greater. In these comparisons, heads made from cobalt chrome had a higher rate of revision


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_11 | Pages 1 - 1
7 Jun 2023
Gaston P Clement N Ohly N Macpherson G Hamilton D
Full Access

In this RCT the primary aim was to assess whether a short (125mm) Exeter V40 stem offered an equivalent hip specific function compared to the standard (150mm) stem when used for cemented total hip arthroplasty (THA). Secondary aims were to evaluate health-related quality of life (HRQoL), patient satisfaction, stem height and alignment, radiographic loosening, and complications between the two stems. A prospective multicentre double-blind randomised control trial was conducted. During a 15-month period, 220 patients undergoing THA were randomised to either a standard (n=110) or short (n=110) stem Exeter. There were no significant (p≥0.065) differences in preoperative variables between the groups. Functional outcomes and radiographic assessment were undertaken at 1- and 2-years. There were no differences (p=0.428) in hip specific function according to the Oxford hip score at 1-year (primary endpoint) or at 2-years (p=0.767) between the groups. The short stem group had greater varus angulation (0.9 degrees, p=0.003) when compared to the standard group and were more likely (odds ratio 2.42, p=0.002) to have varus stem alignment beyond one standard deviation from the mean. There were no significant (p≥0.083) differences in the Forgotten joint scores, EuroQol-5-Dimension, EuroQol-VAS, Short form 12, patient satisfaction, complications, stem height or radiolucent zones at 1 or 2-years between the groups. The Exeter short stem offers equivalent hip specific function, HRQoL, patient satisfaction, and limb length when compared to the standard stem at 2-years post-operation. However, the short stem was associated with a greater rate of varus malalignment which may influence future implant survival


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_4 | Pages 41 - 41
1 Apr 2022
Holleyman R Petheram T Reed M Burton P Malviya A
Full Access

Modular femoral stems offer surgeons great flexibility in biomechanical configuration during total hip replacement (THR) however introduce a taper-trunnion articulation known to be a source of additional wear debris through crevice, fretting and galvanic corrosion with mixed material combinations. This study aimed to investigate the influence of the trunnion bearing surface combination on the revision rate following primary total hip replacement (THR). All patients who underwent THR using an Exeter V40 cemented stainless steel stem and monobloc cemented polyethylene acetabular component (uncemented cups excluded to standardise the acetabular bearing surface and fixation) between January 2003 and December 2019 were extracted from the National Joint Registry for England, Wales, Northern Ireland, and the Isle of Man. The primary exposure was the head substrate used corresponding to the trunnion bearing. Time-to-event was determined by duration of implantation from primary surgery to revision with cases censored at death or end of available follow-up. Multivariable Cox proportional hazard models were used to identify predictors of all cause revision, adjusted for age, sex, American Association of Anaesthesiologists (ASA) grade, body mass index, surgical indication (osteoarthritis or other), and femoral head size. 229,870 THR were identified (66% female, mean age 73.4 years (SD 9.1) with the majority (91%) performed for osteoarthritis of which 4,598 were revised. Mean time from primary to revision or censoring was 6.8 years (SD 4.0). Multivariable modelling showed CoCr/SS trunnions were associated with a significantly higher risk of revision (hazard ratio (HR) 1.31 (95%CI 1.15 to 1.48, p<0.0001) as compared to SS/SS (reference). Both Alumina/SS (HR 0.74 (0.65 to 0.84), p<0.0001) and Zirconia/SS (HR 0.61 (0.49 to 0.74), p<0.0001) were associated with a significantly lower risk of revision. Ceramic heads on an Exeter stem were associated with significantly improved survivorship compared to metal heads in primary THR. CoCr/SS trunnion articulations had the poorest survivorship which may be contributed to by trunnionosis


Bone & Joint Research
Vol. 8, Issue 6 | Pages 246 - 252
1 Jun 2019
Liddle A Webb M Clement N Green S Liddle J German M Holland J

Objectives. Previous studies have evidenced cement-in-cement techniques as reliable in revision arthroplasty. Commonly, the original cement mantle is reshaped, aiding accurate placement of the new stem. Ultrasonic devices selectively remove cement, preserve host bone, and have lower cortical perforation rates than other techniques. As far as the authors are aware, the impact of ultrasonic devices on final cement-in-cement bonds has not been investigated. This study assessed the impact of cement removal using the Orthosonics System for Cemented Arthroplasty Revision (OSCAR; Orthosonics) on final cement-in-cement bonds. Methods. A total of 24 specimens were manufactured by pouring cement (Simplex P Bone Cement; Stryker) into stainless steel moulds, with a central rod polished to Stryker Exeter V40 specifications. After cement curing, the rods were removed and eight specimens were allocated to each of three internal surface preparation groups: 1) burr; 2) OSCAR; and 3) no treatment. Internal holes were recemented, and each specimen was cut into 5 mm discs. Shear testing of discs was completed by a technician blinded to the original grouping, recording ultimate shear strengths. Scanning electron microscopy (SEM) was completed, inspecting surfaces of shear-tested specimens. Results. The mean shear strength for OSCAR-prepared specimens (33.6 MPa) was significantly lower than for the control (46.3 MPa) and burr (45.8 MPa) groups (p < 0.001; one-way analysis of variance (ANOVA) with Tukey’s post hoc analysis). There was no significant difference in shear strengths between control and burr groups (p = 0.57). Scanning electron microscopy of OSCAR specimens revealed evidence of porosity undiscovered in previous studies. Conclusion. Results show that the cement removal technique impacts on final cement-in-cement bonds. This in vitro study demonstrates significantly weaker bonds when using OSCAR prior to recementation into an old cement mantle compared with cement prepared with a burr or no treatment. This infers that care must be taken in surgical decision-making regarding cement removal techniques used during cement-in-cement revision arthroplasty, suggesting that the risks and benefits of ultrasonic cement removal need consideration. Cite this article: A. Liddle, M. Webb, N. Clement, S. Green, J. Liddle, M. German, J. Holland. Ultrasonic cement removal in cement-in-cement revision total hip arthroplasty: What is the effect on the final cement-in-cement bond? Bone Joint Res 2019;8:246–252. DOI: 10.1302/2046-3758.86.BJR-2018-0313.R1


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_8 | Pages 77 - 77
11 Apr 2023
Nguyen K Torkan L Bryant T Wood G Ploeg H
Full Access

This study compared the pullout forces of the initial implantation and the “cement-in-cement” revision technique for short and standard-length (125 mm vs. 150 mm) Exeter. ®. V40 femoral stems used in total hip arthroplasty (THA). The idea that the pullout force for a double taper slip stem is relative to the force applied to the femur and that “cement-in-cement” revision provides the same reproduction of force. A total sample size of 15 femoral stems were tested (Short, n = 6 and Standard, n = 9). 3D printed fixtures for repeatable sample preparation were used to minimize variance during testing. To promote stem subsidence and to simulate an in vivo environment, the samples were placed in an incubator at 37°C at 100% humidity and experienced a constant compressive loading of 1335 N for 14 days. The samples underwent a displacement-controlled pullout test. After the initial pullout test, “cement-in-cement” revision will be performed and tested similar to the initial implantation to observe the efficacy of the revision technique. To compare the pullout forces between the two groups, a Kruskal-Wallis test using a significance level of 0.05 was conducted. The mean maximum pullout force for the short and standard-length femoral stems were 3939 ± 1178 N and 5078 ± 1168 N, respectively. The Kruskal-Wallis test determined no statistically significant difference between the two groups for the initial implantation (p = 0.13). The “cement-in-cement” revision pullout force will be conducted in future testing. This study demonstrated the potential use of short stem designs for THA as it provides similar levels of fixation as the standard-length femoral stem. The potential benefits for using a short stem design would be providing similar load transfer to the proximal femur, preserving proximal metaphyseal femoral bone in primary replacement, and reducing the invasiveness during revision


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_12 | Pages 4 - 4
1 Jun 2017
Liddle A German M Green S Townsend A Webb M Holland J
Full Access

Numerous studies have evidenced cement-in-cement techniques as reliable in revision arthroplasty. The original cement mantle is commonly reshaped to aid accurate placement of the new stem. Ultrasonic devices selectively remove cement, preserve host bone and have lower cortical perforation rates than other techniques. As far as the authors are aware, their impact on final cement-cement bonds has not been investigated. This study assessed the impact of cement removal using OSCAR (Orthosonics System for Cemented Arthroplasty Revision, ORTHOSONICS) on final cement-cement bonds. Twenty-four specimens were manufactured by pouring cement (Simplex P Bone Cement, Stryker) into stainless-steel moulds with a central rod polished to Stryker Exeter V40 specifications. After cement curing, rods were removed and eight specimens allocated to each of three internal surface preparation groups: 1) burr; 2) OSCAR; or 3) no treatment. Internal holes were re-cemented, then each specimen was cut into 5mm discs. Shear testing of discs was completed by a technician blinded to original grouping (Instron 5567, UK), recording ultimate shear strengths. The mean shear strength for OSCAR-prepared specimens (17 MPa, 99% CI 14.9 to 18.6, SD=4.0) was significantly lower than that measured for the control (23 MPa, 99% CI 22.5 to 23.7, SD=1.4) and burr (23 MPa, 99% CI 22.1 to 23.7, SD=1.9) groups (P<0.001, one-way ANOVA with Tukey's post-hoc analysis). There was no significant difference between control and burr groups (P>0.05). Results show that cement removal technique impacts on final cement-cement bonds. This in vitro study shows a significantly weaker bond when using OSCAR prior to re-cementation into an old cement mantle, compared to cement prepared with a burr or no treatment. These results have implications for surgical practice and decision-making about specific cement removal techniques used during cement-in-cement revision arthroplasty, suggesting that the risks and benefits of ultrasonic cement removal need careful consideration


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XI | Pages 24 - 24
1 Apr 2012
Wilkinson A Cooney A Chappell A
Full Access

In recent years the majority of X-ray departments have moved to a digital format of recording and archiving radiographs. These digital images (as with previous ‘films’) have a built in magnification factor (variable with each patient), which, may cause errors in templating for joint replacement surgery. Placing a marker of known size at the same level as the joint in question allows calculation of the magnification. This may help to restore hip offset in total hip replacement. To establish the magnification factor for digital radiographs taken in our unit. To assess the usefulness of marker images in accurate preoperative templating. Preoperative marker radiographs were identified retrospectively. The apparent size of the marker was measured on digital image. This value was used to calculate the magnification of the image. The scaled X-ray was up loaded to a digital templating software programme. This software uses a ‘scaling tool’ to calculate the magnification of the image. The hip joint templating tool was the used to calculate the offset of the proximal femur, this was performed with the calculated magnification and also an assumed magnification of 120%. The recommended offset of Exeter V40 stem was noted for both values. Images were identified for 40 patients with markers. The average magnification was 122% for both PACS and Orthoview with a range 113% – 129% and a standard deviation of 4%. The median value for magnification was 120%. The average change in offset between calculated and estimated magnification was 1.275mm with a maximum change of 3mm. In two cases this difference resulted in a change in the recommended offset (5%). The use of marker radiographs is widely described. In this small series the magnification is the same as previously reported in other studies. The difference in offset between calculated and estimated magnification was relatively small and caused a change in the recommended offset in only two patients. Variation in the use of the templating tool in our software can produce a much greater change in offset. Marker radiographs will only be useful as part of a standardised method of pre-operative templating


Bone & Joint Open
Vol. 5, Issue 6 | Pages 514 - 523
24 Jun 2024
Fishley W Nandra R Carluke I Partington PF Reed MR Kramer DJ Wilson MJ Hubble MJW Howell JR Whitehouse SL Petheram TG Kassam AM

Aims

In metal-on-metal (MoM) hip arthroplasties and resurfacings, mechanically induced corrosion can lead to elevated serum metal ions, a local inflammatory response, and formation of pseudotumours, ultimately requiring revision. The size and diametral clearance of anatomical (ADM) and modular (MDM) dual-mobility polyethylene bearings match those of Birmingham hip MoM components. If the acetabular component is satisfactorily positioned, well integrated into the bone, and has no surface damage, this presents the opportunity for revision with exchange of the metal head for ADM/MDM polyethylene bearings without removal of the acetabular component.

Methods

Between 2012 and 2020, across two centres, 94 patients underwent revision of Birmingham MoM hip arthroplasties or resurfacings. Mean age was 65.5 years (33 to 87). In 53 patients (56.4%), the acetabular component was retained and dual-mobility bearings were used (DM); in 41 (43.6%) the acetabulum was revised (AR). Patients underwent follow-up of minimum two-years (mean 4.6 (2.1 to 8.5) years).


Bone & Joint Open
Vol. 3, Issue 9 | Pages 710 - 715
5 Sep 2022
Khan SK Tyas B Shenfine A Jameson SS Inman DS Muller SD Reed MR

Aims

Despite multiple trials and case series on hip hemiarthroplasty designs, guidance is still lacking on which implant to use. One particularly deficient area is long-term outcomes. We present over 1,000 consecutive cemented Thompson’s hemiarthroplasties over a ten-year period, recording all accessible patient and implant outcomes.

Methods

Patient identifiers for a consecutive cohort treated between 1 January 2003 and 31 December 2011 were linked to radiographs, surgical notes, clinic letters, and mortality data from a national dataset. This allowed charting of their postoperative course, complications, readmissions, returns to theatre, revisions, and deaths. We also identified all postoperative attendances at the Emergency and Outpatient Departments, and recorded any subsequent skeletal injuries.


Bone & Joint Open
Vol. 3, Issue 3 | Pages 196 - 204
4 Mar 2022
Walker RW Whitehouse SL Howell JR Hubble MJW Timperley AJ Wilson MJ Kassam AM

Aims

The aim of this study was to assess medium-term improvements following total hip arthroplasty (THA), and to evaluate what effect different preoperative Oxford Hip Score (OHS) thresholds for treatment may have on patients’ access to THA and outcomes.

Methods

Patients undergoing primary THA at our institution with an OHS both preoperatively and at least four years postoperatively were included. Rationing thresholds were explored to identify possible deprivation of OHS improvement.


Bone & Joint Research
Vol. 10, Issue 10 | Pages 629 - 638
20 Oct 2021
Hayashi S Hashimoto S Kuroda Y Nakano N Matsumoto T Ishida K Shibanuma N Kuroda R

Aims

This study aimed to evaluate the accuracy of implant placement with robotic-arm assisted total hip arthroplasty (THA) in patients with developmental dysplasia of the hip (DDH).

Methods

The study analyzed a consecutive series of 69 patients who underwent robotic-arm assisted THA between September 2018 and December 2019. Of these, 30 patients had DDH and were classified according to the Crowe type. Acetabular component alignment and 3D positions were measured using pre- and postoperative CT data. The absolute differences of cup alignment and 3D position were compared between DDH and non-DDH patients. Moreover, these differences were analyzed in relation to the severity of DDH. The discrepancy of leg length and combined offset compared with contralateral hip were measured.