Vol. 99-B, Issue SUPP_3 | Pages 84 - 84
1 Feb 2017
Coyle R Bas M Rodriguez J Hepinstall M
Full Access

Background

Posterior referencing (PR) total knee arthroplasty (TKA) aims to restore posterior condylar offset. When a symmetric femoral implant is externally rotated (ER) to the posterior condylar axis, it is impossible to anatomically restore the offset of both condyles. PR jigs variously reference medially, laterally, or centrally. The distal femoral cutting jigs typically reference off the more distal medial condyle, causing distal and posterior resection discrepancies. We used sawbones to elucidate differences between commonly used PR cutting jigs with regards to posterior offset restoration.

Materials/Methods

Using 32 identical sawbones, we performed distal and posterior femoral resections using cutting guides from 8 widely available TKA systems. 6 systems used a central-referencing strategy, 1 system used a lateral-referencing strategy, and 1 system used a medial-referencing strategy with implants of asymmetric thickness. Distal femoral valgus resection was set at 5 degrees for all specimens. Rotation was set at 3 degrees for 2 sawbones and 5 degrees for 2 sawbones with each system. We measured the thickness of all bone resections, and compared those values to known implant thickness.