Abstract
In this RCT the primary aim was to assess whether a short (125mm) Exeter V40 stem offered an equivalent hip specific function compared to the standard (150mm) stem when used for cemented total hip arthroplasty (THA). Secondary aims were to evaluate health-related quality of life (HRQoL), patient satisfaction, stem height and alignment, radiographic loosening, and complications between the two stems.
A prospective multicentre double-blind randomised control trial was conducted. During a 15-month period, 220 patients undergoing THA were randomised to either a standard (n=110) or short (n=110) stem Exeter. There were no significant (p≥0.065) differences in preoperative variables between the groups. Functional outcomes and radiographic assessment were undertaken at 1- and 2-years.
There were no differences (p=0.428) in hip specific function according to the Oxford hip score at 1-year (primary endpoint) or at 2-years (p=0.767) between the groups. The short stem group had greater varus angulation (0.9 degrees, p=0.003) when compared to the standard group and were more likely (odds ratio 2.42, p=0.002) to have varus stem alignment beyond one standard deviation from the mean. There were no significant (p≥0.083) differences in the Forgotten joint scores, EuroQol-5-Dimension, EuroQol-VAS, Short form 12, patient satisfaction, complications, stem height or radiolucent zones at 1 or 2-years between the groups.
The Exeter short stem offers equivalent hip specific function, HRQoL, patient satisfaction, and limb length when compared to the standard stem at 2-years post-operation. However, the short stem was associated with a greater rate of varus malalignment which may influence future implant survival.